Summary of Contents for Steris AMSCO 50 Series

  • Page 1 AMSCO 30 Series and ® AMSCO 50 Series ® Reprocessing Sink Operator Manual INSTALLATION AND OPERATOR MANUAL 11002980 Revision C...
  • Page 2 11002980_C Reprocessing Sink Operator Manual...
  • Page 3: Word From Steris

    A summary of the safety precautions to be observed when operating and servicing this equipment follow this section. Do not operate or service equipment until becoming familiar with this information. Any alteration of this equipment not authorized or performed by STERIS Service Engineering will void the warranty and could adversely affect equipment operation.
  • Page 4 11002980_C Reprocessing Sink Operator Manual...
  • Page 5: Table Of Contents

    TABLE OF CONTENTS Section Number Description Page 1 A Word From STERIS............................3 2 Safety Precautions ............................i 2.1 Safety Precautions ..........................i 3 Installation ............................... 3 3.1 Installation Instructions ........................3 3.2 Pre-installation ........................... 4 3.3 Installation Procedure ........................5 4 Operating Instructions ..........................
  • Page 6 11002980_C Reprocessing Sink Operator Manual...
  • Page 7 LIST OF FIGURES Description Page Figure 3-1 Reprocessing Sink..........................3 Figure 3-2 Install Mounting Shanks........................5 Figure 3-3 Support Bracket ........................... 5 Figure 3-4 Mount Support Bracket ........................6 Figure 3-5 Mount Pre-rinse Faucet ........................7 Figure 3-6 Install Braided Lines ..........................8 Figure 3-7 Install Mounting Shank on Pure Water Faucet..................
  • Page 8 viii 11002980_C Reprocessing Sink Operator Manual...
  • Page 9 LIST OF TABLES Description Page Table 6-1 Check for faulty component(s) ......................25 Table 7-1 Replacement Parts and Consumables....................27 Operator Manual Reprocessing Sink 11002980_C...
  • Page 10 11002980_C Reprocessing Sink Operator Manual...
  • Page 11: Safety Precautions

    AMSCO Reprocessing Sink are insured only if STERIS original parts and accessories (or parts and accessories approved by STERIS) are used for service, repairs, and modifications to this equipment. The replacement of spare parts must only be performed by STERIS or STERIS- trained technicians.
  • Page 12 If gap is too large, the tail piece will bottom out as sink is lowered, causing damage. If gap is too small, the tail piece will leak when sink is fully raised. Unplug unit when cleaning lifting columns or cleaning near controller. Causing unit to overheat too many times greatly reduces life of unit and could damage unit, making it inoperable, requiring replacement.
  • Page 13: Installation

    Section 3 — Installation Installation 3.1 Installation Instructions F F i i g g u u r r e e 3 3 - - 1 1 . . R R e e p p r r o o c c e e s s s s i i n n g g S S i i n n k k •...
  • Page 14: Pre-Installation

    Section 3 — Installation – Steel-Toe Shoes • Tools Required: – Crescent Wrench – Pipe Wrench – Rubber Mallet – Level – Phillips Screwdriver – Tape Measure – Saw (to cut ABS tubing) – ABS Primer/Glue – Thread Sealing Tape 3.2 Pre-installation The AMSCO Reprocessing Sink should be moved into area intended for installation and kept 3’-4’...
  • Page 15: Installation Procedure

