Operating Instructions
2
ASSEMBLY AND INSTALLATION
2.1
Measurement and Control System
A typical measurement system consists of:
•
a Alpha COND 1000 process controller
•
a suitable Conductivity electrode with the appropriate Cell constant and
integrated temperature sensor Pt 1000 or Pt 100,
•
an immersion, flow or process assembly
•
a final control element such as pump or valve and
•
a chart recorder.
Flow Assembly
to Dosing Pumps
Pt100/Pt1000
Temperature Sensor
Process Assembly
with Electrode
2.2
Unit Dimensions
The field-tested control panel housing is 96 x 96 mm; with protection class IP 54
(front).
92 + 0.5
92 + 0.5
Mounting Cut-Out
Alpha COND 1000
Alpha-CON1000 Controller
Note: The Taped Corners Have to Be On Top
92
56
32
2
Thermo Scientific Alpha COND 1000
Chart Recorder
Alpha-CON1000
Power Mains
(220/110 VAC)
Flat Gasket (1mm)
(To be Inserted By Customer)
max. 175
Alarm / Siren
System
96
max. 45