Operating Instructions
Proline Promag H 300
Electromagnetic flowmeter
PROFINET over Ethernet-APL
Solutions
Services

Advertisement

loading
Need help?

Need help?

Do you have a question about the Proline Promag H 300 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Endress+Hauser Proline Promag H 300

  • Page 1 Products Solutions Services BA02106D/06/EN/02.24-00 71661691 2024-07-17 Valid as of version 01.00.zz (Device firmware) Operating Instructions Proline Promag H 300 Electromagnetic flowmeter PROFINET over Ethernet-APL advanced physical layer...
  • Page 2 • The manufacturer reserves the right to modify technical data without prior notice. Your Endress+Hauser sales organization will supply you with current information and updates to this manual. Endress+Hauser...
  • Page 3: Table Of Contents

    Proline Promag H 300 PROFINET over Ethernet-APL Table of contents Table of contents 6.1.2 Environmental and process About this document ....6 requirements ....
  • Page 4 Table of contents Proline Promag H 300 PROFINET over Ethernet-APL 8.3.10 User roles and related access 10.5.14 "Build-up index adjustment" wizard . authorization ....57 10.6 Advanced settings .
  • Page 5 13.2 Measuring and test equipment ... 173 13.3 Endress+Hauser services ....173 Repair ......174 14.1 General notes .
  • Page 6: About This Document

    About this document Proline Promag H 300 PROFINET over Ethernet-APL About this document Document function These Operating Instructions contain all the information required in the various life cycle phases of the device: from product identification, incoming acceptance and storage, to installation, connection, operation and commissioning, through to troubleshooting, maintenance and disposal.
  • Page 7: Tool Symbols

    Proline Promag H 300 PROFINET over Ethernet-APL About this document 1.2.4 Tool symbols Symbol Meaning Flat-blade screwdriver Allen key Open-ended wrench 1.2.5 Symbols for certain types of information Symbol Meaning Permitted Procedures, processes or actions that are permitted. Preferred Procedures, processes or actions that are preferred.
  • Page 8: Documentation

    • Device Viewer (www.endress.com/deviceviewer): Enter the serial number from the nameplate • Endress+Hauser Operations app: Enter serial number from nameplate or scan matrix code on nameplate. The following documentation may be available depending on the device version ordered: Document type...
  • Page 9: Safety Instructions

    Proline Promag H 300 PROFINET over Ethernet-APL Safety instructions Safety instructions Requirements for the personnel The personnel for installation, commissioning, diagnostics and maintenance must fulfill the following requirements: ‣ Trained, qualified specialists must have a relevant qualification for this specific function and task.
  • Page 10: Workplace Safety

    Verification for borderline cases: ‣ For special fluids and fluids for cleaning, Endress+Hauser is glad to provide assistance in verifying the corrosion resistance of fluid-wetted materials, but does not accept any warranty or liability as minute changes in the temperature, concentration or level of contamination in the process can alter the corrosion resistance properties.
  • Page 11: Device-Specific It Security

    Proline Promag H 300 PROFINET over Ethernet-APL Safety instructions Device-specific IT security The device offers a range of specific functions to support protective measures on the operator' s side. These functions can be configured by the user and guarantee greater in- operation safety if used correctly.
  • Page 12: Infrastructure Mode

    Safety instructions Proline Promag H 300 PROFINET over Ethernet-APL WLAN passphrase: Operation as WLAN access point A connection between an operating unit (e.g. notebook or tablet) and the device via the WLAN interface (→  66), which can be ordered as an optional extra, is protected by the network key.
  • Page 13: Product Description

    Proline Promag H 300 PROFINET over Ethernet-APL Product description Product description The device consists of a transmitter and a sensor. The device is available as a compact version: The transmitter and sensor form a mechanical unit. Product design N ic...
  • Page 14: Incoming Acceptance And Product

    • Enter the serial numbers from the nameplates in the Device Viewer (www.endress.com/deviceviewer): all the information about the device is displayed. • Enter the serial numbers from the nameplates into the Endress+Hauser Operations app or scan the DataMatrix code on the nameplate with the Endress+Hauser Operations app: all the information about the device is displayed.
  • Page 15: Incoming Acceptance And Product Identification

    Proline Promag H 300 PROFINET over Ethernet-APL Incoming acceptance and product identification 4.2.1 Transmitter nameplate 3 4 5 Order code: Ser. no.: Ext. ord. cd.: Date: A0029192  2 Example of a transmitter nameplate Manufacturer address/certificate holder Name of the transmitter...
  • Page 16: Sensor Nameplate

    Incoming acceptance and product identification Proline Promag H 300 PROFINET over Ethernet-APL 4.2.2 Sensor nameplate Order code: Ser. no.: Ext. ord. cd.: Date: A0029204  3 Example of sensor nameplate Name of sensor Manufacturer address/certificate holder Order code Serial number (Ser. no.) Extended order code (Ext.
  • Page 17: Symbols On The Device

    Proline Promag H 300 PROFINET over Ethernet-APL Incoming acceptance and product identification 4.2.3 Symbols on the device Symbol Meaning WARNING! This symbol alerts you to a dangerous situation. Failure to avoid this situation can result in serious or fatal injury. Please consult the documentation for the measuring instrument to discover the type of potential danger and measures to avoid it.
  • Page 18: Storage And Transport

    Storage and transport Proline Promag H 300 PROFINET over Ethernet-APL Storage and transport Storage conditions Observe the following notes for storage: ‣ Store in the original packaging to ensure protection from shock. ‣ Do not remove protective covers or protective caps installed on process connections.
  • Page 19: Measuring Devices With Lifting Lugs

    Proline Promag H 300 PROFINET over Ethernet-APL Mounting 5.2.2 Measuring devices with lifting lugs CAUTION Special transportation instructions for devices with lifting lugs ‣ Only use the lifting lugs fitted on the device or flanges to transport the device. ‣...
  • Page 20 Mounting Proline Promag H 300 PROFINET over Ethernet-APL A0042131 The device should ideally be installed in an ascending pipe. A0042317 Installation upstream from a down pipe NOTICE Negative pressure in the measuring pipe can damage the liner! ‣ If installing upstream of down pipes whose length h ≥ 5 m (16.4 ft): install a siphon with a vent valve downstream of the device.
  • Page 21 Proline Promag H 300 PROFINET over Ethernet-APL Mounting A0041088 Installation near pumps NOTICE Negative pressure in the measuring tube can damage the liner! ‣ In order to maintain the system pressure, install the device in the flow direction downstream from the pump.
  • Page 22 Mounting Proline Promag H 300 PROFINET over Ethernet-APL Orientation The direction of the arrow on the nameplate helps you to install the measuring device according to the flow direction (direction of medium flow through the piping). Orientation Recommendation Vertical orientation...
  • Page 23: Inlet And Outlet Runs

    Proline Promag H 300 PROFINET over Ethernet-APL Mounting A0028998 EPD electrode for empty pipe detection, available from ≥ DN 15 (½") Measuring electrodes for signal detection Measuring instruments with a nominal diameter < DN 15 (½") do not have an EPD electrode.
  • Page 24: Environmental And Process Requirements

    Mounting Proline Promag H 300 PROFINET over Ethernet-APL 6.1.2 Environmental and process requirements Ambient temperature range Transmitter Standard: –40 to +60 °C (–40 to +140 °F) Local display –20 to +60 °C (–4 to +140 °F), the readability of the display may be impaired at temperatures outside the temperature range.
  • Page 25: Special Mounting Instructions

    Proline Promag H 300 PROFINET over Ethernet-APL Mounting 6.1.3 Special mounting instructions Weather protection cover 280 (11.0) 255 (10.0) 146 (5.75) 134 (5.3) 12 (0.47) 30 (1.18) A0029553  4 Engineering unit mm (in) Hygienic compatibility • When installing in hygienic applications, please refer to the information in the "Certificates and approvals/hygienic compatibility"...
  • Page 26: Turning The Display Module

    Mounting Proline Promag H 300 PROFINET over Ethernet-APL – – 3 mm 4 mm 7 Nm (5.2 lbf ft) A0029993  5 Housing in non-Ex version 1. Depending on the device version: Loosen the securing clamp of the connection compartment cover.
  • Page 27: Post-Mounting Check

