A3 PLUS - A5 - A6 PLUS - L6 SURGICAL SINGLE CART GENERAL WARNINGS • These instructions explain how to correctly use the following dental unit: A3 PLUS (SN.71IF, 71IG) A6 PLUS A3 PLUS CONTINENTAL A5 CONTINENTAL A6 PLUS CONTINENTAL L6 CONTINENTAL (SN.71HU)
A3 PLUS - A5 - A6 PLUS - L6 SURGICAL SINGLE CART WARNING: Class 2 light source according to On (a part of the unit). Standard IEC 62471:2006. Equipment On. WARNING: hand crushing hazard. Equipment OFF. Mark of conformity with technical regulations of Ukraine.
A3 PLUS - A5 - A6 PLUS - L6 SURGICAL SINGLE CART Age: approximately 4 years and older, according to the physical characteristics described below (height and weight) Height: from 1 m. Weight: see "Maximum permissible load on the dental chair" in the DATA SPECIFICATIONS paragraph.
A3 PLUS - A5 - A6 PLUS - L6 SURGICAL SINGLE CART 1.2.2. ENVIRONMENTAL CONDITIONS The equipment is to be installed in rooms that satisfy the following requirements: • temperature between 10 and 40°C; • relative humidity between 30 and 75%;...
A3 PLUS - A5 - A6 PLUS - L6 SURGICAL SINGLE CART 1.3. SAFETY WARNINGS • All equipment is permanently installed. Depending on the type of dental chair the dental unit comes with, refer to the installation SCHEMATICS in paragraph DATA SPECIFICATIONS The CEFLA s.c.
A3 PLUS - A5 - A6 PLUS - L6 SURGICAL SINGLE CART • For operators in Europe: any serious accident occurred in relation to the device must be reported to CEFLA s.c. and to the competent authority of the Member State where the user and/or patient lives.
A3 PLUS - A5 - A6 PLUS - L6 SURGICAL SINGLE CART 1.5. CLEANING AND DISINFECTION Cleaning is the first step required for any disinfection process. Physically scrubbing with detergents and surface-active substances and rinsing with water removes a considerable amount of micro-organisms. If the surface is not first cleaned, the disinfection process cannot be successful.
A3 PLUS - A5 - A6 PLUS - L6 SURGICAL SINGLE CART 1.6. STERILISATION Every instrument supplied is NOT STERILE and must be sterilized in steam autoclave (max. 135°C) before use, without performing any type of chemical sterilization. Sterilization must be performed using appropriate packaging materials, verified during the sterilization process validation.
A3 PLUS - A5 - A6 PLUS - L6 SURGICAL SINGLE CART DESCRIPTION OF THE EQUIPMENT 2.1. IDENTIFICATION PLATES Data given on plate: • Name of manufacturer. • Name of equipment. • Rated voltage. • Type of current. • Rated frequency.
A3 PLUS - A5 - A6 PLUS - L6 SURGICAL SINGLE CART 2.2. DENTAL UNITS Description of the parts. Water unit. Adjustable arm. Dentist’s board. Dentist’s control console. Tray holder (optional). Assistant's board. Assistant’s board in IDRICO SINGOLO version. Assistant’s board control console.
Page 14
A3 PLUS - A5 - A6 PLUS - L6 SURGICAL SINGLE CART Model A5: Model A5 CONTINENTAL (SN. 71IAxxxx) Model A5 INTERNATIONAL (SN. 71IBxxxx) Model A5 CART (SN. 71ICxxxx) Model A5 SINGLE CART (SN. 71IExxxx)
Page 15
A3 PLUS - A5 - A6 PLUS - L6 SURGICAL SINGLE CART Model A5 IDRICO SINGOLO (SN. 71IDxxxx) Model A6 PLUS: Model A6 PLUS CONTINENTAL (SN. 71HWxxxx) Model A6 PLUS INTERNATIONAL (SN. 71HXxxxx) Model L6: Model L6 CONTINENTAL (SN. 71H1xxxx)
Page 16
A3 PLUS - A5 - A6 PLUS - L6 SURGICAL SINGLE CART Model L6 CONTINENTAL, HYBRID version (SN. 71H4xxxx) Model L6 INTERNATIONAL (SN. 71H2xxxx) Model L6 SIDE DELIVERY (SN. 71H3xxxx) Model SURGICAL SINGLE CART (SN. 71HSxxxx)
A3 PLUS - A5 - A6 PLUS - L6 SURGICAL SINGLE CART 2.4. SPECIAL WARNINGS CART models. TIPPING HAZARD. To move the cart, grab the relevant handle (a). When a device has to be moved from one room to another, place the board in the lowest position.
