Bridge CMMS GLOBAL S 05.YY.05, 07.YY.07, 09.YY.08, 12.YY.10 AND 15.YY.10 USER MANUAL H00007842 ISO 14001 ISO 9001 VDA 6.4 H00007841 Rev00 ...
Page 2
Hexagon Manufacturing Intelligence declines all liability for any errors that may appear in the manual.
Reading the PDF Files ....................vi Conventions ............................ vii Safety Information – the Customer’s Responsibilities ..............vii Overview..........................8 The GLOBAL S Measuring Machine ....................9 Messaging Lights ....................... 11 Thermal Compensation System ................11 PULSE ........................11 Intended Use and Unauthorized Use ..................... 12 Main Components ..........................
Page 4
Anti-Tilt Feet and Anti-Vibration Supports ..............78 Steel Wire of the Counterbalancing Cylinder ............. 79 Emergency Buttons and Safety Devices ..............79 Air Supply System ..................... 80 Consumables and Spare Parts ...................... 82 Declaration of Conformity ....................83 iv • GLOBAL S - User Manual...
Preface This manual was written for users of the Hexagon Manufacturing Intelligence (Hexagon in the rest of this manual) GLOBAL S measuring machines. The document provides a general description of the measuring machine and its main technical and design features. The manual also contains instructions on how to use and maintain the measuring machine safely.
Use and Maintenance of the Measuring System – General Safety Regulations This document was written for owners of a Hexagon measuring system; it contains the general safety regulations that users and preventive maintenance engineers must know and follow in order to use and maintain the measuring system safely.
Conventions The following conventions are used in this manual to distinguish between various kinds of information. Danger Indicates a hazardous situation that, if not avoided, will result in death or serious injury. Warning Indicates a hazardous situation that, if not avoided, could result in death or serious injury.
Overview This chapter provides a brief description of the GLOBAL S measuring machine and its main components and its versions. 8 • GLOBAL S - User Manual...
The GLOBAL S Measuring Machine GLOBAL S is a family of measuring machines with Cartesian axes, a gantry architecture and a vertical Z- rail. The machine is available in several versions that differ from one another in precision and performance.
Page 10
GLOBAL S Measuring Machine models Table 1 shows the models of the GLOBAL S measuring machine. Version Model GREEN 05.yy.05 07.yy.07 09.yy.08 12.yy.10 BLUE 05.yy.05 07.yy.07 09.yy.08 12.yy.10 15.yy.10 CHROME 05.yy.05 07.yy.07 09.yy.08 12.yy.10 15.yy.10 Table 1 GLOBAL S Measuring Machine models...
Messaging Lights The Messaging Lights LED indicator kit (optional) can be used to monitor machine state at a distance from the measuring station. Thermal Compensation System The thermal compensation system is of the multisensor type. The temperature sensors are positioned inside the measuring machine on the axes, table and beam. In addition, the user is provided with an external sensor that is to be applied to the workpiece to be measured when the thermal compensation system is active.
Hexagon. It is not to be used, for example, to move or lift parts, to mill or to measure parts of the human body.
Page 13
Figure 1 The GLOBAL S Measuring Machine 5.yy.05 and 7.yy.07 Models Central carriage Anti-vibration support Main carriage (beam) Machine stand Main carriage (right leg) Granite work table Pneumatic control unit Z-rail Y axis bellow Main carriage (left leg) Optional Laser scanner...
Page 14
Figure 2 The GLOBAL S Measuring Machine 09.yy.08, 12.yy.10 and 15.yy.10 Models Central carriage Optional Laser scanner Main carriage (beam) Optional Machine skirt Main carriage (right leg) Granite work table Pneumatic control unit Z-rail (Column) Y axis bellow Main carriage (left leg) Anti-vibration support pedestal 14 •...
Protective Systems Controlling Access to the Danger Zone Figure 2 shows the measuring system without any guards covering the area in which the measuring machine operates (danger zone). As a reminder, when installing the machine and when using the machine, the Customer must fit the measuring machine with an appropriate protective system capable of meeting the specific application and safety requirements and, in particular, those that emerge from the risk analysis for the specific installation.
Figure 3 Laser scanner system Laser scanner The dimensions of the protected areas can be programmed during installation (by Hexagon) and must not be smaller than those shown on the following table. 16 • GLOBAL S - User Manual...
