Warranty 2 Tektronix warrants that this product will be free from defects in materials and workmanship for a period of one (1) year from the date of shipment. If any such product proves defective during this warranty period, Tektronix, at its option, either will repair the defective product without charge for parts and labor, or will provide a replacement in exchange for the defective product.
Avoid Exposed Circuitry. Do not touch exposed connections and components when power is present. Do Not Operate in Wet/Damp Conditions. Do Not Operate in an Explosive Atmosphere. Keep Product Surfaces Clean and Dry. TDP0500 & TDP1000 Technical Reference www.valuetronics.com...
DANGER indicates an injury hazard immediately accessible as you read the marking. WARNING indicates an injury hazard not immediately accessible as you read the marking. CAUTION indicates a hazard to property including the product. The following symbols may appear on the product: TDP0500 & TDP1000 Technical Reference www.valuetronics.com...
Preface Preface This is the Technical Reference Manual for the TDP0500 & TDP1000 differential probes. This manual provides specifications and performance verification procedures for the probes. TDP0500 & TDP1000 Technical Reference www.valuetronics.com...
Specifications The specifications in Tables 1 through 4 apply to a TDP0500 or TDP1000 differential probe installed on a Tektronix DPO7000 oscilloscope. When the probe is used with another oscilloscope, the oscilloscope must have an input impedance of 50 Ω...
Nonoperating: 0-90% RH, tested at +30 to +60 °C See warning that follows. WARNING. To avoid a burn hazard at high ambient temperatures, do not touch the probe with bare hands at nonoperating temperatures above +50 °C . TDP0500 & TDP1000 Technical Reference www.valuetronics.com...
(See Figure 7 on page 6.) Bandwidth limit -3 dB at 1 MHz -3 dB at 10 kHz -3 dB at 100 Hz Propagation delay 6.5 ns from probe tip to output Figure 1: Typical bandwidth (TDP0500) TDP0500 & TDP1000 Technical Reference www.valuetronics.com...
Table 4: Nominal electrical characteristics Input configuration Differential (two inputs, + and - ), with case ground Output coupling DC coupling Voltage ranges 4.25 V and 42 V Terminate output into 50 Ω Termination TDP0500 & TDP1000 Technical Reference www.valuetronics.com...
fidelity. NOTE. All adapter specifications are typical unless otherwise indicated. Straight Pins Tektronix part number: 016-1891-xx Bandwidth: >1.0 GHz 10/90 Rise time: <350 ps Best overall signal fidelity of the available adapters. TDP0500 & TDP1000 Technical Reference www.valuetronics.com...
Page 20
10/90 Rise time: <350 ps This adapter has sharp, adjustable pins that can span up to 0.35 inch apart. They are useful for probing small circuit board features such as vias and narrow traces. TDP0500 & TDP1000 Technical Reference www.valuetronics.com...
Page 21
Tektronix part number: 196-3504-xx Bandwidth: >820 MHz 10/90 Rise time: <430 ps Use this adapter to provide easy access to test points that you frequently check, or that may be difficult to probe with other methods. TDP0500 & TDP1000 Technical Reference www.valuetronics.com...
Page 22
Solder the leads to your test points, spaced up to 5.5 inches apart. CAUTION. To prevent short circuits, solder and dress the adapter leads carefully, and make sure that the adapter pins do not touch other conductors when the adapter is not connected to the probe. TDP0500 & TDP1000 Technical Reference www.valuetronics.com...
Page 23
Usable Bandwidth: <250 MHz Calculated rise time: 1.4 ns Use this adapter for DC and low-frequency measurements. The socket ends plug onto square pins, component leads, and the KlipChip adapters included with the probe. TDP0500 & TDP1000 Technical Reference www.valuetronics.com...
Page 24
Tektronix part number: 206-0364-xx Usable Bandwidth: <100 MHz Calculated rise time: 3.5 ns Use the KlipChip Adapters with the Y-lead adapters. Due to the length of these adapters, they are only recommended for DC and low-frequency measurements. TDP0500 & TDP1000 Technical Reference www.valuetronics.com...