A3 PLUS - A5 - A6 PLUS - L6 SURGICAL SINGLE CART STARTING Press the main switch (f1) Switch (f1) on: • equipment ON; • electric system powered; • water and air system connected; • DISPLAY (d) on. Switch (f1) off: •...
A3 PLUS - A5 - A6 PLUS - L6 SURGICAL SINGLE CART 3.1. CONVERSION OF THE DENTAL UNIT CONFIGURATION FOR LEFT-HANDED OPERATORS (HYBRID models only) To reverse dental unit configuration, proceed as follows: • Lower the dental chair and the backrest until the headrest is lower than the water unit.
A3 PLUS - A5 - A6 PLUS - L6 SURGICAL SINGLE CART DENTAL CHAIR OPERATION Dental chair movements. The dental chair can be operated from the following units: • Dentist's board (a) (see par. 5.). • Multifunction foot control (b) (see par. 5.2.).
Page 22
A3 PLUS - A5 - A6 PLUS - L6 SURGICAL SINGLE CART Dental units model: L6 SIDE DELIVERY Dental units model: L6 CONTINENTAL HYBRID version Dental chair movements: • with the instrument extracted NOT working: manual movements allowed, automatic movements inhibited, but if they are already in progress at the moment of extraction they are not interrupted;...
A3 PLUS - A5 - A6 PLUS - L6 SURGICAL SINGLE CART 4.2. MOVEMENT LOCK DEVICES Use the devices below when movement of the equipment needs to be blocked: • Dental chair control buttons (a) or (c). By pressing any of the dental chair control buttons all movements will be stopped.
A3 PLUS - A5 - A6 PLUS - L6 SURGICAL SINGLE CART 4.4. MOVABLE ARMRESTS (OPTIONAL) Maximum load supported by armrest: 68 kg. Turn movable armrest clockwise to move it downwards, so that the patient can easily get on and off the chair.
A3 PLUS - A5 - A6 PLUS - L6 SURGICAL SINGLE CART DENTIST'S BOARD OPERATION Layout of instruments. The positions of the instruments on the board are determined by the customer at the time of order. Enabling the instruments. • Syringe is always on (see paragraph 5.3.).
Page 26
A3 PLUS - A5 - A6 PLUS - L6 SURGICAL SINGLE CART Instrument recall arm lock device (only version with upper recovery of hoses). The arm can be locked in the instrument extracted position, by taking it at 2/3 from the limit stop (1).
Page 27
A3 PLUS - A5 - A6 PLUS - L6 SURGICAL SINGLE CART Cleaning the dentist's board. Clean the dentist's board using a suitable product (see paragraph 1.4). Removable instrument holder; just slide it out of its seat to remove it.
A3 PLUS - A5 - A6 PLUS - L6 SURGICAL SINGLE CART 5.1. DENTIST’S CONTROL CONSOLE The console consists of a control panel and a DISPLAY that, depending on the model, can be of LCD Touch (A2) or Digit (A1) type.
Page 29
A3 PLUS - A5 - A6 PLUS - L6 SURGICAL SINGLE CART Description of CONSOLE buttons: Button lock Button pressed shortly: recall stored Decreases settable values. position “A”. (pressed and held for at least 5 sec. for Clean function). Button pressed and held: seat up.
Page 30
A3 PLUS - A5 - A6 PLUS - L6 SURGICAL SINGLE CART Warning icons. DISPLAYS show a series of icons giving information on the dental unit operation. Description of the warning icons: Active operator. Dental chair position set manually. Dental chair stored position “A”, “B”, “C” or “D”.
