Ohmic Measurement System Calibration/Adjustment Procedure - Fluke MOLBOX1 PLUS - V 6.00 Operation And Maintenance Manual

Molbloc terminal (ver. 6.00 and higher)
Table of Contents

Advertisement

5.4.1
OHMIC Measurement System Calibration/Adjustment
Procedure
The molbox1+ channels are calibrated individually. The standard calibration procedure
requires at least two simulators and is performed as follows:
 With Channel A selected, press [SPECIAL], <4cal>, <2temp> to read and record the
current values of TA and TM for Channel A, then exit to the main run screen to select
Channel B and repeat the process to read and record the current values of TA and TM
for Channel B.
 With Channel A selected, the [P&T] button twice to display the real time temperature
readings.
 Connect a molbloc simulator to the molbox1+ Channel A and initialize it as if connecting a
molbloc (see Section 3.5.1).
 Enter the average molbloc simulator calibrated temperature and the molbox1+ top row
centered calibration temperature reading (the averaged temperature) for each
temperature point into a spreadsheet. Repeat steps  and  for each simulator.
Calculate the "non-corrected" temperature readings by backing out the as received TA
and TM recorded in Step  above, following:
non-corrected reading = (corrected reading - TA)/TM
 Perform a best fit linear regression to find the offset and slope that best fit the non-
corrected temperature readings to the calibrated molbloc simulator temperatures. Most
commercial spreadsheet products have a standard function to perform a linear
regression. The offset, in the current temperature unit, is the new value of TA, the slope
is the new value of TM.
 Press [SPECIAL], <4cal>, <2temp> to write the new values of TA (in the current unit of
measure) and TM and the new calibration date (YYYYMMDD) for the channel that is
being calibrated.
 Calculate "as left data" for the calibration:
as left reading = (non-corrected reading • new TM) + new TA
 If desired, verify the as left data by rerunning the calibration with the new TA and TM
applied and checking that the disagreements between the molbloc simulators and the
molbox1+ temperature readings are in tolerance.
 Repeat steps  thru  for Channel B.
The value of TA must be entered in the current unit of measure, either °C
or °F, that is displayed by the molbox1+.
If you are not familiar or comfortable with the mathematics used for TA
and TM manipulation, it is recommended you verify the as left readings as
suggested in Step  above to provide practical confirmation of the
integrity of the calculations.
If an out-of-tolerance condition persists, the molbox1+ and/or the molbloc
simulator(s) may need service.
5. MAINTENANCE, ADJUSTMENTS AND CALIBRATION
 Caution
Page 151
© 1995 - 2010 Fluke Corporation, DH Instruments Division

Advertisement

Table of Contents
loading

This manual is also suitable for:

Molbox1+s

Table of Contents