The frequency of this flashing light is in the range of sensitivity for those prone to epileptic seizures. For your safety when using the Aeros you should pay attention to the following types of statements in this User’s Manual: §...
The HunterLab Aeros Colorimetric Spectrophotometer is designed for precision color and appearance measurement. It measures numerical color and related data in absolute and relative terms. HunterLab cannot guarantee the accuracy, completeness, efficacy, and timeliness of the data due to inherent uncertainties in instrumental readings, variations in sample presentation, and potential inconsistencies in human color perception.
Page 4
Aeros Supplemental Manual for EasyMatch QC ver. 2.2...
Aeros Supplemental Manual for EasyMatch QC ver. 2.2 Contents PREFACE ........................2 Copyrights and Trademarks ......................2 Legal Disclaimers: Instrumental – Visual Evaluation ................. 3 Disclaimer of Liability: Utilization of Data, Metadata, and Information ..........3 CONTENTS ........................5 SETTING UP THE AEROS ....................9 Standard Accessories ........................
Page 6
Aeros Supplemental Manual for EasyMatch QC ver. 2.2 TAKE SAMPLE OR STANDARD MEASUREMENT ............23 Selecting Measurement Parameters ....................23 Sensor Manager > Measurement Configuration ................23 Measure Sample/Standard ......................24 Measurements > Read Standard ....................24 Measurements > Read Sample ...................... 25 Visual Inspection ...........................
Page 7
Aeros Supplemental Manual for EasyMatch QC ver. 2.2 90mm Plastic Petri Dish, 10-Pack (L02-1016-781) ................37 Sample Trays ............................37 WHEN YOU NEED ASSISTANCE ................... 39 INDEX ........................41...
Page 8
Aeros Supplemental Manual for EasyMatch QC ver. 2.2...
The Aeros is a continuous, automated, non-contact color measurement instrument with height detection and color compensation. For use in laboratory environments, the Aeros uses LED illumination technology to measure samples and monitor color on a rotating tray. Pour, place and measure single or multiple samples on the tray, with easy cleanup and maintenance.
Figure 2. Calibration Box Selecting a Space for the Aeros The following illustrates a successful installation. Set up the HunterLab Aeros in a laboratory setting with controlled, consistent temperature and humidity. It is recommended that access to the rear connectors be maintained. The selected workspace should be free of drafts and characterized by proper room lighting.
For more information, please refer to SPECIFICATIONS. Cleaning the Aeros Clean the outside surfaces of the Aeros using a soft cloth. Do not spray liquids directly on the instrument. . Care should be taken to avoid degradation of optical surfaces. Refer to MAINTENANCE...
Page 12
Aeros Supplemental Manual for EasyMatch QC ver. 2.2...
Front and Rear USB Connectors There are two USB connectors on the Aeros. The one in the front is typically used for exporting jobs and workspaces, backing up the instrument and updating software. The USB port on the back of the...
Aeros Installation Note: The Aeros should be lifted from under the base plate, near the center of the unit. It should not be carried by grasping any part of the plastic housing. The Aeros is simple to set up and attach to your computer. The following instructions guide you through the initial installation of your Aeros system.
Aeros Supplemental Manual for EasyMatch QC ver. 2.2 Aeros Options and Sample Devices For the latest information, please refer to support.hunterlab.com and Aeros.
Page 16
Aeros Supplemental Manual for EasyMatch QC ver. 2.2...
Aeros Supplemental Manual for EasyMatch QC ver. 2.2 Aeros Installation with EasyMatch QC Install EasyMatch® QC Software Complete the following steps: Log into the system using an account that has ADMINISTRATOR privileges for the PC — network or local. Insert the installation CD into the CD-ROM drive. If the system is setup to automatically run CD programs, the menu will appear and you may skip to Step 5.
Figure 8. No License Warning The SoftKey License is uniquely associated with the sensor serial number and is provided on a thumb drive supplied with EasyMatch QC or via email from HunterLab. Go to HELP > LICENSE REGISTRATION > ACTIVATION.
Page 19
Option #3: Sentinel Key iii. a. If the user has a HunterLab USB hardware key, then it can be used with a new sensor on the same computer. Return to Install the Software, Step 5 (Figure 6) and select the Sentinel Key to continue.
Open EasyMatch QC in the computer. From the SENSOR MENU, select ADD SENSOR. Figure 11. Add Aeros Sensor Select Aeros as the Sensor Type and use the USB (selected as default) to connect Aeros with EasyMatch QC. Click NEXT to search available Aeros.
QC with the Aeros. Sensor Manager > Standardization The Aeros must be standardized on a regular basis to keep it operating properly. Standardization sets the top and bottom of the photometric scale. During standardization, the bottom of the scale is set using a black tile read on the Aeros.