    Section 3 — Installation 3.3 Installation Procedure 1. Install Mounting Shanks (on Pre-rinse Faucet) Insert Brass ½” NPT mounting shank (11001017) into pre-rinse faucet inlets using sealing tape and tighten. • Prevent damage to threads. Avoid cross-threading during assembly. • Insert correct end into the faucet as shown here. F F i i g g u u r r e e 3 3 - - 2 2 .
  • Page 16 Section 3 — Installation 3. Mount Support Bracket Mount support bracket to backsplash using two ¼-20 x ¾” machine screws, two ¼” washer, and two ¼-20 nuts. F F i i g g u u r r e e 3 3 - - 4 4 . . M M o o u u n n t t S S u u p p p p o o r r t t B B r r a a c c k k e e t t 11002980_C Reprocessing Sink Operator Manual...
  • Page 17 Section 3 — Installation 4. Mount Pre-rinse Faucet Mount pre-rinse faucet to backsplash with mounting shanks through holes shown. • Secure pre-rinse faucet using washers and locknuts on back of unit as shown. • Reassemble clamp (removed in STEP 2) to support bracket. F F i i g g u u r r e e 3 3 - - 5 5 .
  • Page 18 Section 3 — Installation 5. Install Lines and Test for Leaks Install both braided lines (11001115) using reducing coupling (11000975) as shown below. F F i i g g u u r r e e 3 3 - - 6 6 . . I I n n s s t t a a l l l l B B r r a a i i d d e e d d L L i i n n e e s s •...
  • Page 19 Section 3 — Installation 6. Install Mounting Shank (on Pure Water Faucet) Insert stainless steel, or tin lined copper ½” NPT mounting shank (11001022) into Pure Water Faucet inlet using sealing tape and tighten. • Prevent damage to threads. Avoid cross-threading during assembly. •...
  • Page 20 Section 3 — Installation 7. Mount Pure Water Faucet Where the pure water faucet is mounted determines the workflow of the unit (left to right, or right to left). Verify desired configuration. • Two holes are provided on backsplash for either configuration (see above) •...
  • Page 21 Section 3 — Installation 8. Install Lines to Pure Water Faucet and Test for Leaks Install stainless elbow (11001069) and braided supply line (11001117) as shown using thread sealing tape. Test connection for leaks. Any leaks need to be addressed before moving to next step. F F i i g g u u r r e e 3 3 - - 9 9 .
  • Page 22 Section 3 — Installation 10. Install Lines to Water Pistol and Air Gun and Test for Leaks Install supply hose connection for air gun using brass elbow (11001067) and air supply line (11001119) as shown using thread sealing tape. Test connection for leaks. Install braided stainless-steel supply line (11001115) for water pistol, using brass coupling (11000974) as shown.
  • Page 23 Section 3 — Installation 13. Level Unit Using leveling feet located under unit, level sink. It is ideal to use front/side top edges of unit to level accurately. Use front and side edges when leveling F F i i g g u u r r e e 3 3 - - 1 1 2 2 . . L L e e v v e e l l U U n n i i t t u u s s i i n n g g L L e e v v e e l l i i n n g g F F o o o o t t Operator Manual Reprocessing Sink 11002980_C...
  • Page 24 Section 3 — Installation 14. Assemble and Adjust Tail Piece to Lever Drain Assembly Ensure gasket is present in metal tail piece adapter. Raise adapter with tailpipe up to lever drain and be sure tailpipe fits as straight as possible so as to not introduce any unnecessary stress to the drain assembly.
  • Page 25 Section 3 — Installation 15. Install Drain and Test for Leaks Configure/Install basin drains. • Components are included to build configuration (below) or can be used to customize connection. (Otherparts may be required and should be purchased locally.) F F i i g g u u r r e e 3 3 - - 1 1 4 4 . . T T a a i i l l P P i i e e c c e e T T o o L L e e v v e e r r D D r r a a i i n n A A s s s s e e m m b b l l i i n n g g Operator Manual Reprocessing Sink 11002980_C...
  • Page 26 Section 3 — Installation • Connect basin drains to existing floor or wall drain in accordance with applicable local code standards. • Configure P-trap and install. NOTE: Coupling and a nipple can be moved to opposite end of "U" piece to lower the drain outlet F F i i g g u u r r e e 3 3 - - 1 1 5 5 .
  • Page 27 Section 3 — Installation 17. Install Seismic Anchors If seismic anchoring system will be installed, use (4) brackets provided (11002934) to mark floor for anchors. If the seismic anchoring system will not be installed, proceed to STEP 17. NOTE: Anchors include their own instructions. Remove 4 brackets (2 per foot) to install anchors, then re-install brackets and secure.
  • Page 28 Press Down button again and hold 3-5 seconds and then release. f. This will reset both columns to the controller and ensures they are both at the same height. g. If this fails to reset the system, contact STERIS Service Representative. CAUTION- POSSIBLE EQUIPMENT DAMAGE HAZARD:: Causing unit to overheat too many times greatly reduces life of unit and could damage unit, making it inoperable, requiring replacement.
  • Page 29 Section 3 — Installation 19. Install Gun Tips Mounting Bracket Peel tape from water pistol attachments bracket (11001513) and fasten to gun tips mounting bracket (11000525). If this requires access to the back of the sink, it should be done before moving the sink into position.
  • Page 30 11002980_C Reprocessing Sink Operator Manual...
  • Page 31: Operating Instructions

    Section 4 — Operating Instructions Operating Instructions 4.1 Operating Instructions Height Adjustment (AMSCO 50 Series Reprocessing Sink) To operate the height adjustment feature, use the up/down switch on the unit located on the front/center F F i i g g u u r r e e 4 4 - - 1 1 . . U U p p / / D D o o w w n n S S w w i i t t c c h h f f o o r r H H e e i i g g h h t t A A d d j j u u s s t t m m e e n n t t •...
  • Page 32 11002980_C Reprocessing Sink Operator Manual...
  • Page 33: Routine Maintenance

    Section 5 — Routine Maintenance Routine Maintenance 5.1 Routine Maintenance Daily Maintenance Procedure: Cleaning The following cleaning and disinfection procedure should be performed at least once per day. Usage may warrant higher cleaning and disinfection frequencies: Refer to facility’s policies and practices related to reprocessing sink usage. CAUTION- PERSONAL INJURY AND/OR EQUIPMENT DAMAGE HAZARD:: Unplug unit when cleaning lifting columns or cleaning near controller.
  • Page 34 11002980_C Reprocessing Sink Operator Manual...
  • Page 35: Troubleshooting

    Section 6 — Troubleshooting Troubleshooting 6.1 Troubleshooting WARNING-PERSONAL INJURY AND/OR EQUIPMENT DAMAGE HAZARD:: Repairs and adjustments to this equipment must be made only by fully qualified service personnel. Maintenance performed by inexperienced, unqualified persons or installation of unauthorized parts could cause personal injury or result in costly equipment damage.
  • Page 36 11002980_C Reprocessing Sink Operator Manual...
  • Page 37: Replacement Parts

    3. Include configuration and serial number of the Reprocessing Sink on your order. 4. Send your order directly to STERIS. NOTE: Use only STERIS authorized parts on this equipment. Use of unauthorized parts will void the warranty. Contact STERIS for recommendations on cleaning products or parts that are not listed below.
  • Page 38 Section 7 — Replacement Parts Table 7-1 Replacement Parts and Consumables (Continued) Description Part Number 1C2410 Prolystica® HP Dye Free & Fragrance Free Enzymatic Manual Cleaner, 2.5 Gallon Prolystica® 2X Concentrate Enzymatic Presoak and Cleaner, 2.5 Gallon 1C3310 Prolystica® 2X Concentrate Enzymatic Presoak and Cleaner, 4x1 Gallon 1C3308 11002980_C Reprocessing Sink...