Page 17
Front side Left side Right side Rear side Df mm(in) Dl mm(in) Dr mm(in) Db mm(in) Model External Internal External Internal External Internal External Internal area area area area area area area area 1000 05.yy.05 (36.8) (23.4) (39.4) (26.1) (34.2) (21.1) (37.6) (24.4)
The Jogbox is a device connected to the control system by means of which the user can perform operations on the measuring system standing close to the machine. 0° 15° 35° 25° Figure 5 Jogbox NJB stand position Screw Stand 18 • GLOBAL S - User Manual...
Page 19
Figure 6 NJB Jogbox keypad Joystick Wire connector Keypad Speed adjuster Machine start Enable button Emergency stop Move the joystick to move the linear axes X, Y and Z of the measuring machine. The controls on the Jogbox may be used to: ...
Page 20
The button must be kept pressed when the JogBox is used to move the axes. If the joystick is moved without holding this button pressed, the axes will not move (this prevents unwanted accidental movement of machine axes). 20 • GLOBAL S - User Manual...
Page 21
Slow The button is used to set the machine axis maximum speed when movement is controlled by Jogbox. Two values can be selected (fastJogVel and slowJogVel, with fastJogVel>slowJogVel). These are configured in the control system. When the control system is in Manual status, the maximum speed of the machine axes is set to the lower of the values between the one set using the SLOW pushbutton and the manualVel setup parameter.
Page 22
The button switches the control system between the Run and Hold states. The LED shows the current setting: LED off. The control system suspends axis movement (hold). LED on. The control system allows axis movement (run). Default setting at control system start-up 22 • GLOBAL S - User Manual...
Workstation The workstation is a device connected to the control system by means of which the user can perform operations on the measuring system standing close to the machine. Warning To prevent the risk of physical injury: Do not sit or stand on workstation. Do not lean on workstation.
Page 24
Probe Accessory Drawer for storage. Lift column keypad can be used by the Operator to raise and lower the Operator workspace. An optional extension table is available for use with the workstation. 24 • GLOBAL S - User Manual...
Page 25
Program up to four memory positions available on the lift column keypad Caution Make sure no obstacles are in the workstations path. Make sure the workstations is not touching any walls or equipment. Make sure all cords are long enough to accommodate the change in height. Once a preset button is pushed, the desk will move to the programmed height.
Figure 11 FDCPanel settings page On this page, click on the selection box to enable/disable the following measuring machine functions: Quadratic relation jogbox Eco Mode / Eco Mode + Auto retract 26 • GLOBAL S - User Manual...
Page 27
Quadratic relation jogbox Enables/disables the quadratic relation between joystick movement and controlled speed. When enabled, this function modifies the joystick sensitivity to allow a more accurate positioning in low speed movements. When deactivated, the joystick is more reactive in high speed movements on long distances. The user can select the required mode according to the most frequency method of use of the joystick (small accurate movements or fast movements on long distances).
Page 28
Auto retract Enables/disables automatic movement performed after taking a point in jog mode. This automatic movement, if enabled, moves the probe away from the part to the distance set in the “Distance” field. 28 • GLOBAL S - User Manual...
Operating Instructions This chapter contains the instructions to be followed to ensure proper, safe, and efficient use of the measuring machine. Operating Instructions • 29...
Piercing of the eyes by a When sharp tools are present, always wear special protective glasses. sharp tool. 30 • GLOBAL S - User Manual...
Page 31
Make all the necessary checks during preventive maintenance: see the “Maintenance” chapter. Contact the Hexagon Customer Service immediately if any errors appear, signalling faults in the mode selector, power circuit, reductions in speed or safety functions in general. ...
+10° (50° F) - +40°C (104° F) Relative humidity >20% <90% non-condensing (*)15.yy.10 Models (**) Optional Table 3 Operating Conditions Notice With machine in standstill the air consumption of the active pneumatic anti- vibration supports (optional) is negligible. 32 • GLOBAL S - User Manual...
Hexagon procedure are strictly respected. See the Technical Data bulletin for the model of the GLOBAL S measuring machine in use to find out the measuring performance that may be obtained when specific conditions for maximum measuring precision and the experimental conditions are satisfied.
Complete the measuring system shutdown procedure by following the instructions given in the user manuals of the control system and the measuring software. Close the shut-off valve of the pneumatic control unit to avoid wasting air and ensure optimum safety conditions. 34 • GLOBAL S - User Manual...