    Proline Promag H 300 PROFINET over Ethernet-APL Mounting – – – 3 mm A0030035 1. Depending on the device version: Loosen the securing clamp of the connection compartment cover. 2. Unscrew the connection compartment cover. 3. Turn the display module to the desired position: max. 8 × 45° in each direction.
  • Page 28: Electrical Connection

    Electrical connection Proline Promag H 300 PROFINET over Ethernet-APL Electrical connection WARNING Live parts! Incorrect work performed on the electrical connections can result in an electric shock. ‣ Set up a disconnecting device (switch or power-circuit breaker) to easily disconnect the device from the supply voltage.
  • Page 29: Endress+Hauser

    Proline Promag H 300 PROFINET over Ethernet-APL Electrical connection Loop resistance 15 to 150 Ω/km Cable inductance 0.4 to 1 mH/km Further details are provided in the Ethernet-APL Engineering Guideline (https://www.ethernet-apl.org). Current output 0/4 to 20 mA Standard installation cable is sufficient.
  • Page 30 Electrical connection Proline Promag H 300 PROFINET over Ethernet-APL Standard cable - customer-specific cable With the following order option, no cable is supplied with the device and must be provided by the customer: Order code for DKX001: Order code 040 for "Cable", option 1 "None, provided by customer, max 300 m"...
  • Page 31: Terminal Assignment

    Proline Promag H 300 PROFINET over Ethernet-APL Electrical connection 7.2.3 Terminal assignment Transmitter: supply voltage, input/outputs The terminal assignment of the inputs and outputs depends on the individual order version of the device. The device-specific terminal assignment is documented on an adhesive label in the terminal cover.
  • Page 32: Connecting The Measuring Instrument

    Electrical connection Proline Promag H 300 PROFINET over Ethernet-APL Connecting the measuring instrument NOTICE An incorrect connection compromises electrical safety! ‣ Only properly trained specialist staff may perform electrical connection work. ‣ Observe applicable federal/national installation codes and regulations. ‣...
  • Page 33 Proline Promag H 300 PROFINET over Ethernet-APL Electrical connection N ic N ic te r te r e rg ö ff e rg ö ff iz e iz e te n te n ri r ri r s io s io te r ö...
  • Page 34 Electrical connection Proline Promag H 300 PROFINET over Ethernet-APL 22 mm 24 mm – A0033984 4. Connect the cable according to the terminal assignment.  Signal cable terminal assignment: The device-specific terminal assignment is documented on an adhesive label in the terminal cover.
  • Page 35: Connecting The Remote Display And Operating Module Dkx001

    Proline Promag H 300 PROFINET over Ethernet-APL Electrical connection 7.3.2 Connecting the remote display and operating module DKX001 The remote display and operating module DKX001 is available as an optional extra →  176.. • The remote display and operating module DKX001 is only available for the following housing version: order code for "Housing": option A "Aluminum, coated"...
  • Page 36: Connection Example In Special Situations

    • Grounding rings can be ordered separately as accessory DK5HR* from Endress+Hauser (does not contain any seals). When ordering make sure that the grounding rings are...
  • Page 37: Special Connection Instructions

    Proline Promag H 300 PROFINET over Ethernet-APL Electrical connection Potential equalization via grounding electrodes on process connection A0028972 Hexagonal-headed bolts of process connection Integrated grounding electrodes O-ring seal Sensor Special connection instructions 7.5.1 Connection examples PROFINET with Ethernet-APL A0047536  8...
  • Page 38: Current Output 4-20 Ma

    Electrical connection Proline Promag H 300 PROFINET over Ethernet-APL Current output 4-20 mA 4...20 mA A0028758  9 Connection example for 4-20 mA current output (active) Automation system with current input (e.g. PLC) Analog display unit: observe maximum load Transmitter 4...20 mA...
  • Page 39: Switch Output

    Proline Promag H 300 PROFINET over Ethernet-APL Electrical connection Switch output A0028760  12 Connection example for switch output (passive) Automation system with switch input (e.g. PLC with a 10 kΩ pull-up or pull-down resistor) Power supply Transmitter: observe input values →  184...
  • Page 40: Hardware Settings

    Electrical connection Proline Promag H 300 PROFINET over Ethernet-APL Status input A0028764  15 Connection example for status input Automation system with status output (e.g. PLC) Power supply Transmitter Hardware settings 7.6.1 Setting the device name A measuring point can be quickly identified within a plant on the basis of the tag name.
  • Page 41: Activating The Default Ip Address

    Proline Promag H 300 PROFINET over Ethernet-APL Electrical connection DIP switch ON/OFF Device name – 3…7 – Serial number of the device: EH-PROMAG300-065 Setting the device name Risk of electric shock when opening the transmitter housing. ‣ Before opening the transmitter housing: ‣...
  • Page 42: Ensuring The Degree Of Protection

    Electrical connection Proline Promag H 300 PROFINET over Ethernet-APL ‣ Disconnect the device from the power supply. A0034499 1. Depending on the housing version, loosen the securing clamp or fixing screw of the housing cover. 2. Depending on the housing version, unscrew or open the housing cover and disconnect the local display from the main electronics module where necessary .
  • Page 43: Endress+Hauser

    Proline Promag H 300 PROFINET over Ethernet-APL Electrical connection Are the mounted cables relieved of tension?  Are all cable glands installed, securely tightened and leak-tight? Cable run with "water trap"  →  42? Is the terminal assignment correct ? ...
  • Page 44: Operation Options

    Operation options Proline Promag H 300 PROFINET over Ethernet-APL Operation options Overview of operation options A0046226 Local operation via display module Computer with web browser or with operating tool (e.g FieldCare, DeviceCare, SIMATIC PDM) Field Xpert SMT70 Mobile handheld terminal Automation system (e.g.
  • Page 45: Structure And Function Of The Operating Menu

    Proline Promag H 300 PROFINET over Ethernet-APL Operation options Structure and function of the operating menu 8.2.1 Structure of the operating menu For an overview of the operating menu for experts: see the "Description of Device Parameters" document supplied with the device →  206...
  • Page 46: Operating Philosophy

    Operation options Proline Promag H 300 PROFINET over Ethernet-APL 8.2.2 Operating philosophy The individual parts of the operating menu are assigned to certain user roles (e.g. operator, maintenance etc.). Each user role contains typical tasks within the device life cycle.
  • Page 47: Access To Operating Menu Via Local Display

    Proline Promag H 300 PROFINET over Ethernet-APL Operation options Access to operating menu via local display 8.3.1 Operational display X X X X X X X 1120.50 A0029346 Operational display Device tag Status area Display range for measured values (up to 4 lines) Operating elements →...
  • Page 48 Operation options Proline Promag H 300 PROFINET over Ethernet-APL Input Symbol Meaning Status input Measurement channel numbers Symbol Meaning Measurement channel 1 to 4  The measurement channel number is displayed only if more than one channel is present for the same measured variable type (e.g. totalizer 1 to 3).
  • Page 49: Navigation View

    Proline Promag H 300 PROFINET over Ethernet-APL Operation options 8.3.2 Navigation view In the submenu In the wizard /../Operation 0091-1 /../Curr. output 1 Access stat.disp Assign curr. Operator Volume flow Locking status Display A0013993-EN A0016327-EN Navigation view Navigation path to current position...
  • Page 50: Locking Procedure

    Operation options Proline Promag H 300 PROFINET over Ethernet-APL Setup Is displayed: • In the menu next to the "Setup" selection • At the left in the navigation path in the Setup menu Diagnosis Is displayed: • In the menu next to the "Diagnostics" selection •...
  • Page 51: Editing View

    Proline Promag H 300 PROFINET over Ethernet-APL Operation options 8.3.3 Editing view Numeric editor +0.000 Xx – A0034250  17 For entering values in parameters (e.g. limit values) Entry display area Input screen Confirm, delete or reject entry Operating elements...
  • Page 52 Operation options Proline Promag H 300 PROFINET over Ethernet-APL Operating key Meaning Enter key • Pressing the key briefly confirms your selection. • Pressing the key for 2 s confirms your entry. Escape key combination (press keys simultaneously) Close the editing view without accepting a change.
  • Page 53: Operating Elements

    Proline Promag H 300 PROFINET over Ethernet-APL Operation options 8.3.4 Operating elements Operating key Meaning Minus key In menu, submenu Moves the selection bar upwards in a picklist In wizards Goes to previous parameter In the text and numeric editor Move the entry position to the left.
  • Page 54 Operation options Proline Promag H 300 PROFINET over Ethernet-APL Calling up and closing the context menu The user is in the operational display. 1. Press the  and  keys for longer than 3 seconds.  The context menu opens.
  • Page 55: Navigating And Selecting From List

