Products described in this manual are manufactured under international patents and one or more of the following U.S. patents: 5,142,483, 5,257,640, 5,331,838, 5,445,035. Other U.S. and international patents pending. DH Instruments, DH, DHI, PG7000, PG7302 and PG7307 are trademarks, registered and otherwise, of DH Instruments, a Fluke Company.
This manual is designed to be used in conjunction with the PG7000 Piston Gauges Operation and Maintenance Manual, to operate a PG7307 Piston Gauge. PG7307 is similar in conception and features to a PG7302 piston gauge. PG7307 differs in that it includes provisions for the application of an independent controlled clearance pressure to the outside surface of the cylinder.
1. INTRODUCTION PRODUCT OVERVIEW PG7307 is a special model in the PG7000 line of piston gauges. PG7307 is identical to PG7302 except for its piston-cylinder modules and mounting post. PG7307 is designed to allow an independent controlled clearance pressure (CCP) to be applied to the outside surface of the cylinder.
1. INTRODUCTION SPECIFICATIONS 1.4.1 GENERAL SPECIFICATIONS Specifications below are specific to PG7307. All other specifications are identical to PG7000 specifications found in the PG7000 Operation and Maintenance Manual (P/N 550099). 13 kg (29 lb) Weight Instrument platform with no mass loaded.
The controlled clearance pressure is introduced through the CCP port on the back of the PG7307 platform. The pressure is applied to the cylinder through the mounting post to a separate circuit in the piston-cylinder module (see Figure 2).
Body oils and acids can permanently etch the surfaces. In normal use, always maintain PG7307 oil piston-cylinders vertical with the O-ring assembly down. Do not invert the assembly and allow oil to run up the piston head into the adjustment mass and cap.
PG7307™ OPERATION AND MAINTENANCE MANUAL To assemble and disassemble a PG7307 piston cylinder module, proceed as follows: so as NOT to apply torque to the piston (7) when removing the cap. When installing the piston cap during reassembly, remember to reinstall the adjustment mass (3).
6. WARRANTY STATEMENT Except to the extent limited or otherwise provided herein, DH Instruments, a Fluke Company (DHI) warrants for one year from purchase, each new product sold by it or one of its authorized distributors, only against defects in workmanship and/or materials under normal service and use. Products which have been changed or altered in any manner from their original design, or which are improperly or defectively installed, serviced or used are not covered by this warranty.