Moving the Machine Axes The machine axes are to be moved by running a part program or using the joystick of the NJB terminal. Caution Avoid moving the machine axes by pushing on their moving parts. Caution Run the AUTOZERO procedure to restore correct Jogbox operation if the NJB terminal is knocked or dropped or if axis position holding instability is found after pressing the ‘’enable’’...
Page 36
AUTOZEROs are required, it is a sign of a defective joystick or of a NJB damaged by a strong impact, drop or abusive operation and the NJB should be replaced. 36 • GLOBAL S - User Manual...
Part weight compensation The GLOBAL S measuring machine applies automatic corrections to the performed measurements to compensate for geometric changes induced by the loads of parts onto the table. To allow part weight compensation to work properly, the user is required to define some information relevant to the specific measurement setup: ...
For each model of measuring machine, this chapter indicates: Lengths of the machine axis strokes Overall dimensions Weight of the measuring machine Maximum weight of the part Dimensions and drilling pattern of the work table 38 • GLOBAL S - User Manual...
Overall Dimensions and Mechanical Characteristics of the Measuring Machine Figure 15 Overall Dimensions of the Machine - 05.yy.05 Models Notice Figure 15 measurements shown in mm only. Overall Dimensions and Mechanical Characteristics • 39...
Page 40
(5.5) (27.2) (0.8) (500) 1190 05.07.05 (1378) (46.9) (6.2) (34.1) (31.5) (5.2) (27.2) (1.0) (500) Table 7 Dimensions of Work Table, Distances between Supporting Points, Position of Inserts and Weights - 05.yy.05 Models 40 • GLOBAL S - User Manual...
Page 41
Figure 16 Overall Dimensions of the Machine - 07.yy.07 Models Notice Figure 16 measurements shown in mm only. Overall Dimensions and Mechanical Characteristics • 41...
Page 42
(lb) kg (lb) 1650 1255 07.10.07 (27.6) (65.0) (7.9) (16.9) (37.4) (2.0) (1984) (2767) Table 9 Dimensions of Work Table, Distances between Supporting Points, Position of Inserts and Weights - 07.yy.07 Models 42 • GLOBAL S - User Manual...
Page 43
Figure 17 Overall Dimensions of the Machine - 09.yy.08 Models (with DC241 Controller) Notice Figure 17 measurements shown in mm only. Overall Dimensions and Mechanical Characteristics • 43...
Page 44
1350 1800 2270 09.20.08 (27.6) (106.7) (9.1) (26.7) (53.1) (5.0) (3968) (5004) Table 11 Dimensions of Work Table, Distances between Supporting Points, Position of Inserts and Weights - 09.yy.08 Models (with DC241 Controller) 44 • GLOBAL S - User Manual...
Page 45
Figure 18 Overall Dimensions of the Machine - 09.yy.08 Models (with DC800C or DC800 I/O-Ready Controller) Notice Figure 18 measurements shown in mm only. Overall Dimensions and Mechanical Characteristics • 45...
Page 46
09.20.08 (28.5) (111.4) (10.8) (27.6) (57.1) (5.0) (3968) (7275) Table 13 Dimensions of Work Table, Distances between Supporting Points, Position of Inserts and Weights - 09.yy.08 Models (with DC800C or DC800 I/O-Ready Controller) 46 • GLOBAL S - User Manual...
Page 47
Figure 19 Overall Dimensions of the Machine - 12.yy.10 Models Notice Figure 19 measurements shown in mm only. Overall Dimensions and Mechanical Characteristics • 47...
Page 48
(12566) 3980 1000 1980 2250 7600 12.30.10 (24.6) (156.7) (15.0) (39.4) (78.0) (0.9) (4960) (16755) Table 15 Dimensions of Work Table, Distances between Supporting Points, Position of Inserts and Weights - 12.yy.10 Models 48 • GLOBAL S - User Manual...
Page 49
Figure 20 Overall Dimensions of the Machine – 15.yy.10 Models Notice Figure 20 measurements shown in mm only. Overall Dimensions and Mechanical Characteristics • 49...
Page 50
(14617) 15.30.10 3980 1000 1980 2250 8865 (24.6) (156.7) (13.8) (39.4) (78.0) (0.9) (4960) (19544) Table 17 Dimensions of Work Table, Distances between Supporting Points, Positions of Inserts and Weights – 15.yy.10 Models 50 • GLOBAL S - User Manual...
