Summary of Contents for Ohkura VM7000A

  • Page 1 VM7000A PAPERLESS RECORDER OPERATION MANUAL WXPVM70mnA0001E March., 2016(Rev.12) Copyright © 2009-2016 Ohkura Electric Co., Ltd. All Rights Reserved.
  • Page 2: To Use This Equipment Safely

    To use this equipment safely Thank you for purchasing our VM7000A Paperless Recorder. Please installs it, operates it and prepares it, after this manual is often read, and it understands enough. ・ There is danger where the accident and the trouble occur when handling is mistaken.
  • Page 3 The symbols below are used on this instrument for the cautioning information. Symbols used on the instrument This shows “Caution for handling”. This symbol is used on the parts need to reference the instruction manual for saving human body and the instrument. This shows “Protective grounding”.
  • Page 4 Do not put your foot on the installed instrument or get on it, because it is ● dangerous. Only our serviceman or persons authorized by OHKURA are allowed to remove ● Maintenance and take the inner module, the main unit and printed circuit boards apart.
  • Page 5: Table Of Contents

    <CONTENTS> 1. INTRODUCTION .................... 1-1 Paperless Recorder ......................... 1-1 Accessory check ..........................1-1 When temporarily keeping it ......................1-1 Confirmation of model number and specification ................1-2 Handling SD card ..........................1-3 2. NAMES AND FUNCTIONS OF PARTS ............2-1 Names and functions of parts ......................
  • Page 6 7.11 Setting the REC/CALC ........................7-18 7.12 Copy the channel setting ........................7-21 7.13 Setting the F value ........................... 7-22 7.14 Setting the timer of channel ......................7-23 7.15 Setting the group name ........................7-24 7.16 Setting the displayed channel ......................7-25 7.17 Select the ON/OFF of graph type .....................
  • Page 7 8.24 “SD / Param” list ..........................8-29 8.25 “Comm.” list ............................8-29 8.26 “Device / Other” list .......................... 8-31 8.27 "Security" list ............................ 8-33 8.28 Engineering ............................8-33 9. MAINTENANCE ..................... 9-1 Inspection ............................9-1 Recommended replacement cycle of parts ..................9-1 Troubleshooting ..........................
  • Page 8: 1. Introduction

    1. INTRODUCTION 1.1 Paperless Recorder This Recorder displays measured data in real time on the liquid crystal display. It is a paperless ① type that is also capable of saving the measured data to a SD memory card (hereinafter referred to as SD card).
  • Page 9: Confirmation Of Model Number And Specification

    1.4 Confirmation of model number and specification The marking plate to which the model number has been described is on the case. Please confirm this instrument is a specification the same as the order referring to the table below. 9 10 V...
  • Page 10: Handling Sd Card

    1.5 Handling SD card Correspondence SD card is as follows. ・Panasonic corporation 1 32GB ~ ・SanDisk corporation 32GB ~ There is no SD card in this equipment. Please buy it in the computer shop etc. CAUTION SD card on the market is sold having formatted it usually, therefore, it is not necessary to ・...
  • Page 11: 2. Names And Functions Of Parts

    2. NAMES AND FUNCTIONS OF PARTS Names and functions of parts ①Display Unit ②Button operation part cover Knob Display area ① The LCD is provided with touch panel. Display the measurement data and other various Parameter set screens. Touch the surface to set data. Button operation part cover ②...
  • Page 12 ⑤MENU button ③Status display lamp ⑥FUNC button ④REC button ⑦SD card slot Status display lamp ③ Allow the power ON/OFF, LCD (display) ON/OFF and record status to be displayed. Lamp ON (highlighted) :Power ON, recording suspended Lamp blinks(1 sec ON/1 sec OFF) :Power ON, recording in progress Lamp blinks(high speed) :Power ON, SD card writing Lamp OFF...
  • Page 13: Set Of O-Ring For Waterproof

    2.2 Set of O-ring for waterproof When the factory is shipped, O-ring for the waterproof is not installed. When the waterproof and dustproof uses it by the necessary environment, please install it according to the figure below. O-ring for waterproof It is not abnormal though the opening and shutting operation of the cover CAUTION becomes very hard if O-ring for the waterproof is installed.
  • Page 14: 3. Installation

    3. INSTALLATION 3.1 Installation place Install place This equipment is a structure that is installed in the panel and used. Please choose and install the following places. Place without vibration and impact. (Vibration:10 60Hz 0.2m/s Impact:The impact is ・ ~ not allowed.) Place where dust and oily smoke are few.
  • Page 15: 4. Wiring

    4. WIRING 4.1 Terminal stand array and LAN connector The terminal stand is one row in the uppermost part, and 4 rows or less are in the lower side for the analog input and option. The part of “Terminal No.41 49”...
  • Page 16: Wiring For Power Supply

