Teledyne DALSA Rad-icon 2022 Detector User Manual CameraLink Model December 30, 2015 www.teledynedalsa.com/ndt...
Page 2
Please feel free to contact the global TD-PI sales office for ord ering, obtaining m ore d etailed inform ation or in case you encou nter any p roblem w hen op erating the d etector: Teledyne D ALSA – Rad-icon Imaging 3400 Garrett Drive Santa Clara, CA 95054...
Page 3
Readout modes ....................23 Synchronization modes .................... 24 Free running mode, Full Field of View ............... 24 External trigger mode, Full Field of View ............25 Snap shot mode ....................26 Shuttering ......................... 27 Diagnostics ....................... 28 Teledyne DALSA Rad-icon 2022 Detector...
Page 4
Set Offset ......................44 Get Offset ......................44 Dark Current after Extended Standby ..............45 Mode switches ......................45 Error Handling ......................45 Mechanical Interface ____________________________________________ 46 Appendix A ____________________________________________________ 47 Revision History ________________________________________________ 47 Rad-icon 2022 Detector Teledyne DALSA...
Page 5
To ensu re p ersonnel and p atient safety, it is necessary that a rad iation m eter w ill be u sed to check for rad iation leakage after installation of the d igital cam era in the u ser‟s system . Teledyne DALSA Rad-icon 2022 Detector...
Page 6
The d etector can influ ence the p erform ance of very sensitive equ ip m ent . Strong EMC d istu rbances can influ ence the d etector ‟s p erform ance. Rad-icon 2022 Detector Teledyne DALSA...
Page 7
N ote: For transp ort and storage, the stand ard p rod u ct p ackaging is assu m ed to be in p lace w hich inclu d es a sealed anti-static shield bag. Teledyne DALSA Rad-icon 2022 Detector...
Page 9
Program m able exp osu re tim e, line rate and trigger m od es. Typical Applications N on Destru ctive Testing Electronics Insp ection Com p u ted Tom ograp hy (CT) X-ray Crystallograp hy Teledyne DALSA Rad-icon 2022 Detector...
Page 10
One 2-p in Lem o connector for p ow er Lem o EGG.2B.302 Mating connector: straight typ e: FGG.2B.302.CLAD52 One 4-p in Lem o connector for triggering Lem o EGG.2B.304 Mating connector: straight typ e: FGG.2B.304.CLAD52 Cam Link2 (fu ll) Cam Link1 (base) Rad-icon 2022 Detector Teledyne DALSA...
Page 11
The d etector requ ires a single voltage inp u t (+12V/ m ax 3A). There is a p rotection for m isconnection, via a resettable fu se. When the fu se has been activated (isolating), Teledyne DALSA Rad-icon 2022 Detector...
Page 12
The trigger circu itry insid e the d etector is show n in the figu re below . 4pin LEMO Connector Trigger Output Collector + BAT54J 100p ESD protection 4 x ESD12VS1UA Trigger Output Emitter - TCMT1600 FPGA Trigger Input Anode + 100p Trigger Input Cathode - TCMT1600 Shield Rad-icon 2022 Detector Teledyne DALSA...
Page 13
The recom m end ed LED forw ard current is 5m A, w hich is obtained w ith 5V input voltage. Absolute m axim um input voltage is 10.0V. Do not apply an input voltage w ith reversed polarity. Teledyne DALSA Rad-icon 2022 Detector...
Page 14
Connect a p ow er cable from the d etector to a +12V p ow er su p p ly. Software Installation Refer to the next chap ter for instru ctions on installing and ru nning the d etector. Rad-icon 2022 Detector Teledyne DALSA...
Page 15
A link to d ow nload the softw are w ill be em ailed to you a fter su bm itting the ap p lication. Com p lete the d ow nload p rocess and ru n the installation file (e.g. “Sap eraLT80xSDKSetu p .exe”). Teledyne DALSA Rad-icon 2022 Detector...
Page 16
3096. The actu al form at of 2064 colu m ns by 2236 row s is d efined in the ROI settings tab . Rad-icon 2022 Detector Teledyne DALSA...
