NOVA EXAM Floorstand, Wall, Table and Ceiling Instructions for Use & Maintenance 5000468 A06 Instructions for Use...
Page 2
NOVA EXAM Task & Examination Light BASE MODEL MOUNT AREA BLANK (United States, 120V) T (table) 01 (Japan, 100V) FL (floor) 02 (United Kingdom, 240V) W (wall) 03 (Europe, 230V) C (ceiling) 25 (Spain, 230V) 26 (Australia, 240V) Example; NXFL is a floor model for the US NXC25 is a ceiling mounted model for Spain, includes a Spanish IFU...
Technical Description The NOVA EXAM LED light is a durable, attractive light designed to deliver high-quality illumination in the modern hospital or clinic for examination application. Its design offers excellent dual brightness, color rendition, large field size, and convenient positioning with easy-grip handles.
Page 5
Classification Type of protection against electric shock………………………………………………..Class 1 Allowable leakage current……………………………………………...does not exceed 200µa Reliability of earth protection………………………………………...does not exceed 0.1 ohm Mode of operation………………………………………………………………..…….continuous Protection against explosion hazards……………............…....……...not to be used in presence of flammable anesthetics Protection against hazardous parts and ingress of liquids…………………….………...IP20 Degree of mobility…...……portable (floor stand models move on 4 casters [2 locking])…...
4. Be sure that nothing has been removed from the cartons before they are checked by the individual in charge. 1.1.5 If a Shortage or Damage Occurs: 1. You, the receiver, not Burton Medical, is responsible for filing any claim(s) with the delivering carrier within five days after receipt of the shipment.
Wall Mount Assembly 1.2.1 Burton Medical recommends a wall mounting height of 55-60 inches (140-152.4 cm) from floor plane. 1.2.2 Attach the provided Wall Mount to a suitable vertical surface, it should go into wood support that may need to be constructed. It should not be an- chored to drywall alone.
1.5 Floor stand Operation 1.5.1 Plug the power cord into a grounded convenience socket. Note: Turn the light “On” by pressing the power switch located at the base of the head to either left or right depending on the intensity required and adjust the arm as required.
LED module replacement must be replaced by an Burton Medical appointed tech- nician. 2.2.1 Turn the power switch off (O) and disconnect the fixture from the mains supply.
2.3 Preventive Maintenance Checklists Check Corrective Action Weekly Check overall operation of the fixture: Do the switch and LED’s operate properly? If the answer to any of these ques- tions is NO, do not use the product. Does the light head stay in position (no drift) Consult with your maintenance per- when the arm is moved up and down or sonnel before operating the light.
Replacement Parts 5.1 Power Supply..........4000173 5.2 Handle............5000451 5.3 Switch.............4000182 5.4 Fuse T 0.5 Amp L…........0007027 5.5 Locking Caster..........4000232 5.6 LED Module……….........6000266 3. Symbols and Warnings Symbols This symbol on the power switch is “OFF”. The rocker switch, when postioned to the “O” side, turns off power to the light. This symbol on the power switch is “ON”.
Troubleshooting Guide Symptoms Probable Cause Remedy Not plugged in. Plug into a suitable hospital-grade grounded socket. Switch not “on” ( I or II ). Turn switch on ( I ), left or right LED failed. Replace with new LED module Fuses failed: Replace with new fuse(s).
Warranty Information For complete, detailed information on your specific product’s warranty coverage, please visit: http://www.burtonmedical.com/wp-content/themes/burtonmedical/literature/warranty.pdf...
This product was designed and assembled in the U.S.A. by: Burton Medical LLC 2300 W Windsor Ct., STE C Addison, IL 60101 U.S.A. Tel.: (800) 444-9909 Fax: (800) 765-1770 Web: www.burtonmedical.com...