450 volt / 5 ns rise time (std), 400 volt / 2.5 ns rise time (opt), 8 ns to 100 ns pulse width pulse generator with ieee 488.2 / rs-232 / ethernet control (56 pages)
Summary of Contents for AVTECH ELECTROSYSTEMS AVMP-5-P-PS-TUE1
Page 1
FAX: 800-561-1970 (USA & Canada) or +1-613-686-6679 (Intl) OTTAWA, ONTARIO U.S.A. 13669-0265 CANADA K2C 3H4 info@avtechpulse.com http://www.avtechpulse.com/ INSTRUCTIONS MODEL AVMP-5-P-PS-TUE1 +2 TO +10 VOLTS, 0.5 TO 5 MHz HIGH SPEED DUAL-CHANNEL PULSE GENERATOR WITH < 500 ps RISE AND FALL TIMES SERIAL NUMBER: ____________...
WARRANTY Avtech Electrosystems Ltd. warrants products of its manufacture to be free from defects in material and workmanship under conditions of normal use. If, within one year after delivery to the original owner, and after prepaid return by the original owner, this Avtech product is found to be defective, Avtech shall at its option repair or replace said defective item.
TABLE OF CONTENTS WARRANTY........................2 TECHNICAL SUPPORT....................2 TABLE OF CONTENTS....................3 INTRODUCTION......................5 ORIGINAL QUOTATION AND SPECIFICATIONS............6 REGULATORY NOTES....................8 FCC PART 18.......................... 8 EC DECLARATION OF CONFORMITY..................8 DIRECTIVE 2002/95/EC (RoHS).....................9 DIRECTIVE 2002/96/EC (WEEE)....................9 INSTALLATION......................10 VISUAL CHECK........................10 POWER RATINGS........................ 10 CONNECTION TO THE POWER SUPPLY................10 PROTECTION FROM ELECTRIC SHOCK................11 ENVIRONMENTAL CONDITIONS..................12...
INTRODUCTION The AVMP-5-P-PS-TUE1 is a high performance customized dual-channel pulse generator. Two main outputs (A and B) are generated. The outputs are synchronous unipolar positive outputs with B 180 degrees out of phase with respect to A. The output amplitudes are controlled by a single (shared) ten-turn locking dial. The amplitude may be adjustable from <...
Technische Universiteit Eindhoven XXXXX, We are pleased to quote as follows: Quote number: 14754 Model number: AVMP-5-P-PS-TUE1 Output amplitude: +2V to +10V, adjustable. The output amplitude is controlled by a locking ten-turn dial. Load resistance: Must be 50 Ohms. Number of outputs: Two (A and B). The outputs are synchronous unipolar positive outputs with B 180 degrees out of phase with respect to A.
Page 7
Please call or email me if I can be of further assistance. Thank you for your interest in our products! Dr. Michael J. Chudobiak Chief Engineer Avtech Electrosystems Ltd. ---------------------- since 1975 --- PO Box 265 ph: 888-670-8729 or 613-686-6675...
This instrument does not normally require regular maintenance to minimize interference potential. However, if loose hardware or connectors are noted, they should be tightened. Contact Avtech (info@avtechpulse.com) if you require assistance. EC DECLARATION OF CONFORMITY Avtech Electrosystems Ltd. P.O. Box 5120, LCD Merivale Ottawa, Ontario Canada K2C 3H4 declare that this pulse generator meets the intent of Directive 89/336/EEC for Electromagnetic Compatibility.
and that this pulse generator meets the intent of the Low Voltage Directive 72/23/EEC as amended by 93/68/EEC. Compliance pertains to the following specifications as listed in the official Journal of the European Communities: EN 61010-1:2001 Safety requirements for electrical equipment for measurement, control, and laboratory use DIRECTIVE 2002/95/EC (RoHS) This instrument is exempt from Directive 2002/95/EC of the European Parliament and...
INSTALLATION VISUAL CHECK After unpacking the instrument mainframe and the transformer module, examine to ensure that they have not been damaged in shipment. Visually inspect all connectors, knobs, and handles. Confirm that a power cord and an instrumentation manual (this manual), are with the instrument.
Destination Region Description Option Manufacturer Part Number United Kingdom, Hong Kong, BS 1363, -AC00 Qualtek 370001-E01 Singapore, Malaysia 230V, 50 Hz AS 3112:2000, Australia, New Zealand -AC01 Qualtek 374003-A01 230-240V, 50 Hz Continental Europe, Korea, European CEE 7/7 -AC02 Qualtek 364002-D01 Indonesia, Russia “Schuko”...
