Error Messages; Table 39. Error Messages - Fluke MOLBOX1 PLUS - V 6.00 Operation And Maintenance Manual

Molbloc terminal (ver. 6.00 and higher)
Table of Contents

Advertisement

4.3.3

Error Messages

molbox1+ always replies to a command. If the command is incorrect or contains invalid data,
an error number is returned in the form "ERR# n" where n is an integer number that
represents a specific error.
Table 31 is a list of the possible error numbers and the error description for each.
REPLY
ERROR
ERR# 0
ERR# 1
ERR# 2
ERR# 3
ERR# 4
ERR# 5
ERR# 6
ERR# 7
ERR# 8
ERR# 9
ERR# 10
ERR# 11
ERR# 12
ERR# 13
ERR# 14
ERR# 15
ERR# 16
ERR# 17
ERR# 18
ERR# 21
ERR# 22
ERR# 23
ERR# 24
ERR# 25
ERR# 26
ERR# 27
ERR# 28
ERR# 29
ERR# 30
ERR# 31
ERR# 32
ERR# 34
ERR# 35
ERR# 36
ERR# 37
ERR# 38
ERR# 39
ERR# 40
ERR# 41
ERR# 42
ERR# 43
ERR# 44
ERR# 45
ERR# 46
ERR# 47
ERR# 48
ERR# 49
This allows for easy error trapping by the host computer.

Table 39. Error Messages

DESCRIPTION
"The molbox1+ is in high security level and cannot accept
"User defined coefficient cannot be 0"
"External device not detected"
"External device improperly configured"
"Numeric argument missing or out of range"
"Missing or improper command argument(s)"
"External device timeout error"
"MFC not defined or selected"
"Command missing argument"
"Text detected in numeric field"
"Averaging cycle not started"
"Selected gas not available"
"Command not yet available"
"Option not available or installed"
"COM port failed to initialize"
"Internal device #1 timeout error"
"Internal device #2 timeout error"
"Fatal mass flow calculation error"
"Incompatible device detected""
"Selected range not available"
"Temperature change was too great"
"Offset was excessive. Check tare"
Page 115
© 1995 - 2010 Fluke Corporation, DH Instruments Division
4. REMOTE OPERATION
remote commands"
"OK"
"molbloc flow is too great"
"Text argument is too long"
"Unknown command"
"System overpressured"
"User unit not defined"
"MFC Malfunction"
"User device not defined"
"Pressure is not stable"
"molbloc not detected"
"RPT out of calibration"
"Busy averaging"
"molbloc EEPROM is full"
"molbloc gas not found"
"molbloc is write protected"
"molbloc write error"
"MFC unit mismatch"
"molbloc verify error"
"molbloc read error"
"molbloc invalid"
"PRT measurement error"
"Incorrect MFC mode"
"Entry already exists"
"Argument(s) not allowed"
"The leak is too large"
"Possible upstream leak"