Endress+Hauser Proline Promag H 500 Operating Instructions Manual
Endress+Hauser Proline Promag H 500 Operating Instructions Manual

Endress+Hauser Proline Promag H 500 Operating Instructions Manual

Electromagnetic flowmeter profinet over ethernet-apl
Hide thumbs Also See for Proline Promag H 500:
Table of Contents

Advertisement

Quick Links

BA02103D/06/EN/02.24-00
71661683
2024-07-17
Valid as of version
01.00.zz (Device firmware)
advanced physical layer
Products
Operating Instructions
Proline Promag H 500
Electromagnetic flowmeter
PROFINET over Ethernet-APL
Solutions
Services

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Proline Promag H 500 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Endress+Hauser Proline Promag H 500

  • Page 1 Products Solutions Services BA02103D/06/EN/02.24-00 71661683 2024-07-17 Valid as of version 01.00.zz (Device firmware) Operating Instructions Proline Promag H 500 Electromagnetic flowmeter PROFINET over Ethernet-APL advanced physical layer...
  • Page 2 • The manufacturer reserves the right to modify technical data without prior notice. Your Endress+Hauser sales organization will supply you with current information and updates to this manual. Endress+Hauser...
  • Page 3: Table Of Contents

    Proline Promag H 500 PROFINET over Ethernet-APL Table of contents Table of contents About this document ....6 Mounting ......21 Document function .
  • Page 4 Table of contents Proline Promag H 500 PROFINET over Ethernet-APL Hardware settings ..... 65 Commissioning ....108 7.7.1...
  • Page 5 13.2 Measuring and test equipment ... 202 13.3 Endress+Hauser services ....202 Repair ......203 14.1 General notes .
  • Page 6: About This Document

    About this document Proline Promag H 500 PROFINET over Ethernet-APL About this document Document function These Operating Instructions contain all the information required in the various life cycle phases of the device: from product identification, incoming acceptance and storage, to installation, connection, operation and commissioning, through to troubleshooting, maintenance and disposal.
  • Page 7: Tool Symbols

    Proline Promag H 500 PROFINET over Ethernet-APL About this document 1.2.4 Tool symbols Symbol Meaning Torx screwdriver Phillips head screwdriver Open-ended wrench 1.2.5 Symbols for certain types of information Symbol Meaning Permitted Procedures, processes or actions that are permitted. Preferred Procedures, processes or actions that are preferred.
  • Page 8: Documentation

    • Device Viewer (www.endress.com/deviceviewer): Enter the serial number from the nameplate • Endress+Hauser Operations app: Enter serial number from nameplate or scan matrix code on nameplate. The following documentation may be available depending on the device version ordered: Document type...
  • Page 9: Safety Instructions

    Proline Promag H 500 PROFINET over Ethernet-APL Safety instructions Safety instructions Requirements for the personnel The personnel for installation, commissioning, diagnostics and maintenance must fulfill the following requirements: ‣ Trained, qualified specialists must have a relevant qualification for this specific function and task.
  • Page 10: Workplace Safety

    Verification for borderline cases: ‣ For special fluids and fluids for cleaning, Endress+Hauser is glad to provide assistance in verifying the corrosion resistance of fluid-wetted materials, but does not accept any warranty or liability as minute changes in the temperature, concentration or level of contamination in the process can alter the corrosion resistance properties.
  • Page 11: Device-Specific It Security

    Proline Promag H 500 PROFINET over Ethernet-APL Safety instructions Device-specific IT security The device offers a range of specific functions to support protective measures on the operator' s side. These functions can be configured by the user and guarantee greater in- operation safety if used correctly.
  • Page 12: Infrastructure Mode

    Safety instructions Proline Promag H 500 PROFINET over Ethernet-APL WLAN passphrase: Operation as WLAN access point A connection between an operating unit (e.g. notebook or tablet) and the device via the WLAN interface (→  92), which can be ordered as an optional extra, is protected by the network key.
  • Page 13: Product Description

