Belluscura X-PLO2R Troubleshooting Manual

Portable oxygen concentrator
Hide thumbs Also See for X-PLO2R:

Advertisement

Quick Links

X-PLO
R™
2
PORTABLE OXYGEN
CONCENTRATOR
TROUBLESHOOTING
GUIDE
Rev. LBL 5050-03
PAGE 1 OF 13
X-PLO
R™ POC Troubleshooting Guide
2

Advertisement

loading
Need help?

Need help?

Do you have a question about the X-PLO2R and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Belluscura X-PLO2R

  • Page 1 X-PLO R™ PORTABLE OXYGEN CONCENTRATOR TROUBLESHOOTING GUIDE Rev. LBL 5050-03 PAGE 1 OF 13 X-PLO R™ POC Troubleshooting Guide...
  • Page 2: Shutting Down

    Stated in the IFU Message Displayed on Possible Causes Equipment Page 18 -19 Displayed in System (Internal Use) Provider Actions Text *Not exact just Condition/Action/ an example* Explanation On/Off has been pressed for two The patient has pushed the power button for 2 seconds seconds.
  • Page 3 Stated in the IFU Page 18 -19 Message Displayed in Displayed on System Possible Causes Equipment Condition/Action/Explanation Text *not exact just an (Internal Use) Provider Actions example Battery power is low, with less than Battery Low – Charge the battery with 20% of charge remaining.
  • Page 4 Stated in the IFU Page 18 -19 Message Displayed Displayed on System Possible Causes Equipment Condition/Action/Explanation in Text *not exact just an (Internal Use) Provider Actions example Concentrator has insufficient battery Battery Empty – Attach power supply or power to produce oxygen. Attach Attach Power external power supply or exchange swap for a charged...
  • Page 5: Replace Battery

    Stated in the IFU Page 18 -19 Message Displayed in Displayed on Possible Causes Equipment Condition/Action/Explanation Text System (Internal Use) Provider Actions *Not exact just an example* If in warranty RMA Battery will not charge should be initiated. Battery error has occurred. Stop but should work as a IF out of warranty a new using the battery and switch to a...
  • Page 6 Stated in the IFU Page 18 -19 Message Displayed in Text Displayed on Possible Causes Equipment Condition/Action/Explanation System (Internal Use) Provider Actions *Not exact just an example* POC temperature is too high. Ensure air intake and outlet vents Patient has incorrectly placed the system in the bag. Ensure system is put into the bag have clear access and particle filters are clean.
  • Page 7 Stated in the IFU Page 18 -19 Message Displayed Displayed on Possible Causes Equipment Condition/Action/Explanation in Text System (Internal Use) Provider Actions *Not exact just an example* Move to a warmer area. System could take some time to warm-up. POC temperature is too cold. Try moving Internal temperature has System Cold to a warmer environment.
  • Page 8 Stated in the IFU Page 18 -19 Message Displayed in Displayed on System Possible Causes Equipment Condition/Action/Explanation Text *Not exact just an (Internal Use) Provider Actions example* Replace the sieve cartridge as soon Replace the sieve Replace Sieve as possible. Contact your cartridge.
  • Page 9: External Power

    Condition/Action/Explanation Text *Not exact just an (Internal Use) Provider Actions example* Ensure all power supplies are fully assembled Two AC cords are not connected. (AC cord to AC brick). RMA should be initiated after above troubleshooting. Complete a full power down cycle by shutting the system down, removing the power supply from the battery, removing the battery from Power cycling could reset the alarms.
  • Page 10 Stated in the IFU Page 18 -19 Message Displayed in Displayed on System Possible Causes Equipment Condition/Action/Explanation Text *Not exact just an (Internal Use) Provider Actions example* NO BREATH DETECTED – CHECK CANNULA Check cannula and ensure it is fully seated on the cannula barb.
  • Page 11 Stated in the IFU Page 18 -19 Message Displayed in Displayed on System Possible Causes Equipment Condition/Action/Explanation Text *Not exact just an (Internal Use) Provider Actions example* Insert cartridge, without dust cap. No sieve cartridge is inserted, O2 concentration is below 21%.
  • Page 12 Stated in the IFU Page 18 -19 Message Displayed in Displayed on System Possible Causes Equipment Condition/Action/Explanation Text *Not exact just an (Internal Use) Provider Actions example* Complete a full power cycle. Ensure the power supply is fully plugged into the battery and outlet.