Diagnostic DM-600 IHB User Manual

Automatic upper arm blood pressure and pulse monitor

Advertisement

Quick Links

USER MANUAL
DM-600 IHB
AUTOMATIC UPPER ARM
BLOOD PRESSURE AND PULSE MONITOR
0197
5129

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the DM-600 IHB and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Diagnostic DM-600 IHB

  • Page 1 USER MANUAL DM-600 IHB AUTOMATIC UPPER ARM BLOOD PRESSURE AND PULSE MONITOR 0197 5129...
  • Page 2: Table Of Contents

    CONTENTS 1. INTRODUCTION......... . . 1 1.1 Device features .
  • Page 3: Introduction

    Thank you for buying the blood pressure and pulse monitor DM-600 IHB.. The model can be used with irregular pulse. If the device detects irregular pulse, the symbol   . appears on the display. In such a case, it is advisable to consult your physician.
  • Page 4: Electromagnetic Interference

    Self-measurement means control and not diagnosis and treatment. Unusual values should always be consulted with your doctor. You should under no circumstances change the doses of medications prescribed by The displayed pulse rate is not suitable for controlling the operating ...
  • Page 5 it is recommended to perform regular blood pressure self-measurements. This allows for detecting any changes in the value of blood pressure at an early stage and respond accordingly. If the patient is undergoing treatment for hypertension/hypotension, regular measurements should be taken at a specific time of day and the results recorded, and then presented to the doctor.
  • Page 6: Pressure Monitor Design

    3. PRESSURE LCD display MONITOR DESIGN power supply socket air tube connector MEMORY button START/STOP TIME button button plug connecting the CUFF air tube with the blood (Applied part type BF) pressure monitor Arm circumference range: (insert in the air tube connector) 22-42 cm air tube metal buckle...
  • Page 7: Starting The Device

     min. 600mA We recommend using only the power adapter supplied by the  manufacturer, model Diagnostic ZID 6-1 (100~240 V, 50/60Hz, 6 V, 1000 mA (1 A)) If the device is defective, unplug the power supply or the power ...
  • Page 8: User Selection And Setting Date And Time

    Do not touch the power adapter with a wet hand.  Do not tangle the wires during usage.  The power adapter is added to the set optionally  power (charged additionally) supply 4.4 User selection and setting date and time User selection: The blood pressure monitor allows you to track blood pressure readings of 2 users.
  • Page 9: Taking Measurements

    10. Set the exact time, i.e. minutes, using the MEMORY button. 11. Press the TIME button: the measurement unit will start flashing. 12. Press the MEMORY button to set the unit of measurement (mmHg or kPa). 13. After completing the settings, press TIME. Now the settings are confirmed and the timer starts running.
  • Page 10: Wrapping The Cuff

    In the case of repeated measurements there is a build-up of blood in  the arm, leading to incorrect results. For this reason, the correct blood pressure measurement should be carried out after a 5 minute break or after lifting the arm, to facilitate the outflow of the accumulated blood (after at least 3 minutes).
  • Page 11: Body Posture During Measurement

    5.4 Body posture during measurement Relax, rest the elbow on the table with palm facing The arm should be up; the cuff should be at heart level. Accuracy of positioned at the same height readings may be reduced if the cuff is not wrapped as the heart.
  • Page 12: Memory

    6 MEMORY Internal memory stores up to 2x120 of measurement results. 1) Recalling results from memory To access memory, press the MEMORY button.  The device will display the average score of 3 most recent  measurements, the symbol   without the measurement number will appear (FIG.
  • Page 13: Error Messages

    7. ERROR MESSAGES If an error occurs during the measurement, the reading will be interrupted and an error code displayed. Error code Possible cause ERR ERR 1 No pulse detected. ERR 2 Measurement results affected by interference. Cause: there was an arm movement during measurement. ERR 3 Inflation of the cuff has taken too long.
  • Page 14: Maintenance

    Diagnosis S.A. Free infoline 800 11 30 70. 9. WARRANTY The DM-600 IHB blood pressure monitor is covered by a 2-year warranty starting from the purchase date. The warranty does not cover damage due to improper handling, accidents, non-compliance with the user manual, or changes made in the device by third parties.
  • Page 15: Symbols

    Take care to prevent children from using the device without  supervision; some parts of the device are small and may be swallowed. Use only original elements supplied by the manufacturer. The  use of other elements may reduce the level of safety. 11.
  • Page 16: Technical Data

    12. TECHNICAL DATA Measurement Oscillometric Digital Display method LCD display Measurement range Pressure 30 to 280 mmHg (± 1 mmHg) Pulse: 40 do 200 beats per minute Measurement accuracy Pressure: ±3mmHg Pulse: ±5% reading Inflating Automatic pumping device Deflating Automatically through air valve Memory function 2 x 120 measurements with date and time Power supply...
  • Page 17 Guidelines and manufacturer's declaration - electromagnetic emissions The devices are intended for use in the electromagnetic environment as described below. The customer or the user of the device should assure that the device is used in such an environment Fulfillment Emission test Guidelines regarding electromagnetic environment of requirements...
  • Page 18 GUIDELINES AND MANUFACTURER'S DECLARATION REGARDING ELECTROMAGNETIC IMMUNITY The device is intended for use in the electromagnetic environment specified below. The customer or the user of the DEVICE should assure that it is used in such an environment. Immunity test Test level, IEC 60601 Compatibility Electromagnetic environment - guidelines standard...
  • Page 19 Recommended distance between portable and mobile radio communication equipment and the DEVICE The [DEVICE or SYSTEM] is intended for use in the electromagnetic environment in which the interference caused by the emission of radio waves is controlled. The buyer or the user of the [DEVICE or SYSTEM] can help prevent electromagnetic interference by keeping a minimum distance between portable and mobile radio communication equipment (transmitters) and the [DEVICE or SYSTEM], as recommended below, according to the maximum output power of the communication equipment.
  • Page 20: Warranty Card

    DATE OF SALE ..........WARRANTY TERMS Diagnosis S.A. grants a warranty: 24 months for DIAGNOSTIC blood pressure monitors and cuffs (excluding  pump assembly) Hardware defects revealed during the warranty period will be rectified free of charge within 21 days.