Cleaning The Collector And Pump Tubing - BD PureWick Urine Collection System Getting Started

Advertisement

Cleaning the collector
and pump tubing
Instructions for cleaning the collector tubing with elbow connector and pump tubing
Initial Rinse:
Disconnect the collector tubing from the
elbow connector and disconnect the elbow connector
from the lid. Rinse the disconnected tubing thoroughly
by running cool tap water through the inside of the
tubing. While rinsing, remove any excess dirt by wiping
the outside of the tubing and elbow connector with
low lint disposable wipes. While rinsing, flush the
tubing with cool tap water to remove any excess dirt.
Cleaning:
Prepare a soapy solution by mixing 1
teaspoon (approximately 5 mL) of dish soap with
approximately 4 L of cool tap water. Completely
submerge the tubing and elbow connector in the
solution and allow it to soak for a minimum of ten (10)
minutes. At the beginning of the soak time, flush the
inside of the tubing with the prepared solution. While
submerged, use a soft brush (e.g., toothbrush) to brush
all accessible areas of the tubing and elbow connector
for a minimum of one (1) minute to remove any visible
signs of debris or dirt.
Rinse:
Rinse the tubing and elbow connector
thoroughly with cool tap water until there is no visible
sign of the soap cleaning solution. While rinsing, flush
the inside of the tubing with the tap water.
Visual inspection:
Inspect the tubing to ensure that all
debris and dirt have been removed. If there is any sign
of debris or dirt, repeat the steps above until there is no
sign of debris or dirt remaining on or inside the tubing.
Disinfection:
Fully submerge the tubing and
elbow connector in 70% isopropyl alcohol
(IPA). Allow it to soak for a minimum of ten (10)
minutes. Flush the IPA through the tubing, prior
to starting the ten (10) minute time.
Rinse:
Rinse the tubing and elbow connector
thoroughly with cool tap water, ensuring that
the inside of the tubing is flushed with water.
Drying:
Dry the tubing and elbow connector
outer surface with a clean low lint towel or cloth
and allow the inside of the tubing to air dry for a
minimum of thirty (30) minutes at room
temperature.