Performance Verification Performance Verification Use the following procedures to verify the warranted specifications of the TDP0500 and TDP1000 Differential Probes. Before beginning these procedures, photocopy the test record and use it to record the performance test results. (See Table 7 on page 36.) The recommended calibration interval is one year.
Performance Verification Special Adapters Some of the adapters used in these procedures are available only from Tektronix. These adapters are described on the following pages. TekVPI Calibration Fixture The TekVPI Calibration Fixture, Tektronix part number 067-1701-xx, connects between the host instrument and the probe under test.
Page 28
Probe Calibration Fixture (Optional) The Probe Calibration Fixture, Tektronix part number 067-0419-xx, provides a means to check the rise time specification of the probe at the 42 V range. (See Figure 11.) This check is not required to complete the performance verification of the probe, but is provided for users who want to check the probe rise time at the higher voltage range.
6. Allow the probe and test equipment to warm up for 20 minutes at an ambient temperature of 20 °C to 30 °C. 7. Photocopy the test record and use it to record the test results. (See Table 7 on page 36.) Perform the verification procedures in order. TDP0500 & TDP1000 Technical Reference www.valuetronics.com...
9. Set the input voltage on the DC source to approximately 5 V. Record the actual voltage as V 10. Record the output voltage as V 11. Calculate the attenuation twice, using the values from the measurements as follows: TDP0500 & TDP1000 Technical Reference www.valuetronics.com...
Page 31
23. Verify that the output returns to approximately 0 volts. (This is a functional check; there is no specified performance limit.) 24. Press the DC Reject button on the probe to turn DC reject off. 25. Keep the output connections for the next procedure. TDP0500 & TDP1000 Technical Reference www.valuetronics.com...
10. Measure and record the output voltage as V 11. Calculate attenuation using the formula in step 7. 12. Verify that the attenuation is in the range of 47.5 to 52.5. Record the results in the test record. TDP0500 & TDP1000 Technical Reference www.valuetronics.com...
Page 33
22. Verify that the attenuation is in the range of 4.75 to 5.25. Record the results in the test record. 23. Disconnect the probe from the test setup and connect it to any oscilloscope channel to keep the probe at operating temperature. TDP0500 & TDP1000 Technical Reference www.valuetronics.com...
BNC adapter if necessary. The setup should now appear as shown. (See Figure 15.) 8. Normalize the network analyzer to remove the loss through the TekVPI Calibration Fixture. Figure 15: Normalize the signal path for the following tests TDP0500 & TDP1000 Technical Reference www.valuetronics.com...
Page 35
3. Connect the probe to the TekVPI Calibration Fixture. 4. Connect the probe head to the DM pins on the Probe Tip to BNC adapter. (See Figure 16.) 5. Set the probe for 4.25 V, full bandwidth, and DC reject off. TDP0500 & TDP1000 Technical Reference www.valuetronics.com...
Page 36
Performance Verification Figure 16: Setup for the analog bandwidth and CMRR tests 6. Read the amplitude at 500 MHz for the TDP0500 or 1 GHz for the TDP1000. The use of the marker function, (if equipped), will simplify resolving the bandwidth.
Page 37
13. Verify that the -3 dB bandwidth is approximately 1 MHz. (This is a functional check only. There is no specified performance limit for the bandwidth limit function.) 14. Cycle the bandwidth limit button on the probe back to full bandwidth. 15. Retain the setup for the next test. TDP0500 & TDP1000 Technical Reference www.valuetronics.com...
14. Verify that the CMRR is less than the values in the 42 V column of Table 6, and record the results in the test record. 15. Disconnect the probe from the TekVPI Calibration Fixture and connect it to any channel on the oscilloscope to keep the probe at operating temperature. TDP0500 & TDP1000 Technical Reference www.valuetronics.com...