    Proline Promag H 500 PROFINET over Ethernet-APL Product description Product description The measuring system consists of a transmitter and a sensor. The transmitter and sensor are mounted in physically separate locations. They are interconnected by connecting cables. Product design Two versions of the transmitter are available.
  • Page 14 Product description Proline Promag H 500 PROFINET over Ethernet-APL 3.1.2 Proline 500 Signal transmission: analog Order code for "Integrated ISEM electronics", option B "Transmitter" For use in applications required to meet special requirements due to ambient or operating conditions. As the electronics are located in the transmitter, the device is ideal in the event of: •...
  • Page 15: Incoming Acceptance And Product

    • Enter the serial numbers from the nameplates in the Device Viewer (www.endress.com/deviceviewer): all the information about the device is displayed. • Enter the serial numbers from the nameplates into the Endress+Hauser Operations app or scan the DataMatrix code on the nameplate with the Endress+Hauser Operations app: all the information about the device is displayed.
  • Page 16: Transmitter Nameplate

    Incoming acceptance and product identification Proline Promag H 500 PROFINET over Ethernet-APL 4.2.1 Transmitter nameplate Proline 500 – digital Order code: Ser. no.: Ext. ord. cd.: Date: A0029194  3 Example of a transmitter nameplate Name of the transmitter Manufacturer address/certificate holder...
  • Page 17 Proline Promag H 500 PROFINET over Ethernet-APL Incoming acceptance and product identification Proline 500 3 4 5 Order code: Ser. no.: Ext. ord. cd.: Date: A0029192  4 Example of a transmitter nameplate Manufacturer address/certificate holder Name of the transmitter Order code Serial number (Ser.
  • Page 18: Sensor Nameplate

    Incoming acceptance and product identification Proline Promag H 500 PROFINET over Ethernet-APL 4.2.2 Sensor nameplate Order code: Ser. no.: Ext. ord. cd.: Date: A0029204  5 Example of sensor nameplate Name of sensor Manufacturer address/certificate holder Order code Serial number (Ser. no.) Extended order code (Ext.
  • Page 19: Symbols On The Device

    Proline Promag H 500 PROFINET over Ethernet-APL Incoming acceptance and product identification 4.2.3 Symbols on the device Symbol Meaning WARNING! This symbol alerts you to a dangerous situation. Failure to avoid this situation can result in serious or fatal injury. Please consult the documentation for the measuring instrument to discover the type of potential danger and measures to avoid it.
  • Page 20: Storage And Transport

    Storage and transport Proline Promag H 500 PROFINET over Ethernet-APL Storage and transport Storage conditions Observe the following notes for storage: ‣ Store in the original packaging to ensure protection from shock. ‣ Do not remove protective covers or protective caps installed on process connections.
  • Page 21: Measuring Devices With Lifting Lugs

    Proline Promag H 500 PROFINET over Ethernet-APL Mounting 5.2.2 Measuring devices with lifting lugs CAUTION Special transportation instructions for devices with lifting lugs ‣ Only use the lifting lugs fitted on the device or flanges to transport the device. ‣...
  • Page 22 Mounting Proline Promag H 500 PROFINET over Ethernet-APL A0042131 The device should ideally be installed in an ascending pipe. A0042317 Installation upstream from a down pipe NOTICE Negative pressure in the measuring pipe can damage the liner! ‣ If installing upstream of down pipes whose length h ≥ 5 m (16.4 ft): install a siphon with a vent valve downstream of the device.
  • Page 23 Proline Promag H 500 PROFINET over Ethernet-APL Mounting A0041088 Installation near pumps NOTICE Negative pressure in the measuring tube can damage the liner! ‣ In order to maintain the system pressure, install the device in the flow direction downstream from the pump.
  • Page 24 Mounting Proline Promag H 500 PROFINET over Ethernet-APL Orientation The direction of the arrow on the nameplate helps you to install the measuring device according to the flow direction (direction of medium flow through the piping). Orientation Recommendation Vertical orientation...
  • Page 25: Inlet And Outlet Runs

