Wavetek 9105 Operator's Manual

Universal calibration system

Advertisement

Quick Links

O
'
G
PERATOR
S
UIDE
Model 9100
Universal Calibration System
www.valuetronics.com

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the 9105 and is the answer not in the manual?

Questions and answers

Summarization of Contents

Interconnections
External Leads
Details on how external leads affect specification degradation.
Model 9105 Lead Set
Information on the Model 9105 Lead Set for optimal connections.
Connection Accessories
Description of accessories for specific connection purposes.
DC Voltage Calibration
3.1 Interconnections
Connection diagram for DC Voltage calibration.
3.2 The 9100 as a Fixed Source
Procedure for using the 9100 as a fixed DC voltage source for calibration.
3.2.1 9100 and UUT Setup (Fixed Source)
Steps for setting up the 9100 and UUT for DC Voltage calibration.
3.2.2 Sequence of Operations (Fixed Source)
Step-by-step guide for calibrating DC Voltage using the 9100.
3.3 The 9100 as an Adjustable Source
Procedure for using the 9100 as an adjustable DC voltage source for calibration.
3.3.1 9100 and UUT Setup (Adjustable Source)
Setup steps for the 9100 and UUT for adjustable DC Voltage calibration.
3.3.2 Sequence of Operations (Adjustable Source)
Step-by-step guide for adjustable DC Voltage calibration.
AC Voltage Calibration
4.1 Interconnections
Connection diagram for AC Voltage calibration.
4.2 The 9100 as a Fixed Source
Procedure for using the 9100 as a fixed AC voltage source for calibration.
4.2.1 9100 and UUT Setup (Fixed Source)
Steps for setting up the 9100 and UUT for AC Voltage calibration.
4.2.2 Sequence of Operations (Fixed Source)
Step-by-step guide for calibrating AC Voltage using the 9100.
4.3 The 9100 as an Adjustable Source
Procedure for using the 9100 as an adjustable AC voltage source for calibration.
4.3.1 9100 and UUT Setup (Adjustable Source)
Setup steps for the 9100 and UUT for adjustable AC Voltage calibration.
4.3.2 Sequence of Operations (Adjustable Source)
Step-by-step guide for adjustable AC Voltage calibration.
DC Current Calibration
5.1 Interconnections
Connection schemes for DC Current UUT calibration.
5.1 Interconnections (Contd.)
Additional interconnection details for DC Current calibration.
5.1 Interconnections (Contd.)
Notes on positioning current sensing clamp in coils for calibration.
5.2 The 9100 as a Fixed Source
Procedure for using the 9100 as a fixed DC current source for calibration.
5.2.1 9100 and UUT Setup (Fixed Source)
Steps for setting up the 9100 and UUT for DC Current calibration.
5.2.2 Sequence of Operations (Fixed Source)
Step-by-step guide for calibrating DC Current using the 9100.
5.3 The 9100 as an Adjustable Source
Procedure for using the 9100 as an adjustable DC current source for calibration.
5.3.1 9100 and UUT Setup (Adjustable Source)
Setup steps for the 9100 and UUT for adjustable DC Current calibration.
5.3.2 Sequence of Operations (Adjustable Source)
Step-by-step guide for adjustable DC Current calibration.
AC Current Calibration
6.1 Interconnections
Connection schemes for AC Current UUT calibration.
6.1 Interconnections (Contd.)
Additional interconnection details for AC Current calibration.
6.1 Interconnections (Contd.)
Notes on positioning current sensing clamp in coils for AC calibration.
6.2 The 9100 as a Fixed Source
Procedure for using the 9100 as a fixed AC current source for calibration.
6.2.1 9100 and UUT Setup (Fixed Source)
Steps for setting up the 9100 and UUT for AC Current calibration.
6.2.2 Sequence of Operations (Fixed Source)
Step-by-step guide for calibrating AC Current using the 9100.
6.3 The 9100 as an Adjustable Source
Procedure for using the 9100 as an adjustable AC current source for calibration.
6.3.1 9100 and UUT Setup (Adjustable Source)
Setup steps for the 9100 and UUT for adjustable AC Current calibration.
6.3.2 Sequence of Operations (Adjustable Source)
Step-by-step guide for adjustable AC Current calibration.
Resistance Calibration
7.1 Interconnections
General connection scheme for UUT Resistance calibration.
7.2 The 9100 as a Fixed Source
Procedure for using the 9100 as a fixed Resistance source for calibration.
7.2.1 9100 and UUT Setup (Fixed Source)
Steps for setting up the 9100 and UUT for Resistance calibration.
