QUICK REFERENCE GUIDE - CLEANING AND DISINFECTION
Caylis
Reactive system with low air loss
ALLOWED DISINFECTANTS
Disinfectant
Isopropyl alcohol solution ≤ 70%
Chlorine solution ≤ 1% (10,000 ppm
Quaternary ammonium solution 1.92%
(1920ppm)
Acceptable pH 7-10 only
Quaternary ammonium solution (3-10%)
Hydrogen peroxide solution, <7% (pH<9)
Chlorine concentrations may vary from 250 ppm to 10,000 ppm depending on local policy and contamination
1
status.
NOTE
Rinse thoroughly with clean water, and allow to dry, to remove residual chemicals before storage.
WARNING
To prevent eye and skin damage, always use protective glasses, gloves and apron. If contact occurs,
rinse with plenty of water. If eyes or skin become irritated, contact a physician. Always read the IFU and
the safety data sheet (SDS) of the disinfectant.
To prevent electrical shock, always disconnect the product from the power source before cleaning and
inspecting.
To prevent cross-contamination, always follow the cleaning and disinfection instructions in this IFU.
To prevent eye or skin irritation, never disinfect in the presence of a patient.
Equipment needed
Protective equipment (protective glasses, gloves and apron)
•
Neutral detergent (cleaning solution)
•
Allowed disinfectants
•
A clean bowl with warm water
•
•
Single use or washable clean cloths
Detergent / disinfection wipes can be used as an alternative in line with the compatible chemical list provided.
•
Tube-set
Pump
cover
●
●
1
)
●
●
08.CY.02EN_01 • 2024-09 • Page 1/4
Internal mattress
system
●
●
●
●
●
●
●
●
●
Bottom
top cover
cover
●
●
●
●
●
●
●
●
●
●