Dear Customer,
Thank you for choosing a Hanna Instruments Product.
Please read this instruction manual carefully before using the instrument. If you need
HI 711
additional technical information, do not hesitate to e-mail us at tech@hannainst.com.
Preliminary examination:
Total Chlorine
Please examine this product carefully. Make sure that the instrument is not damaged. If
any damage occured during shipment, please notify your Dealer.
Each HI 711 meter is supplied complete with:
• Two Sample Cuvettes and Caps
• Six powder reagents for Total Chlorine
• 1 x 1.5V AAA Battery
• Instruction Manual
Range
Resolution
Accuracy
Typical EMC Dev.
Light Source
Light Detector
Method
Environment
Battery Type
Auto-Shut off
Dimensions
Weight
w w w . h a n n a i n s t . c o m
For more details about spare parts and accessories
see "Accessories".
Technical specifications:
0.00 to 3.50 ppm
0.01 ppm
±0.03 ppm ±3% of reading @ 25°C
±0.01 ppm
Light Emitting Diode @ 525 nm
Silicon Photocell
Adaptation of USEPA method 330.5. The reaction
between the chlorine and DPD reagent causes a pink
tint in the sample.
0 to 50°C (32 to 122°F);
max 95% RH non-condensing
1 x 1.5V AAA
After 2 minutes of non-use and 10 seconds after reading.
81.5 x 61 x 37.5 mm (3.2 x 2.4 x 1.5")
64 g (2.25 oz.).
Functional description:
1. Cuvette cap.
2. Cuvette with cap.
3. Cuvette holder.
4. Liquid Crystal Display.
5. Button
Errors and warnings:
Light High: There is too much light to perform a measurement.
Please check the preparation of the zero cuvette.
Light Low: There is not enough light to perform a measurement.
Please check the preparation of the zero cuvette.
Inverted cuvettes: The sample and the zero cuvette are inverted.
Under range: A blinking "0.00" indicates that the sample absorbs
less light than the zero reference. Check the procedure and make sure
you use the same cuvette for reference (zero) and measurement.
Over Range: A flashing value of the maximum concentration indicates
an over range condition. The concentration of the sample is beyond
the programmed range: dilute the sample and re-run the test.
Battery low: The battery must be replaced soon.
Dead battery: This indicates that the battery is dead and must be
replaced. Once this indication is displayed, normal operation of the
instrument will be interrupted. Change the battery and restart the
meter.

Advertisement

loading
Need help?

Need help?

Do you have a question about the HI 711 and is the answer not in the manual?

Questions and answers

Summary of Contents for Hanna Instruments HI 711

  • Page 1 Please examine this product carefully. Make sure that the instrument is not damaged. If 3. Cuvette holder. any damage occured during shipment, please notify your Dealer. Each HI 711 meter is supplied complete with: 4. Liquid Crystal Display. • Two Sample Cuvettes and Caps 5.
  • Page 2 “bAd” then ”bAt” each for 1 second and turns off. • Remove the cuvette, open it and add the Hanna Instruments reserves the right to modify the design, construction and To restart the instrument, the battery must be replaced with a fresh one.