    Proline Promag H 500 PROFINET over Ethernet-APL Mounting A0028998 EPD electrode for empty pipe detection, available from ≥ DN 15 (½") Measuring electrodes for signal detection Measuring instruments with a nominal diameter < DN 15 (½") do not have an EPD electrode.
  • Page 26: Environmental And Process Requirements

    Mounting Proline Promag H 500 PROFINET over Ethernet-APL 6.1.2 Environmental and process requirements Ambient temperature range Transmitter • Standard: –40 to +60 °C (–40 to +140 °F) • Optional: –50 to +60 °C (–58 to +140 °F) (Order code for "Test, certificate", option JN "Ambient temperature of transmitter –50 °C (–58...
  • Page 27 Proline Promag H 500 PROFINET over Ethernet-APL Mounting [mbar] 8 m/s 7 m/s 6 m/s 5 m/s 4 m/s 3 m/s max. 8° 2 m/s 1 m/s d / D A0029002 Length of connecting cable Proline 500 – digital transmitter Lengths of connecting cable →...
  • Page 28: Special Mounting Instructions

    Mounting Proline Promag H 500 PROFINET over Ethernet-APL 6.1.3 Special mounting instructions Weather protection cover 213 (8.4) 203 (8.0) A0029552  7 Weather protection cover for Proline 500 – digital; engineering unit mm (in) 280 (11.0) 255 (10.0) 146 (5.75) 134 (5.3)
  • Page 29: Preparing The Measuring Device

    Proline Promag H 500 PROFINET over Ethernet-APL Mounting For sensor For flanges and other process connections: Use a suitable mounting tool. 6.2.2 Preparing the measuring device 1. Remove all remaining transport packaging. 2. Remove any protective covers or protective caps present from the sensor.
  • Page 30 • Grounding rings can be ordered separately as an accessory from Endress+Hauser →  236. When ordering make sure that the grounding rings are compatible with...
  • Page 31: Mounting The Transmitter Housing: Proline 500 - Digital

    Proline Promag H 500 PROFINET over Ethernet-APL Mounting A0028971  9 Installing grounding rings Hexagonal-headed bolts of process connection O-ring seals Grounding ring or plastic disk (spacer) Sensor 1. Loosen the 4 or 6 hexagonal-headed bolts (1) and remove the process connection from the sensor (4).
  • Page 32: Wall Mounting

    Mounting Proline Promag H 500 PROFINET over Ethernet-APL NOTICE Excessive tightening torque applied to the fixing screws! Risk of damaging the plastic transmitter. ‣ Tighten the fixing screws as per the tightening torque: 2.5 Nm (1.8 lbf ft) ø 20…70 TX 25 ø...
  • Page 33: Mounting The Transmitter Housing: Proline 500

    Proline Promag H 500 PROFINET over Ethernet-APL Mounting 2. Insert wall plugs into the drilled holes. 3. Screw in the fixing screws slightly. 4. Fit the transmitter housing over the fixing screws and mount in place. 5. Tighten the fixing screws.
  • Page 34: Turning The Transmitter Housing: Proline 500

    Mounting Proline Promag H 500 PROFINET over Ethernet-APL 20 70 … ( 0.79 to 2.75) SW 13 A0029057  13 Engineering unit mm (in) 6.2.6 Turning the transmitter housing: Proline 500 To provide easier access to the connection compartment or display module, the transmitter housing can be turned.
  • Page 35: Post-Mounting Check

    Proline Promag H 500 PROFINET over Ethernet-APL Mounting – – – 3 mm A0030035 1. Depending on the device version: Loosen the securing clamp of the connection compartment cover. 2. Unscrew the connection compartment cover. 3. Turn the display module to the desired position: max. 8 × 45° in each direction.
  • Page 36: Electrical Connection