    4.2 Wiring for power supply Warning Please energize to this equipment after doing the protection earth without fail for the ● electric shock prevention. Please do not cut the protective earth, and please do not remove connecting wires of the ●...
  • Page 17: Wiring For Analog Input

    4.3 Wiring for analog input CAUTION Notes of input wire ● ・ Please do not mix the noise about the input wiring. Moreover, the use of an effective shield line or twist line is recommended to the noise in the input wiring. ・...
  • Page 18: Wiring For Com Alm

    4.4 Wiring for COM ALM COM ALM can be used as an alarm output of measurements etc. terminal [COM ALM ] terminal is “Terminal No. 55 56”. COM ALM ~ ※ Please connect neither “A terminal (output terminal)” nor “C terminal (COM terminal)” oppositely. The circuit might be damaged when not correctly connected.
  • Page 19: Wiring For Di/Do (Option)

    4.6 Wiring for DI/DO (Option) DI/DO becomes a connector joint. DI/DO connector [Pin array] Pin No. Signal name Pin No. Signal name DO10 DO11 DO12 DI_COM DO_COM DI_COM DO_COM DI_COM DO_COM DI_COM DO_COM DI_COM DO_COM DI_COM DO_COM DI_COM DO_COM DI_COM DO_COM...
  • Page 20 The DI/DO cable (WMSU0468A01: cable length 1m, WMSU0468A02: cable length 3m) is available as an option. For details, please contact to state of the purchase, or our salesperson. Refer to below table for signal arrangement of the cable. DI cable DO cable Pin No.
  • Page 21 [DI schematic diagram] +12V DI_COM No-voltage contact input (9 point), terminals (DI_COM) of each channel are common. Rating:Photo-coupler drive 12V DC about 3mA/1 point [DO schematic diagram] Load +5~30V DO_ COM Open collector output (12 point) , terminals (DO_COM) of each channel are common. Rating:30V DC 20mA/1 point [Reference]...
  • Page 22: Wiring For Rs-485 (Option)

    CAUTION Please put terminator (Rt=200Ω) on the host side. In view of the host, please install terminator ● (Rt=200Ω) in VM7000A in the distance most. The connection is 32 or less including the host computer. ● The length of the cable is 1.2km or less.
  • Page 23: 5. Operation

    5. OPERATION 5.1 Before operating Please confirm the installation of SD card (Refer item 1.5), wiring (Refer item 4) before it operates. Afterwards, please confirm the various parameter settings (Refer item 7). 5.2 Start and stop of record The record begins when REC button is pushed. When the record is stopped, the REC button is pushed again, and it Stop key touches on the confirmation screen.
  • Page 24: 6. Display Function

    6. DISPLAY FUNCTION 6.1 Basic composition of data display screen ④Graph type ①Group screen name ③Change trend ②Clock display ⑤Change display ⑥Scale display ⑦Data display area ⑩Event flag display area ⑧Measured value display area ⑭Sub record display ⑬Main record display ⑨Event display area ⑮Internal memory ⑪SD card loading /...
  • Page 25 Data display area ⑦ Allow the Real time trend display, Historical trend display, Bar graph display, Digital display and Event history / communication history display to be displayed. (Refer item 6.2 to 6.6) Display measured data at the position of the cursor when Historical trend display. Measured value display area ⑧...
  • Page 26 Comment:Arbitrary comments can be registered on a trend. (It supports since recorder version 1.20.) (A comment can be set up only on a historical-trend screen. A comment is not directly displayed on a trend screen. Refer item 6.7 for comment function.) Scale:The scale can temporarily be changed.
  • Page 27: Real Time Trend Display Of Measured Data

    6.2 Real time trend display of measured data [Explanation] The measured data can be display in graph. The vertical or horizontal trend directions can be selected by touching to GRPH key. The refreshment cycles of graph synchronizes record cycles. Horizontal trend Vertical trend Time (H : M : S)
  • Page 28: Digital Display Of Measured Data

    6.4 Digital display of measured data Explanation] [ The measured data can be display in digital graph. [Operation] The measured data can be changed digital graph by touching the GRPH key several times. Channel No. ①Unit ②Display of measured value ③Alarm No.
  • Page 29: Historical Trend Display

    6.5 Historical trend display [Explanation] The past data of currently recoding data or the data saved in the past can be read and displayed. [Operation] Press the REAL key on the Real time trend display, and the following display appears. ①Cursor date ②Cursor ⑦Measured value...
  • Page 30 File select key ⑥ The data saved in the past can be read and displayed. Folder group File group When you select an arbitrary folder from among the “Folder group”, the file data included in the folder is displayed in the file group. When select an arbitrary file, and touching the OK key, on the historical trend screen, the data preserved in the past is read and displayed.
  • Page 31: Event History / Communication History Display