    Proline Promag H 300 PROFINET over Ethernet-APL Operation options 8.3.6 Navigating and selecting from list Different operating elements are used to navigate through the operating menu. The navigation path is displayed on the left in the header. Icons are displayed in front of the individual menus.
  • Page 56: Calling Up Help Text

    Operation options Proline Promag H 300 PROFINET over Ethernet-APL The direct access code consists of a 5-digit number (at maximum) and the channel number, which identifies the channel of a process variable: e.g. 00914-2. In the navigation view, this appears on the right-hand side in the header of the selected parameter.
  • Page 57: User Roles And Related Access Authorization

    Proline Promag H 300 PROFINET over Ethernet-APL Operation options Ent. access code Invalid or out of range input value Min:0 Max:9999 A0014049-EN For a description of the editing view - consisting of the text editor and numeric editor - with symbols →  51, for a description of the operating elements →  53 8.3.10...
  • Page 58: Enabling And Disabling The Keypad

    Operation options Proline Promag H 300 PROFINET over Ethernet-APL 2. Enter the access code.  The -symbol in front of the parameters disappears; all previously write- protected parameters are now re-enabled. 8.3.12 Enabling and disabling the keypad lock The keypad lock makes it possible to block access to the entire operating menu via local operation.
  • Page 59: Requirements

    Proline Promag H 300 PROFINET over Ethernet-APL Operation options 8.4.2 Requirements Computer hardware Hardware Interface CDI-RJ45 WLAN Interface The computer must have a RJ45 The operating unit must have a WLAN interface. interface. Connection Standard Ethernet cable Connection via Wireless LAN.
  • Page 60: Connecting The Device

    Operation options Proline Promag H 300 PROFINET over Ethernet-APL Measuring device: Via CDI-RJ45 service interface Device CDI-RJ45 service interface Measuring device The measuring device has an RJ45 interface. Web server Web server must be enabled; factory setting: ON  For information on enabling the Web server →  64...
  • Page 61 Proline Promag H 300 PROFINET over Ethernet-APL Operation options IP address 192.168.1.XXX; for XXX all numerical sequences except: 0, 212 and 255 → e.g. 192.168.1.213 Subnet mask 255.255.255.0 Default gateway 192.168.1.212 or leave cells empty Via WLAN interface Configuring the Internet protocol of the mobile terminal NOTICE If the WLAN connection is lost during the configuration, settings made may be lost.
  • Page 62: Logging On

    Operation options Proline Promag H 300 PROFINET over Ethernet-APL 2. Enter the IP address of the web server in the address line of the web browser: 192.168.1.212  The login page appears. 2 3 4 Device name: Device tag: Signal Status:...
  • Page 63: User Interface

    Proline Promag H 300 PROFINET over Ethernet-APL Operation options 8.4.5 User interface A0029418 Function row Local display language Navigation area Header The following information appears in the header: • Device name • Device tag • Device status with status signal →  144 •...
  • Page 64: Disabling The Web Server

    Operation options Proline Promag H 300 PROFINET over Ethernet-APL Working area Depending on the selected function and the related submenus, various actions can be performed in this area: • Configuring parameters • Reading measured values • Calling up help text •...
  • Page 65: Access To The Operating Menu Via The Operating Tool

    Proline Promag H 300 PROFINET over Ethernet-APL Operation options 3. If no longer needed: Reset the modified properties of the Internet protocol (TCP/IP) →  60. If communication with the web server was established via the default IP address 192.168.1.212, DIP switch no. 10 must be reset (from ON → OFF). Afterwards, the IP address of the device is active again for network communication.
  • Page 66: Service Interface

    Operation options Proline Promag H 300 PROFINET over Ethernet-APL Service interface Via service interface (CDI-RJ45) A point-to-point connection can be established to configure the device onsite. With the housing open, the connection is established directly via the service interface (CDI-RJ45) of the device.
  • Page 67 Proline Promag H 300 PROFINET over Ethernet-APL Operation options A0034570 Transmitter with integrated WLAN antenna Transmitter with external WLAN antenna LED lit constantly: WLAN reception is enabled on measuring device LED flashing: WLAN connection established between operating unit and measuring device Computer with WLAN interface and web browser (e.g.
  • Page 68: Fieldcare

    FieldCare Function range FDT-based (Field Device Technology) plant asset management tool from Endress+Hauser. It can configure all smart field units in a system and helps you manage them. By using the status information, it is also a simple but effective way of checking their status and condition.
  • Page 69: Devicecare

    DeviceCare Function range Tool for connecting and configuring Endress+Hauser field devices. The fastest way to configure Endress+Hauser field devices is with the dedicated "DeviceCare" tool. Together with the device type managers (DTMs) it presents a convenient, comprehensive solution. Innovation brochure IN01047S Source for device description files →...
  • Page 70: Simatic Pdm

    Operation options Proline Promag H 300 PROFINET over Ethernet-APL 8.5.4 SIMATIC PDM Function range Standardized, vendor-independent program from Siemens for the operation, configuration, maintenance and diagnosis of intelligent field devices via the PROFINET protocol. Source for device description files →  71...
  • Page 71: System Integration

    Proline Promag H 300 PROFINET over Ethernet-APL System integration System integration Overview of device description files 9.1.1 Current version data for the device Firmware version 01.00.zz • On the title page of the manual • On the transmitter nameplate • Firmware version Diagnostics →...
  • Page 72: File Name Of The Manufacturer-Specific Device Master File (Gsd)

    System integration Proline Promag H 300 PROFINET over Ethernet-APL 9.2.1 File name of the manufacturer-specific device master file (GSD) Example of the name of a device master file: GSDML-V2.43-EH-PROMAG_300_500_APL_yyyymmdd.xml GSDML Description language V2.43 Version of the PROFINET specification Endress+Hauser PROMAG...
  • Page 73: Cyclic Data Transmission

    Proline Promag H 300 PROFINET over Ethernet-APL System integration Cyclic data transmission 9.3.1 Overview of the modules The following graphic shows which modules are available to the device for cyclic data transfer. Cyclic data transfer is performed with an automation system.
  • Page 74: Data Structure

    System integration Proline Promag H 300 PROFINET over Ethernet-APL Selection: input variable Slot Sub-slot Input variables Volume flow • Volume flow • Mass flow • Corrected volume flow • Flow velocity • Temperature • Electronics temperature • Buildup index • Current input 1 •...
  • Page 75 Proline Promag H 300 PROFINET over Ethernet-APL System integration Selection: device function, binary input, slot 81 Slot Sub-slot Device function Status (meaning) Partially filled pipe detection • 0 (device function not active) • 1 (device function active) Low flow cut off...
  • Page 76 System integration Proline Promag H 300 PROFINET over Ethernet-APL Selection: input variable Slot Sub-slot Input variables Volume Data structure Volume Totalizer Control input data Byte 1 Byte 2 Byte 3 Byte 4 Byte 5 Measured value: floating point number (IEEE 754) Status Status coding →...
  • Page 77 Proline Promag H 300 PROFINET over Ethernet-APL System integration Totalizer Control module Transmit totalizer value from the measuring device to the automation system. The Totalizer Control module cyclically transmits a selected totalizer value, including the status, from the measuring device to the automation system. The totalizer value is depicted in the first four bytes in the form of a floating point number as per the IEEE 754 standard.
  • Page 78: Binary Output Module

    System integration Proline Promag H 300 PROFINET over Ethernet-APL Slot Sub-slot Compensation value Temperature Density Data structure Output data of Analog Output Byte 1 Byte 2 Byte 3 Byte 4 Byte 5 Measured value: floating point number (IEEE 754) Status Status coding →...
  • Page 79: Status Coding

    Proline Promag H 300 PROFINET over Ethernet-APL System integration Slot Sub-slot Device function Status (meaning) Reserved Reserved Only available with the Heartbeat application package Selection: device function, binary output, slot 211 Slot Sub-slot Device function Status (meaning) Flow override • 0 (disable device function) •...
  • Page 80: Factory Setting

    System integration Proline Promag H 300 PROFINET over Ethernet-APL Status Coding (hex) Meaning GOOD - Maintenance 0xA4 to 0xA7 The measured value is valid. required Maintenance of the device due in the near future. GOOD - Maintenance 0xA8 to 0xAB The measured value is valid.
  • Page 81: System Redundancy S2