Page 17
COM p orts. You w ill be asked to reboot the system after saving settings. Serial com m u nication can be m ad e u sing follow ing settings: Bau d rate: 115200 Data bits: 8 Parity: N one Stop bits: 1 Flow control: none Teledyne DALSA Rad-icon 2022 Detector...
Page 18
Select the correct d evice and check the vid eo statu s to verify if the d ete ctor is connected to the system . Pixel Clock, Frame V alid and Line V alid signals all have to be p resent: Load the correct fram e grabber configu ration file (available on requ est). Rad-icon 2022 Detector Teledyne DALSA...
Page 19
To stop a Grab p rocess the u ser m u st p ress the Freeze bu tton. After the Freeze bu tton is p ressed the d etector ignores any ad d itional triggers and no ad d itional im ages are acqu ired . Teledyne DALSA Rad-icon 2022 Detector...
Page 20
On p osition 1032 (cou nting from 0) a d u m m y colu m n m ay be inserted . A bu tting row is N OT inserted . If requ ired , this shou ld be d one in softw are after the im age is acqu ired . Rad-icon 2022 Detector Teledyne DALSA...
Page 21
Please refer to the d atasheet for exact fu ll w ell cap acity valu es. See chap ter Fu ll Well selection for m ore com m and d etails. Teledyne DALSA Rad-icon 2022 Detector...
Page 22
X-ray sou rce. The rising ed ge of the synchronization signal d eterm ines the start of the vertical blanking w ind ow . Rad-icon 2022 Detector Teledyne DALSA...
Page 23
1548 (cou nting from row 0). In the horizontal d irection a d u m m y colu m n can be inserted in Colu m n 1032 (cou nting from 0). Teledyne DALSA Rad-icon 2022 Detector...
Page 24
≥ 15.5 ms B: readout time 30 ms 15 ms C: Extended Exposure / A – B A – B Xray window Figu re 3, Tim ing d iagram 'FreeRu nning, Fu ll Field of View ' Rad-icon 2022 Detector Teledyne DALSA...
Page 25
For the trigger ou tp u t this m eans that the op tocou p ler is acting as a sw itch: When trigger ou tp u t is active (see d iagram s above) then the sw itch is cond u cting. Teledyne DALSA Rad-icon 2022 Detector...
Page 26
B – A (trigger active high – B – A (trigger active high – reset time) reset time) D: Trigger active low > 31 ms > 15.5 ms Figure 5, Timing diagram 'Snapshot Mode, Full Field of View' Rad-icon 2022 Detector Teledyne DALSA...
Page 27
X-rays from the p u lse are not incid ent on the d etector d u ring read ou t. Incorrect u se w ill cau se im aging artifacts that show u p as very u neven illu m ination w ithin the im ages. Teledyne DALSA Rad-icon 2022 Detector...
Page 28
LED states. Color Of Status LED Meaning Orange Pow er ap p lied , Stand by m od e Green Detector is op erational N o p ow er Table 1, LED status Rad-icon 2022 Detector Teledyne DALSA...
Page 29
1s. The p ow er consu m p tion in Stand by m od e is less than 50% of t he p ow er consu m p tion d u ring Pow erOn m od e. Please also see Dark Cu rrent after Extend ed Stand by. Teledyne DALSA Rad-icon 2022 Detector...
Page 30
The follow ing p aram eter conventions are u sed in the m anu al: = integer valu e = real nu m ber m = m em ber of a set = string Exam p le: to retu rn the cu rrent d etector settings gcp <CR> Rad-icon 2022 Detector Teledyne DALSA...
Page 31
For qu ick help , the cam era can be qu eried seeking list of all ava ilable Purpose com m and s and p aram eters. Access Type Read Only Syntax help <CR> Parameter N one Return Value Example Teledyne DALSA Rad-icon 2022 Detector...
Page 32
To id entify the m anu factu rer of the d etector Read Only Access Type Syntax gcn<CR> N one Parameter Return Value Teled yne Dalsa: Manu factu rer User Enters – gcn<CR> Detector Retu rns – Example Teled yne Dalsa: Manu factu rer Rad-icon 2022 Detector Teledyne DALSA...