    6.6 Event history / communication history display [Explanation] When a specific event is generated in the data recording now, it is possible to make a mark. The history of LAN communication is preserved. [Operation] The event history display can be displayed by touching the GRPH key several times. And, the screen can be changed communication history display by touching the DISP key several times.
  • Page 32: Comment List Display

    6.7 Comment list display [Explanation] Arbitrary comments are registered to a trend, and it can jump to the time which registered the comment from the comment list screen. (It supports since recorder version 1.20.) A comment can be registered only on a historical-trend screen. A comment is not directly ※...
  • Page 33 ● A comment list is referred to, and it jumps to a comment point. If the GRPH key of a historical-trend screen is touched, it will move to a comment list screen. The comment displayed on the comment list screen is touched and chosen, and cursor is moved to a comment point by touching a Jump key.
  • Page 34: 7. Setting And Checking Parameters

    7. SETTING AND CHECKING PARAMETERS 7.1 Operational mode [Explanation] This Paperless recorder can do a more detailed setting by setting the “Operation mode” to “Advanced mode” on the parameter setting screen and the system setting screen. [Operation] System key ⇒Device/Other key ⇒Mode key Touch the MENU button ⇒...
  • Page 35: Setting And Checking

    7.2 Setting and checking Main menu ① The menu display can be displayed by pressing MENU button. The parameter setting display can be displayed by select “Parameter” touching. Back key when touching, it returns to the trend screen. Back key Main menu Parameter setting Sub menu...
  • Page 36: Outline Of Parameter Setting Procedure

    7.3 Outline of parameter setting procedure Data display screen Menu screen MENU Page Parameter Input CH Input … 7-7 Input setting screen … 7-9 Scaling Scaling setting screen Display Display setting screen … 7-11 Scale setting screen … 7-13 Scale Alarm value Alarm value setting screen …...
  • Page 37: Basic Operation Of Setting Screens

    7.4 Basic operation of setting screens [Explanation] The basic operation of the setting screens is classified in the following methods. To move a set item, the corresponding item is touched. Item into which set content whenever touching changes ● In this case, OFF key when touching, it changes with Instant value key Average key ⇒...
  • Page 38 When selecting contents to set from list. ● The item displayed in the list is selected touching. The scroll bar is displayed when there is a selection item that exceeds the size of the screen. The display can be changed by touching arrow key ( ) or sliding scroll bar.
  • Page 39 Character entry box ① The input character is displayed. By touching the [ Character entry box], an input position is arbitrarily movable. ① Alphabet key ② The capital letters and small letters can be changed. Numerical value key ③ The numerical value input can be changed. Symbol key ④...
  • Page 40: Setting The Input Spec

    7.5 Setting the input spec [Explanation] For select of the input types for each channel (thermocouple, resistance bulb and DC voltage input), and the presence of Burn Out function can be set. Note: When the recorder is in recording, the setting cannot be changed. [Operation]...
  • Page 41 Burn Out ③ Setting the Burn Out function.(The record swings over to the 0% side or the 100% side) Burn Out function can only be set when the input type is “TC” and “mV”. ④ The method of making amends for the temperature of the terminal of the thermo-couple range is selected.
  • Page 42: Setting The Scaling

    7.6 Setting the scaling [Explanation] Set the “scaling” and “Square root” when the input kind is “DC voltage” or “DC current”, and set the “Measure range”, “Scale range”, “Decimal point” and “Unit”. Note: Scaling can only be displayed when the input type is “DC voltage” or “DC current”. Note: When the recorder is in recording, the setting cannot be changed.
  • Page 43 Scale range ④ This item can only be set when the “ Scaling” is “ON” or “Square root ON”. ② The lower bound value (left) and the upper bound value (right) of the scaling range to “Measure value” are set. (-32000 to 32000 digit) Decimal point ⑤...
  • Page 44: Setting The Display Method Of Channel

    7.7 Setting the display method of channel [Explanation] The tag name, the explanation, and the display color of each channel are set, and a set content is displayed on the trend screen. Note: When the recorder is in recording, the setting cannot be changed. [Operation]...
  • Page 45 Display color ④ The display color of each channel is set, and displays on the trend screen. Please select an arbitrary display color from the display color selection screen (figure below). The part where the setting is reflected becomes “Trend display”, “Scale display”, and “Measurements display area”.
  • Page 46: Setting The Scale Of Channel

    7.8 Setting the scale of channel [Explanation] Range of scale, and Partitions of each channel are set. Note: When the recorder is in recording, the setting cannot be changed. [Operation] Select the Input CH key Scale key on the Parameter. ⇒...
  • Page 47 Scale No. ③ You can display up to three scales, each on one level. Configure the scale No. setting to specify which level each channel should be displayed on. [One-level display] Scale No. 1: CH01, CH02, CH03, CH04, CH05, CH06, CH07, CH08, and CH09, Scale No.
  • Page 48: Setting The Alarm Value

