Measuring Frequency

XW Warning
To avoid electrical shock, disregard the bar graph for
frequencies > 1 kHz. If the frequency of the measured
signal is > 1 kHz, the bar graph is unspecified.
The Meter measures the frequency of a signal. The trigger level is
0 V, 0 A AC for all ranges.
AC/DC Voltage Frequency
Hz
HOLD
MIN MAX
RANGE
V
V
+
To exit frequency, press YELLOW button or turn the rotary
switch.
In frequency, the bar graph shows the AC/DC voltage or AC
current accurately up to 1 kHz.
Select progressively lower ranges using manual ranging for a
stable reading.
AC Current Frequency
x 2
Hz
HOLD
MIN MAX
RANGE
mA
A
+
AIK09F.EPS

Using the Bar Graph

The bar graph is like the needle on an analog Meter. It has an
overload indicator (>) to the right and a polarity indicator ( ) to the
left.
Because the bar graph updates about 40 times per second, which
is 10 times faster than the digital display, the bar graph is useful
for making peak and null adjustments and for observing rapidly
changing inputs.
The bar graph is disabled when measuring capacitance or
temperature. In frequency, the bar graph accurately indicates the
voltage or current up to 1 kHz.
The number of lit segments indicates the measured value and is
relative to the full-scale value of the selected range.
In the 60 V range, for example (see below), the major divisions on
the scale represent 0, 15, 30, 45, and 60 V. An input of 30 V
lights the negative sign and the segments up to the middle of the
scale.
True RMS Multimeters
Using the Bar Graph
AIK11F.EPS
9