    Proline Promag H 300 PROFINET over Ethernet-APL System integration System redundancy S2 A redundant layout with two automation systems is necessary for processes that are in continuous operation. If one system fails the second system guarantees continued, uninterrupted operation. The measuring device supports S2 system redundancy and can communicate with both automation systems simultaneously.
  • Page 82: Commissioning

    Commissioning Proline Promag H 300 PROFINET over Ethernet-APL Commissioning 10.1 Post-mounting and post-connection check Before commissioning the device: ‣ Make sure that the post-installation and post-connection checks have been performed successfully. • Checklist for "Post-mounting" check→  27 • Checklist for "Post-connection check" →  42 10.2...
  • Page 83: Configuring The Measuring Instrument

    Proline Promag H 300 PROFINET over Ethernet-APL Commissioning 10.5 Configuring the measuring instrument The Setup menu with its guided wizards contains all the parameters needed for standard operation. X X X X X X X 20.50 Main menu Display language English Display/operat.
  • Page 84: Defining The Tag Name

    Commissioning Proline Promag H 300 PROFINET over Ethernet-APL ‣ Pulse/frequency/switch output →  95 1 to n ‣ Relay output 1 to n →  101 ‣ →  103 Low flow cut off ‣ Empty pipe detection →  104 ‣...
  • Page 85: Parameter Overview With Brief Description

    Proline Promag H 300 PROFINET over Ethernet-APL Commissioning "APL port" submenu Navigation "Setup" menu → Communication → APL port ‣ APL port →  85 IP address (7263) Subnet mask (7265) →  85 Default gateway (7264) →  85 MAC address (7262) →...
  • Page 86: Setting The System Units

    Commissioning Proline Promag H 300 PROFINET over Ethernet-APL Parameter overview with brief description Parameter Description User entry / User interface Factory setting IP address Enter the IP address of the measuring 4 octet: 0 to 255 (in the 192.168.1.212 device. particular octet) Subnet mask Displays the subnet mask.
  • Page 87 Proline Promag H 300 PROFINET over Ethernet-APL Commissioning Conductivity unit →  87 Temperature unit →  87 →  87 Mass flow unit Mass unit →  87 Density unit →  87 Corrected volume flow unit →  88 →...
  • Page 88 Commissioning Proline Promag H 300 PROFINET over Ethernet-APL Parameter Prerequisite Description Selection Factory setting Corrected volume flow unit – Select corrected volume flow Unit choose list Country-specific: unit. • Nl/h • Sft³/h Result The selected unit applies to: Corrected volume flow parameter (→...
  • Page 89: Configuration Of The Analog Inputs

    Proline Promag H 300 PROFINET over Ethernet-APL Commissioning 10.5.4 Configuration of the Analog Inputs The Analog inputs submenu guides the user systematically to the individual Analog input 1 to n submenu. From here you get to the parameters of the individual analog input.
  • Page 90: Displaying The I/O Configuration

    Commissioning Proline Promag H 300 PROFINET over Ethernet-APL 10.5.5 Displaying the I/O configuration The I/O configuration submenu guides the user systematically through all the parameters in which the configuration of the I/O modules is displayed. Navigation "Setup" menu → I/O configuration ‣...
  • Page 91 Proline Promag H 300 PROFINET over Ethernet-APL Commissioning Navigation "Setup" menu → Current input ‣ Current input 1 to n Terminal number →  91 Signal mode →  91 →  91 0/4 mA value 20 mA value →  91 Current span →...
  • Page 92: Configuring The Status Input

    Commissioning Proline Promag H 300 PROFINET over Ethernet-APL 10.5.7 Configuring the status input The Status input submenu guides the user systematically through all the parameters that have to be set for configuring the status input. Navigation "Setup" menu → Status input 1 to n ‣...
  • Page 93 Proline Promag H 300 PROFINET over Ethernet-APL Commissioning Terminal number →  93 Signal mode →  93 →  93 Process variable current output Current range output →  94 Lower range value output →  94 Upper range value output →...
  • Page 94 Commissioning Proline Promag H 300 PROFINET over Ethernet-APL Parameter Prerequisite Description User interface /  Factory setting Selection / User entry Current range output – Select current range for • 4...20 mA NE Depends on country: process value output and (3.8...20.5 mA) • 4...20 mA NE upper/lower level for alarm •...
  • Page 95: Configuring The Pulse/Frequency

    Proline Promag H 300 PROFINET over Ethernet-APL Commissioning 10.5.9 Configuring the pulse/frequency/switch output The Pulse/frequency/switch output wizard guides you systematically through all the parameters that can be set for configuring the selected output type. Navigation "Setup" menu → Advanced setup → Pulse/frequency/switch output ‣...
  • Page 96 Commissioning Proline Promag H 300 PROFINET over Ethernet-APL Parameter overview with brief description Parameter Prerequisite Description Selection / User Factory setting interface / User entry Operating mode – Define the output as a pulse, • Pulse Pulse frequency or switch output. • Frequency • Switch Terminal number –...
  • Page 97 Proline Promag H 300 PROFINET over Ethernet-APL Commissioning Minimum frequency value →  98 Maximum frequency value →  98 →  98 Measuring value at minimum frequency Measuring value at maximum →  98 frequency Failure mode →  98 Failure frequency →...
  • Page 98 Commissioning Proline Promag H 300 PROFINET over Ethernet-APL Parameter Prerequisite Description Selection / User Factory setting interface / User entry Minimum frequency value The Frequency option is Enter minimum frequency. 0.0 to 10 000.0 Hz 0.0 Hz selected in the Operating mode parameter (→  95)
  • Page 99 Proline Promag H 300 PROFINET over Ethernet-APL Commissioning Configuring the switch output Navigation "Setup" menu → Pulse/frequency/switch output ‣ Pulse/frequency/switch output 1 to n Operating mode →  99 Terminal number →  99 Signal mode →  99 Switch output function →...
  • Page 100: Low Flow Cut Off

    Commissioning Proline Promag H 300 PROFINET over Ethernet-APL Parameter Prerequisite Description Selection / User Factory setting interface / User entry Switch output function The Switch option is selected Select function for switch • Off in the Operating mode output. • On parameter. • Diagnostic behavior •...
  • Page 101: 10.5.10 Configuring The Relay Output

    Proline Promag H 300 PROFINET over Ethernet-APL Commissioning Parameter Prerequisite Description Selection / User Factory setting interface / User entry Switch-on delay • The Switch option is Define delay for the switch-on 0.0 to 100.0 s 0.0 s selected in the Operating of status output.
  • Page 102 Commissioning Proline Promag H 300 PROFINET over Ethernet-APL Switch state →  103 Powerless relay status →  103 Parameter overview with brief description Parameter Prerequisite Description User interface /  Factory setting Selection / User entry Terminal number – Shows the terminal numbers •...
  • Page 103: 10.5.11 Configuring The Low Flow Cut Off

    Proline Promag H 300 PROFINET over Ethernet-APL Commissioning Parameter Prerequisite Description User interface /  Factory setting Selection / User entry Switch-on delay In the Relay output function Define delay for the switch-on 0.0 to 100.0 s 0.0 s parameter, the Limit option is of status output.
  • Page 104: 10.5.12 Configuring Empty Pipe Detection

    Commissioning Proline Promag H 300 PROFINET over Ethernet-APL 10.5.12 Configuring empty pipe detection The measuring instruments are calibrated with water (approx. 500 µS/cm) at the factory. For liquids with a lower conductivity, it is advisable to perform a new full pipe adjustment onsite.
  • Page 105: 10.5.13 Configuring Flow Damping

    Proline Promag H 300 PROFINET over Ethernet-APL Commissioning 10.5.13 Configuring flow damping The Configure flow damping wizard guides the user systematically through the parameters, depending on the selected scenario: • Configuration of damping for the application To configure flow damping for the specific requirements of the process application.
  • Page 106: 10.5.14 "Build-Up Index Adjustment" Wizard

    Commissioning Proline Promag H 300 PROFINET over Ethernet-APL Parameter overview with brief description Parameter Description Selection / User interface Factory setting Scenario Select the applicable scenario. • Replace old device Configure damping for • Configure damping for application application • Restore factory settings Old device Select the measuring device to replace.
  • Page 107 Proline Promag H 300 PROFINET over Ethernet-APL Commissioning Navigation "Expert" menu → Sensor → Build-up index adjustment ‣ Build-up index adjustment Prerequisites →  107 Progress →  107 →  107 Build-up index reference value E 1 Signal to noise ratio →...
  • Page 108: Advanced Settings

