1. SAFETY INFORMATION 1.1 Warning Self-diagnosis and undertaking treatment based on the results of measurements may be dangerous. Follow the advice of your doctor or qualified healthcare professional. People suffering from arrhythmia, diabetes, circulation problems or medical conditions associated with stroke should use the device in accordance with the recommendations of the doctor.
The batteries have a positive / negative polarity. If the battery does not connect well with the device, do not try to insert it forcibly. Do not use the Luer-type lock. If Luer-type lock connectors are used in the construction of the tubes, there is a risk of accidentally connecting to an intravascular flow system, allowing the pumping of air into the blood vessels.
Keep the device out of reach of infants, young children and people who are not able to give informed consent. This product is suitable for blood pressure measurement at home and for use by authorized medical personnel at hospital. WARNING!Read the attached user manual.
DISPLAY Systolic pressure Blood pressure classification indicator Diastolic according to the World pressure Health Organization (WHO) Designation pulse of units Heart symbol time Memory symbol Low battery Average symbol symbol 3. BEFORE TAKING READINGS 3.1 Batteries 3.1.1 Installation and replacement 1.
400mA; We recommend using only the power adapter supplied by the manufacturer, model Diagnostic ZID 6-1 (100~240 V, 50/60Hz, 6 V, 1000 mA (1 A)) If the device is defective, unplug the power supply or the power cor.
b) Set the two digits indicating year. c) Press the SET button and continue to set the values for month d) Repeat steps a) to c) to set the month, day, hour and minutes 2. Przeliczanie jednostek (mmHg na kPa) a) Pressing the MEM button, automatically changes the unit as shown.
4.2 Adjusting the cuff 1. Insert the air tube in the socket on the left side of the device. 2. Insert the end of the cuff under the metal buckle, with the velcro facing out. 3. Wrap the cuff approximately 2-3 cm above the elbow. For best results, wrap the cuff on bare skin, at heart level.
4.4 Taking measurement After installing the batteries and wrapping the cuff, the device is ready to take measurement: 1. To ensure that the results are most accurate, during taking the reading, relax, do not smoke, do not take deep breaths, do not talk loudly and try not to move.
Note: Holding the MEM button longer will erase all data from the memory. Deleting data from memory: a) You must go into memory mode. b) Press and hold down the MEM button until “---” is displayed. c) All data will be deleted, it is not possible to delete a single record. d) Press the ON/OFF button to exit the memory mode and turn off the device.
6. TROUBLESHOOTING In the event of irregularities during operation, please refer to the following: PROBLEM HOW TO REPAIR After replacing batteries and 1. Check battery polarity. turning on the device, nothing 2. If you still cannot turn on the device, reinstall or replace the displays batteries.
7. SPECIFICATION Description Automatic upper arm blood Model Diagnostic S-500 pressure monitor Display Digital display LCD Measurement Oscillometric method principle Measurement Pressure: 0~280 mmHg Accuracy Pressure: ±3mmHg (±0.4 kPa) range Pulse rate: 40~180 beats/min Pulse: ±5% Memory of 90 measurements...
Page 15
Guidelines and manufacturer's declaration - electromagnetic emissions The devices are intended for use in the electromagnetic environment as described below. The customer or the user of the device should assure that the device is used in such an environment. Fulfillment Guidelines regarding electromagnetic Emission test of requirem...
Page 16
Guidelines and manufacturer's declaration regarding electromagnetic immunity With respect to the EQUIPMENT or SYSTEMS, which do not serve as LIFE SUPPORTING SYSTEMS GUIDELINES AND MANUFACTURER'S DECLARATION REGARDING ELECTROMAGNETIC IMMUNITY The device is intended for use in the electromagnetic environment specified below. The customer or the user of the DEVICE should assure that it is used in such an environment.
Recommended spacing between portable and mobile radio communication equipment and the DEVICE in relation to DEVICES that do not serve as LIFE SUPPORTING SYSTEMS Recommended spacing between portable and mobile radio communication equipment and the DEVICE The [DEVICE or SYSTEM] is intended for use in the electromagnetic environment in which the interference caused by the emission of radio waves is controlled.
9.2 What is high blood pressure? High blood pressure also known as HBP or hypertension is a commonly misunderstood medical condition. It is believed that people with hypertension are tense, nervous or hyperactive, but pressure has nothing to do with personality. A person can in fact be peaceful, relaxed and at the same time suffer from hypertension.
Page 19
9.3 What is morning hypertension (sudden morning surge in blood pressure)? Morning high blood pressure or a surge in blood pressure in the morning is referred to as the average of weekly morning blood pressure reading, measured 1 to 2 hours after the waking up, the value of which exceeds 135/85 mm Hg.
EXPLANATION OF SYMBOLS START/STOP Pause and turn on The product complies with the requirements of the European Union 0197 Rev. Systolic pressure in mmHg Date of the last revision Diastolic pressure in mmHg Manufacturer Pulse rate Manufacturing date Note on installing battery Serial number Direct current Read the user manual...
DATE OF SALE ..........WARRANTY TERMS Diagnosis S.A. grants a warranty: 24 months for DIAGNOSTIC S-500 blood pressure monitor and cuffs (excluding pump assembly) 12 months for pressure monitor accessories ...