5. Connect the SMA cable from Channel 8 to the SMA connector on the TekVPI Adapter. 6. Turn on Channel 8 and set the vertical scale to 50 mV/div. Figure 17: Test system rise time setup TDP0500 & TDP1000 Technical Reference www.valuetronics.com...
Page 40
13. Use the oscilloscope measurement capability to display rise time. Increase the stability of the pulse edge measurement by using averaging, if available. Rise time is determined from the 10% and 90% amplitude points on the waveform. Record the rise time as t TDP0500 & TDP1000 Technical Reference www.valuetronics.com...
Page 41
200 ps/div while maintaining the edge view. For a more stable measurement display, turn on averaging. 20. Adjust the position controls on the sampling oscilloscope to display the entire leading edge waveform. TDP0500 & TDP1000 Technical Reference www.valuetronics.com...
Page 42
) that you measured in step 13, and the test system and probe rise time (t ) that you measured in step 21, calculate the probe-only rise time using the formula shown. Example: This completes the performance verification. An optional rise time check for the 42 V range follows. TDP0500 & TDP1000 Technical Reference www.valuetronics.com...
8. Connect a short SMA cable from the other SMA connector on the probe calibration fixture to the high voltage output connector on the pulse generator. 9. Connect the trigger out from the pulse generator to the trigger in on the sampling oscilloscope. TDP0500 & TDP1000 Technical Reference www.valuetronics.com...
Page 44
. This measurement at the 42 V range is typically about 400 ps. 21. Perform the calculation with the two measured rise times and the formula below. A typical result will be about 330 ps. TDP0500 & TDP1000 Technical Reference www.valuetronics.com...
Page 45
Performance Verification Figure 20: Probe rise time setup (42 V range) TDP0500 & TDP1000 Technical Reference www.valuetronics.com...
Page 46
4.25 V range 50 dB _____________ 30 kHz 45 dB _____________ 1 MHz 25 dB _____________ 250 MHz 42 V range 55 dB _____________ 30 kHz 50 dB _____________ 1 MHz 18 dB _____________ 250 MHz TDP0500 & TDP1000 Technical Reference www.valuetronics.com...
24.) 2. Allow all test equipment to warm up for 20 minutes in an environment that is within the environmental conditions listed in the specifications section. Perform the verification procedures in order. TDP0500 & TDP1000 Technical Reference www.valuetronics.com...
Analog Bandwidth Normalize the Setup 1. Set the synthesizer for 1 GHz/10 dBm for the TDP1000 probe and 500 MHz/10 dBm for TDP0500 probe. 2. Set the spectrum analyzer: Center frequency at 500 MHz for the TDP0500 probe, and 1 GHz for the TDP1000 probe.
Page 49
Figure 22: Probe tip test connections 8. Set the probe to 4.25 V range, DC reject off, full bandwidth. 9. Adjust the reference level of the spectrum analyzer to display a signal on screen. TDP0500 & TDP1000 Technical Reference www.valuetronics.com...
Page 50
3. For example, if the reference level from step 3 is 9 dBm, and the level that you measured in this step is -26 dBm, then the difference between the two measurements is 35 dBm, which is within the probe specification. 14. Keep the setup for the next procedure (CMRR). TDP0500 & TDP1000 Technical Reference www.valuetronics.com...
16. Repeat the procedure for the remaining frequencies listed in the table. CMRR CMRR Test Differential Mode Common Mode @4.25 V @42 V Frequency Measurements Measurements (Calculated) (Calculated) 4.25 V 42 V 42 V 4.25 V 30 kHz 1 MHz 250 MHz TDP0500 & TDP1000 Technical Reference www.valuetronics.com...
Quick Start User Manual for a list of replaceable accessories for your probe. If your probe does not meet the specifications tested in the Performance Verification, you can send the probe to Tektronix for repair. Follow the procedure below to prevent damage to the probe during shipping.
Need help?
Do you have a question about the TDP0500 and is the answer not in the manual?
Questions and answers