The CREATE SAMPLE PROFILE button will allow the Aeros to look at the specimen and select the optimum fixed distance from the turntable. When this is pressed, the Sensor will move to the Top position.
Aeros Supplemental Manual for EasyMatch QC ver. 2.2 Figure 18. Sensor > Measurement Configuration Measure Sample/Standard From the MEASUREMENTS MENU, you may read either a standard or a sample. You may also average data. The functions available through the MEASUREMENTS MENU are described in the remainder of this chapter.
Aeros Supplemental Manual for EasyMatch QC ver. 2.2 Measurements > Read Sample The MEASUREMENTS > READ SAMPLE command initiates reading of a sample. Place the sample before initiating the READ SAMPLE command. If Prompt for sample name was checked in OPTIONS > NAMING CONVENTIONS, you will then be prompted to ENTER A NAME for the sample (as well as a Product ID and Extra ID, if you wish) and to choose the standard with which the sample should be linked.
Page 26
Aeros Supplemental Manual for EasyMatch QC ver. 2.2...
Aeros Supplemental Manual for EasyMatch QC ver. 2.2 Aeros Maintenance and Testing The Aeros is engineered to be virtually maintenance free. This section outlines the few parts of the sensor that are to be maintained for the instrument to function properly.
HunterLab Order Processing Department, or contact your local HunterLab representative to order a replacement Diagnostics on the Aeros Diagnostics are included with the Aeros EasyMatch QC. To initiate, select DIAGNOSTICS from the SENSOR MENU. THE GREEN TILE TEST, REPEATABILITY TEST and HARDWARE CHECKS are the available tests.
Page 29
Attach the green tile and press NEXT to continue. Ten readings of the green tile are made and averaged. The average reading and its Pass/Fail status will be shown. Confirm that the result is PASS. The Pass/Fail tolerance for the Aeros is ± 0.30.
Page 30
Aeros Supplemental Manual for EasyMatch QC ver. 2.2 Figure 26. Screen Display for Test Results Figure 27. Green Tile Test Results...
Aeros Supplemental Manual for EasyMatch QC ver. 2.2 Repeatability Test The repeatability test is initiated by standardizing and leaving the white tile in view. Each white tile reading is reported along with a pass/fail evaluation. Remove all samples from the instrument and attach the standardization box. The sensor will move to the top.
Page 32
Aeros Supplemental Manual for EasyMatch QC ver. 2.2 Figure 31. Repeatability Readings When the test is complete, results are shown and output as a report. Figure 32. Repeatability Screen Results...
The operating conditions (temperature and humidity ranges) are given in the Operating Conditions section below. Note: Do not leave Aeros in an area where temperature or humidity extremes are possible. Operating Conditions Storage Temperature (3weeks) -20°C to 65°C (-5°F to 150°F)
Aeros Supplemental Manual for EasyMatch QC ver. 2.2 System Power 100 – 240 VAC, 47 – 63 Hz to universal power supply @ 24 VDC/3.75A 90W Conditions Of Illumination And Viewing Full spectrum balanced LED system array; 5 year typical LED...
Aeros Supplemental Manual for EasyMatch QC ver. 2.2 Measurement Color Data, Spectral Plot, EZ View, Color Plot, Trend Plot, Spectral Data Views Data Illuminants A, C, D50, D55, D65, D75, F02, F07, F11, TL84, ULT30, ULT35 Observers 2° and 10°...
Aeros Supplemental Manual for EasyMatch QC ver. 2.2 Regulatory Notice Aeros Sample Accessories HunterLab Flash Drive (A10-1013-423) 2 GB USB 2.0 flash drive formatted for customer use with customized code for Setup backup and CSV Datalog export. USB Flexible Keyboard (A13-1014-294) This 88 character keyboard allows the user to enter IDs directly into the instrument.
Aeros Supplemental Manual for EasyMatch QC ver. 2.2 Sample Holder, Offset, 3.5" Dish (D02-1016-773) Black Anodized Aluminum Sample Dish for use with offset 3.5" petri dish holder to measure spice blends/seasoning mixtures/powder. Sample Holder 145mm and 90mm Petri Dishes (D04-1018-680) Sample Holder provides easy and repeatable positioning of 145 mm and 90 mm Petri dishes.
Page 38
Aeros Supplemental Manual for EasyMatch QC ver. 2.2...
Additionally, our global support website offers 24/7 assistance with a library of information on various color measurement and appearance topics such as applications, instrument operation, and troubleshooting. The HunterLab global support website is located at support.hunterlab.com. For personalized assistance, go to support.hunterlab.com...