A3 PLUS - A5 - A6 PLUS - L6 SURGICAL SINGLE CART 5.1.1. MAIN SETTINGS 5.1.1.1. HYGIENE CYCLE SETTING • Press button (H1) on the assistant's board to open the HYGIENE CYCLE SETTING menu: Hygiene cycle selection. Button pressed shortly: FLUSHING cycles.
A3 PLUS - A5 - A6 PLUS - L6 SURGICAL SINGLE CART 5.1.1.1.2. LONG FLUSHING CYCLE SETTING (LCD Touch console only) We recommend performing this operation at the beginning of each working day. Performing the cycle. • Press button (H1) on the assistant's board to select the FLUSHING cycle.
A3 PLUS - A5 - A6 PLUS - L6 SURGICAL SINGLE CART 5.1.1.2. STOPWATCH (LCD Touch console only) • Press the button (C) to access the STOPWATCH function: Stopwatch setting The countdown can be set from a minimum of 10 seconds to a maximum of 10 minutes.
A3 PLUS - A5 - A6 PLUS - L6 SURGICAL SINGLE CART 5.1.4. PROGRAMMING THE DENTAL CHAIR POSITIONS A, B, C and D This setting is specific for each single operator. Proceed as follows: • Adjust the dental chair into the desired position with the manual movement buttons (L, M, N, O).
A3 PLUS - A5 - A6 PLUS - L6 SURGICAL SINGLE CART 5.1.7. CONSOLE CONTROL PANEL LOCKING BUTTON Press this button to enable/disable the control panel and the LCD Touch DISPLAY (if any) in order to easily carry out console cleaning operations.
Page 36
A3 PLUS - A5 - A6 PLUS - L6 SURGICAL SINGLE CART Joystick for dental chair movements (3). It controls the following movements: Dental chair seat up. Dental chair backrest up. Dental chair seat down. Dental chair backrest down. To stop movement, release the control joystick.
A3 PLUS - A5 - A6 PLUS - L6 SURGICAL SINGLE CART "PUSH-PEDAL” FOOT CONTROL 5.2.2. Description of the parts. Handle. Control lever. Dental chair movement control. Chip-air/Patient rinse position recall control. Water Clean System/Automatic dental chair return control. Spray operation LED.
Page 38
A3 PLUS - A5 - A6 PLUS - L6 SURGICAL SINGLE CART Joystick for dental chair movements (3). It controls the following movements: Dental chair seat up. Dental chair backrest up. Dental chair seat down. Dental chair backrest down. To stop movement, release the control.
A3 PLUS - A5 - A6 PLUS - L6 SURGICAL SINGLE CART 5.2.3. "POWER PEDAL" FOOT CONTROL Description of the parts. Handle. Foot control. Dental chair movement control. Chip-air control or activation/deactivation of instrument spray function. Water Clean System control activation/deactivation instrument spray function.
Page 40
A3 PLUS - A5 - A6 PLUS - L6 SURGICAL SINGLE CART Left-hand button (4). With instrument removed • Short press: enables/disables the spray function. • Long press (at least 2 seconds): sends a jet of air to the instruments (Chip air);...
A3 PLUS - A5 - A6 PLUS - L6 SURGICAL SINGLE CART 5.2.4. WIRELESS FOOT CONTROL Warnings for use. • Avoid keeping the WIRELESS foot control in proximity of other RF sources, such as wireless LAN cards, other radio devices, home RF devices, microwave ovens.
Page 42
A3 PLUS - A5 - A6 PLUS - L6 SURGICAL SINGLE CART Recharging the battery. When the batteries in the WIRELESS foot control need to be recharged. Proceed as follows: • Open the door (1) on the back of the pedal and connect the charging cable (2).
A3 PLUS - A5 - A6 PLUS - L6 SURGICAL SINGLE CART 5.3. SYRINGE Description of the instrument. Spout. Handle. Syringe release button. Air button. Water button. Hot/cold selector (only for 6-function syringes). Hot/cold indicator LED (only for 6-function syringes).
