Problem
e.g.
Measurement or
inflation does not
start instantly.
Readings from
repeated mea-
surements differ
considerably.
Prior to a mea-
surement, the
position sensor
icon on the display
does not work.
Reason
Error message lower than
64:
System error
Right measurement
position was not reached.
Blood pressure is a
fluctuating value.
For healthy adults,
deviations of 10 to
20 mmHg are possible.
Cuff is not properly fitted.
Readings were not con-
sistently taken at heart
level.
Talking, coughing,
laughing, moving etc.
when measuring will
influence the reading.
Position sensor is de-
activated.
Solution
Have the product checked at a
Braun Service Centre.
Repeat measurement as shown
and described under «Measuring»
or deactivate the position sensor.
none
Ensure that the appliance is fitted
as follows:
• on top of the slim side of the
forearm
• with the display on top
• with the battery compartment
facing the user's body
• approx. 2 cm away from the
thumb root
For each measurement, proceed
exactly as shown and described
under «Measuring».
While measuring, relax, keep still,
do not move, or speak.
Refer to the chapter «Settings»
to activate the position sensor.
11