7.2.2 Sequence of Operations (Fixed Source)
Step-by-step guide for calibrating Resistance using the 9100.
7.3 The 9100 as an Adjustable Source
Procedure for using the 9100 as an adjustable Resistance source for calibration.
7.3.1 9100 and UUT Setup (Adjustable Source)
Setup steps for the 9100 and UUT for adjustable Resistance calibration.
7.3.2 Sequence of Operations (Adjustable Source)
Step-by-step guide for adjustable Resistance calibration.
Conductance Calibration
8.1 Interconnections
General connection scheme for UUT Conductance calibration.
8.2 The 9100 as a Fixed Source
Procedure for using the 9100 as a fixed Conductance source for calibration.
8.2.1 9100 and UUT Setup (Fixed Source)
Steps for setting up the 9100 and UUT for Conductance calibration.
8.2.2 Sequence of Operations (Fixed Source)
Step-by-step guide for calibrating Conductance using the 9100.
8.3 The 9100 as an Adjustable Source
Procedure for using the 9100 as an adjustable Conductance source for calibration.
8.3.1 9100 and UUT Setup (Adjustable Source)
Setup steps for the 9100 and UUT for adjustable Conductance calibration.
8.3.2 Sequence of Operations (Adjustable Source)
Step-by-step guide for adjustable Conductance calibration.
Frequency Calibration
9.1 Interconnections
Connection diagram for Frequency calibration.
9.2 The 9100 as a Fixed Source
Procedure for using the 9100 as a fixed Frequency source for calibration.
9.2.1 9100 and UUT Setup (Fixed Source)
Steps for setting up the 9100 and UUT for Frequency calibration.
9.2.2 Sequence of Operations (Fixed Source)
Step-by-step guide for calibrating Frequency using the 9100.
9.3 The 9100 as an Adjustable Source
Procedure for using the 9100 as an adjustable Frequency source for calibration.
9.3.1 9100 and UUT Setup (Adjustable Source)
Setup steps for the 9100 and UUT for adjustable Frequency calibration.
9.3.2 Sequence of Operations (Adjustable Source)
Step-by-step guide for adjustable Frequency calibration.
Mark/Period Calibration
10.1 Interconnections
Connection diagram for Mark/Period calibration.
10.2 The 9100 as a Fixed Source
Procedure for using the 9100 as a fixed Mark/Period source for calibration.
10.2.1 9100 and UUT Setup (Fixed Source)
Steps for setting up the 9100 and UUT for Mark/Period calibration.
10.2.2 Sequence of Operations (Fixed Source)
Step-by-step guide for calibrating Mark/Period using the 9100.
10.3 The 9100 as an Adjustable Source
Procedure for using the 9100 as an adjustable Mark/Period source for calibration.
10.3.1 9100 and UUT Setup (Adjustable Source)
Setup steps for the 9100 and UUT for adjustable Mark/Period calibration.
10.3.2 Sequence of Operations (Adjustable Source)
Step-by-step guide for adjustable Mark/Period calibration.
% Duty Calibration
11.1 Interconnections
Connection diagram for % Duty calibration.
11.2 The 9100 as a Fixed Source
Procedure for using the 9100 as a fixed % Duty source for calibration.
11.2.1 9100 and UUT Setup (Fixed Source)
Steps for setting up the 9100 and UUT for % Duty calibration.
11.2.2 Sequence of Operations (Fixed Source)
Step-by-step guide for calibrating % Duty using the 9100.
11.3 The 9100 as an Adjustable Source
Procedure for using the 9100 as an adjustable % Duty source for calibration.
11.3.1 9100 and UUT Setup (Adjustable Source)
Setup steps for the 9100 and UUT for adjustable % Duty calibration.
11.3.2 Sequence of Operations (Adjustable Source)
Step-by-step guide for adjustable % Duty calibration.
Auxiliary Calibration
12.1 Auxiliary Functions
Overview of auxiliary functions: Capacitance, Temperature, Logic.
12.1.1 Function Grouping and Selection
How to access and select auxiliary functions via the 'Aux' key.
12.1.2 Calibration Procedures
Instructions for calibrating UUTs in auxiliary functions are in respective sections.
Capacitance Calibration
13.1 Interconnections
Connection scheme for UUT Capacitance calibration.
13.2 The 9100 as a Fixed Source
Procedure for using the 9100 as a fixed Capacitance source for calibration.
13.2.1 9100 and UUT Setup (Fixed Source)
Steps for setting up the 9100 and UUT for Capacitance calibration.
13.2.2 Sequence of Operations (Fixed Source)
Step-by-step guide for calibrating Capacitance using the 9100.
13.3 The 9100 as an Adjustable Source
Procedure for using the 9100 as an adjustable Capacitance source for calibration.
