Process pressure measurement
pressure transmitter with ceramic and metal sensors
overload-resistant and function-monitored; communication via
hart, profibus pa or foundation fieldbus (96 pages)
Pressure transmitter with ceramic and metal sensors overload-resistant and function-monitored; communication via hart profibus pa or foundation fieldbus (84 pages)
Summary of Contents for Endress+Hauser HART Cerabar S PMC71
Page 1
Products Solutions Services BA00271P/00/EN/20.22-00 71556690 2022-01-06 Valid as of software version: 02.30.zz Operating Instructions Cerabar S PMC71, PMP71, PMP75 Process pressure measurement HART...
Page 2
The manufacturer reserves the right to modify technical data without prior notice. Your Endress+Hauser distributor will supply you with current information and updates to this manual. Endress+Hauser...
This may be a connections. potential equalization line or a star grounding system depending on national or company codes of practice. 1.2.3 Tool symbols Symbol Meaning Allen key A0011221 Open-ended wrench A0011222 Endress+Hauser...
Registered label of E.I. Du Pont de Nemours & Co., Wilmington, USA ® TRI-CLAMP Registered label of Ladish & Co., Inc., Kenosha, USA ® HART Registered trademark of the FieldComm Group, Austin, USA ® GORE-TEX Trademark of W.L. Gore & Associates, Inc., USA Endress+Hauser...
The manufacturer is not liable for damage caused by improper or unintended use. Clarification for borderline cases: In the case of special fluids and fluids used for cleaning, Endress+Hauser is glad to provide assistance in clarifying the corrosion resistance of wetted materials, but does not accept any warranty or liability.
It meets general safety standards and legal requirements. It also conforms to the EC directives listed in the device-specific EC Declaration of Conformity. Endress+Hauser confirms this by affixing the CE mark to the device.
Scope of delivery The scope of delivery comprises: • Cerabar S pressure transmitter • For devices with the "HistoROM/M-DAT" option: CD-ROM with Endress+Hauser operating program • Optional accessories Documentation supplied: • The BA00271P and BA00274P Operating Instructions are available on the Internet.
The devices comply with the applicable standards and regulations as listed in the EC Declaration of Conformity and thus comply with the statutory requirements of the EC Directives. Endress+Hauser confirms the successful testing of the device by affixing to it the CE mark.
The measuring instrument must be stored in a dry, clean area and protected against damage from impact (EN 837-2). Storage temperature range: See Technical Information. Installation conditions 4.2.1 Installation dimensions For dimensions, please refer to "Mechanical construction" section in TI00383P. Endress+Hauser...
• It is possible to rotate the housing by up to 380° to ensure optimum readability of the on- site display. → ä 19, "Rotating the housing". • Endress+Hauser provides a mounting bracket for installation on pipes or walls. → ä 16, "Wall and pipe mounting (optional)".
• The device must be installed as follows in order to comply with the cleanability requirements of the ASME-BPE (Part SD Cleanability): Pressure measurement in gases ➀ ➁ P01-PMx7xxxx-11-xx-xx-xx-001 Fig. 1: Measuring arrangement for pressure measurement in gases Cerabar S Shutoff device Endress+Hauser...
Page 13
• Damping of pressure shocks • The defined water column only causes minimal (negligible) measurement errors and minimal (negligible) thermal effects on the device For technical data (e.g. materials, dimensions or order numbers), see the accessory document SD01553P. Endress+Hauser...
The zero point shift can be corrected. • Do not clean or touch the process membrane of the diaphragm seal with hard or pointed objects. • Do not remove the protection on the process membrane until just before installation. Endress+Hauser...
The seal is not allowed to press against the process membrane as this could affect the measurement result. ‣ Ensure that the seal is not touching the process membrane. A0017743 Fig. 5: Process membrane Seal 4.4.4 Installation with heat insulation – PMC71 high temperature version and PMP75 See Technical Information. Endress+Hauser...