    Electrical connection Proline Promag H 500 PROFINET over Ethernet-APL Electrical connection WARNING Live parts! Incorrect work performed on the electrical connections can result in an electric shock. ‣ Set up a disconnecting device (switch or power-circuit breaker) to easily disconnect the device from the supply voltage.
  • Page 37 Proline Promag H 500 PROFINET over Ethernet-APL Electrical connection Loop resistance 15 to 150 Ω/km Cable inductance 0.4 to 1 mH/km Further details are provided in the Ethernet-APL Engineering Guideline (https://www.ethernet-apl.org). Current output 0/4 to 20 mA Standard installation cable is sufficient.
  • Page 38 Electrical connection Proline Promag H 500 PROFINET over Ethernet-APL Choice of connecting cable between the transmitter and sensor Depends on the type of transmitter and the installation zones A0032477 Proline 500 digital transmitter Proline 500 transmitter Promag sensor Non-hazardous area Hazardous area: Zone 2;...
  • Page 39 Proline Promag H 500 PROFINET over Ethernet-APL Electrical connection Cable lengths for use in Non-hazardous area, Hazardous area: Zone 1; Cross-section Hazardous area: Zone 2; Class I, Division 1 Class I, Division 2 1.50 mm (AWG 15) 300 m (900 ft) 180 m (540 ft) 2.50 mm...
  • Page 40: Terminal Assignment

    Electrical connection Proline Promag H 500 PROFINET over Ethernet-APL A0029151  15 Cable cross-section Electrode cable Coil current cable Core Core insulation Core shield Core jacket Core reinforcement Cable shield Outer jacket Operation in zones of severe electrical interference The measuring system meets the general safety requirements →  235 and EMC specifications →...
  • Page 41: Device Plug Pin Assignment

    Proline Promag H 500 PROFINET over Ethernet-APL Electrical connection Order code for "Input; output 1", option RB "PROFINET with Ethernet-APL" Order code Cable entry/connection "Electrical connection" L, N, P, U M12 plug × 1 – 7.2.5 Device plug pin assignment...
  • Page 42: Preparing The Measuring Device

    Electrical connection Proline Promag H 500 PROFINET over Ethernet-APL NOTICE In systems without potential matching, the multiple grounding of the cable shield causes mains frequency equalizing currents! Damage to the bus cable shield. ‣ Only ground the bus cable shield to either the local ground or the protective ground at one end.
  • Page 43: Preparing The Connecting Cable: Proline 500 - Digital

    Proline Promag H 500 PROFINET over Ethernet-APL Electrical connection 7.2.8 Preparing the connecting cable: Proline 500 – digital When terminating the connecting cable, pay attention to the following points: ‣ For cables with fine-wire cores (stranded cables): Fit the cores with ferrules.
  • Page 44 Electrical connection Proline Promag H 500 PROFINET over Ethernet-APL Transmitter Electrode cable Coil current cable 70 (2.76) 80 (3.15) 50 (1.97) 17 (0.67) 50 (1.97) 8 (0.3) 10 (0.4) 8 (0.3) A0029544 A0029543 Sensor Electrode cable Coil current cable 70 (2.76) 80 (3.15)
  • Page 45: Connecting The Measuring Instrument: Proline 500 - Digital

    Proline Promag H 500 PROFINET over Ethernet-APL Electrical connection Connecting the measuring instrument: Proline 500 - digital NOTICE An incorrect connection compromises electrical safety! ‣ Only properly trained specialist staff may perform electrical connection work. ‣ Observe applicable federal/national installation codes and regulations.
  • Page 46 Electrical connection Proline Promag H 500 PROFINET over Ethernet-APL Connecting the sensor connection housing via terminals For the device version with the order code for "Sensor connection housing": Option A "Aluminum coated" 10 (0.4) 3 mm TX 20 TX 20...
  • Page 47 Proline Promag H 500 PROFINET over Ethernet-APL Electrical connection Connecting the sensor connection housing via terminals For the device version with the order code for "Sensor connection housing": Option B "Stainless, hygienic" 10 (0.4) 8 mm 22 mm 24 mm A0029613 1.
  • Page 48 Electrical connection Proline Promag H 500 PROFINET over Ethernet-APL Connecting the sensor connection housing via the connector For the device version with the order code for "Sensor connection housing": Option C "Ultra-compact hygienic, stainless" A0029615 1. Connect the protective ground.
  • Page 49 Proline Promag H 500 PROFINET over Ethernet-APL Electrical connection Connecting the connecting cable to the transmitter TX 20 TX 20 10 (0.4) 22 mm 24 mm A0029597 1. Loosen the 4 fixing screws on the housing cover. 2. Open the housing cover.
  • Page 50: Connecting The Transmitter