    Commissioning Proline Promag H 300 PROFINET over Ethernet-APL 10.6 Advanced settings The Advanced setup submenu with its submenus contains parameters for specific settings. Navigation to the "Advanced setup" submenu X X X X X X X 20.50 0104-1 Main menu...
  • Page 109: Using The Parameter To Enter The Access Code

    Proline Promag H 300 PROFINET over Ethernet-APL Commissioning ‣ WLAN settings →  114 ‣ Electrode cleaning cycle →  116 ‣ →  117 Heartbeat setup ‣ Configuration backup →  118 ‣ Administration →  119 10.6.1 Using the parameter to enter the access code Navigation "Setup"...
  • Page 110 Commissioning Proline Promag H 300 PROFINET over Ethernet-APL Navigation "Setup" menu → Advanced setup → Totalizer 1 to n ‣ Totalizer 1 to n Assign process variable →  110 Unit totalizer →  110 →  110 Totalizer operation mode Failure mode →...
  • Page 111: Carrying Out Additional Display Configurations

    Proline Promag H 300 PROFINET over Ethernet-APL Commissioning 10.6.4 Carrying out additional display configurations In the Display submenu you can set all the parameters associated with the configuration of the local display. Navigation "Setup" menu → Advanced setup → Display ‣...
  • Page 112 Commissioning Proline Promag H 300 PROFINET over Ethernet-APL Parameter overview with brief description Parameter Prerequisite Description Selection / User Factory setting entry Format display A local display is provided. Select how measured values • 1 value, max. size 1 value, max. size are shown on the display.
  • Page 113 Proline Promag H 300 PROFINET over Ethernet-APL Commissioning Parameter Prerequisite Description Selection / User Factory setting entry 0% bargraph value 3 A selection was made in the Enter 0% value for bar graph Signed floating-point Country-specific: Value 3 display parameter. display. number •...
  • Page 114: Wlan Configuration

    Commissioning Proline Promag H 300 PROFINET over Ethernet-APL Parameter Prerequisite Description Selection / User Factory setting entry Separator A local display is provided. Select decimal separator for • . (point) . (point) displaying numerical values. • , (comma) Backlight One of the following Switch the local display •...
  • Page 115 Proline Promag H 300 PROFINET over Ethernet-APL Commissioning Connection state →  116 Received signal strength →  116 Parameter overview with brief description Parameter Prerequisite Description Selection / User Factory setting entry / User interface WLAN – Switch WLAN on and off. • Disable Enable •...
  • Page 116: Performing Electrode Cleaning

    Commissioning Proline Promag H 300 PROFINET over Ethernet-APL Parameter Prerequisite Description Selection / User Factory setting entry / User interface Connection state – Displays the connection status. • Connected Not connected • Not connected Received signal strength – Shows the received signal • Low High strength.
  • Page 117: Performing Heartbeat Basic Setup

    Proline Promag H 300 PROFINET over Ethernet-APL Commissioning Parameter Prerequisite Description Selection / User Factory setting entry / User interface ECC interval For the following order code: Specify the interval between 0.5 to 168 h 0.5 h "Application package", option one cleaning cycle and the EC "ECC electrode cleaning"...
  • Page 118: Configuration Management

    Commissioning Proline Promag H 300 PROFINET over Ethernet-APL Parameter overview with brief description Parameter Description User entry / Selection Factory setting Plant operator Enter the plant operator. Max. 32 characters such as – letters, numbers or special characters (e.g. @, %, /) Location Enter the location.
  • Page 119: Function Scope Of The "Configuration Management" Parameter

    Proline Promag H 300 PROFINET over Ethernet-APL Commissioning Parameter Description User interface / Selection Factory setting Backup state Shows the current status of data saving or • None None restoring. • Backup in progress • Restoring in progress • Delete in progress •...
  • Page 120 Commissioning Proline Promag H 300 PROFINET over Ethernet-APL Using the parameter to define the access code Complete this wizard to specify an access code for the Maintenance role. Navigation "Setup" menu → Advanced setup → Administration → Define access code ‣...
  • Page 121: Simulation

    Proline Promag H 300 PROFINET over Ethernet-APL Commissioning Parameter overview with brief description Parameter Description Selection Factory setting Device reset Reset the device configuration - either • Cancel Cancel entirely or in part - to a defined state. • To delivery settings •...
  • Page 122 Commissioning Proline Promag H 300 PROFINET over Ethernet-APL Switch state 1 to n →  123 Device alarm simulation →  123 →  123 Diagnostic event category Diagnostic event simulation →  123 Parameter overview with brief description Parameter...
  • Page 123: Protecting Settings From Unauthorized Access

    Proline Promag H 300 PROFINET over Ethernet-APL Commissioning Parameter Prerequisite Description Selection / User Factory setting entry Switch state 1 to n The On option is selected in Select status of the relay • Open Open the Switch output simulation output for the simulation.
  • Page 124 Commissioning Proline Promag H 300 PROFINET over Ethernet-APL 3. Enter the access code again in the Confirm access code parameter (→  120) to confirm.  The  symbol appears in front of all write-protected parameters. • Disabling parameter write protection via access code →  57.
  • Page 125: Write Protection Via Write Protection Switch

    Commissioning Via Web browser, FieldCare, DeviceCare (via CDI-RJ45 service interface), fieldbus You can only obtain a reset code from your local Endress+Hauser service organization. The code must be calculated explicitly for every device. 1. Note down the serial number of the device.
  • Page 126 Commissioning Proline Promag H 300 PROFINET over Ethernet-APL 2. Setting the write protection (WP) switch on the main electronics module to the OFF position (factory setting) disables hardware write protection.  No option is displayed in the Locking status parameter →  127. On the local display, the ...
  • Page 127: Operation

    Proline Promag H 300 PROFINET over Ethernet-APL Operation Operation 11.1 Reading off the device locking status Device active write protection: Locking status parameter Operation → Locking status Function scope of the "Locking status" parameter Options Description None The access authorization displayed in the Access status parameter applies →...
  • Page 128: Process Variables" Submenu

    Operation Proline Promag H 300 PROFINET over Ethernet-APL 11.4.1 "Process variables" submenu The Process variables submenu contains all the parameters needed to display the current measured values for each process variable. Navigation "Diagnostics" menu → Measured values → Process variables ‣...
  • Page 129: Totalizer

    Proline Promag H 300 PROFINET over Ethernet-APL Operation Parameter Prerequisite Description User interface Corrected conductivity One of the following conditions is met: Displays the conductivity currently Positive floating-point • Order code for "Sensor option", option corrected. number CI "Medium temperature Dependency measurement"...
  • Page 130: Input Values" Submenu

    Operation Proline Promag H 300 PROFINET over Ethernet-APL 11.4.3 "Input values" submenu The Input values submenu guides you systematically to the individual input values. Navigation "Diagnostics" menu → Measured values → Input values ‣ Input values ‣ Current input 1 to n →...
  • Page 131: Output Values

    Proline Promag H 300 PROFINET over Ethernet-APL Operation Parameter overview with brief description Parameter Description User interface Value status input Shows the current input signal level. • High • Low 11.4.4 Output values The Output values submenu contains all the parameters needed to display the current measured values for every output.
  • Page 132 Operation Proline Promag H 300 PROFINET over Ethernet-APL Navigation "Diagnostics" menu → Measured values → Output values → Pulse/frequency/switch output 1 to n ‣ Pulse/frequency/switch output 1 to n →  132 Output frequency Pulse output 1 to n →  132 Switch state →...
  • Page 133: Adapting The Measuring Device To The Process Conditions

    Proline Promag H 300 PROFINET over Ethernet-APL Operation 11.5 Adapting the measuring device to the process conditions The following are available for this purpose: • Basic settings using theSetup menu (→  83) • Advanced settings using theAdvanced setup submenu (→  108) 11.6...
  • Page 134: Function Range Of "Reset All Totalizers" Parameter

