Invacare® Polaris™EX™CPAP with SoftX™
Assembly, Installation and Operating Instructions
Model Number: ISP3000
DO NOT LEAVE THIS GUIDE WITH THE PATIENT
NOTE: Check all parts for shipping damage. In case of damage, DO NOT use. Contact 
carrier/Invacare for further instructions.
To ensure the safe operation of the Invacare Polaris EX CPAP, these instructions 
MUST be followed:
DO NOT operate this equipment without first reading and understanding
this instruction sheet. If you are unable to understand the warnings, cautions
and instructions, contact your healthcare professional, dealer or technical
personnel if applicable before attempting to operate this equipment -
otherwise, injury or damage may occur.
To ensure proper operation of the compliance clock, the Invacare Polaris EX
CPAP should be left plugged in for 12 hours after initial programming.
Control Panel
LCD Display Window
Heated Humidifier
Indicator Light
Heated Humidifier
On/Off Button
BUTTON
ENTER
HEATED HUMIDIFIER
UP
DOWN
For more information regarding Invacare products, parts, and services,
Part No 1114775
Safety Summary
WARNING
CAUTION
SYMBOL
FIGURE 1 Control Panel
please visit www.invacare.com
1
Polaris EX Power
Indicator Light
Enter Button
Up Button
Down Button
INDICATED FUNCTION
ENTER OR SELECTION
HEATED HUMIDIFIER ON/OFF
UP OR INCREASE OR YES
DOWN OR DECREASE OR NO
Invacare® Polaris™EX™CPAP with SoftX™

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Polaris EX ISP3000 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Invacare Polaris EX ISP3000

  • Page 1 - otherwise, injury or damage may occur. CAUTION To ensure proper operation of the compliance clock, the Invacare Polaris EX CPAP should be left plugged in for 12 hours after initial programming. Control Panel...
  • Page 2: Error Code Display

    FIGURE 4. S T A FIGURE 4 NOTE: The LCD window brightness is dimmed when no button is pressed for 30 seconds.  To restore original brightness press any button. NOTE: Heated Humidifier On/Off button is only active when the option is installed. Error Code Display NOTE: For this procedure, refer to FIGURE 5. 1. If an internal component failure is detected, the blower will immediately shut  off. “ERROR CODE” and a 3‐digit number will appear in the LCD window. For  example, if the error code is 532, then “ERROR CODE 532” will be displayed on the  LCD display. 2. Unplug the unit and plug it back in. If the error code still appears in the LCD  window, contact Invacare technical support. E R R 5 3 2 FIGURE 5 Error Code Display Invacare® Polaris™EX™CPAP with SoftX™ Part No 1114775...
  • Page 3: Programming Instructions

    2. “CPAP PRESSURE” will appear on the LCD display window and the current  CPAP setting will be highlighted. 3. Press the Up ( ) or Down ( ) buttons to increase or decrease CPAP setting by  one cmH 4. Press the Enter button to confirm the CPAP setting. Fine Pressure Adjustment NOTE: The Fine Pressure Adjustment feature allows the clinician to adjust the therapy  pressures in intervals of 0.1 cmH O over a range of ‐1 cmH O to +1 cmH O. The default  setting for this feature is 0.0 cmH O and will default to this setting if the CPAP therapy  pressures are changed. 1. Enable the Clinician Set‐up mode (if not already in Clinician Set‐up mode).  Refer to Enabling the Clinician Set‐Up Mode. 2. Repeatedly press the Enter button until the display shows “Fine Press Adj”. 3. Press the Up ( ) or Down ( ) buttons to increase or decrease the fine pressure  adjustment in intervals of 0.1 cmH 4. Press the Enter button to set the selection. Part No 1114775 Invacare® Polaris™EX™CPAP with SoftX™...
  • Page 4 5. Press the Up ( ) or Down ( ) button to increase or decrease the RAMP start  pressure as desired. 6. Press the Enter button to confirm the RAMP start pressure. Adjusting SoftX (Exhalation Unloading) Setting or Change to Standard CPAP 1. Enable the Clinician Set‐up mode (if not already in Clinician Set‐up mode).  Refer to Enabling the Clinician Set‐Up Mode on page 3. 2. Press the Enter button until SoftX (exhalation unloading) setting is displayed  on the LCD display window. 3. Press the Up ( ) or Down ( ) button to scroll through the SoftX settings. 4. Press the Enter button to set the SoftX setting. SoftX adjustment may also be  made via the User mode. Refer to Enabling The User Settings Mode on page 7. Invacare® Polaris™EX™CPAP with SoftX™ Part No 1114775...
  • Page 5: Setting The Current Date