Mounting Cerabar S PMC71, PMP71, PMP75 with 4...20 mA HART 4.4.5 Wall and pipe mounting (optional) Endress+Hauser offers a mounting bracket for installation on pipes or walls (for pipe diameters from 1 ¼" to 2"). ø42...60 (1.65...2.36) 122 (4.8) 140 (5.51) 158 (6.22)
Mount the housing on a wall or pipe using the mounting bracket (item 7). In the case of pipe mounting, the nuts on the bracket must be tightened uniformly with a torque of at least 5 Nm (3.69 lbf ft). Mount the cable with a bending radius (r) ≥120 mm (4.72 in). Endress+Hauser...
Page 18
For the "U1 Prepared for diaphragm seal mount" version in feature 70 "Process connection; Material" in the order code up to, and including, 40 bar sensors (600 psi), Endress+Hauser recommends welding on the diaphragm seal as follows: the total welding depth of the fillet weld is 1 mm (0.04 in) at an external diameter of 16 mm (0.63 in).
The covers for the terminal compartment and electronics compartment are hooked into the housing and closed with a screw in each case. These screws should be finger-tightened (2 Nm (1.48 lbf ft)) to the stop to ensure that the covers sit tightly. Endress+Hauser...
Mounting Cerabar S PMC71, PMP71, PMP75 with 4...20 mA HART Post-installation check After installing the device, carry out the following checks: • Are all screws firmly tightened? • Are the housing covers screwed down tight? Endress+Hauser...
Minimum supply voltage = 10.5 V DC, jumper is inserted in accordance with the illustration. Minimum supply voltage = 11.5 V DC, jumper is inserted in "Test" position. Devices with integrated overvoltage protection are labeled OVP (overvoltage protection) here. Endress+Hauser...
Page 22
Connection of devices with M12 plug PIN assignment for M12 plug Meaning Signal + Not assigned Signal – Ground A0011175 5.1.3 Cable version connection – gnye 4...20 mA P01-PMx4xxxx-04-xx-xx-xx-010 Fig. 11: rd = red, bk = black, gnye = green/yellow Endress+Hauser...
(20 to 14 AWG) • External ground terminal: 0.5 to 4 mm (20 to 12 AWG) 5.2.3 Cable specification • Endress+Hauser recommends using twisted, shielded two-wire cables. • Outer cable diameter: 5 to 9 mm (0.2 to 0.35 in) Endress+Hauser...
Page 24
Ex systems as standard. 5.2.6 Connecting Field Xpert SFX100 Compact, flexible and robust industry handheld terminal for remote configuration and for obtaining measured values via the HART current output (4-20 mA). For details, refer to Operating Instructions BA00060S/04/EN. Endress+Hauser...
USB port. This allows remote operation of the transmitter using Endress+Hauser' s FieldCare operating program. Power is supplied to the Commubox via the USB port. The Commubox is also suitable for connecting to intrinsically safe circuits. → See Technical Information TI00404F for further information.
• Are all screws firmly tightened? • Are the housing covers screwed down tight? As soon as voltage is applied to the device, the green LED on the electronic insert lights up for a few seconds or the connected on-site display lights up. Endress+Hauser...
• Rapid and safe commissioning using Quick Setup menus Parameter Measured value display Identification Function name Value number Header line Main line Unit Information line Symbol – Bargraph Bargraph Editing modes Operating keys Selection options Value that can be edited Current measured value P01-xMx7xxxx-07-xx-xx-xx-001 Endress+Hauser...
Operating keys on the outside of the device under the Operating keys protective flap Slot for optional display Slot for optional HistoROM ® /M-DAT module DIP switch for locking/unlocking parameters relevant to the measured values DIP switch for damping on/off Green LED to indicate value being accepted Endress+Hauser...
– You are in the menu at a selection level: each time you press the keys simultaneously, you go up a level in the menu. Note: The terms function group, level and selection level are explained → ä 33, "Menu structure". Endress+Hauser...
Observe the Observe the Observe the input limits. input limits. input limits. Please note warning on Page → ä 42, "Commissioning". Endress+Hauser...
Page 31
SET LRV and SET URV are parameter names that are used for the on-site display or remote operation, such as FieldCare. WARNING Changing the measuring mode can affect the calibration data! This situation can result in product overflow. ‣ Check the calibration data if the measuring mode is changed. Endress+Hauser...
Page 32
(EMPTY value (FULL CALIB. CALIB. Please note warning on Page → ä 42, "Commissioning". Parameter name used for the on-site display or remote control, such as FieldCare. Endress+Hauser...