    Operation Proline Promag H 300 PROFINET over Ethernet-APL 11.6.2 Function range of "Reset all totalizers" parameter Options Description Cancel No action is executed and the user exits the parameter. Reset + totalize Resets all totalizers to 0 and restarts the totaling process. This deletes all the previously aggregated flow values.
  • Page 135 Proline Promag H 300 PROFINET over Ethernet-APL Operation Data logging →  136 Logging delay →  136 →  136 Data logging control Data logging status →  136 Entire logging duration →  136 Parameter overview with brief description...
  • Page 136 Operation Proline Promag H 300 PROFINET over Ethernet-APL Parameter Prerequisite Description Selection / User Factory setting entry / User interface Assign channel 4 The Extended HistoROM Assign a process variable to For the picklist, see application package is logging channel. Assign channel 1 available.
  • Page 137: Diagnostics And Troubleshooting

    Proline Promag H 300 PROFINET over Ethernet-APL Diagnostics and troubleshooting Diagnostics and troubleshooting 12.1 General troubleshooting For local display Error Possible causes Remedial action Local display is dark, but signal output is within The cable of the display module is not plugged in...
  • Page 138 Diagnostics and troubleshooting Proline Promag H 300 PROFINET over Ethernet-APL For access Error Possible causes Remedial action Write access to parameter not possible. Hardware write protection is enabled. Set the write protection switch on the main electronics module to the OFF position →...
  • Page 139: Diagnostic Information Via Light Emitting Diodes

    Proline Promag H 300 PROFINET over Ethernet-APL Diagnostics and troubleshooting For system integration Error Possible causes Remedy The PROFINET device name is not A device name containing one or Specify a correct device name displayed correctly and contains more underscores has been (without underscores) via the coding.
  • Page 140 Diagnostics and troubleshooting Proline Promag H 300 PROFINET over Ethernet-APL Color Meaning Port 1 active: Not connected or no connection established. PROFINET with White Connection available, no active communication Ethernet-APL Flashing white Connection with active communication Port 2 active: Not connected or no connection established.
  • Page 141: Diagnostic Information On Local Display

    Proline Promag H 300 PROFINET over Ethernet-APL Diagnostics and troubleshooting 12.3 Diagnostic information on local display 12.3.1 Diagnostic message Faults detected by the self-monitoring system of the measuring device are displayed as a diagnostic message in alternation with the operational display.
  • Page 142: Diagnostic Information

    Diagnostics and troubleshooting Proline Promag H 300 PROFINET over Ethernet-APL Diagnostic behavior Symbol Meaning Alarm • Measurement is interrupted. • Signal outputs and totalizers assume the defined alarm condition. • A diagnostic message is generated. Warning • Measurement is resumed.
  • Page 143: Calling Up Remedial Measures

    Proline Promag H 300 PROFINET over Ethernet-APL Diagnostics and troubleshooting 12.3.2 Calling up remedial measures X X X X X X X X X X X X X X 20.50 S801 Supply voltage Menu Diagnostic list Diagnostics 1 S801 Supply voltage...
  • Page 144: Calling Up Remedy Information

    Diagnostics and troubleshooting Proline Promag H 300 PROFINET over Ethernet-APL A0031056 Status area with status signal Diagnostic information Remedial measures with service ID In addition, diagnostic events which have occurred can be shown in the Diagnostics menu: • Via parameter →  166 •...
  • Page 145: Calling Up Remedy Information

    Proline Promag H 300 PROFINET over Ethernet-APL Diagnostics and troubleshooting Xxxxxx/…/…/ Device name: Xxxxxxx Mass flow: kg/h 12.34 Device tag: Xxxxxxx Volume flow: 12.34 m /h ³ Status signal: Function check (C) Xxxxxx Diagnostics 1: C485 Simu... Remedy information: Deactivate...
  • Page 146: Adapting The Diagnostic Information

    Diagnostics and troubleshooting Proline Promag H 300 PROFINET over Ethernet-APL 12.6 Adapting the diagnostic information 12.6.1 Adapting the diagnostic behavior Each item of diagnostic information is assigned a specific diagnostic behavior at the factory. The user can change this assignment for specific diagnostic information in the Diagnostic behavior submenu.
  • Page 147: Overview Of Diagnostic Information

    Proline Promag H 300 PROFINET over Ethernet-APL Diagnostics and troubleshooting Supported status information Status Coding (hex) BAD - Maintenance alarm 0x24 to 0x27 BAD - Process related 0x28 to 0x2B BAD - Function check 0x3C to 0x3F UNCERTAIN - Initial value...
  • Page 148 Diagnostics and troubleshooting Proline Promag H 300 PROFINET over Ethernet-APL Diagnostic information Remedy instructions Influenced measured variables Short text 083 Memory content inconsistent 1. Restart device • Conductivity 2. Restore S-DAT data • Corrected conductivity Measured variable status 3. Replace S-DAT •...
  • Page 149: Diagnostic Of Electronic

    Proline Promag H 300 PROFINET over Ethernet-APL Diagnostics and troubleshooting Diagnostic information Remedy instructions Influenced measured variables Short text 170 Coil resistance faulty Check ambient and process temperature • Conductivity • Corrected conductivity Measured variable status • Measured values • Density...
  • Page 150 Diagnostics and troubleshooting Proline Promag H 300 PROFINET over Ethernet-APL Diagnostic information Remedy instructions Influenced measured variables Short text 242 Firmware incompatible 1. Check firmware version • Conductivity 2. Flash or replace electronic module • Corrected conductivity Measured variable status •...
  • Page 151 Proline Promag H 300 PROFINET over Ethernet-APL Diagnostics and troubleshooting Diagnostic information Remedy instructions Influenced measured variables Short text 271 Main electronics faulty 1. Restart device • Conductivity 2. Replace main electronic module • Corrected conductivity Measured variable status • Measured values •...
  • Page 152 Diagnostics and troubleshooting Proline Promag H 300 PROFINET over Ethernet-APL Diagnostic information Remedy instructions Influenced measured variables Short text 276 I/O module faulty 1. Restart device • Conductivity 2. Change I/O module • Corrected conductivity Measured variable status • Measured values •...
  • Page 153 Proline Promag H 300 PROFINET over Ethernet-APL Diagnostics and troubleshooting Diagnostic information Remedy instructions Influenced measured variables Short text 311 Sensor electronics (ISEM) faulty Maintenance required! • Conductivity Do not reset device • Corrected conductivity Measured variable status • Measured values •...
  • Page 154 Diagnostics and troubleshooting Proline Promag H 300 PROFINET over Ethernet-APL Diagnostic information Remedy instructions Influenced measured variables Short text 361 I/O module 1 to n faulty 1. Restart device • Conductivity 2. Check electronic modules • Corrected conductivity Measured variable status 3.
  • Page 155 Proline Promag H 300 PROFINET over Ethernet-APL Diagnostics and troubleshooting Diagnostic information Remedy instructions Influenced measured variables Short text 376 Sensor electronics (ISEM) faulty 1. Replace sensor electronic module • Conductivity (ISEM) • Corrected conductivity Measured variable status [from the factory] 2.
  • Page 156: Diagnostic Of Configuration

    Diagnostics and troubleshooting Proline Promag H 300 PROFINET over Ethernet-APL Diagnostic information Remedy instructions Influenced measured variables Short text 383 Memory content Reset device • Conductivity • Corrected conductivity Measured variable status • Measured values • Density Quality Good • Electronics temperature Quality substatus •...
  • Page 157 Proline Promag H 300 PROFINET over Ethernet-APL Diagnostics and troubleshooting Diagnostic information Remedy instructions Influenced measured variables Short text 431 Trim 1 to n required Carry out trim – Measured variable status Quality Good Quality substatus Coding (hex) 0x80 to 0x83...
  • Page 158 Diagnostics and troubleshooting Proline Promag H 300 PROFINET over Ethernet-APL Diagnostic information Remedy instructions Influenced measured variables Short text 442 Frequency output 1 saturated 1. Check frequency output settings – 2. Check process Measured variable status Quality Good Quality substatus...
  • Page 159 Proline Promag H 300 PROFINET over Ethernet-APL Diagnostics and troubleshooting Diagnostic information Remedy instructions Influenced measured variables Short text 484 Failure mode simulation active Deactivate simulation • Conductivity • Corrected conductivity Measured variable status • Density • Electronics temperature Quality Good •...
  • Page 160 Diagnostics and troubleshooting Proline Promag H 300 PROFINET over Ethernet-APL Diagnostic information Remedy instructions Influenced measured variables Short text 492 Frequency output 1 to n simulation active Deactivate simulation frequency output – Measured variable status Quality Good Quality substatus Coding (hex)
  • Page 161 Proline Promag H 300 PROFINET over Ethernet-APL Diagnostics and troubleshooting Diagnostic information Remedy instructions Influenced measured variables Short text 496 Status input 1 to n simulation active Deactivate simulation status input – Measured variable status Quality Good Quality substatus Coding (hex)
  • Page 162 Diagnostics and troubleshooting Proline Promag H 300 PROFINET over Ethernet-APL Diagnostic information Remedy instructions Influenced measured variables Short text 530 Electrode cleaning active Switch off electrode cleaning • Conductivity • Corrected conductivity Measured variable status • Measured values • Density...
  • Page 163: Diagnostic Of Process