ENVIRONMENTAL CONDITIONS This instrument is intended for use under the following conditions: 1. indoor use; 2. altitude up to 2 000 m; 3. temperature 5 °C to 40 °C; 4. maximum relative humidity 80 % for temperatures up to 31 °C decreasing linearly to 50 % relative humidity at 40 °C;...
FUSES This instrument contains four fuses. All are accessible from the rear-panel. Two protect the AC prime power input, and two protect the internal DC power supplies. The locations of the fuses on the rear panel are shown in the figure below: Fuses #1 and #2 Fuse #4 Fuse #3...
DC FUSE REPLACEMENT The DC fuses may be replaced by inserting the tip of a flat-head screwdriver into the fuse holder slot, and rotating the slot counter-clockwise. The fuse and its carrier will then pop out. FUSE RATINGS The following table lists the required fuses: Recommended Replacement Part Nominal Fuses...
FRONT PANEL CONTROLS 1) POWER Switch . This is the main power switch. When turning the instrument on, there may be a delay of several seconds before the instrument appears to respond. 2) OVERLOAD Indicator . When the instrument is powered, this indicator is normally green, indicating normal operation.
REAR PANEL CONTROLS 1. AC POWER INPUT . An IEC-320 C14 three-pronged recessed male socket is provided on the back panel for AC power connection to the instrument. One end of the detachable power cord that is supplied with the instrument plugs into this socket. 2.
GENERAL INFORMATION BASIC TEST ARRANGEMENT The AVMP-5-P-PS-TUE1 should be tested with a dual-channel sampling oscilloscope with a bandwidth of at least 10 GHz to properly observe the high-speed waveform. (The cables and attenuators must also have a bandwidth of at least 10 GHz.) A typical test...
Page 18
to operate with 50% input and output duty cycles. The input impedance is 50 Ohms. The relationship between the IN, OUT A, and OUT B pulses is shown below: 50% DUTY CYCLE IN +3 to +5V (TTL HIGH) 0V (TTL LOW) PROPAGATION DELAY AMPLITUDE OUT A...
PREVENTING DAMAGE The AVMP-5-P-PS-TUE1 may fail if triggered at a PRF greater than 5 MHz. This unit is designed to operate into a load impedance of 50 Ohms and the output stage will be damaged if it is operated into an open circuit (or any other high impedance).
MECHANICAL INFORMATION TOP COVER REMOVAL If necessary, the interior of the instrument may be accessed by removing the four Phillips screws on the top panel. With the four screws removed, the top cover may be slid back (and off). Always disconnect the power cord and allow the instrument to sit unpowered for 10 minutes before opening the instrument.
MAINTENANCE REGULAR MAINTENANCE This instrument does not require any regular maintenance. On occasion, one or more of the four rear-panel fuses may require replacement. All fuses can be accessed from the rear panel. See the “FUSES” section for details. CLEANING If desired, the interior of the instrument may be cleaned using compressed air to dislodge any accumulated dust.
WIRING DIAGRAMS WIRING OF AC POWER, STYLE 1 M a i n s c i r c u it s - h a z a r d o u s l i v e . D o n o t a t t e m p t a n y r e p a i r s o n t h i s i n s t r u m e n t M o u n t w i t h M 2 .
WIRING OF AC POWER, STYLE 2 M a i n s c i r c u it s - h a z a r d o u s l i v e . D o n o t a t t e m p t a n y r e p a i r s o n t h i s i n s t r u m e n t b e y o n d t h e f u s e r e p l a c e m e n t p r o c e d u r e s d e s c r i b e d i n t h e m a n u a l .
PCB 158N - LOW VOLTAGE POWER SUPPLY, 2/3 8 3 0 8 3 5 F U S E H O L D E R 6 4 0 4 4 5 - 6 T P 6 T P 3 T E S T - L O O P T E S T - L O O P S 1 A S 1 B , O R D C...
PCB 158N - LOW VOLTAGE POWER SUPPLY, 3/3 R 2 1 1 . 5 K o r 1 . 8 K O Y C A P B A N K R 2 6 1 5 K C 2 3 1 0 0 0 u F , 3 5 V ( P 5 1 6 9 - N D ) L 6 2 7 1 1 H V W A R N I N G R 1 7...
Need help?
Do you have a question about the AVMP-5-P-PS-TUE1 and is the answer not in the manual?
Questions and answers