    7.9 Setting the alarm value [Explanation] When abnormality occurs in measurements recording, warning is occurred. Moreover, the warning units relay number of the option is specified, and the warning output destination is set. [Operation] Select the Input CH key Alarm value key on the Parameter. ⇒...
  • Page 49 Output ③ The warning units relay number of the option is specified, and the warning output destination is set. The selection item changes by connecting the relay and the transistor. Common : The main body has been equipped normally. Relay : When the “Relay card (Option)”...
  • Page 50: Setting The Alarm Action

    7.10 Setting the alarm action [Explanation] “Hysteresis” and “Alarm delay” of each channel are set. Then, when measurements stay near the alarm value, occurred/release of alarm is consecutive and it is made not to happen. Note: When the recorder is in recording, the setting cannot be changed. [Operation]...
  • Page 51: Setting The Rec/Calc

    7.11 Setting the REC/CALC [Explanation] “Input filter”, “Record type”, “Offset”, and “Gain” of each channel are set. Note: When the recorder is in recording, the setting cannot be changed. [Operation] Select the Input CH key REC/CALC key on the Parameter. ⇒...
  • Page 52 OK, Cancel key ⑥ The each setting for changing contents is saved by the OK key. And, the “Parameter” can be returned not change by the Cancel key. No selected range ⑦ The gray color is displayed that cannot be selected now. In this case, can be selected by presses OK key (or Cancel key) returning to “Parameter”.
  • Page 53 ●The specification of offset and gain This recorder can set offset (shift value) and gain (inclination). The graph when setting offset and a gain is shown.(following) Ex) Channel type: 0-5mV Scaling range: 0-100% Change of output value by Offset Offset = 50 Offset = 0 Input(0-5mV) Change of output value by Gain...
  • Page 54: Copy The Channel Setting

    7.12 Copy the channel setting [Explanation] The setting value of each channel is copied onto another channel. Note: When the recorder is in recording, cannot be copy of channel. [Operation] Select the Input CH key Copy key on the Parameter. ⇒...
  • Page 55: Setting The F Value

    7.13 Setting the F value [Explanation] “Reference temp.”, “Z value (Extinction value of bacterium by heating sterilization)” and “Start temp.” of each channel are set. Note: When the recorder is in recording, the setting cannot be changed. [Operation] Select the Calc. CH key F value key on the Parameter.
  • Page 56: Setting The Timer Of Channel

    7.14 Setting the timer of channel [Explanation] To execute reset of the input value at the set interval time, “T1 timer (sec)”, “T2 timer (min)” of each channel is set. Note: When the recorder is in recording, the setting cannot be changed. [Operation]...
  • Page 57: Setting The Group Name

    7.15 Setting the group name [Explanation] Name of group that displays it on trend screen, switch of channel No. and tag display, and switch of ON/OFF of group display are set. Note: When the recorder is in recording, the setting cannot be changed. [Operation]...
  • Page 58: Setting The Displayed Channel

    7.16 Setting the displayed channel [Explanation] The combination of channel No. that each group displays in the measured value display area of the trend screen is set. Note: When the recorder is in recording, the setting cannot be changed. [Operation] Select the Display key Group CH key on the Parameter.
  • Page 59: Select The On/Off Of Graph Type

    7.17 Select the ON/OFF of graph type [Explanation] The switch of display/non-display of “Horizontal trend”, “Vertical trend”, “Bar graph”, and “Digital display” of each group displayed on the trend screen is set. (Refer item 6.2 6.6 for each trend ~ screen.) Note: When the recorder is in recording, the setting cannot be changed.
  • Page 60: Setting The Automatic Change Function Of A Display Group

    7.18 Setting the automatic change function of a display group [Explanation] In a realtime trend screen, a group is automatically switched at a fixed interval. (It supports since recorder version 1.30.) [Operation] Select the Display key Auto display key on the Parameter. ⇒...
  • Page 61: Setting The Record Cycle Of Main Record

    7.19 Setting the record cycle of main record [Explanation] “Record type of main record”, “File record cycle” and “ON/ OFF of file overwrite function” is set. Note: When the recorder is in recording, the setting cannot be changed. [Operation] Select the Record key Setting (Main) key on the Parameter.
  • Page 62: Setting The Record Cycle Of Sub Record

    7.20 Setting the record cycle of sub record [Explanation] “Record type of sub record”, “Sub pre record” and “Sub record timing” is set. Note: When the recorder is in recording, the setting cannot be changed. [Operation] Select the Record key Setting (Sub) key on the Parameter.
  • Page 63 Sub record timing ③ The sub recorded timing is set. (A sub record alone cannot be recorded. Please confirm it is previously recording before it records.) Func key:The sub record begins when the FUNC button is pushed. System Settings “Device / Other” ⇒ “FUNC key” is “OFF” only when you can choose. ※...
  • Page 64: Setting The Schedule