    Proline Promag H 300 PROFINET over Ethernet-APL Diagnostics and troubleshooting 12.7.4 Diagnostic of process Diagnostic information Remedy instructions Influenced measured variables Short text 803 Loop current 1 faulty 1. Check wiring – 2. Change I/O module Measured variable status Quality...
  • Page 164 Diagnostics and troubleshooting Proline Promag H 300 PROFINET over Ethernet-APL Diagnostic information Remedy instructions Influenced measured variables Short text 834 Process temperature too high Reduce process temperature • Conductivity • Corrected conductivity Measured variable status [from the factory] • Density •...
  • Page 165 Proline Promag H 300 PROFINET over Ethernet-APL Diagnostics and troubleshooting Diagnostic information Remedy instructions Influenced measured variables Short text 937 Sensor symmetry 1. Eliminate external magnetic field near • Conductivity sensor • Corrected conductivity Measured variable status [from the factory] 2.
  • Page 166: Pending Diagnostic Events

    Diagnostics and troubleshooting Proline Promag H 300 PROFINET over Ethernet-APL Diagnostic information Remedy instructions Influenced measured variables Short text 962 Pipe empty 1. Perform full pipe adjustment • Conductivity 2. Perform empty pipe adjustment • Corrected conductivity Measured variable status [from the factory] 3.
  • Page 167: Diagnostics List

    Proline Promag H 300 PROFINET over Ethernet-APL Diagnostics and troubleshooting Parameter overview with brief description Parameter Prerequisite Description User interface Actual diagnostics A diagnostic event has occurred. Shows the current occured diagnostic Symbol for diagnostic event along with its diagnostic behavior, diagnostic code information.
  • Page 168: 12.10.2 Filtering The Event Logbook

    Diagnostics and troubleshooting Proline Promag H 300 PROFINET over Ethernet-APL / ../Eventlist I1091 Config. change I1157 Mem.err. ev.list 0d01h19m10s F311 Electr. failure A0014008-EN  28 Using the example of the local display • A maximum of 20 event messages can be displayed in chronological order.
  • Page 169 Proline Promag H 300 PROFINET over Ethernet-APL Diagnostics and troubleshooting Info number Info name I1092 HistoROM backup deleted I1137 Electronics changed I1151 History reset I1155 Reset electronics temperature I1156 Memory error trend I1157 Memory error event list I1256 Display: access status changed...
  • Page 170: Resetting The Measuring Device

    Diagnostics and troubleshooting Proline Promag H 300 PROFINET over Ethernet-APL Info number Info name I1712 New flash file received I1725 Sensor electronic module (ISEM) changed I1726 Configuration backup failed 12.11 Resetting the measuring device The entire device configuration or some of the configuration can be reset to a defined state with the Device reset parameter (→...
  • Page 171 Proline Promag H 300 PROFINET over Ethernet-APL Diagnostics and troubleshooting Parameter overview with brief description Parameter Description User interface Factory setting Device tag Shows name of measuring point. Character string comprising Promag numbers, letters and special characters Serial number Shows the serial number of the measuring Max.
  • Page 172: 12.13 Firmware History

    "Manufacturer' s information" document. The manufacturer' s information is available: • In the Download Area of the Endress+Hauser web site: www.endress.com → Downloads • Specify the following details: • Product root: e.g. 5H3B The product root is the first part of the order code: see the nameplate on the device.
  • Page 173: Maintenance

    Replacement seals (accessory part) →  206 13.2 Measuring and test equipment Endress+Hauser offers a variety of measuring and testing equipment, such as Netilion or device tests. Your Endress+Hauser Sales Center can provide detailed information on the services. List of some of the measuring and testing equipment: →  178 13.3...
  • Page 174: Repair

    • The measuring devices have a modular design. • Spare parts are grouped into logical kits with the associated Installation Instructions. • Repairs are carried out by Endress+Hauser Service or by appropriately trained customers. • Certified devices can only be converted to other certified devices by Endress+Hauser Service or at the factory.
  • Page 175: Disposal

    Proline Promag H 300 PROFINET over Ethernet-APL Repair 14.5 Disposal If required by the Directive 2012/19/EU on waste electrical and electronic equipment (WEEE), the product is marked with the depicted symbol in order to minimize the disposal of WEEE as unsorted municipal waste. Do not dispose of products bearing this marking as unsorted municipal waste.
  • Page 176: Accessories

    Various accessories, which can be ordered with the device or subsequently from Endress +Hauser, are available for the device. Detailed information on the order code in question is available from your local Endress+Hauser sales center or on the product page of the Endress+Hauser website: www.endress.com.
  • Page 177: For The Sensor

    Proline Promag H 300 PROFINET over Ethernet-APL Accessories 15.1.2 For the sensor Accessories Description Adapter set Adapter connections for installing a Promag H instead of a Promag 30/33 A or Promag 30/33 H (DN 25). Consists of: • 2 process connections •...
  • Page 178: Service-Specific Accessories

    • As a downloadable DVD for local PC installation. Netilion lloT ecosystem: Unlock knowledge Endress+Hauser ' s Netilion lloT ecosystem enables you to optimize your plant performance, digitize workflows, share knowledge and improve collaboration. Based on decades of experience in process automation, Endress+Hauser offers the process industry an lloT ecosystem that enables you to gain useful insights from data.
  • Page 179: Technical Data

    Proline Promag H 300 PROFINET over Ethernet-APL Technical data Technical data 16.1 Application The measuring device is intended only for the flow measurement of liquids with a minimum conductivity of 5 µS/cm. Depending on the version ordered, the measuring device can also measure potentially explosive, flammable, poisonous and oxidizing media.
  • Page 180 Technical data Proline Promag H 300 PROFINET over Ethernet-APL Recommended Nominal diameter Factory settings flow min./max. full Full scale value Pulse value Low flow cut off scale value current output (~ 2 pulse/s) (v ~ 0.04 m/s) (v ~ 0.3/10 m/s) (v ~ 2.5 m/s)
  • Page 181 Proline Promag H 300 PROFINET over Ethernet-APL Technical data Recommended Nominal diameter Factory settings flow Low flow min./max. full Full scale value cut off Pulse value scale value current output (v ~ (~ 2 pulse/s) (v ~ 0.3/10 m/s) (v ~ 2.5 m/s) 0.04 m/s...
  • Page 182: Status Input

    Technical data Proline Promag H 300 PROFINET over Ethernet-APL Open-circuit voltage ≤ 28.8 V (active) Possible input variables • Temperature • Density Status input Maximum input values • DC –3 to 30 V • If status input is active (ON): R >3 kΩ...
  • Page 183: Output

    Proline Promag H 300 PROFINET over Ethernet-APL Technical data 16.4 Output Output signal PROFINET with Ethernet-APL Device use Device connection to an APL field switch The device may only be operated according to the following APL port classifications: • If used in hazardous areas: SLAA or SLAC •...
  • Page 184 Technical data Proline Promag H 300 PROFINET over Ethernet-APL Damping Configurable: 0 to 999.9 s Assignable measured • Volume flow variables • Mass flow • Corrected volume flow • Flow velocity • Conductivity • Corrected conductivity • Temperature • Electronics temperature...
  • Page 185 Proline Promag H 300 PROFINET over Ethernet-APL Technical data Switching delay Configurable: 0 to 100 s Number of switching Unlimited cycles Assignable functions • Disable • On • Diagnostic behavior • Limit value: • Disable • Volume flow • Mass flow •...
  • Page 186 Technical data Proline Promag H 300 PROFINET over Ethernet-APL The following inputs and outputs are available for assignment: • Choice of current output: 4 to 20 mA (active), 0/4 to 20 mA (passive) • Pulse/frequency/switch output • Choice of current input: 4 to 20 mA (active), 0/4 to 20 mA (passive) •...
  • Page 187 Proline Promag H 300 PROFINET over Ethernet-APL Technical data Local display Plain text display With information on cause and remedial measures Backlight Red lighting indicates a device error. Status signal as per NAMUR recommendation NE 107 Interface/protocol • Via digital communication: PROFINET with Ethernet-APL •...
  • Page 188: Power Supply

