Jumper HAA21 User Manual
Jumper HAA21 User Manual

Jumper HAA21 User Manual

Bluetooth electronic blood pressure monitor

Advertisement

s
Bluetooth Electronic Blood Pressure Monitor
HAA21 User Manual
Version: V1.0
Revision Date: 2023.10

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the HAA21 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Jumper HAA21

  • Page 1 Bluetooth Electronic Blood Pressure Monitor HAA21 User Manual Version: V1.0 Revision Date: 2023.10...
  • Page 2 Blood Pressure Monitor. Intended use The Electronic Blood Pressure Monitor is intended to measure the systolic and diastolic blood pressure as well as the pulse rate of adult person via non-invasive oscillometric technique at medical facilities or at home. FCC ID:2ADYL-HAA21...
  • Page 3 Safety Precautions The warnings and illustrations shown in the User Manual enable you to use the product safely and correctly, thus preventing you and others from being injured, specifically as follows: Legend, mark and meaning Warning message Anti-electric shock degree is Type BF of the application part When the product life expires and the end users discard the products, send them to the designated...
  • Page 4 Quick Start Guide Avoid smoking, eating, drinking caffeinated drinks or exercising for 30 minutes before taking measurement. 1. Sit upright in a chair with both feet on the floor. 2. Remove tight fitting clothing from your upper arm along with any thick clothing. 3.
  • Page 5 Overview Structure ①Air Jack ⑤Volume Switch ②User Button ⑥USB External Power Supply ③Power Button Interface ④Memory Button Features Dual users and guest mode  199 Readings * 2 Users  Voice Reminder Function  Average of Last 3 Readings  Irregular Heartbeat Detector ...
  • Page 6: Packing List

    Display Packing List Name Quantity Electronic Blood Pressure Monitor Cuff Dry Battery (AAA) User Manual Zipped Carrying Bag...
  • Page 7 Preparation 2.1 Battery Installation a) Open the battery cover as shown in the picture. b ) Place 4 AAA dry batteries. Pay attention battery electrode indication. 2.2 Battery power indication and replacement After the product is turned on, if low power symbol appears on the screen, the measurement cannot be performed, and the battery must be replaced.
  • Page 8 2.4 Cuff The applicable arm circumference range of the cuff is 22-36cm or 22-42cm. Connection: Insert the air plug of cuff air tube into air jack of the Electronic Blood Pressure Monitor. Figure as shown above.
  • Page 9: Function Setting Mode

    Function Setting Mode 3.1 User mode Switch between “User 1” 、“User 2” and “Guest” can be processed in the shutdown state or upon the measurement completion, by short press on the “User ” button. (Note: In the shutdown state, press “User ”...
  • Page 10 After entering the “Date” setting mode, 1- 1 will flash on the screen. Press the “Memory ” button and the value will be incremented among 1 and 31. Press the “User ” button for “Date” value confirmation and enter the “Time” setting mode. After entering the “Time”...
  • Page 11 3.3 Unit setting Either mmHg or KPa will show up on the screen, upon entering the “Unit” setting mode by pressing the “User ” button (after the “Minute” setting mode). Display of mmHg indicates that the mmHg unit is selected; display of KPa indicates that the kPa unit is selected. Unit will be switched upon every press on the “Memory ”...
  • Page 12 Correct Method of Use 4.1 How to use the cuff (1) Place the cuff flat on the table, leave the hook & loop downwards, and pass the end of cuff through the metal ring to form a loop. The hook &...
  • Page 13 (3) The cuff should be wrapped on the upper arm comfortably, with tight space for two fingers. Before measurement, remove tight fitting clothing from your upper arm along with any thick clothing. (4) Place the lower arm flat on the desktop, leaving the center of palm naturally upwards, sitting upright, and ensuring the center of cuff and the heart are at the same level.
  • Page 14 (1) Press the “Power ” button, and the device will return to zero automatically, the air pump will start to inflate the cuff, and the screen will display the change of the pressure in the cuff. (2) When reaching the stable pressure upon inflation, the air pump will stop the inflation, and the pressure in the cuff will be reduced gradually and displayed on the screen.
  • Page 15 that is 25 % less or 25 % more than the average rhythm detected while your monitor is measuring blood pressure) (3) If you suffer from an irregular heartbeat, measurements taken with this device should be evaluated with your doctor. 4.3 Using memory function This blood pressure monitor is available for 2 users with 199 memories per user (The data of guest mode is not stored).
  • Page 16 4.5 Classification standard for blood pressure condition There is no definition of hypotension yet. Generally, if the systolic blood pressure is less than 90mmHg (12kPa), it is hypotension.
  • Page 17 Contraindications, Precautions, Warnings & Prompt Instruction (1) Warning: Keep the device out of the reach of the children under 12 and people who can’t express their intention. When children of 12~18 use the device, they should be accompanied by the adults. Pregnant women shall use under the guidance of doctors.
  • Page 18 (2) Precautions: Do not repair, disassemble or modify the Electronic Blood Pressure Monitor without permission. Do not collide or drop the host to avoid collision or strong impact on the device. Do not mix old and new batteries of different brands for use. Indoor temperature, environment, noise, user's body position, speech or exercise might affect the blood pressure measurements.The body movement, magnetic field and improper wearing of cuff and...
  • Page 19: Troubleshooting