Work Table Material Granite Flatness Compliant to DIN 876/III Part fixing M8 x 1.25 threaded inserts. Thread length 20 mm (0.8 in). Drilling pattern, see the overall view drawings for the specific model in this chapter. Column Counterbalancing System Type Pneumatic and adjustable.
Description This chapter provides a general description of the measuring machine and its components, and was written to help the operator to perform the operating and maintenance operations. The chapter discusses the following topics: Structure and main components of the machine ...
The optical scale of the Z axis is applied to the Z-rail. At the bottom end of the Z-rail there is the flange for attaching the measuring head. 54 • GLOBAL S - User Manual...
Page 55
Figure 22 Components of the Measuring Machine 5.yy.05 and 7.yy.07 Models Z-rail Anti-vibration support Z-tower Machine stand Central carriage Granite Main carriage (beam) Anti-vibration support Main carriage (right leg) Work table Pneumatic control unit Main carriage (left leg) Description • 55...
Page 56
Figure 23 Components of the Measuring Machine 09.yy.08, 12.yy.10 and 15.yy.10 Models Z-rail Anti-vibration support pedestal Z-tower Optional Laser scanner Central carriage Granite Main carriage (beam) Work table Main carriage (right leg) Main carriage (left leg) Pneumatic control unit 56 • GLOBAL S - User Manual...
Axis Position Transduction System Each axis is equipped with a linear optical transducer consisting of an optical scale and a position reader, which micrometrically detects the position of the axis and reads the movement. When an axis moves, the position reader generates electrical pulses which it sends to the control system. The control system counts the pulses and converts them into the corresponding axis movement.
Page 58
Figure 24 Axis Position Transduction System Optical reader (Z axis) Optical scale (X axis) Optical scale (Z axis) Optical scale (Y axis) Optical reader (X axis) Optical reader (Y axis) 58 • GLOBAL S - User Manual...
Axis Driving System The movements of each axis are executed by a cog belt system driven by a direct current motor. The moving part of the axis is fixed, by means of a drive unit, to the closed-loop cog belt attached to two pulleys (one drive and one return pulley).
Page 60
Driving belt (Y axis) – Reduction unit (Y axis) – 09.yy.08, 12.yy.10, 15.yy.10 Models 05.yy.05, 07.yy.07 Models Reduction unit (Y axis) – Driving belt (X axis) 09.yy.08, 12.yy.10,15.yy.10 Models Reduction unit (X axis) 60 • GLOBAL S - User Manual...
Axis Sliding System The axis sliding system consists of air bearings that ensure the support and friction–free movement of the moving parts of the axes (main carriage, central carriage and Z-rail). The moving parts are supported by a flow of compressed air, which comes from the pneumatic control unit. The flow supplies the bearings so as to form an air cushion between the bearings and the guideways.
Page 62
Air bearing (X axis) – Air bearing on left leg (Y axis) – 07.yy.07 Models 05.yy.05 Models Air bearing (Z axis) – Air bearing on left leg (Y axis) – 07.yy.07 Models 07.yy.07 Models 62 • GLOBAL S - User Manual...
Page 63
Figure 28 Axis Sliding System 09.yy.08, 12.yy.10, 15.yy.10 Models Air bearing on right leg (Y axis) – Air bearing (X axis) – 09.yy.08, 12.yy.10, 15.yy.10 Models 09.yy.08, 12.yy.10, 15.yy.10 Models Air bearing on left leg (Y axis) – Air bearing (Z axis) – 09.yy.08, 12.yy.10, 15.yy.10 Models 09.yy.08, 12.yy.10, 15.yy.10 Models Description •...
The air supply system consists of the pneumatic control unit and the supply circuits and its function is to: supply the machine axis air bearing system. supply the pneumatic cylinder for counterbalancing the weight of the Z-rail. Figure 29 Air Supply System Diagram 64 • GLOBAL S - User Manual...
Pneumatic Control Unit The pneumatic control unit filters, regulates and distributes compressed air to the circuits supplying the various appliances. The control unit is situated at the rear of the supporting bench. The filtering unit consists of a pair of self-cleaning filters (primary and secondary filter). There is also a pressure gauge and a manual pressure regulator for measuring and regulating the air pressure at the input of the air bearing supply circuit, respectively.
