1. Pre-Check
Note: To perform these tests, gas supply cylinders or gas supply shutoff valves are required in order to isolate the
gas supply form the device. Attempting to perform these tests with central pipeline supplied gas without a
local shut off mechanism is not recommended.
WARNING: Proper inspection and maintenance of this device is essential to prevent gas leaks. All
hoses, fittings, and connections should be inspected regularly, and all leaks should be repaired
immediately.
WARNING: If precheck test cannot be executed successfully, do not use this device and contact
distributor.
Field Leak Test
To test the device for leaks in the field, ensure the
N
O Key Lock is in the ON position. For adjustable
2
1
models, set the %N
required for 50/50 model)
Ensure gas supply lines are installed correctly.
Open the supply valves to pressurize the system
2
and confirm the O
2
are within the green bands.
3
Turn off the supply valves and start a stopwatch.
After approximately 2 minutes, confirm that the
4
pressure on the O
2
has not dropped below 40 psi.
If either pressure has dropped below 40 psi, check
the demand valve to ensure the quick-connect
fitting is secure and the valve is firmly pushed into
the delivery port. Inspect all hoses and connections
for leaks; if any hoses are leaking, ensure the
5
threads are tightened and the hose crimp is intact.
Repair or replace any damaged or leaking hoses
immediately.
If
authorized
distributor
troubleshooting.
Emergency Air Intake Valve Test
To test the emergency air intake valve, turn the N
1
the %N
O knob to 0% (no setting required for 50/50 model).
2
Ensure the gas supply lines are installed correctly. Open the supply valves to pressurize the system and
2
confirm the O
and N
2
3
Connect a new medical breathing circuit with mask or mouthpiece to the device.
4
Take a breath of 100% O
5
Turn the O
gas supply valve OFF.
2
While monitoring the O
6
the resistance against inhalation once the O
If the resistance against inhalation only slightly increases as the gauge drops, the emergency air intake valve
7
is functioning properly. If it becomes significantly difficult to inhale from the mask or mouthpiece, the emergency
air intake valve may be malfunctioning; if so, contact your authorized distributor for service and troubleshooting.
FM-1533, Rev -
13JAN2025
Nitronox Plus Quick Start Guide
O knob to 50% (no setting
2
and N
O inlet pressure gauges
2
and N
O inlet pressure gauges
2
leaks
persist,
contact
for
service
O inlet pressure gauges are within the green band.
2
from the device. Note the level of resistance against inhalation.
2
inlet pressure gauge, continue breathing through the mask or mouthpiece. Observe
2
Failsafe Test
To test the failsafe mechanism, ensure the N
key lock is in the ON position. For adjustable
1
models, set the %N
required for 50/50 model).
Ensure gas supply lines are installed. Open the
supply valves to pressurize the system and
2
confirm the O
within the green band and the mixture pressure
gauge is within the green band.
3
Turn OFF the O
While observing the O
manually activate the demand valve by pushing
on the back of the demand valve (where it says:
4
"PRESS TO TEST") until the inlet pressure gauge
reads 0 psi. The O
activate briefly.
Continue manually activating the demand valve.
Confirm that the mixture pressure gauge drops to
5
0 psi and that no gas flow comes out of the
your
connection port while the O
and
If the mixture pressure does not deplete or the
device is able to flow gas while the O
6
off, contact your authorized distributor for service
and troubleshooting.
O key lock to the OFF position. For adjustable models, set
2
supply gauge depletes to 0 psi.
2
Parker Hannifin Corporation
245 Township Line Road
Hatfield, PA 19440 USA
Office 215-723-4000
O knob to 50% (no setting
2
and N
O inlet pressure gauges are
2
2
supply valve.
2
inlet pressure gauge,
2
low pressure alarm may
2
supply remains off.
2
Page 1 of 4
O
2
supply is
2

Advertisement

loading
Need help?

Need help?

Do you have a question about the Porter Nitronox Plus and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Parker Porter Nitronox Plus

  • Page 1 Parker Hannifin Corporation 245 Township Line Road Hatfield, PA 19440 USA Office 215-723-4000 Nitronox Plus Quick Start Guide 1. Pre-Check Note: To perform these tests, gas supply cylinders or gas supply shutoff valves are required in order to isolate the gas supply form the device.
  • Page 2 Parker Hannifin Corporation 245 Township Line Road Hatfield, PA 19440 USA Office 215-723-4000 Alarm Function Test To test the device alarms, ensure the N O Key Lock is in the ON position. For adjustable models, set the O knob to 50% (no setting required for 50/50 model).
  • Page 3 Parker Hannifin Corporation 245 Township Line Road Hatfield, PA 19440 USA Office 215-723-4000 3. Cleaning The Nitronox Plus must be cleaned between each use in order to prevent the spread of infections. Cleaning of the device has been validated with Super Sani-Cloth™ Germicidal wipes.
  • Page 4 Parker Hannifin Corporation 245 Township Line Road Hatfield, PA 19440 USA Office 215-723-4000 WARNING: Do not change the connection fitting type or diameter of the supply hoses. The Diameter Indexed Safety System (DISS) is designed to prevent misconnection of N O and O supply lines.