    Troubleshooting * If you have troubles using the device, please check the following instruction. Problem Possible cause How to correct Replace new batteries or use Low battery the USB for power supply. No display when press the The polarities of Install the batteries in correct Power button batteries are...
  • Page 20 The result of blood “ Er 3 ” Please remeasure in the right pressure is displayed way. abnormal. “Er P” Please check the cuff, tie it up The inflation fails displayed and try again. The inflation pressure is greater “Er H” Please measure again.
  • Page 21: Storage And Maintenance

    Storage & Maintenance Keep the device away from direct sunlight, extreme temperatures,  humidity or moisture. Use a dry, soft cloth to clean the device, or if desired, use a cloth  lightly dampened with water. Do not use corrosive cleaner, benzene, thinner or other volatile ...
  • Page 22: Specifications

    Specifications Product name Bluetooth Electronic Blood Pressure Monitor Model HAA21 Display mode Digital display mode Measuring mode Oscillographic method Measuring body Upper arm part Pressure value 0-295mmHg(0kPa-39.3kPa) Measuring range Pulse value 40-199 pulse beats/min Pressure value ±3mmHg(±0.4kPa) Static measurement accuracy Pulse value ±...
  • Page 23 Screen size 101mm (length) *71mm (width) Upper-arm-type cuff (measured arm circumference Cuff 22-36cm or 22-42cm) Cuff, User Manual, Dry Battery*4, Annexure Zipped Carrying Bag, Quick Guide High-performance dry battery can be used for about Battery life 300 times at normal temperature Service life 5 years Date of production...
  • Page 24 EMC Information-Guidance and Manufacture’s Declaration 1* WARNING: Use of this equipment adjacent to or stacked with other equipment should be avoided because it could result in improper operation. If such use is necessary, this equipment and the other equipment should be observed to verify that they are operating normally.”...
  • Page 25 Harmonic emissions Not applicable IEC 61000-3-2 Voltage fluctuations/ Not applicable flicker emissions IEC 61000-3-3 Table 2 declaration - electromagnetic immunity Immunity test IEC 60601 test level Compliance level Electrostatic ±8 kV contact ±8 kV contact discharge (ESD) ±2 kV, ±4 kV, ±8 kV, ±15 ±2 kV, ±4 kV, ±8 IEC 61000-4-2 kV air...
  • Page 26 0 % UT; 0.5 cycle At 0°, Voltage dips, short Not applicable 45°, 90°, 135°, 180°, 225°, interruptions 270°and 315° voltage variations on 0 % UT; 1 cycle and power supply input 70 % UT; 25/30 cycles Single phase: at 0° lines 0 % UT;...
  • Page 27 declaration - IMMUNITY to proximity fields from RF wireless communications equipment Immu IEC60601 test level Compli nity ance Test Modulation Maxim Immunit test level frequency y level power Radiat 385 MHz **Pulse 1.8W 27 V/m 27 V/m ed RF Modulation: 18Hz 61000 450 MHz...
  • Page 28 2450 MHz **Pulse 28 V/m 28 V/m Modulation: 217Hz 5240 MHz **Pulse 0.2 W 9 V/m 9 V/m 5500 MHz Modulation: 5785 MHz 217Hz Note* - As an alternative to FM modulation, 50 % pulse modulation at 18 Hz may be used because while it does not represent actual modulation, it would be worst case.
  • Page 29 guarantee that interference will not occur in a particular installation. If this equipment does cause harmful interference to radio or television reception, which can be determined by turning the equipment off and on, the user is encouraged to try to correct the interference by one or more of the following measures: -- Reorient or relocate the receiving antenna.
  • Page 30 After-sale Service After-sale service unit: Shenzhen Jumper Medical Equipment Co., Ltd. Address: D Building, No. 71, Xintian Road, Fuyong Street, Baoan, Shenzhen, Guangdong, China Tel: +86-755-26696279 E-mail: info@jumper-medical.com Website: www.jumper-medical.com Postal Code: 518103 Authorized European Representative: MedPath GmbH Mies-van-der-Rohe-Strasse 8, 80807 Munich, Germany Shenzhen Jumper Medical Equipment Co., Ltd.