    NOTE: There is no AM or PM adjustment. The hour value is based on a 24 hr. clock, e.g.,  6:33 PM is 18:33. 4. Press the Enter button to set the hour (HH) value to the displayed value. The  minute (MM) value will flash. 5. Repeat STEPS 3 and 4 to set the minute (MM) value. 6. A confirmation screen will appear. 7. Press the Up ( ) button to update the unit’s real time clock with the new date  and time. The compliance start date will not be affected. Clearing the Compliance Data 1. Enable the Clinician Set‐up mode (if not already in Clinician Set‐up mode).  Refer to Enabling the Clinician Set‐Up Mode on page 3. 2. Press the Enter button until the display shows the “CLEAR COMPLIANCE?”. 3. If the Up ( ) button is pressed, all compliance data will be cleared and the  following message will be displayed for two seconds, “COMPLIANCE DATA  ERASED!”. 4. To retain the compliance data, press the Down ( ) button. The following  message will be displayed for two seconds, “COMPLIANCE DATA  MAINTAINED!”. Part No 1114775 Invacare® Polaris™EX™CPAP with SoftX™...
  • Page 6 1. Ensure “Standby” appears in the LCD display window. 2. Press the Up ( ) and Down ( ) buttons at the same time for four seconds or  longer to start the Clinician mode. 3. A menu will appear with access to the Clinician Set‐up mode and Full  Compliance mode. 4. Access Full Compliance mode by pressing the Down ( ) button. NOTE: If no button is pressed for two minutes or more, the unit automatically exits the  Full Compliance mode. No changes will be saved. Viewing the Full Compliance Data While in Full Compliance mode, press the Enter button to cycle through the  following data: • Average Hours (last five days) • Percent Compliance (last five days) • Total Days (calendar) • Total (Compliance) Hours (accumulated) • Average Hours in use per day • Percentage of Compliance Days Invacare® Polaris™EX™CPAP with SoftX™ Part No 1114775...
  • Page 7 Exiting the Full Compliance Mode At any time, press and hold the Up ( ) and Down ( ) buttons at the same time  for approximately two seconds to exit Full Compliance mode. NOTE: If no button is pressed for two minutes or more, the unit automatically exits Full  Compliance mode and returns to Standby mode. No changes will be saved. Enabling The User Settings Mode 1. While in Standby mode, press and hold either the Up ( ) or Down ( ) button  for approximately four seconds or longer. A menu will appear with access to  the User Settings mode and the Quick Code mode. 2. Press the Up ( ) button until the software version number is displayed to  access the User Settings mode.  3. Press the Enter button to cycle through the Set‐up mode settings. 4. Press the Up ( ) and Down ( ) buttons to change settings. 5. After a change is made, press the Enter button to confirm the change and  proceed to the next setting. NOTE: If no button is pressed for two minutes or more, the unit automatically exits to  Standby mode. No changes will be saved. Part No 1114775 Invacare® Polaris™EX™CPAP with SoftX™...
  • Page 8 Adjusting the SoftX (Exhalation Unloading) Setting or Change to Standard CPAP 1. Enable the User Settings mode. Refer to Enabling The User Settings Mode on  page 7. 2. Press the Enter button until the SoftX setting is displayed. The current SoftX  setting will be highlighted. 3. Press the Up ( ) or Down ( ) buttons to scroll through the SoftX settings. NOTE: The unit can be set to Standard CPAP (without SoftX). Standard CPAP is an  option within the SoftX settings. 4. Press the Enter button to accept the new SoftX setting. Exiting the User Settings Mode 1. If any settings have been changed, a confirmation screen will appear asking if  the changes should be saved. To save the changes, press the Up ( ) button. The  following message will be displayed for two seconds, “CHANGES  CONFIRMED!”. 2. To discard the changes, press the Down ( ) button. The following message  will be displayed for two seconds, “CHANGES IGNORED!”.  Invacare® Polaris™EX™CPAP with SoftX™ Part No 1114775...
  • Page 9 7 1 0 0 1 FIGURE 7 Percentage of Compliance 3. Total Usage Days will be displayed with the resolution of whole day such as  “0014”. Total Usage Days will be displayed in four digits. If there are less than  four digits (e.g., 67), Total Usage Days will be displayed with leading zero(s)  (e.g., 0067, or 0009). Refer to FIGURE 8. A R T : D D / Y Y 7 1 0 0 1 FIGURE 8 Total Usage Days NOTE: The maximum Total Days is approximately 2740 days (which equates to 7.5 years). Part No 1114775 Invacare® Polaris™EX™CPAP with SoftX™...
  • Page 10 2. To enable the heated humidifier, press and release the Heated Humidifier  ON/OFF ( ) button. The Heated Humidifier Indicator light will illuminate,  and “HEAT” will be displayed in the lower left‐hand corner of the LCD display. • If the operating mode is Standby, enter Therapy or Ramp modes to enable  heated humidifier. Otherwise, the heated humidifier will preheat the water  for 20 minutes and then disable. • The CPAP must be in the Therapy or Ramp operating mode for the heater  humidifier to be continuously enabled, with “HEAT” being displayed on  the LCD display. Disabling the Heated Humidifier To disable the humidifier heater, press and release the Heated Humidifier ON/OFF  ) button. Enabling the Heated Humidifier Set-up Mode 1. The Set‐up mode for the humidifier heater can be entered at any time during  the Standby, Therapy, or Ramp mode. To enter, press the Heated Humidifier  ON/OFF ( ) button for approximately four seconds or longer, until the  “TEMP. SETTING” appears. NOTE: The default temperature setting is 105° F. Invacare® Polaris™EX™CPAP with SoftX™ Part No 1114775...
  • Page 11: Completing The Setup

