User Manual

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the ORP Meter and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for bluelab ORP Meter

  • Page 1 ORP Meter User Manual...
  • Page 2: Table Of Contents

    4.3 Hydrating the ORP probes and pH 2.0 Bluelab ORP Meter ......4 probes ..........12 2.1 Overview .......... 4 5.0 Storing the Bluelab ORP Meter .. 12 2.2 Display ..........4 2.3 Preparing for use ......5 6.0 Troubleshooting & FAQ ....13 Insert Batteries ........
  • Page 3: Introduction & Overview

    1x KCL storage solution sachet 1.3 How it works The Bluelab ORP Meter is designed to be able to measure either ORP (mV) or pH. What is being measured is determined by the probe attached and the user selecting the appropriate mode. The Bluelab ORP Meter has two press buttons; power and...
  • Page 4: Bluelab Orp Meter

    2.0 Bluelab ORP Meter 2.1 Overview ORP/pH probe connector Display Bluelab®® ORP Probe Switch mode/Calibrate Power button button 2.2 Display Reading display Low battery mV unit indicator pH unit Successful pH calibration indicators...
  • Page 5: Preparing For Use

    Connecting a probe Both the ORP Probe and pH Probe connect to the meter in the same way. Choose the probe you wish to use, then attach the probe to the Bluelab ORP Meter via the BNC fitting. Note: The Bluelab pH Probe is sold separately.
  • Page 6: Remove The Storage Cap

    Note: A minimum of 30 seconds is recommended, though it may take several minutes, depending on the solution for the measurement to stabilize. Calibrate the pH If measuring pH, calibrate the Bluelab ORP Meter by following the instructions in section 3.0 of this manual.
  • Page 7: Ph Calibration

    Note: ORP probes do not require calibration as the Meter is factory calibrated. IMPORTANT This must be done before the Bluelab ORP Meter is used for the first time to measure pH. 3.0 pH Calibration pH calibration is required before first use and then at least monthly to ensure readings are accurate.
  • Page 8: How To Perform A Ph Calibration

    If after an hour the meter has not been calibrated with a second calibration point the calibration indicators disappear and the ORP Meter reverts to an uncalibrated state. Calibration is required. Now you can calibrate to pH 4.0 and/or pH...
  • Page 9: Ph Calibration Tips

    ‘ ’ will appear on the screen. Rinse the pH probe tip in fresh water. The ORP Meter is now calibrated and ready to measure pH. Probes require cleaning and calibrating at least every 30 days, more frequent with high use.
  • Page 10: Probe Care

    The lifetime of ORP and pH probes depends on the environment it is used in and the way that it is treated. To receive a long life from your Bluelab ORP or pH Probe, please ensure you follow the guide below.
  • Page 11: Cleaning Bluelab Orp Probes And Ph Probes

    After cleaning, use the probe straight away, or place the storage cap on the probe tip. Always ensure the cap contains enough Bluelab pH Probe KCl Storage Solution to cover the probe tip.
  • Page 12: Hydrating The Orp Probes And Ph Probes

    5.1. 5.0 Storing the Bluelab ORP Meter Store the Bluelab ORP Meter in a cool, dry and clean place when not in use. Keep out of direct sunlight. Keep Bluelab ORP Meter out of direct sunlight to prevent irreparable damage to the LCD reading display.
  • Page 13: Troubleshooting & Faq

    Gently swirl the probe in the solution for a inaccurate Probe not gently swirled during stabilization minimum of 30 seconds. Contaminated Bluelab pH Probe / glassware Clean probe (see section 4.2) then calibrate. not clean. Hydrate probe in KCl storage solution for 24 Bridge contaminated blocked or dry.
  • Page 14: Frequently Asked Questions

    6.2 Frequently asked questions Question Answer There are white crystals on the No, these are Potassium Chloride KCl crystals which can form if the hydration solution end of my probe, is it broken or has evaporated. Wash these off under tap water, then refill the storage cap with KCl storage solution to rehydrate the probe overnight before use.
  • Page 15: Specifications

    0 - 50°C / 32 - 122°F Operating environment (Do not use or store outside of this temperature range) Power source 2 x AAA Alkaline batteries User Manual languages English *10mV in negative range -1000 - 0mV Note Bluelab pH Probe sold separately.
  • Page 16: Accessories & Warranty

    7.0 & pH 4.0, KCl 8.3 ORP/pH Probe KCl Storage Solution The best solution to store and hydrate your Bluelab ORP and pH products. Bluelab pH Probe KCl Storage Solution increases response time and maximizes the life of Bluelab pH probes.
  • Page 17: Get In Touch

    If you need assistance or advice - we’re here to help you. Email: support@Bluelab.com Looking for specifications or technical advice? Visit us online at Bluelab.com or facebook.com/Bluelabofficial Bluelab Corporation Limited 8 Whiore Avenue, Tauriko Business Estate Tauranga 3110, New Zealand Register your product online at Bluelab.com/product-registration...

This manual is also suitable for:

Metorp