A3 PLUS - A5 - A6 PLUS - L6 SURGICAL SINGLE CART 5.4. TURBINE Connecting the handpiece and changing the drill. Refer to the specific instructions provided with the handpiece. Use. The instrument is supplied NON-sterile. Carefully read the operating instructions supplied with the instrument before attempting to sterilize.
A3 PLUS - A5 - A6 PLUS - L6 SURGICAL SINGLE CART When the peristaltic pump is enabled, the other sources of liquid supply are disabled 5.4.1. TURBINE (SINGLE WATER UNIT models) Use. The instrument is supplied NON-sterile. Carefully read the operating instructions supplied with the instrument before attempting to sterilize.
A3 PLUS - A5 - A6 PLUS - L6 SURGICAL SINGLE CART 5.5. ELECTRIC MICROMOTOR Coupling the handpieces and changing the drill. Refer to the specific instructions provided with the micromotor and various handpieces. Use. The instrument is supplied NON-sterile.
Page 47
A3 PLUS - A5 - A6 PLUS - L6 SURGICAL SINGLE CART Selecting the micromotor working programmes (LCD TOUCH console only). It is possible to set 4 micromotor working programmes. Each working programme saves the following data: • Operating mode.
A3 PLUS - A5 - A6 PLUS - L6 SURGICAL SINGLE CART 5.5.1. RESTORATIVE OPERATING MODE Characteristics. • Speed range: 100 to 40,000 Rpm (1:1 handpiece). • Reduction ratio selection. • Display of the instantaneous maximum speed. • Direct selection of the set working programmes.
Page 49
A3 PLUS - A5 - A6 PLUS - L6 SURGICAL SINGLE CART Drill rotation at maximum torque. Select the drill rotation mode when the set maximum torque is reached: Rotation lock. Stop normal rotation and reverse rotation. Stop normal rotation, short reverse rotation and resumption of normal rotation.
A3 PLUS - A5 - A6 PLUS - L6 SURGICAL SINGLE CART 5.5.3. ELECTRIC MICROMOTOR (SINGLE WATER UNIT models) Use. The instrument is supplied NON-sterile. Carefully read the operating instructions supplied with the instrument before attempting to sterilize. Operating times: work 5 min., rest 5 min.
A3 PLUS - A5 - A6 PLUS - L6 SURGICAL SINGLE CART 5.6. SCALER Connecting the handpiece and insert. Refer to the specific instructions provided with the handpiece. • Before attempting to connect the handpiece, make sure the contacts are perfectly dry. Blow air from the syringe, if necessary, to dry.
Page 52
A3 PLUS - A5 - A6 PLUS - L6 SURGICAL SINGLE CART Fibre optic brightness adjustment. • Place the instrument in its work position. • To adjust fibre optic brightness, press the button (F) (for at least 2 seconds). • Adjust beam brightness by pressing “increase” or “decrease” buttons.
A3 PLUS - A5 - A6 PLUS - L6 SURGICAL SINGLE CART Warnings for use. • The instrument is supplied NOT STERILE and must be sterilized before use (see paragraph 1.6.). Carefully read the operating instructions supplied with the instrument before attempting to sterilize.
A3 PLUS - A5 - A6 PLUS - L6 SURGICAL SINGLE CART 5.7. T LED CURING LIGHT Intended use. The T-LED curing light is a medical device for polymerization of light-cured dental materials (dental pit and fissure sealants, dental restorative materials, dental cements and bonding agents) used in various sectors of the dental practice (preventive dentistry, orthodontics, prosthetics).
Page 55
A3 PLUS - A5 - A6 PLUS - L6 SURGICAL SINGLE CART General description of the lamp. Lamp handle. Rotary end. Optical conductor (10 mm). Anti-reflective cone. Power cord. Control panel. • The curing light can be used in different configurations (wand, gun or any intermediate position) as it is more convenient for the user.
