Introduction
®
The GE Telaire
7001 CO
/Temperature monitor (shown in Figure 1
2
below) is an easy to use hand-held instrument, which provides stable and
highly accurate readings due to GE Telaire patented dual beam NDIR
technology. The monitor, equipped with a 0-4 V output, is ideal for long-
term monitoring and recording.
Figure 1: GE Telaire 7001
Startup Procedure
The following procedures describe Battery Operation, Low Battery and
AC Power Operation.
Battery Operation
For portable use, the monitor will operate up to 70 hours on 4 AA
industrial alkaline batteries. The batteries can be installed by removing
the cover from the battery compartment and following the diagram
displayed on the back of the monitor (see Rear View in Figure 1 above).
Low Battery
The low battery signal flashes when less than 30 minutes of battery life
remains. At this point, the batteries should be replaced or the AC adapter
should be used as a substitute. If ordinary operation continues, the
monitor will become inoperable and only the LOW BATT signal will
display on the LCD.
AC Power Operation
The sensor is shipped with a 6V DC, 500 mA AC/DC adapter. To use the
adapter, connect the plug into the back of the monitor and plug the
transformer into any standard wall outlet.
Note:
It is important to use the adapter supplied with the GE
Telaire 7001. Using the wrong adapter may cause
damage to the monitor. Also, the battery operation will
not function as a backup during a power loss.
Temperature Monitor
CO
/
2
®
Telaire
7001 CO
Power-Up Procedure
To power-up the monitor, do as follows:
1.
Press the Power Button and a 2 second delay will occur before the
display becomes visible.
2.
A duration of 10 seconds will elapse before displaying the current
CO
readings.
2
3.
WARM-UP will display for approximately 1 minute. During this
time, adjustments cannot be made to the sensor.
Display Features and Modes
WARM-UP - Indicates a 1 minute warm-up
ON LINE - Indicate when a PC is communicating to the sensor via
RJ45 port
Normal Operating Mode - After warm-up the sensor will stabilize
and display the current conditions
Adjustment Modes
By pressing the mode button, you can scroll through the adjustment
modes. Once the desired mode is displayed, press Enter to make
adjustments. Press Enter again to save and leave the adjustment mode.
The following are the adjustment modes:
ELEVATION - Used to compensate for elevation changes.
CALIBRATION - Used when calibrating.
TEMPERATURE -Used for temperature calibration
OUTSIDE - To manually input CO
rate.
CALIBRATION IN PROGRESS - Displays during calibration.
Display Features
CO
Readings (Upper Display) - Remain visible at all times
2
Temperature and Ventilation Rates (Lower Display) - The Up/
Down arrows allow you to toggle through the Temperature and
Ventilation modes. When pressing the Up arrow, the display will go
through the following sequence:
Temp °C > Temp °F > Vent Rate l/p/s > Vent Rate cfm/p >
Blank
US Standard to Metric Conversion - The Temperature, Ventila-
tion rates, and Elevation Readings can be viewed in US Standard or
Metric Readings. The Temperature is converted from Fahrenheit
(°F) to Celsius (°C); the Ventilation Rates are converted from Cubic
Feet Per Minute Per Person (cfm/p) to Liters Per Second (l/p/s) and
the Feet (ft) to Meters (m).
and Temperature Monitor
2
User Instructions
levels for the CFM ventilation
2
T62292-007
October 2006
Page 1 of 5

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Telaire 7001 and is the answer not in the manual?

Questions and answers

Summary of Contents for GE Telaire 7001

  • Page 1 To power-up the monitor, do as follows: below) is an easy to use hand-held instrument, which provides stable and highly accurate readings due to GE Telaire patented dual beam NDIR Press the Power Button and a 2 second delay will occur before the technology.
  • Page 2 Ventilation Rate (cfm/p) The ventilation rate is defined as the value that represents how much Before using the GE Telaire 7001 monitor, it is important to make the outside air is being introduced on a CFM per person basis. elevation correction outlined below.
  • Page 3: Temperature Adjustment

    View™ Real-Time Graphing/Calibration Software 2080 - If you are calibrating to ambient conditions, make sure the sensor is This software allows the GE Telaire 7001 Monitor to log directly to displaying a stable reading. Avoid breathing in the area of the a Windows compatible PC and graph concentrations in real time monitor.
  • Page 4: Specifications

    Specifications Performance - Temperature Channel The GE Telaire 7001 Monitor meets all the specifications listed below. Temperature Range: Method • Voltage output - 32 to 104 °F (0 to 40 °C) Dual Beam Absorption Infrared™ • Display - 32 to 122 °F (0 to 50 °C)
  • Page 5: Power Supply

    Power Supply Warranty Repairs GE Sensing will repair Telaire product that fails to meet the terms provided for in the Return and Warranty Policy Statement (See, http:// Battery Type www.gesensing.com/service/brochures.htm).Warranty period shall start from date of manufacture and be based on product category and type of equipment as specified in Table 1: Product Warranty Periods.

This manual is also suitable for:

Telaire t7001Telaire t7001sk

Table of Contents