Page 33
To id entify the d etector serial nu m ber Access Type Read Only Syntax gcs<CR> Parameter N one Return Value nnnnnn: Serial N u m ber User Enters – gcs<CR> Detector Retu rns – Example 123: Serial N u m ber Teledyne DALSA Rad-icon 2022 Detector...
Page 34
Purpose d etector Access Type Read Only Syntax gcv<CR> Parameter N one Return Value vvv.w w .xx.yyy.zzz: Firm w are Version User Enters – gcv<CR> Detector Retu rns – Example 135.00.00.003.003: Firm w are Version Rad-icon 2022 Detector Teledyne DALSA...
Page 35
Access Type Read Only Syntax gd m <CR> Parameter N one “0:User”, “1: Maintenance” or Return Value “2: Prod u ction” User Enters – gd m <CR> Detector Retu rns – 0: User Example Teledyne DALSA Rad-icon 2022 Detector...
Page 36
Access Type Read Only Syntax gp m <CR> Parameter N one “0: N orm al” or “1: Stand by” Return Value User Enters – gp m <CR> Detector Retu rns – Example 0: N orm al Rad-icon 2022 Detector Teledyne DALSA...
Page 37
4: Alternating 3FFF and 0000 5: Fram e cou nter 6: Constant valu e of 0x00BF 7: Constant valu e of 0x00BF Parameter D efault Value 0 0: Sensor Data w ith Low Full Well Return Value Example Teledyne DALSA Rad-icon 2022 Detector...
Page 38
Parameter N one <o1> <o2>: Im age sou rce Return Value See svs com m and for <o1> and <o2> User Enters – gvs<CR> Detector Retu rns – Example 0 0: Im age sou rce Rad-icon 2022 Detector Teledyne DALSA...
Page 39
Return Value User Enters – gfw <CR> Detector Retu rns – Example 1: H igh Fu ll Well N ote: to select a fu ll w ell u se the svs com m and . Teledyne DALSA Rad-icon 2022 Detector...
Page 40
<CR> Parameter N one “0: Free Ru nning”, “1: External Trigger” or Return Value “2: Snap shot m od e” User Enters – gtm <CR> Detector Retu rns – Example 0: Free Ru nning Rad-icon 2022 Detector Teledyne DALSA...
Page 41
Access Type Read Only Syntax gvm <CR> Parameter N one “0: Fu ll Resolu tion” or Return Value “1: Flexible ROI” User Enters – gvm <CR> Detector Retu rns – Example 0: Full Resolu tion Teledyne DALSA Rad-icon 2022 Detector...
Page 42
Get Binning Mod e Access Type Read Only gbn<CR> Syntax Parameter N one “1: Binning enabled “ or “0: Binning d isabled ” Return Value User Enters – gbn<CR> Detector Retu rns – Example 0: Binning d isabled Rad-icon 2022 Detector Teledyne DALSA...
Page 43
Parameter N one Return Value <tim e>: Extend ed Exp osu re Valu e [m s]: User Enters – get<CR> Detector Retu rns – Example 30: Extend ed Exp osu re Valu e [m s] Teledyne DALSA Rad-icon 2022 Detector...
Page 44
“1: Offset 1 p ct 95”, “2: Offset 3 p ct 9” or “3: Offset 6 p ct 25” Return Value User Enters – got<CR> Detector Retu rns – Example 0: Offset 0 p ct Rad-icon 2022 Detector Teledyne DALSA...
Page 45
„Error: Incorrect nu m ber of p aram eters‟. When an incorrect valu e is send to the d etector, the d etector w ill resp ond w ith a resp onse like „Error: Invalid Im age Sou rce‟. Teledyne DALSA Rad-icon 2022 Detector...
Page 46
User Manual Mechanical Interface 221 x 204 mm Active Area 235 mm 226.0 mm 50.0 mm 11 mm 55.6 mm 289 mm 12 mm 95 mm 141 mm 174 mm 190 mm 206 mm Rad-icon 2022 Detector Teledyne DALSA...
Page 47
User Manual Appendix A Revision History Revision Change Descrip tion Revision Date N u m ber Draft release June 29, 2015 Upd ated connectors, softw are, m echanical d raw ing Dec. 30, 2015 Teledyne DALSA Rad-icon 2022 Detector...