    Exiting the Set-up Mode Press the Enter button to accept the humidifier heater plate setting and return to  the previous operating mode. The humidifier heater will be enabled with HEAT  displayed on the LCD display. • If the previous operating mode was Standby, then the humidifier heater will  preheat the water for 20 minutes. If either Therapy or Ramp mode is not  entered within this time, the humidifier heater will become disabled, and  “HEAT” will not be displayed on the LCD display. • If the previous operating mode was Therapy, the humidifier heater will be  enabled with “HEAT” being displayed on the LCD display. • If the previous operating mode was Ramp, the ramp time will start from the  beginning and the humidifier heater will be enabled with “HEAT” being  displayed on the LCD display. NOTE: If the Set‐up mode for the humidifier heater is enabled and no button is pressed for  30 seconds or more, the unit will automatically exit to the previous operating mode and no  changes will be saved. Completing the Setup Invacare recommends the following necessary equipment for use with the Polaris  ™ ® ™ EX CPAP system: Invacare Twilight  Mask and SleepNet  IQ  Gel Mask, 6 ft.  smooth inner bore tubing (22 mm ID). Part No 1114775 Invacare® Polaris™EX™CPAP with SoftX™...
  • Page 12 ™ and ®. Elyria, Ohio USA Sanford, FL USA All trademarks are owned by or 44036-2125 32773 licensed to Invacare Corporation 800-333-6900 800-832-4707 unless otherwise noted. Sleepnet is a registered trademark of Innovative Medical Systems. Inc.

Table of Contents