    7.21 Setting the schedule [Explanation] “ON/OFF of schedule function” (Record is started or stopped periodically.), “Start time”, “End time” and “A day of the week” is set. Note: When the recorder is in recording, the setting cannot be changed. [Operation] Select the Record key Schedule key on the Parameter.
  • Page 65: Setting The Unit

    7.22 Setting the unit [Explanation] “Making”, “Edit”, and “Delete”, etc. of the unit displayed on the trend screen are operated. Note: When the recorder is in recording, the unit cannot be made. [Operation] Select the Others key Unit key on the Parameter. ⇒...
  • Page 66: Setting The Message

    7.23 Setting the message [Explanation] When “alarm is occurred and cleared” or “DI ON/OFF”, the message is displayed. Note: When the recorder is in recording, the setting cannot be changed. [Operation] Select the Others key Message key on the Parameter. ⇒...
  • Page 67: Setting The Di Function

    7.24 Setting the DI function [Explanation] DI function is set. Note: When the recorder is in recording, the setting cannot be changed. [Operation] Select the Others key DI key on the Parameter. ⇒ ①DI No. ②DI function DI No. ① The set DI No.
  • Page 68: Counts The Progress Time

    7.25 Counts the progress time [Explanation] Progress time is counted according to conditions. The counted time can be outputted as a value, if "PASS function" is used. (Refer item 2.3 of “PARAMETER LOADER INSTRUCTION MANUAL” for "PASS function".) Note: When the recorder is in recording, the setting cannot be changed. [Operation]...
  • Page 69: Initialization Of Parameter

    7.26 Initialization of parameter [Explanation] Parameter data is initialed. Note: When the recorder is in recording, the parameter cannot be initialized. [Operation] Select the Others key Param initial key on the Parameter. ⇒ When Param initial key is selected, the confirmation screen (above figure) is displayed. The parameter data is initialized when selecting YES key.
  • Page 70 ●Parameter data initial value list ・Input CH ・Display Input Input type :±10V Group name Display name :DISP_GRP_1 :DISP_GRP_2 Scaling Scaling :ON :DISP_GRP_3 Meas. range(L) :-10.00 :DISP_GRP_4 Meas. range(H) :10.00 :DISP_GRP_5 Scale range(L) :0.00 :DISP_GRP_6 Scale range(H) :100.00 :DISP_GRP_SUB Decimal point :2 TAG disp set :Channel No.
  • Page 71: Using The Wizard Function

    7.27 Using the wizard function [Explanation] Necessary minimum set item can continuously be set by using the wizard function. Moreover, when the Mode of the System is only “Normal”, this item is displayed. (Refer item 7.1 for Mode key.) Note: When the recorder is in recording, the wizard function cannot be used. [Operation]...
  • Page 72: Input Ch" List

    7.28 “Input CH” list [Explanation] Various settings concerning the input channel are done. Refer item 7.5 7.12 for details. ~ [Reference] When the operational mode is only an advanced mode, the item that “ is attached to “Advanced” ○” is displayed. Refer item 7.1 for “Mode”. There are setting and item not displayed according to the number of input channels and set of the other content.
  • Page 73 [Scale] Item Setting contents Advanced Channel Select the channel number. Range of scale Input the range of scale. Scale No. Set the scale No. Input the value of partitions. When “0” is input, it displays it according Partitions ○ to the scale display automatically. [Alarm value]...
  • Page 74: Calc. Ch" List

    7.29 “Calc. CH” list [Explanation] Various settings concerning the calc. channel are done. Refer item 7.5 7.14 for details. ~ The operational expression and the more details setting of the operation channel can be set with the parameter loader software. Setting it with the main body of the recorder becomes only the part. Please refer the “PARAMETER LOADER INSTRUCTION MANUAL”...
  • Page 75 [Alarm action] Item Setting contents Advanced Channel Select the channel number. ○ Hysteresis ( ) Input the value of Hysteresis. ○ Input the value of Alarm delay. Alarm Delay(sec) ○ [REC/CALC] Item Setting contents Advanced Channel Select the channel number. ○...
  • Page 76: Display" List

    7.30 “Display” list [Explanation] In the “Display”, it is possible to set it variously concerning the display of the measuring data. Refer item 7.15 7.18 for details. ~ [Group name] Item Setting contents Advanced Group Select the group number. Display name Set the display name. TAG disp set Select the TAG display set.
  • Page 77: Record" List

    7.31 “Record” list [Explanation] In the record setting, it is possible to set it variously concerning the record of the measuring data. Refer item 7.19 7.21 for details. ~ [Setting(Main)] Item Setting contents Advanced Record Cycle Select the record cycle. File rec cycle Select the file rec cycle.
  • Page 78: Others" List

    7.32 “Others” list [Explanation] Various settings concerning the others are done. Refer item 7.22 7.27 for details. ~ [Unit] Item Setting contents Advanced Unit The edit display of new unit can be moved by touching “Add”. [Message] Item Setting contents Advanced Message No.
  • Page 79: 8. Setting And Checking Systems