The knob can be accessed by removing the right side guard of the secondary carriage (Figure 31). Figure 31 Z-rail Weight Counterbalancing System Cylinder 05.yy.05, 07.yy.07 Models Steel wire Regulating knob - Z-rail 09.yy.08, 12.yy.10, 15.yy.10 Models Regulating knob – 66 • GLOBAL S - User Manual...
Page 67
The pneumatic circuit for counterbalancing the Z-rail is completed by the pressure switch S1, which is situated below the pressure regulator and measures the pressure at the input of the counterbalancing cylinder. The pressure switch S1 gives its consent to open the solenoid valve that controls the inlet of air to the air bearing supply circuit, only if the precision value inside the counterbalancing cylinder lies within the range set.
Maintenance This chapter provides a description of the preventive maintenance procedures indicated by Hexagon to ensure the measuring machine works in a safe and continuous fashion. Before proceeding to carry out any maintenance work, the maintenance staff must be familiar with all the contents of this chapter.
General Maintenance Information The GLOBAL S measuring machines were designed and built in such a way as to ensure reliable operation and minimum maintenance requirements. The maintenance operations must in any case be performed by trained engineers, i.e. by persons with technical know-how or experience, or who have received specific instructions about the maintenance operations to be performed.
Also turn off the air supply except when it must be possible to move the moving parts of the machine, so as to avoid unexpected movements. 70 • GLOBAL S - User Manual...
Accessing Components Covered by Guards Where necessary, before carrying out maintenance operations with the machine open, remove the guards covering the components that are to be repaired so that the work can be done comfortably and safely. Warning The machine must be turned off and the air supply must also be turned off before guards may be removed and maintenance work may be carried out.
Page 72
Figure 33 Fixed Guards – 07.yy.07 Models 72 • GLOBAL S - User Manual...
To remove the fixed guards of the machine, simply remove the screws that secure these components to the machine structure. According to version and model, the GLOBAL S measuring machine may be provided with side skirts. Having completed the maintenance work, reassemble the guards using all the screws and washers removed during disassembly.
(the use of other products may cause damage to the paintwork). Prevent the detergent liquid from coming into contact with the other components of the measuring machine (for example, guideways, optical scales and racks). 76 • GLOBAL S - User Manual...
Carefully inspect and grease the steel wire of the counterbalancing cylinder. Call the Hexagon customer service if there evident signs of wear are found. Check the axis reduction and driving belts. Call the Hexagon customer service if there are evident signs of wear. Caution Never attempt to service components of the drive system.
(see “Overall Dimensions and Mechanical Characteristics of the Measuring Machine” on page 39). If any faults are found, stop the measuring system and call the Hexagon customer service immediately. Caution If any faults in the adjustable feet or the anti-vibration supports that could jeopardize the stability of the machine are found, stop the measuring system immediately.
Use KLÜBER MICROLUBE GL 261 grease or an equivalent class NGLI 1 grease (ISO 3448). Warning If the steel wire shows signs of wear, call the Hexagon customer service immediately to have it replaced. Do not use the measuring machine until the wire has been replaced.
This section contains the instructions for the preventive maintenance of the air supply system of the measuring machine. Figure 37 Pneumatic central unit Pressure regulator Primary filter Air pressure gauge On-off valve Secondary filter Air inlet 80 • GLOBAL S - User Manual...
Page 81
Checking and Regulating the Pressure of the Air Supply System (Air Pressure on Air Bearings) With the frequency indicated in the maintenance schedule and after carrying out maintenance work on the pneumatic control unit filters, check and, if necessary, regulate the working pressure rating (0,45 MPa) using the pneumatic control unit pressure regulator.
Table 18 Consumables Warning Isopropyl alcohol and heptanes are products subject to safety regulations. They are to be used in strict respect of the applicable laws in force and the manufacturer’s safety regulations. 82 • GLOBAL S - User Manual...
Declaration of Conformity Declaration of Conformity – Applies to European Union The Global CMM has been designed and manufactured in compliance with the essential safety requirements stated in Directive 2006/42/EC Machinery (MD) Directive 2014/30/EU Electromagnetic Compatibility (EMC) Directive 2014/35/EU Low Voltage Directive References to the relevant harmonised standards used, or references to the specifications in relation to which conformity is declared: Safety of Machinery - General Principles for Design –...
Need help?
Do you have a question about the GLOBAL S and is the answer not in the manual?
Questions and answers