Fluke 283 FC Calibration Information Manual
Fluke 283 FC Calibration Information Manual

Fluke 283 FC Calibration Information Manual

True-rms 1500v multimeter
Hide thumbs Also See for 283 FC:

Advertisement

Introduction
To prevent possible electrical shock, fire, or personal injury, read all safety
information before you use the Product.
The 283 FC True-RMS 1500V Multimeter (the Meter or Product) is a True-RMS Digital Multimeter.
This manual contains the verification and calibration adjustment procedures for the Product. For
complete operating instructions, routine maintenance procedures, and replaceable parts, see
the Users Manual at www.fluke.com.
Contact Fluke
Fluke Corporation operates worldwide. For local contact information, go to our website:
www.fluke.com.
To register your product, or to view, print, or download the latest manual or manual supplement,
go to our website: www.fluke.com/productinfo.
Safety Information
General Safety Information is in the printed Safety Information document that ships with the
Product and at www.fluke.com. More specific safety information is listed where applicable.
A Warning identifies hazardous conditions and procedures that are dangerous to the user. A
Caution identifies conditions and procedures that can cause damage to the Product or the
equipment under test.
November 2024
Fluke Corporation. All rights reserved.
©2024
Specifications are subject to change without notice.
All product names are trademarks of their respective companies.
True-RMS 1500V Multimeter
Calibration Information
XW Warning
283 FC
1

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the 283 FC and is the answer not in the manual?

Questions and answers

Summary of Contents for Fluke 283 FC

  • Page 1 To prevent possible electrical shock, fire, or personal injury, read all safety information before you use the Product. The 283 FC True-RMS 1500V Multimeter (the Meter or Product) is a True-RMS Digital Multimeter. This manual contains the verification and calibration adjustment procedures for the Product. For complete operating instructions, routine maintenance procedures, and replaceable parts, see the Users Manual at www.fluke.com.
  • Page 2: Specifications

    283 FC Calibration Information Specifications For complete specifications, refer to the Users Manual at www.fluke.com. Performance Tests XW Warning To prevent possible electrical shock, fire, or personal injury, do not perform the performance test procedures unless the Product is fully assembled.
  • Page 3 The function performance tests and calibration adjustment require the equipment in Table XW Warning To prevent possible electrical shock, fire, or personal injury, see the Safety Information document for the 5560A Calibrator available at www.fluke.com. Table 1. Required Equipment Recommended Equipment Measurement Function...
  • Page 4 283 FC Calibration Information Figure 1. Performance Test Connections Table 2. Performance Tests Test (Switch Range Input Lower Limit Upper Limit Units Position) 5 V 45 Hz 4.947 5.053 5 V 1 kHz 4.897 5.103 6.000 V 3 V 45 Hz 2.967...
  • Page 5 True-RMS 1500V Multimeter Performance Tests Table 2. Performance Tests (cont.) Test (Switch Range Input Lower Limit Upper Limit Units Position) 999.9Hz 1 V 900 Hz 898.9 901.1 99.99kHz 5 V 50 kHz 49.93 50.07 Volts AC, Frequency 0.01 V 0.007 0.013 2.994 3.006...
  • Page 6 283 FC Calibration Information Table 2. Performance Tests (cont.) Test (Switch Range Input Lower Limit Upper Limit Units Position) 10 nF 1000 nF 900 nF 10.00 μF 9 μF 8.87 9.13 100.0 μF 30 μF 29.6 30.6 Capacitance 300 μF μF...
  • Page 7: Before Calibration Adjustment

    V dc UPPER Wireless Verification To verify the Product wireless connection and that Fluke Connect connects to the FC APP, see the Users Manual. Before Calibration Adjustment Before the Product calibration process, you must put the Product into calibration mode and enter your password.
  • Page 8: Calibration Adjustment

    283 FC Calibration Information Enter the Password Enter the Product password: 1. Push . The adjustment counter shows, for example: N000. 2. Push .The display shows ????. Enter the factory default password: 1234. The top row of buttons corresponds to each number: a.
  • Page 9: Maintenance And Parts

    600 V C-08 60 V, 60 Hz C-09 600 Ω C-10 6 kΩ C-11 60 kΩ C-12 600 kΩ C-13 6 MΩ C-14 30 MΩ Maintenance and Parts For routine maintenance and replaceable parts, see the Users Manual available from www.fluke.com.
  • Page 10 Lifetime Limited Warranty Each Fluke 20, 70, 80, 170, 180 and 280 Series DMM will be free from defects in material and workmanship for its lifetime. As used herein, “lifetime” is defined as seven years after Fluke discontinues manufacturing the product, but the warranty period shall be at least ten years from the date of purchase.

Table of Contents