Page 1
AIM-50 ™ INSTALLATION MANUAL WALL MOUNT VERSION PERFORMANCE RELIABILITY VALUE The Right Light EXAMINATION...
Page 3
Installation Manual ™ AIM-50 Wall Mount The following models are covered in this manual: A50W AIM-50™ Wall mount version 120 V A51W AIM-50™ Wall mount version 100 V A52W AIM-50™ Wall mount version 240 V A53W AIM-50™ Wall mount version 230 V This product was designed and assembled in the U.S.A.
This manual contains detailed information on the above, but responsibility for effective installation ultimately rests with skilled and qualified contractors. Products should at all times be handled by qualified staff; it is the responsibility of the user to ensure this is the case. AIM-50™ Installation Manual − Wall Mount Version 5000145 A01...
Dear Installer, Congratulations on your purchase of the AIM-50™ Examination Light! The installation manual gives instructions on how to install the AIM-50™ Examination Light Wall Mount Version. For instructions on operation, maintenance and further description of the product please refer to the Operation & Maintenance manual.
The installation shall meet requirements contained in the latest issues of the NEC and NFPA, and other local and national building codes; as they apply to hospitals, clinics, medical offices, and the like. AIM-50™ Installation Manual − Wall Mount Version 5000145 A01...
Equipment anchorage diagrams are supplied with these instructions to help with the installation (see page 14). The diagrams were prepared by a California-licensed Structural Engineer. If the lights are installed accordingly, the system will meet requirements of the State Seismic Codes. AIM-50™ Installation Manual − Wall Mount Version 5000145 A01...
10.0 ft (3050 mm) 91 in (2311 mm) 67 in (1702 mm) 108 in (2743 mm) Range of Motion Figure 1: Vertical range of motion Figure 2: Horizontal range of motion AIM-50™ Installation Manual − Wall Mount Version 5000145 A01...
If a Shortage or Damage Occurs: 1. You, the receiver, not Burton, are responsible for filing any claim(s) with the delivering carrier within five (5) days after receipt of the shipment. 2. If damage or shortage occurs in transit, the delivering carrier is required by law to make notation of a shortage or damage.
Figure 3: Contractor supplied mounting plate (dimensions are inches [mm]) 2. Assemble the mounting plate to the bearing structure of the wall. 3. Feed the wires coming through the central hole in the mounting plate into the wall mount assembly. AIM-50™ Installation Manual − Wall Mount Version 5000145 A01...
Operation & Maintenance manual. 1. Route conduit and electrical lines to wall mount. Use AWG 16 (1,5mm 2. Wire the wall mount as shown in the illustration above (120V version is illustrated). AIM-50™ Installation Manual − Wall Mount Version 5000145 A01...
The wires that go to the light fixture are labeled “to light fixture”. 100V / 230V / 240V versions: Install wall switch per local codes. Wall switch is not furnished with the product. AIM-50™ Installation Manual − Wall Mount Version 5000145 A01...
7. Make sure the screw is fully seated against the shaft on the wall mount (d). The arm should rotate smoothly. 8. Insert the plug cap (b) into the shaft until it locks into position. 9. Fasten the cross-slotted screw (a) to secure the plug cap. AIM-50™ Installation Manual − Wall Mount Version 5000145 A01...
4. Unite the electrical connectors. Figure 3 Figure 4 Make sure the plastic bearing (i) is on the yoke and there is grease in the groove (h). The connectors are not shown in figure 3. AIM-50™ Installation Manual − Wall Mount Version 5000145 A01...
Page 15
The light head should rotate smoothly and not pull out of the arm. Figure 6 10. Thread the brake screw (b) into the collar as shown in figure 6 and tighten it until the rotating joint has the desired friction. AIM-50™ Installation Manual − Wall Mount Version 5000145 A01...
Declaration of acceptance: It is hereby certified that the supporting wall/ceiling and the anchoring of the AIM-50™ suspension system is safe and adequately strong. Project:...
Equipment Anchorage Diagrams Wall mount Version - overview AIM-50™ Installation Manual − Wall Mount Version 5000145 A01...
Page 18
Wall mount Version – plan at wall AIM-50™ Installation Manual − Wall Mount Version 5000145 A01...
Page 20
BURTON MEDICAL 21100 Lassen Street Chatsworth, CA 91311 USA Phone: 800-444-9909, 818-701-8700 Fax: 800-765-1770, 818-701-8725 www.burtonmedical.com The Right Light...