    8. SETTING AND CHECKING SYSTEMS Outline of system setting procedure Page SD remove … 8-2 SD card writing screen SD/Param System SD format SD card format screen … 8-3 Param save Setting value saving screen … 8-4 Param load Setting value loading screen …...
  • Page 80: Remove The Sd Card

    Remove the SD card [Explanation] Data is written to the SD card, and it enters the state that the SD card can be detached. When selecting SD remove key with the SD card has not been inserted in the recorder, the error ※...
  • Page 81: Format The Sd Card

    Format the SD card [Explanation] The data of the SD card is formatted. When selecting SD format key with the SD card has not been inserted in the recorder, the error ※ message is displayed. Please select again after inserting the SD card in the recorder. Note: When the recorder is in recording, the SD card cannot be formatted.
  • Page 82: Save The Set Value

    Save the set value [Explanation] A set value is preserved on the SD card. When selecting Param save key with the SD card has not been inserted in the recorder, the error ※ message is displayed. Please select again after inserting the SD card in the recorder. [Operation]...
  • Page 83: Load The Set Value

    Load the set value [Explanation] The set value preserved in the past is read from the SD card. When selecting Param load key with the SD card has not been inserted in the recorder, the error ※ message is displayed. Please select again after inserting the SD card in the recorder. Note: When the recorder is in recording, the set value cannot be loaded.
  • Page 84: Setting The Ip Address

    Setting the IP address [Explanation] IP address etc. necessary for Ethernet are set. The following can be done by setting “Ethernet and FTP”. The record file preserved on the SD card of the recorder can be downloaded directly to the ●...
  • Page 85 Example of setting IP Address ● The Ethernet setting screen (above figure) is displayed. Input an arbitrary numerical value to “IP Address” and “Subnet Mask”. (In this case, IP address is set to “192.168.0.1”, and the Subnet Mask is set to “255.255.255.0”.) Input an arbitrary numerical value to “Default GW”...
  • Page 86: Setting The Keep Alive Function

    Setting the keep alive function [Explanation] It is regularly confirmed whether the network is connected by setting “Keep alive function”. Moreover, the cycle when “Keep alive function” is executed is set. Note: When the recorder is in recording, the setting cannot be changed. [Operation]...
  • Page 87: Setting The Sntp Client Function

    Setting the SNTP client function [Description] Use the SNTP client function to receive the time data from the time server on a regular basis to correct the recorder's time. [Operation] On the system setting screen, select Comm. key SNTP1 key to open the SNTP function setting ⇒...
  • Page 88: Setting The Time Zone And Sntp Server Function

    Setting the time zone and SNTP server function [Description] Set the time acquisition function ON/OFF and the time zone (standard time). [Operation] On the system setting screen, select Comm. key SNTP2 key to open the SNTP function setting ⇒ screen. ①...
  • Page 89: Setting The Ftp User

    8.10 Setting the FTP user [Explanation] The FTP user's making, the edit, and the deletion, etc. are operated. The following can be done by setting “Ethernet and FTP”. The record file preserved on the SD card of the recorder can be downloaded directly to the ●...
  • Page 90 User name ② When an Add key is selected, the FTP user setting screen (above figure) is displayed. Please input an arbitrary user name from the character input screen, and set it with the OK key. Password ③ The password is set to the made user name. (The space character cannot be used for the password.
  • Page 91: Setting Modbus1 (Rs485 Communication Setting, Timeout, Etc.)

    8.11 Setting Modbus1 (RS485 communication setting, timeout, etc.) [Description] Set the Modbus station number, reception timeout function, parity, stop bit and communication speed. Note) No settings can be changed during the recording process. [Operation] On the system setting screen, select Comm. key Modbus1 key to open the Modbus1 setting ⇒...
  • Page 92: Setting Modbus2 (Communication Type)

    8.12 Setting Modbus2 (communication type) [Description] Set the Modbus communication type and master/slave operations. The setting items vary depending on the communication types and master/slave settings. For details about the master functions, see the COMMUNICATION FUNCTION OPERATION MANUAL. Note) No settings can be changed during the recording process. [Operation] On the system setting screen, select Comm.
  • Page 93 ① Selection of Modbus operation Select the Modbus operation. Modbus TCP (Ethernet): Modbus TCP is used for communication protocols. Only the slave can be selected for the operation. Modbus TCP+RTU: Add RS-485 to the available communication protocols. (Enabled only if the RS-485 option is implemented) For this setting, additionally set the operation on the ModbusRTU side.
  • Page 94: Setting The Modbus Master

