Parker Matrx DIGITAL MDM Cabinet Mount Installation Instructions
Parker Matrx DIGITAL MDM Cabinet Mount Installation Instructions

Parker Matrx DIGITAL MDM Cabinet Mount Installation Instructions

Hide thumbs Also See for Matrx DIGITAL MDM:

Advertisement

Parker Hannifin Corporation
Porter Instrument
245 Township Line Road
Hatfield, PA 19440 USA
Tel: 215-723-4000
Ref. 10550800 Rev. F

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Matrx DIGITAL MDM and is the answer not in the manual?

Questions and answers

Summary of Contents for Parker Matrx DIGITAL MDM

  • Page 1 Parker Hannifin Corporation Porter Instrument 245 Township Line Road Hatfield, PA 19440 USA Tel: 215-723-4000 Ref. 10550800 Rev. F...
  • Page 2 European Communities should contact the Authorized Representative listed below regarding any Medical Device Directive (MDD) inquiries. Direct inquiries may be made to the following addresses: Porter Regulatory Affairs: EU Authorized Representative: Parker Hannifin Corporation Parker Hannifin Ltd Porter Instrument Instrumentation Products Division...
  • Page 3 SAFETY SUMMARY WARNINGS The Matrx by Parker DIGITAL MDM® mixer is a specialized medical device and should be operated ONLY by a professional trained in the use of nitrous oxide. Nitrous oxide and oxygen mixtures should only be administered by or under the supervision of a medical professional.
  • Page 4: Table Of Contents

    Table of Contents Notes and Contact Information ................iv Introduction ......................1 Intended Use ......................1 User and Installer Responsibility ................1 System Notes ......................1 NFPA Requirements and Warnings................1 Section 1 Installation Instructions ..................3 Description ......................... 3 Planning ........................
  • Page 5 Table of Illustrations Figure 1. Outside View of a Typical Digital MDM® Cabinet Mount with Flush Mounted Bag Tee/Scavenging Valve ............... 3 Figure 2. Installation of Demand Valves ................... 4 Figure 3. Digital MDM® Mixer Head Configurations ..............5 Figure 4. Screw and Tee Nut Adjustment .................
  • Page 6: Notes And Contact Information

    Notes and Contact Information Unit Serial Number Date Purchased Dealer Name Dealer Phone Number Notes: Contact Information: Parker Hannifin Corporation Porter Instrument 245 Township Line Road Hatfield, PA 19440 USA Tel: 215-723-4000...
  • Page 7: Introduction

    Diameter Indexed Safety System (DISS). System Notes Matrx by Parker recommends the use of a permanently 3. Type K or L seamless, copper tubing or standard mounted oxygen analyzer in the patient circuit to weight brass piping is required.
  • Page 8 U.S. Gas Color Coding All products in this manual are manufactured according to the standard ANSI (USA) color coding. For products manufactured with other gas color-coding standards, call Matrx by Parker Customer Service at 1-888-723- 4001 for assistance. Color Oxygen...
  • Page 9: Installation Instructions

    This manual includes installation instructions 99C; see the Introduction for details. All equipment for an optional Matrx by Parker Scavenging System, as must be installed in accordance with these well as an optional gas outlet station. Figure 1 illustrates guidelines, as well as all local codes and a typical installation.
  • Page 10: Pre-Installation

     The outlet station can be surface mounted to the 4. Solder a Matrx shutoff valve to each gas line that will connect to the mixer head and the scavenging cabinet surface, flush mounted in the cabinet control valve; see Figures 2 and 9. For soldering surface, or mounted inside the cabinet.
  • Page 11: Figure 3. Digital Mdm® Mixer Head Configurations

    Figure 3. Digital MDM® Mixer Head Configurations. If an oxygen outlet station is to be installed next to the mixer head, it will be supplied assembled to the bezel. 1. Position the installation template at the desired location on the cabinet surface. 2.
  • Page 12: Bag Tee And Scavenging Control Valve Installation

    Figure 5. Digital MDM® Mixer Head Installation 11. With the (4) screws, secure the mixer head, the bezel, and mounting shelf to the cabinet surface. See Figures 4 and 5. 12. Tighten the (4) screws and snap the screw caps over the screw heads.
  • Page 13: Figure 7. Surface Mounting The Bag Tee And Scavenging Control Valve

    4. Drill the (2) 3/4” holes for the gas hoses. 7. Orient the hose barbs (they turn) parallel to the mounting plate. 5. From inside the cabinet, insert the (4) tee nuts into 8. Align the mounting plate with the (4) holes, and the (4) mounting holes.
  • Page 14: J Hook Installation

    Secure with the (4) screws. For gas compatibility, all hoses must be supplied by Matrx by Parker. These hoses are braided 7. Tighten the (4) screws and snap the screw caps to prevent occlusion when bent.
  • Page 15: Rubber Goods Installation

    Figure 9. Connections Schematic (some components may not be part of your installation). Rubber Goods Installation Install the rubber goods. See Figure 10. Figure 10. Installation of Rubber Goods.
  • Page 16: Installation Of Power Cord And Protective Earth (Ground) Connection

    Figure 11. Installation of Power Cord Installation of Power Cord and Protective Earth (ground) connection. Install the power cord at the rear of the mixer head. See Figure 11. Connect the power cord to the electrical outlet. WARNING For safety a Protective Earth (ground) connection must be made through power cord 3rd wire or Ground Lug (see Figure 11.) WARNING...
  • Page 17: Specifications

    If additional gas outlet stations are installed, complete the If the gas pipeline test fails at any step, the pipeline test procedures listed in the Matrx by Parker Outlet Stations must be repaired and tested again, starting at step 1.
  • Page 18: Section 4 Return Information And Warranty

    Matrx by Parker shall dealer. Units for repair should be sent to Matrx by Parker constitute a warranty by Matrx by Parker or give rise to any and packaged in the original shipping container if possible.
  • Page 19: Replacement Parts

    Section 5 Replacement Parts The following replacement parts may be obtained from Digital MDM®: Matrx Part No. Description 20183200 Power Cord (USA) 10545100 Instruction Manual 63882537 Fuse: 2.5 amp, 250 volt, slo-blo (time delay) 5 mm by 20 mm (Note: international 220-240 volt units require two fuses) 91515087 Autoclavable nitrous scavenging assembly Adult...
  • Page 20 Matrx Part No. Description Hoses Black Scavenging hose, 1/4” I.D., 10’ long 91525143 Bag tee hose, 5/16” I.D., 10’ long 91525144 Ohio Male Quick Connect 92305126 Ohio Quick Connect male X DISS female Oxygen 92305127 Ohio Quick Connect male X DISS female Nitrous Oxide 92515063 Ohio Quick Connect male X 1/4 NPT...