13.3.1 9100 and UUT Setup (Adjustable Source)
Setup steps for the 9100 and UUT for adjustable Capacitance calibration.
13.3.2 Sequence of Operations (Adjustable Source)
Step-by-step guide for adjustable Capacitance calibration.
Thermocouple Temperature Calibration
14.1 Interconnections
General connection scheme for Thermocouple Temperature calibration.
14.2 The 9100 as a Fixed Source
Procedure for using the 9100 as a fixed Thermocouple source for calibration.
14.2.1 9100 and UUT Setup (Fixed Source)
Steps for setting up the 9100 and UUT for Thermocouple calibration.
14.2.2 Sequence of Operations (Fixed Source)
Step-by-step guide for calibrating Thermocouple Temperature using the 9100.
14.3 The 9100 as an Adjustable Source
Procedure for using the 9100 as an adjustable Thermocouple source for calibration.
14.3.1 9100 and UUT Setup (Adjustable Source)
Setup steps for the 9100 and UUT for adjustable Thermocouple calibration.
14.3.2 Sequence of Operations (Adjustable Source)
Step-by-step guide for adjustable Thermocouple calibration.
RTD Temperature Calibration
15.1 Interconnections
General connection scheme for UUT RTD Temperature calibration.
15.2 The 9100 as a Fixed Source
Procedure for using the 9100 as a fixed RTD Temperature source for calibration.
15.2.1 9100 and UUT Setup (Fixed Source)
Steps for setting up the 9100 and UUT for RTD Temperature calibration.
15.2.2 Sequence of Operations (Fixed Source)
Step-by-step guide for calibrating RTD Temperature using the 9100.
15.3 The 9100 as an Adjustable Source
Procedure for using the 9100 as an adjustable RTD Temperature source for calibration.
15.3.1 9100 and UUT Setup (Adjustable Source)
Setup steps for the 9100 and UUT for adjustable RTD Temperature calibration.
15.3.2 Sequence of Operations (Adjustable Source)
Step-by-step guide for adjustable RTD Temperature calibration.
Logic Pulses Calibration
16.1 Interconnections
Connection diagram for Logic Pulses calibration.
16.2 The 9100 as a Fixed Source
Procedure for using the 9100 as a fixed Logic Pulses source for calibration.
16.2.1 9100 and UUT Setup (Fixed Source)
Steps for setting up the 9100 and UUT for Logic Pulses calibration.
16.2.2 Sequence of Operations (Fixed Source)
Step-by-step guide for calibrating Logic Pulses using the 9100.
16.3 The 9100 as an Adjustable Source
Procedure for using the 9100 as an adjustable Logic Pulses source for calibration.
16.3.1 9100 and UUT Setup (Adjustable Source)
Setup steps for the 9100 and UUT for adjustable Logic Pulses calibration.
16.3.2 Sequence of Operations (Adjustable Source)
Step-by-step guide for adjustable Logic Pulses calibration.
Logic Levels Calibration
17.1 Interconnections
Connection diagram for Logic Levels calibration.
17.4 The 9100 as a Fixed Source
Procedure for using the 9100 as a fixed Logic Levels source for calibration.
17.4.1 9100 and UUT Setup (Fixed Source)
Steps for setting up the 9100 and UUT for Logic Levels calibration.
17.4.2 Sequence of Operations (Fixed Source)
Step-by-step guide for calibrating Logic Levels using the 9100.
17.5 The 9100 as an Adjustable Source
Procedure for using the 9100 as an adjustable Logic Levels source for calibration.
17.5.1 9100 and UUT Setup (Adjustable Source)
Setup steps for the 9100 and UUT for adjustable Logic Levels calibration.
17.5.2 Sequence of Operations (Adjustable Source)
Step-by-step guide for adjustable Logic Levels calibration.
Error Messages
18.1 Error Messages Overview
General overview of error messages displayed on the Model 9100.
Type A Fatal System Errors
Description of fatal system errors and required actions.
Type B General Operating Errors
Description of general operating errors and required actions.
Type C Overload Errors
Description of overload errors (current, voltage, temp) and actions.
Type D Calibration Errors
Description of calibration errors and required actions.
Type E Cold Junction Compensation Errors
Description of cold junction compensation errors and actions.
Type F Configuration Errors
Description of configuration errors and required actions.
18.2 Wavetek Service Center Support
Guidance on contacting Wavetek Service Center for support and fault diagnosis.
18.3 Alphabetic List of Error Messages
Alphabetical listing of error messages and their types.