Page 56
A3 PLUS - A5 - A6 PLUS - L6 SURGICAL SINGLE CART Operation. • The instrument is supplied non-sterile. • Before use, sterilise the fibre optics, the anti-reflective cone and disinfect the light handpiece. • Put the fiber optic (c) in its housing until it clicks.
Page 57
A3 PLUS - A5 - A6 PLUS - L6 SURGICAL SINGLE CART Maximum curing thickness. The maximum curing thickness with every single cycle is 3 millimetres (refer to the instructions of the composite material used as well). This thickness must not be exceeded as the layer may not be completely cured.
A3 PLUS - A5 - A6 PLUS - L6 SURGICAL SINGLE CART 5.8. C-U2 DENTAL CAMERA The C-U2 dental camera system, complete with an extremely lightweight ergonomic handpiece, is specially designed for simple and well-conceived usability in examining the oral cavity. Auto-exposure and fixed focus features provide easy operation. It is designed to allow the dentist to more efficiently show and explain to patients all oral conditions and reasons for planned treatment.
Page 59
A3 PLUS - A5 - A6 PLUS - L6 SURGICAL SINGLE CART Connecting the handpiece. • Attach the handpiece of camera C-U2 (a) to the end of the hose and tighten the ring nut (b). Make sure the hose is firmly screwed onto the handpiece.
Page 60
A3 PLUS - A5 - A6 PLUS - L6 SURGICAL SINGLE CART Setting operation in single image or multiple image mode (LCD TOUCH console) • With the camera set to LIVE mode, briefly press the button (A) to change from viewing a single image to multiple images (and vice versa).
Page 61
A3 PLUS - A5 - A6 PLUS - L6 SURGICAL SINGLE CART VIEW function. • With the camera in FREEZE mode, by pressing the touch button (A) repeatedly, you can scroll down the pages with frozen images. Handpiece status An optical guide, illuminated by a multicolour LED indicator, found in the area near the control button (g), shows handpiece status as per the table given...
A3 PLUS - A5 - A6 PLUS - L6 SURGICAL SINGLE CART 5.8.1. C-U2 INTRAORAL CAMERA (DIGIT console) Using the camera (DIGIT console). • Place the instrument in its work position. Now camera is activated and is in LIVE mode (monitor shows "moving"...
Page 63
A3 PLUS - A5 - A6 PLUS - L6 SURGICAL SINGLE CART Operation in single image mode. • The "live" image appears on the monitor and a special icon (1) is shown in the top right-hand corner when the camera is set to LIVE and single-image mode.
A3 PLUS - A5 - A6 PLUS - L6 SURGICAL SINGLE CART 5.8.2. C-U2 INTRAORAL CAMERA (SINGLE WATER UNIT models) Using the camera (DIGIT console). • Place the instrument in its work position. Now camera is activated and is in LIVE mode (monitor shows "moving"...
Page 65
A3 PLUS - A5 - A6 PLUS - L6 SURGICAL SINGLE CART Multiple image mode operation. • The "live" image appears on the monitor when the camera is set to LIVE and multiple image mode. • Pressing the touch button (g) on the handpiece (or activating the foot control), you can freeze the image, which is immediately shown on the monitor.
A3 PLUS - A5 - A6 PLUS - L6 SURGICAL SINGLE CART 5.9. ELECTRONIC APEX LOCATOR The instrument is used to locate the apex of the root canal during endodontic treatments. The apex localisation is useful to obtain the work length together with the visible reference rubber stopper, which is manually positioned on the probe inserted in the canal.
Page 67
A3 PLUS - A5 - A6 PLUS - L6 SURGICAL SINGLE CART Root canal length detection. • Do not use the ELECTRONIC APEX LOCATOR in case of teeth with canals obstructed or too large, presence of blood or saliva, fracture, broken crown or metal crown.
A3 PLUS - A5 - A6 PLUS - L6 SURGICAL SINGLE CART 5.10. ZEN-Xi INTEGRATED SENSOR The integrated ZEN-Xi sensor is a medical device used for the acquisition of intraoral X-rays in electronic format through the interface with a Personal Computer.