Select English with or "–". P01-xxxxxxxx-19-xx-xx-xx-033 1. Select to confirm. A in front of the menu text indicates the option that is currently active. (English is now selected as the menu language.) 2. Jump to the next item with . P01-xxxxxxxx-19-xx-xx-xx-034 Endress+Hauser...
Use to save the new value and exit editing mode. See next graphic. P01-xxxxxxxx-19-xx-xx-xx-031 The new value for the damping is now 30.0 s. – Jump to the next parameter with . – Use or to return to editing mode. P01-xxxxxxxx-19-xx-xx-xx-032 Endress+Hauser...
Page 35
Use the key to assign the value (400 mbar (6 psi)) to the GET URV parameter. The device confirms the calibration and jumps back to the parameter, here GET URV (see next graphic). P01-xxxxxxxx-19-xx-xx-xx-037 Switch to the next parameter with . P01-xxxxxxxx-19-xx-xx-xx-035 Endress+Hauser...
During the analysis, the messages "W702, HistoROM data not consistent" and "W706, Configuration in HistoROM and device not identical" can occur. For measures, see Page → ä 50, "Messages". Endress+Hauser...
MODE and the parameters in the POSITION ADJUSTMENT and PROCESS CONNECTION ® group are loaded into the device by the HistoROM /M-DAT. The device is restarted. ® Before removing the HistoROM /M-DAT again from the electronic insert, disconnect the device from the supply voltage. Endress+Hauser...
® Wait approx. 20 seconds. Configuration data are loaded from the HistoROM /M-DAT to the device. The device is restarted. ® Before removing the HistoROM /M-DAT from the electronic insert once again, disconnect the device from the supply voltage. Endress+Hauser...
Endress+Hauser operating program The FieldCare operating program is an Endress+Hauser asset management tool based on FDT technology. With FieldCare, you can configure all Endress+Hauser devices as well as devices from other manufacturers that support the FDT standard. You can find hardware and software requirements on the Internet: www.endress.com →...
ENTER RESET CODE parameter (menu path: (GROUP SELECTION →) OPERATING MENU → OPERATION). There are various reset codes for the device. The following table illustrates which parameters are reset by the particular reset codes. Operation must be unlocked to reset parameters (→ ä 39, Locking/unlocking operation). Endress+Hauser...
Page 41
Operation Any customer-specific configuration carried out at the factory is not affected by a reset (customer-specific configuration remains). If you want to change the customer-specific configuration set at the factory, please contact Endress+Hauser Service. Reset code Description and effect 1846 Display reset –...
• "Post-connection check" checklist → ä 26 . Selecting the language and measuring mode 7.3.1 On-site operation The LANGUAGE and MEASURING MODE parameters are on the 1st selection level. The following measuring modes are available: • Pressure • Level Endress+Hauser...
VALUE should be corrected. To correct the MEASURED VALUE to 0.0 mbar, you must enter the value 2.2 here. (The following applies: MEASURED VALUE = MEASURED VALUE – CALIB. OFFSET) – MEASURED VALUE (after entry for calib. offset) = 0.0 mbar – The current value is also corrected. Factory setting: Endress+Hauser...
For on-site operation, see also → ä 29, "Function of operating elements – on-site display connected" and → ä 33, "On-site operation – on-site display connected". Endress+Hauser...
• For a detailed parameter description and examples of parameters, see the Operating Instructions BA00274P "Cerabar S/Deltabar S/Deltapilot S, Description of device functions". WARNING Changing the measuring mode can affect the calibration data! This situation can result in product overflow. ‣ Check the calibration data if the measuring mode is changed. Endress+Hauser...
Page 46
% must not be in direct proportion to the measured pressure. A linearization table must be entered for the 2nd measured variable. The 2nd measured variable is assigned to the 1st mea- sured variable by means of this table. Endress+Hauser...
DAMPING VALUE Enter damping time (time constant τ). The damping affects the speed at which all subsequent elements, such as the on-site display, measured value and current output react to a change in the pressure. Endress+Hauser...
