Download Print this page

Operational Considerations - Agilent Technologies PN 8510-16 Product Note

Controlling test port output power flatness

Advertisement

Operational Considerations

Calibration time
Table 5. Typical calibration times (in minutes) versus number of calibration points with maximum leveled source power
Number of Points
83620B/83621B
with 8514B
801
30
401
15
201
8
101
4
51
2
1 Warning—RF Unleveled
No IF Found
83631B
with 8515A
with 8515A
32
40
16
20
8
10
4
5
2
2
IF Overload
Source
83651B
with 8517B
60
30
15
8
4
Power control with flatness correction
7