A3 PLUS - A5 - A6 PLUS - L6 SURGICAL SINGLE CART ASSISTANT’S BOARD OPERATION Description of the parts. Assistant’s board: it can hold 2 suction tubes and 3 instruments, of which 1 is dynamic. Control panel used to operate the dental chair and water unit.
Page 70
A3 PLUS - A5 - A6 PLUS - L6 SURGICAL SINGLE CART Dental units model: A5 IDRICO SINGOLO Dental units model: A6 PLUS CONTINENTAL, A6 PLUS INTERNATIONAL Dental units model: L6 CONTINENTAL L6 INTERNATIONAL L6 SIDE DELIVERY Dental units model: L6 CONTINENTAL HYBRID version Cleaning the sliding rollers.
A3 PLUS - A5 - A6 PLUS - L6 SURGICAL SINGLE CART 6.1. ASSISTANT’S BOARD CONSOLE Standard version. IDRICO SINGOLO model version. Description of the buttons: Dental unit POWER SAVING button (with relevant warning LED) Operating light on/off button Water-to-cup button...
A3 PLUS - A5 - A6 PLUS - L6 SURGICAL SINGLE CART 6.2. INSTRUMENTS ON ASSISTANT’S BOARD All the instruments applied to the assistant's board maintain the same operation as those on the dentist's board. In particular: • Syringe, see paragraph 5.3.
A3 PLUS - A5 - A6 PLUS - L6 SURGICAL SINGLE CART 6.3. SUCTION TUBES Suction starts when lifting the suction tube holder off its support. To adjust suction power, use the lever (a) located on the suction tube holder handle.
A3 PLUS - A5 - A6 PLUS - L6 SURGICAL SINGLE CART 6.4. TRAY HOLDER The tray holder (a) is made of stainless steel and can easily be removed from its support. Tray holder support can rotate both clockwise and counter-clockwise, so as to be positioned in the most convenient position for the operator.
A3 PLUS - A5 - A6 PLUS - L6 SURGICAL SINGLE CART WATER UNIT OPERATION 7.1. FILL CUP AND BOWL The bowl can be turned 305° on the water unit by hand or using a motor drive (optional). The bowl and the water-to-cup spout can be removed to make cleaning operations easier.
Page 76
A3 PLUS - A5 - A6 PLUS - L6 SURGICAL SINGLE CART • Press and hold button (A) to exit the settings and confirm the changes made. The LED (L1) will turn off after exiting the settings. Bowl flushing time setting.
Page 77
A3 PLUS - A5 - A6 PLUS - L6 SURGICAL SINGLE CART Changing the water to bowl controller with water delivered to cup. The controller that delivers water to the bowl when the button (B1) is pressed can be turned on/off.
Page 78
A3 PLUS - A5 - A6 PLUS - L6 SURGICAL SINGLE CART Spout, bowl and bowl filter removal. • Pull up the spout (l) and take it off. • Pull up the filter (q) and its cover (p) from the bowl to remove them.
A3 PLUS - A5 - A6 PLUS - L6 SURGICAL SINGLE CART 7.2. S.H.S. SYSTEM (SIMPLIFIED HYGIENIZATION SYSTEM) Description of the system. This system features a tank (a) suitable to contain the independent water supply, preferably distilled water. The tank has a total capacity of 1.8 litres and feeds: •...
Fit tank back into its seat inside dental unit. 7.2.1. MANUAL S.H.S. SYSTEM Fitted on models A3 PLUS, A5 SINGLE CART, SURGICAL SINGLE CART. System features a tank (a) for distilled water. The tank has a total capacity of 1.8 litres.
A3 PLUS - A5 - A6 PLUS - L6 SURGICAL SINGLE CART 7.2.2. MANUAL DISINFECTION CYCLE WITH S.H.S. SYSTEM The S.H.S. system allows performing a manual disinfection cycle of the water ducts of all the instruments on the dentist's board and of the syringe located on the assistant's board.