Page 48
– LEVEL SELECTION "Level Easy Pressure" and CALIBRATION MODE "Wet" – LEVEL SELECTION "Level Standard", LEVEL MODE "Linear" and CALIBRATION MODE "Wet" For on-site operation, see also → ä 29, "Function of operating elements – on-site display connected" and → ä 33, "On-site operation – on-site display connected". Endress+Hauser...
(1) free from contamination and water. P01-PMC71xxx-17-xx-xx-xx-001 Cleaning instructions Endress+Hauser provides flushing rings as an accessory to enable cleaning of the process membrane without removing the transmitter from the process. For further information, please contact your local Endress+Hauser Sales Center. 8.1.1...
• If the device detects a defect in the on-site display during initialization, special error messages are generated. For error messages, see → ä 57, "On-site display error messages". • For support and further information, please contact Endress+Hauser Service. • See also "Repair", "Repair of Ex-certified devices" and "Spare parts". Code...
Page 51
– Main electronics defect. – Replace main electronics. Warning Function C>Linearization – The linearization table is not – Add to or correct linearization (W602) check (C) curve not monotone increasing or decreasing table. Then accept linearization monotonically. table again. Endress+Hauser...
Page 52
– Main electronics defect. – Replace main electronics. 705 (A705) Alarm Failure (F) B>Measurement – Fault in the main electronics. – Briefly disconnect device from error the power supply. – Main electronics defect. – Replace main electronics. Endress+Hauser...
Page 53
– Replace sensor with a suitable sensor. – Unsuitable download carried out. – Check configuration and perform download again. 713 (A713) Alarm Function B>100% POINT – The sensor was replaced. – Carry out calibration again. check (C) level out of edit limits Endress+Hauser...
Page 54
– Perform reset (code 2710) and check (C) FULL CALIB. out of been changed. recalibrate the device. edit limits 723 (A723) Alarm Function B>MAX. FLOW out – FLOW-MEAS. TYPE has been – Carry out calibration again. check (C) of edit limits changed. Endress+Hauser...
Page 55
Factory the Pmax ALARM WINDOW – Change value for Pmax setting: parameter. ALARM WINDOW if necessary. Warning C (See also Operating Instructions BA00274P, parameter description Pmax ALARM WINDOW or these Operating Instructions) Endress+Hauser...
Page 56
– Select a device with a suitable measuring range. 741 (A741) Alarm Function B>TANK HEIGHT – LEVEL MIN or LEVEL MAX has – Perform reset (code 2710) and check (C) out of edit limits been changed. recalibrate the device. Endress+Hauser...
Replace on-site display. Initialization, VU Electr. Defect A114 Initialization, VU Electr. Defect A281 Initialization, VU Checksum Err. A110 Initialization, VU Checksum Err. A112 Initialization, VU Checksum Err. A171 Initialization Supply voltage too low. Set supply voltage to the correct value. Endress+Hauser...
122 for "Sensor connection number such as 613 for "Simulation is number such as 731 for "Pmax ALARM error, incorrect data". active". WINDOW undershot". Menu path: (GROUP SELECTION → OPERATING MENU →OUTPUT Menu path: (GROUP SELECTION →) OPERATING MENU → MESSAGES Endress+Hauser...
Page 59
SET MAX ALARM parameter. Note: When using the case "special", the behavior is limited to an over/underpressure in a range LRL -10% up to LRL -30% and URL +10% up to URL +30%. Factory setting: ALT. CURR. OUTPUT: Normal/NE43 Endress+Hauser...
(→ ä 50). Once you have suppressed this message using the key, the message with the next highest priority is displayed. You can use the key to suppress each message, one after the other. The ALARM STATUS parameter continues to display all the messages present. Endress+Hauser...
Explosion hazard! When repairing Ex-certified devices, please note the following: • Repairs to Ex-certified devices must be carried out by Endress+Hauser Service or by specialist personnel according to national regulations. • Relevant standards, national hazardous area regulations and Safety Instructions and Certificates must be observed.
"Russian" is included as a menu language by default. The menu language "Nederlands" is no longer supported. 06.2014 02.20.zz HART7 protocol revision has been implemented. 10.2017 02.30.zz Improvement of safety confirmation and menus in FieldCare and HART handheld terminal. Endress+Hauser...
Need help?
Do you have a question about the HART Cerabar S PMC71 and is the answer not in the manual?
Questions and answers