AMSCO® 600 ALUS
Automatic Load and Unload System (ALUS)
QUICK REFERENCE GUIDE
Attaching a fixed or variable height transfer carriage to the ALUS Table,
Sterilizer, or Pass-Through Loading Cart Return Chamber
To avoid damage to the ALUS table end bracket and front panels,
ensure locking lever is turned to the left, green label side.
1
Some labels may be black instead of green and red
If using a Variable Height Transfer Carriage, adjust height of Transfer Carriage
2
using preset memory buttons.
Approach sterilizer straight on, NOT at an angle, aligning notch on Loading Car
3
with arrow above sterilizer door.
Turn locking lever to the right, (red label side), and pull back on
Transfer Carriage handles to ensure it is properly engaged and
4
locked to Sterilizer, ALUS table or Pass-Through.
Before unloading the Loading Car, visually verify Transfer Carriage
is properly attached. Do not attach Transfer Carriage to sterilizer
front panel or pass-through front panel gap.
5
Engage locks on Transfer Carriage rear casters and unload Loading Car.
6
Turn locking lever back to the left, green label side to undock Transfer Carriage.
Resetting the ALUS table After "Failure to Unload" Alarm
1
Turn manual operation key to the right, wait 15 seconds.
2
Turn manual operation key to the left, wait 15 seconds.