Cleaning and disinfection

Advertisement

Quick Links

QUICK REFERENCE GUIDE - CLEANING AND DISINFECTION
Caylis
Reactive system with low air loss
ALLOWED DISINFECTANTS
Disinfectant
Isopropyl alcohol solution ≤ 70%
Chlorine solution ≤ 1% (10,000 ppm
Quaternary ammonium solution 1.92%
(1920ppm)
Acceptable pH 7-10 only
Quaternary ammonium solution (3-10%)
Hydrogen peroxide solution, <7% (pH<9)
Chlorine concentrations may vary from 250 ppm to 10,000 ppm depending on local policy and contamination
1
status.
NOTE
Rinse thoroughly with clean water, and allow to dry, to remove residual chemicals before storage.
WARNING
To prevent eye and skin damage, always use protective glasses, gloves and apron. If contact occurs,
rinse with plenty of water. If eyes or skin become irritated, contact a physician. Always read the IFU and
the safety data sheet (SDS) of the disinfectant.
To prevent electrical shock, always disconnect the product from the power source before cleaning and
inspecting.
To prevent cross-contamination, always follow the cleaning and disinfection instructions in this IFU.
To prevent eye or skin irritation, never disinfect in the presence of a patient.
Equipment needed
Protective equipment (protective glasses, gloves and apron)
Neutral detergent (cleaning solution)
Allowed disinfectants
A clean bowl with warm water
Single use or washable clean cloths
Detergent / disinfection wipes can be used as an alternative in line with the compatible chemical list provided.
Tube-set
Pump
cover
1
)
08.CY.02EN_01 • 2024-09 • Page 1/4
Internal mattress
system
Bottom
top cover
cover

Advertisement

loading
Need help?

Need help?

Do you have a question about the Caylis and is the answer not in the manual?

Questions and answers

Summary of Contents for Arjo Caylis

  • Page 1 QUICK REFERENCE GUIDE - CLEANING AND DISINFECTION Caylis Reactive system with low air loss ALLOWED DISINFECTANTS Tube-set Internal mattress Bottom Disinfectant Pump top cover cover system cover ● ● ● ● ● Isopropyl alcohol solution ≤ 70% ● ● ●...
  • Page 2 6. Use water and a clean cloth to wipe off all traces Clean and disinfect of cleaning solution. (33 steps) CAUTION Prepare the pump 1. Switch off the pump unit. Disconnect the pump Never clean the pump by pouring water from the power source.
  • Page 3 19. Use a clean cloth with warm water and neutral Clean the mattress detergent (cleaning solution) to wipe down the cells, bottom cover and securing straps of the NOTE mattress. Make sure the cleaning solution fully The top cover should be assessed for covers all the cells while wiping down.
  • Page 4 33. If required, disinfect the securing straps and base of the mattress using the same method as above. The straps should be unclipped before cleaning for ease of access. ArjoHuntleigh AB Hans Michelsensgatan 10 211 20 Malmö, Sweden 08.CY.02EN_01 • 2024-09 • Page 4/4 www.arjo.com...