    Technical data Proline Promag H 300 PROFINET over Ethernet-APL System redundancy support System redundancy S2 (2 AR with 1 NAP) Device profile PROFINET PA profile 4 (Application interface identifier API: 0x9700) Manufacturer ID Device type ID 0xA43C Device description files (GSD,...
  • Page 189: Performance Characteristics

    Proline Promag H 300 PROFINET over Ethernet-APL Technical data Power consumption Transmitter Max. 10 W (active power) switch-on current Max. 36 A (<5 ms) as per NAMUR Recommendation NE 21 Current consumption Transmitter • Max. 400 mA (24 V) • Max. 200 mA (110 V, 50/60 Hz; 230 V, 50/60 Hz) Power supply failure •...
  • Page 190: Volume Flow

    Technical data Proline Promag H 300 PROFINET over Ethernet-APL Maximum measurement o.r. = of reading error Maximum permissible error under reference operating conditions Volume flow • ±0.5 % o.r. ± 1 mm/s (0.04 in/s) • Optional: ±0.2 % o.r. ± 2 mm/s (0.08 in/s) Fluctuations in the supply voltage do not have any effect within the specified range.
  • Page 191: Electrical Conductivity

    Proline Promag H 300 PROFINET over Ethernet-APL Technical data –10 –20 –30 ⁰ ¹ ² ³ ⁴ ⁵ [µS/cm] A0042279  30 Measurement error (standard) –10 –20 –30 ⁰ ¹ ² ³ ⁴ ⁵ ⁶ [ S/cm] µ A0047944  31 Measurement error (optional: order code for "Calibrated conductivity measurement", option CW)
  • Page 192: Mounting

    The device is suitable for use outdoors and indoors with a relative humidity of 4 to 95 %. Operating height According to EN 61010-1 • ≤ 2 000 m (6 562 ft) • > 2 000 m (6 562 ft) with additional overvoltage protection (e.g. Endress+Hauser HAW Series) Degree of protection Transmitter •...
  • Page 193: Internal Cleaning

    Proline Promag H 300 PROFINET over Ethernet-APL Technical data Vibration broad-band random, according to IEC 60068-2-64 • 10 to 200 Hz, 0.003 g • 200 to 2 000 Hz, 0.001 g • Total: 1.54 g rms Shock half-sine, according to IEC 60068-2-27...
  • Page 194: Pressure-Temperature Ratings

    Technical data Proline Promag H 300 PROFINET over Ethernet-APL Pressure-temperature For an overview of the pressure-temperature ratings for the process connections, see ratings the Technical Information Pressure tightness Liner: PFA Nominal diameter Limit values for absolute pressure in [mbar] ([psi]) for medium temperatures:...
  • Page 195: Window Material

    Proline Promag H 300 PROFINET over Ethernet-APL Technical data Nominal diameter Weight [mm] [in] [kg] [lbs] 1/12 10.4 5/32 10.4 5/16 10.4 ½ 10.1 12.1 1 ½ 15.0 16.1 – 17.9 19.2 10.0 22.1 15.4 34.0 17.8 39.3 Measuring tube...
  • Page 196 Technical data Proline Promag H 300 PROFINET over Ethernet-APL Seals Order code for "Housing": Option B "Stainless, hygienic": EPDM and silicone Cable entries/cable glands Order code for "Housing", option A "Aluminum, coated" The various cable entries are suitable for hazardous and non-hazardous areas.
  • Page 197 Proline Promag H 300 PROFINET over Ethernet-APL Technical data Accessories Protective cover Stainless steel, 1.4404 (316L) External WLAN antenna • Antenna: ASA plastic (acrylonitrile styrene acrylate) and nickel-plated brass • Adapter: Stainless steel and nickel-plated brass • Cable: Polyethylene • Plug: Nickel-plated brass •...
  • Page 198: 16.11 Operability

    Technical data Proline Promag H 300 PROFINET over Ethernet-APL Stainless steel process connections: • With O-ring seal: ≤ 1.6 µm (63 µin) • With aseptic seal: Ra = 0.76 µm (31.5 µin) Optional: Ra = 0.38 µm (15 µin) electropolished (All data refer to parts in contact with the medium) 16.11 Operability...
  • Page 199 Proline Promag H 300 PROFINET over Ethernet-APL Technical data Via remote display and operating module DKX001 The remote display and operating module DKX001 is available as an optional extra →  176.. • The remote display and operating module DKX001 is only available for the following housing version: order code for "Housing": option A "Aluminum, coated"...
  • Page 200 Technical data Proline Promag H 300 PROFINET over Ethernet-APL Supported operating tools Different operating tools can be used for local or remote access to the measuring device. Depending on the operating tool used, access is possible with different operating units and via a variety of interfaces.
  • Page 201: Historom Data Management

    Proline Promag H 300 PROFINET over Ethernet-APL Technical data • Export the Heartbeat verification report (PDF file, only available with the Heartbeat Verification →  205 application package) • Flash firmware version for device firmware upgrade, for example • Download driver for system integration •...
  • Page 202: 16.12 Certificates And Approvals

    The device meets the legal requirements of the applicable EU Directives. These are listed in the corresponding EU Declaration of Conformity along with the standards applied. Endress+Hauser confirms successful testing of the device by affixing to it the CE mark. UKCA marking The device meets the legal requirements of the applicable UK regulations (Statutory Instruments).
  • Page 203: Pharmaceutical Compatibility

    Proline Promag H 300 PROFINET over Ethernet-APL Technical data Sanitary compatibility • 3-A SSI 28-06 or more recent • Confirmation by affixing the 3-A logo for measuring devices with the order code for "Additional approval", option LP "3-A". • The 3-A approval refers to the measuring device.
  • Page 204 Directive a) PED/G1/x (x = category) or b) PESR/G1/x (x = category) on the sensor nameplate, Endress+Hauser confirms compliance with the "Essential Safety Requirements" a) specified in Annex I of the Pressure Equipment Directive 2014/68/EU or b) Schedule 2 of Statutory Instruments 2016 No. 1105.
  • Page 205: 16.13 Application Packages

    The application packages can be ordered with the device or subsequently from Endress+Hauser. Detailed information on the order code in question is available from your local Endress+Hauser sales center or on the product page of the Endress+Hauser website: www.endress.com.
  • Page 206: Supplementary Documentation

    For an overview of the scope of the associated Technical Documentation, refer to the following: • Device Viewer (www.endress.com/deviceviewer): Enter the serial number from the nameplate • Endress+Hauser Operations app: Enter serial number from nameplate or scan matrix code on nameplate. Standard documentation Brief Operating Instructions...
  • Page 207: Special Documentation

    Proline Promag H 300 PROFINET over Ethernet-APL Technical data Contents Documentation code INMETRO Ex d/Ex de XA01518D INMETRO Ex ec XA01519D NEPSI Ex d/Ex de XA01520D NEPSI Ex nA XA01521D EAC Ex d/Ex de XA01656D EAC Ex nA XA01657D JPN Ex d...
  • Page 208: Index

    Index Proline Promag H 300 PROFINET over Ethernet-APL Index Connection examples, potential equalization ..35, 36 Connection preparations ..... . 31 Access authorization to parameters Connection tools .
  • Page 209 Proline Promag H 300 PROFINET over Ethernet-APL Index In the navigation view ..... 49 Fitted electrodes ......197 Display values Flow direction .
  • Page 210 Index Proline Promag H 300 PROFINET over Ethernet-APL Measured values Navigation path (navigation view) ....49 Calculated ......179 Navigation view Measured .
  • Page 211 Proline Promag H 300 PROFINET over Ethernet-APL Index Network diagnostics (Submenu) ....86 Remote operation ......199 Process variables (Submenu) .
  • Page 212 Index Proline Promag H 300 PROFINET over Ethernet-APL Status signals ......141, 144 In the status area of the local display ... 47 Storage concept .
  • Page 213 Proline Promag H 300 PROFINET over Ethernet-APL Index W@M Device Viewer ......14 Weight Transport (notes) ......18 Wizard Build-up index adjustment .