    8.13 Setting the Modbus master [Description] With this device as the Modbus master, carry out the settings of the device to be connected with the general-purpose Modbus slave. This is used when the communication type in section 8.12 is Master (General).
  • Page 95 ④ Selection of data format Select how the acquired data is handled. INT16: Acquired as a signed 2-byte data (-32768 to 32767). UINT16: Acquired as an unsigned 2-byte data (0 to 65535). INT32 (BIG): Acquired as a signed 4-byte data (-2147483648 to 2147483647). The data is recognized as a big endian.
  • Page 96: Setting The Remote Ai

    8.14 Setting the remote AI. [Description] With this device as the Modbus master, perform the settings for connecting with a separately-sold remote AI. This is used when the Master (Remote AI) is set for the communication type in section 8.12. For the specifications of the separately-sold remote AI, see the REMOTE AI INSTRUCTION MANUAL.
  • Page 97: Setting The Lcd Sleep Time

    8.15 Setting the LCD sleep time [Explanation] LCD is automatically turned off when there is no operation in the main body for a fixed time. [Operation] Select the Device/Other key ⇒ LCD key on the System. ①Sleep time ②Act. brightness ③Sleep brightness ④Alarm Recovery Sleep time...
  • Page 98: Setting The Clock

    8.16 Setting the clock [Explanation] The date of the recorder is changed by manual operation. Note: When the recorder is in recording, the setting cannot be changed. [Operation] Select the Device/Other key ⇒ Clock key on the System. ①Current date ②Setting date Current date ①...
  • Page 99: Setting The Func Key

    8.17 Setting the FUNC key [Explanation] The function of the “Switch of the screen”, “Capture”, “Display of message”, and “ON/OFF of sub record” is allocated in the FUNC button. [Operation] Select the Device/Other key ⇒ FUNC key on the System. ①Function key Function key ①...
  • Page 100: Setting The File Format

    8.18 Setting the file format [Explanation] The file format of record data is set. Note: When the recorder is in recording, the setting cannot be changed. [Operation] Select the Device/Other key ⇒ File format key on the System. ①File format File format ①...
  • Page 101: Setting The Jump Menu

    8.19 Setting the jump menu [Explanation] Shortcut keys are registered up to 8 in the menu screen. Note: When the recorder is in recording, the setting cannot be changed. [Operation] Select the Device/Other key ⇒ Jump menu key on the System. ①Jump menu group ②Menu group Jump menu group...
  • Page 102: Setting The Operation Mode

    8.20 Setting the operation mode [Explanation] This Paperless recorder can do a more detailed setting by setting the “Operation mode” to “Advanced mode” on the parameter setting screen and the system setting screen. Note: When the recorder is in recording, the setting cannot be changed. [Operation]...
  • Page 103: Setting The Language

    8.21 Setting the language [Explanation] “Language format” and “Date format” of the recorder are set. Note: When the recorder is in recording, the setting cannot be changed. [Operation] Select the Device/Other key ⇒ Language key ⇒on the System. Moreover, the Language key can be selected directly from the MENU.
  • Page 104: Display The System Information

    8.22 Display the system information [Explanation] The Version and Serial No. of recorder is displayed. [Operation] Select the Device/Other key ⇒ Version key on the System. ①Version ②Serial No. Version ① The Version of recorder is displayed. Serial No. ② The Serial No.
  • Page 105: Setting The Key Lock Function

    8.23 Setting the key lock function [Explanation] The setting change screen and front button of a recorder are locked, and a key can be restricted. (It supports since recorder version 1.20.) [Operation] Select the Security key ⇒ Key lock key on the System. ①Key lock ②Password ③Menu...
  • Page 106 ●The example of setting of a key lock A key lock setting screen is displayed. The setting or cancellation method of a key lock is set up. (In this case, it is set as “Auto + manual”) A password is set up. (In this case, it is set as “12345”.) The key lock level in a “Menu”...
  • Page 107: Sd / Param" List

    8.24 “SD / Param” list [Explanation] Various settings concerning the SD card are done. Refer item 8.2 8.5 for details. ~ [SD remove] Please execute it before taking out the SD card. Item Setting contents Advanced The current measuring data is written to the SD card while recording SD remove and the SD card can be taken out.
  • Page 108 [SNTP1] Items Settings Advanced Current time and Displays the current time and date. date Select ON/OFF of the SNTP client function. If ON is set, time data is received from SNTP server and time calibration is carried out. SNTP client When the correction of the time is attempted using the SNTP function during the function recording process, correction is not performed if there is a difference of ±5 minutes or...
  • Page 109: Device / Other" List

    [ModbusMas.] * Enabled only when the Master (General) is set for the master/slave setting. Items Settings Advanced Communication Select the ID for connection. Slave ID Set the slave ID for ModbusRTU. Relative Address Set the communication address of a slave device. Data format Select how the acquired data is handled.
  • Page 110 [Jump menu] Item Setting contents Advanced The item added to the menu display is selected. [Mode] Item Setting contents Advanced Operation mode Select the operation mode. [Language] Item Setting contents Advanced Language Select the language. Date format Select the date format. Decimal symbol Select the decimal symbol.
  • Page 111: Security" List

