Contents General ..................3 Safety notices ................3 The Erlab guarantee ..............4 Recommendation for implementation/positioning ..........5 Mounting the unit ..............6 Start up ..................11 Admissible weights on worksurfaces ....... 14 Cleaning and maintenance ........... 14...
Captair Smart fume hood throughout its life cycle. The E.S.P. program (Erlab Safety Program) was set up to guarantee your safety. We would remind you that it is important to have the safety parameters validated before using the device for the first time and whenever it is used for a different application.
® of filter usage time. In order to benefit from Erlab extension of warranty offer, the following conditions shall be respected: • The registration and/or the connection of the product shall be performed within the twelve months from the purchase date ;...
Captair Bio 320 Instructions & User’s Manual Recommendation for implementation/positioning Necessary space and dimensions - according to the EN 14175-5 standard 1,4 m. 2 m. 1 m. 3 m. 1 m. Pilier min 15 cm 0,30 m. 3 m. min 2,7 m...
Page 7
Captair Bio 320 Instructions & User’s Manual Resin work surfac IO115020002 Work surface - stainless steel 304L IO115050002 Back panel Z7PIDRI7003 Left side panel Z7PIDVI712 Right side panel Z7PIDVI716 (1)Z3CEILINCBIO320 (2) Top part (1), control panel (2) and UV lamp...
Page 9
Captair Bio 320 Instructions & User’s Manual D x 1 I x 7 E x 1 J x 1...
Page 10
O x 2 Captair Bio Power up the UV lamp using the switch located on the control panel. The UV lamp can exclusively be activated when the front door is closed. The UV timer decontamination cycle can be runned from 0 to 30 minutes. * UV lamp : At 5,000 hours, it will have lost 15% / At 8,000 hours, it needs to be replaced.
Stainless steel Trespa Top Lab ® PLUS work top Captair Bio 320 Smart 35 kg 35 kg Cleaning and maintenance Mechanical item checks Hinges: Hinges must be properly attached and must allow any items on the front side of the workstation to be instantly and easily lifted upward.
Page 12
About Erlab The Erlab Research and Development laboratory Since 1968, Erlab has been a specialist, inventor and world leader in ductless, zero-emission filtering fume hoods for laboratories to provide total safety in chemical handling. Erlab filtration We provide technologies to protect laboratory staff from inhaling chemicals. This is made possible thanks to our Research and Development (R&D) department, which has continuously improved our filtration technology for more than 50 years.
Need help?
Do you have a question about the Captair Bio 320 and is the answer not in the manual?
Questions and answers