A3 PLUS - A5 - A6 PLUS - L6 SURGICAL SINGLE CART 7.3. W.H.E. SYSTEM (WATER HYGIENIZATION EQUIPMENT) The W.H.E. system ensures a safe, physical, separation of the dental unit water system from the water mains thanks to a water free fall section (in compliance with EN 1717).
Page 83
A3 PLUS - A5 - A6 PLUS - L6 SURGICAL SINGLE CART Dental units of A5 and A6 PLUS model: • Open the water side unit cover (see paragraph 7.7.). • Turn the tank (a). • Remove plug (k) and pour disinfectant fluid inside tank until it is full.
A3 PLUS - A5 - A6 PLUS - L6 SURGICAL SINGLE CART 7.4. BIOSTER DISINFECTION AUTOMATIC SYSTEM Description of the system. This system allows carrying out an automatic disinfection cycle of the water circuits of the following instruments: • All instruments that require water for their use are placed on the dentist's board.
Page 85
LCD TOUCH console only: disinfectant fluid residence time in the water ducts can be set (see paragraph 5.1.1.1.3.). • Dental units of A3 PLUS, A5 and A6 PLUS model: position the special container (d) on the bowl and insert the instrument hoses to be disinfected.
• Use the DISPLAY to access the "FLUSHING cycle setting" menu, and set cycle time (see paragraph 5.1.1.1.). • Dental units of A3 PLUS, A5 and A6 PLUS model: position the special container (d) on the bowl. Dental units of L6 model: remove the cover (b).
Page 87
A3 PLUS - A5 - A6 PLUS - L6 SURGICAL SINGLE CART Warning messages on DISPLAY. If the system detects a malfunction, a warning message will be DISPLAYED (see paragraph 10).
The cycles are automatic and are recommended after each operation (STANDARD cycle) and at the end of the day (INTENSIVE cycle) to complete the cleaning and disinfection operations of the dental unit. The INTENSIVE cycle is not available if the S.H.S. system is active. Dental units of A3 PLUS, A5 and A6 PLUS models: Tank for detergent fluid. Couplings for suction tubes.
A3 PLUS - A5 - A6 PLUS - L6 SURGICAL SINGLE CART Dental units of L6 model: • Open the water side unit cover (see paragraph 7.7.). • Rotate the tank (c). • Remove plug (k) and pour disinfectant fluid inside tank until it is full.
A3 PLUS - A5 - A6 PLUS - L6 SURGICAL SINGLE CART 7.8. WATER UNIT LED (optional) Fitted on models L6 only. One or two LED rings (l) can be mounted on the water unit to signal the operating status of the dental unit.
A3 PLUS - A5 - A6 PLUS - L6 SURGICAL SINGLE CART ACCESSORIES 8.1. OPERATING LAMP The operating light comes in 4 models: Lamp with halogen light source, model VENUS PLUS. Lamp with LED light source, VENUS PLUS model, L version.
A3 PLUS - A5 - A6 PLUS - L6 SURGICAL SINGLE CART 8.4. AIR/WATER/230V QUICK-COUPLINGS Power outlet: 220-240 Vac 50/60 Hz 10 A or 100-120 Vac 50/60 Hz 15 A according to IEC/EN 60320-2-2/F. Air quick coupling: pressure 6 Bar.
A3 PLUS - A5 - A6 PLUS - L6 SURGICAL SINGLE CART MAINTENANCE Preventive maintenance CEFLA s.c., the manufacturer of the medical devices, in accordance with applicable standards IEC 60601-1 - 2007, IEC 62353, MDR 2017/745, as amended and supplemented, for medical devices, underlines that the following scheduled maintenance checks specified in the Technical Service Manual are to be carried out by authorised personnel at least once every 12 months: •...
A3 PLUS - A5 - A6 PLUS - L6 SURGICAL SINGLE CART 9.2. DRAINING CONDENSATE Perform this operation every day before starting work. Proceed as follows: • Remove the condensate drain cock (a). • Place a container under condensate drain cock (a).