    8.27 "Security" list [Explanation] Various settings security of system is done. Refer item 8.23 for details. [Key lock] Item Setting contents Advanced Key lock The setting method of a key lock is set up. Password A password is set up.(A blank character can't be used for a password) The level of the key lock in Menu screen is set up.
  • Page 112: 9. Maintenance

    9. MAINTENANCE 9.1 Inspection In order to effectively use the instrument, please inspect periodically. Inspection item Inspection description Remark Measurement value - Correct measurement value has been displayed? Display - The brightness of the backlight appropriate? - Is there lack and blur on the display? Recording data - Measurement value indication and data storing have been correctly?
  • Page 113: Troubleshooting

    9.3 Troubleshooting While in using this instrument, an error message may be displayed on event display area of the bottom of a screen, and the event history. The table below lists error messages and steps to be taken. ● Error messages If one of the errors listed below occurs, an error message is displayed in the event message display area of the bottom of a screen.
  • Page 114: 10. Specification

    10. SPECIFICATION 10.1 Basic specification •Number of inputs:Selection from 6,9,12 points.(Immediately after purchase) •Input circuit:Inputting mutual insulation. •Measuring period:100mm sec •Input type: Direct voltage, direct current (shunt resistance of necessity), TC and RTD •Changing of input type: Setting from set menu displayed with front side MENU button. •Burnout function:...
  • Page 115: Measurement Range

    10.2 Measurement range Max. Code Type Measuring range Measurement accuracy Notes resolution -10.00 ~ +10.00 10μV 0.00 ~ +20.00 10μV 10μV 0.00 ~ +50.00 -0.200 ~ +0.200 -1.000 ~ +1.000 10mV -10.00 ~ +10.00 ±(0.1% F.S.+1digit) 0.000 ~ +5.000 0.01mA 4.00 ~...
  • Page 116: Display Part

    10.3 Display part •Indicator: 5.7 inch TFT color LCD(320×240 dot) The touch panel, back light is applied. The brilliance control is possible. The pixel that always lights partially or doesn't light the liquid crystal display might exist. And, it is not a breakdown though the irregularity of brightness might be caused. Please acknowledge it beforehand.
  • Page 117: Record Function

    10.5 Record function •External recording medium: SD memory card(It corresponds to the SD/SDHC standard) •Internal memory:About 100MB •Record capacity: SD standard:2GB or less SDHC standard:32GB or less •Record method: The record begins by turning on the REC button. It records by the new file name at the time of each record beginning.
  • Page 118: Alarm Function

    •SD memory card: Confirmed operation SD memory card: ・Panasonic company 1 32GB ~ ・Sandisk company 32GB ~ Please buy it in the computer shop etc. •Data format: It is possible to select it from either of method of the binary or binary + CSV.(It is not possible to change while recording.
  • Page 119: Power Supply Part

    10.8 Power supply part • Rated supply voltage 240V AC :100 ~ • Range of use voltage 264V AC :85 ~ • Power supply frequency:50/60Hz(Sharing) • Power consumption: Power consumption Power-supply voltage Usually LCD OFF ※ 100V AC Under 15VA Under 12VA 240V AC Under 25VA...
  • Page 120: Compatible Specification

    10.12 Compatible specification • CE: EMC instruction:EN61326-1 compatible Low Voltage Directive:EN61010-1 agreement • Dustproof and waterproof standard: IEC60529 IP65 conforming (Front Panel only) 10.13 Transportation and storage conditions • Temperature: 60℃ - ~ • Humidity:5 90%RH ~ • Vibration:Under 10 60Hz 2.45m/s ~...
  • Page 121: Support Software

    •DI input: Dry contact input (9 points), Common terminal is shared. Ratings:Photo-coupler drive 12V DC about 3mA/1 point The following control is possible according to the contact input.(ON/OFF pulse time over 0.5 sec) Start/stop of main and sub record operation ①...
  • Page 122: Dimension

    10.16 Dimension unit: mm <Panel cutting> unit: mm 10-9...
  • Page 123: Folder Composition Of Sd Card

    10.17 Folder composition of SD card When the SD card is inserted in the main body of the Recorder, the “Recorder” folder is automatically made. “Recorder” folder contents are as follows. [SD card folder composition] [Recorder]-+-[Cap] +-[Data] +-[Etc] +-[Prm] 1) Cap The image taken by capture is preserved by bmp.
  • Page 124: Option Item

    ・Sub record management file It is a file that records “Trend data (sub) record start time”, “Stop time”, and “Link information”. File name:xxxx_YYMMDDHHmm.ds Extension:ds ・Sub record trend file It is a measurement data file divided at the file record cycle (sub). File name:xxxx_YYMMDDHHmm.dst Extension:dst 3) Etc...