A3 PLUS - A5 - A6 PLUS - L6 SURGICAL SINGLE CART 9.3. SUCTION FILTER CLEANING This operation should be done daily at the end of work. BIOLOGICAL HAZARD Always wear goggles and gloves to prevent contact with infected material when cleaning the suction filters.
A3 PLUS - A5 - A6 PLUS - L6 SURGICAL SINGLE CART 9.4. SUCTION LINES The suction lines must be disinfected using a product suitable for this purpose. For cleaning of the suction system, it is recommended to use STER 3 PLUS (CEFLA s.c.) diluted in a 6% solution (equivalent to 60 ml of product in 1 litre of hot water).
Always wear gloves to prevent contact with infected material when carrying out the following operation. Dental units of A3 PLUS, A5 and A6 PLUS models: • Move the dental chair fully up. • Open the water side unit cover (see paragraph 7.7.).
A3 PLUS - A5 - A6 PLUS - L6 SURGICAL SINGLE CART 9.6. CLEANING THE TURBINE RETURN AIR FILTER Monthly check the oil container filter (g) present in the turbine’s return air line. If necessary, replace the filter element (code 97290014).
In any case, operation should be checked by the manufacturer's authorised personnel once a year. 9.10. CATTANI GRAVITY AMALGAM SEPARATOR Applied only on models A3 PLUS, A5 and A6 PLUS. Draining the separator’s container. • Move the dental chair completely up so that as much liquid as possible can be drained from the container.
A3 PLUS - A5 - A6 PLUS - L6 SURGICAL SINGLE CART 10. WARNING/ERROR MESSAGES Two types of notices can be displayed on the console display: warning messages (Wxxx) and error messages (Exxx). Warning messages (Wxxx) indicate situations that require user intervention.
Page 101
A3 PLUS - A5 - A6 PLUS - L6 SURGICAL SINGLE CART Warning Cause Solution A movement has been interrupted or a movement request cannot Remove the obstacle; if the problem persists, contact the W400 be implemented since the safety device of the foot board under technical service department.
A3 PLUS - A5 - A6 PLUS - L6 SURGICAL SINGLE CART 11. TECHNICAL DATA A3 PLUS 97042227 A6 PLUS Installation plan: 97042225 SURGICAL 97042046 SINGLE CART Technical manual: 97071317 Spare parts catalogue: 97023177 A3 PLUS 110 kg A6 PLUS...
Page 103
A3 PLUS - A5 - A6 PLUS - L6 SURGICAL SINGLE CART SURGICAL 1580x780x1130(h) SINGLE CART A3 PLUS 1520x730x1030(h) Dental chair packaging A6 PLUS dimensions: 1520x730x1030(h) A3 PLUS 140 kg A5 CART 140 + 45 kg A5 SINGLE 45 kg...
Page 104
A3 PLUS - A5 - A6 PLUS - L6 SURGICAL SINGLE CART IEC 61000-4-2 relevant humidity should be at least 30%. 27 V/m at 385 MHz Portable and mobile RF communications equipment Proximity fields from RF 28 V/m at 450, 810, 870, 930,...
Page 105
A3 PLUS - A5 - A6 PLUS - L6 SURGICAL SINGLE CART For transmitters having a maximum nominal output power not listed above, the recommended distance d in metres (m) can be determined using the corresponding equation applicable to the transmitter frequency where P is the maximum output power of the transmitter in Watt (W) according to the transmitter Manufacturer.
A3 PLUS - A5 - A6 PLUS - L6 SURGICAL SINGLE CART 12. DENTAL UNIT MAINTENANCE SCHEDULE REFERENCE WHEN PART ACTION PARAGRAPH Condensate drain cock. Eliminate condensate from air ducts. See paragraph 9.2. 2-minute flushing of the ducts of the instruments used Spray ducts See paragraph 7.5...
Page 121
A3 PLUS - A5 - A6 PLUS - L6 SURGICAL SINGLE CART REFERENCE WHEN PART ACTION PARAGRAPH Clean with suitable disinfectant in compliance with manufacturer's recommendations. Removable instrument hoses. See paragraph 5. Spray product on disposable soft paper. Do not use acids or abrasive products.