    Electrical connection Proline Promag H 500 PROFINET over Ethernet-APL 7.3.2 Connecting the transmitter A0028200 Terminal connection for supply voltage Terminal connection for signal transmission, input/output Terminal connection for signal transmission, input/output Terminal connection for connecting cable between sensor and transmitter Terminal connection for signal transmission, input/output;...
  • Page 51 Proline Promag H 500 PROFINET over Ethernet-APL Electrical connection 7. Plug in the RJ45 connector. 8. Firmly tighten the cable glands.  This concludes the connection process. Connecting the supply voltage and additional inputs/outputs TX 20 TX 20 10 (0.4)
  • Page 52 Electrical connection Proline Promag H 500 PROFINET over Ethernet-APL 3 (0.12) A0029598  17 Engineering unit mm (in) 1. Use a flat-blade screwdriver to press down on the slot between the two terminal holes. 2. Remove the cable end from the terminal.
  • Page 53: Integrating The Transmitter Into A Network

    Proline Promag H 500 PROFINET over Ethernet-APL Electrical connection 7.3.3 Integrating the transmitter into a network This section only presents the basic options for integrating the device into a network. For information on the procedure to follow to connect the transmitter correctly →  45.
  • Page 54: Connecting The Measuring Instrument: Proline 500

    Electrical connection Proline Promag H 500 PROFINET over Ethernet-APL Connecting the measuring instrument: Proline 500 NOTICE An incorrect connection compromises electrical safety! ‣ Only properly trained specialist staff may perform electrical connection work. ‣ Observe applicable federal/national installation codes and regulations.
  • Page 55 Proline Promag H 500 PROFINET over Ethernet-APL Electrical connection Connecting the sensor connection housing via terminals For the device version, order code for "Sensor connection housing": Option B: stainless, hygienic 10 (0.4) 8 mm 22 mm 24 mm A0029617 1. Release the securing screw of the housing cover.
  • Page 56 Electrical connection Proline Promag H 500 PROFINET over Ethernet-APL Attaching the connecting cable to the transmitter 10 (0.4) 3 mm 22 mm 24 mm A0029592 1. Loosen the securing clamp of the connection compartment cover. 2. Unscrew the connection compartment cover.
  • Page 57: Connecting The Transmitter

    Proline Promag H 500 PROFINET over Ethernet-APL Electrical connection 7.4.2 Connecting the transmitter A0026781 Terminal connection for supply voltage Terminal connection for signal transmission, input/output Terminal connection for signal transmission, input/output or terminal connection for network connection via service interface (CDI-RJ45)
  • Page 58 Electrical connection Proline Promag H 500 PROFINET over Ethernet-APL 10 (0.4) mm (in) – A0051111 7. Push the cable through the cable entry. To ensure tight sealing, do not remove the sealing ring from the cable entry. 8. Strip the cable and cable ends and connect to terminals 26-27. In the case of stranded cables, also fit ferrules.
  • Page 59 Proline Promag H 500 PROFINET over Ethernet-APL Electrical connection 5. Firmly tighten the cable glands.  This concludes the cable connection process. 6. Close the terminal cover. 7. Fit the display module holder in the electronics compartment. 8. Screw on the connection compartment cover.
  • Page 60: Integrating The Transmitter Into A Network

    Electrical connection Proline Promag H 500 PROFINET over Ethernet-APL 7.4.3 Integrating the transmitter into a network This section only presents the basic options for integrating the device into a network. For information on the procedure to follow to connect the transmitter correctly →  54.
  • Page 61: Connection Example, Standard Scenario

    • Grounding rings can be ordered separately as accessory DK5HR* from Endress+Hauser (does not contain any seals). When ordering make sure that the grounding rings are...
  • Page 62: Special Connection Instructions

    Electrical connection Proline Promag H 500 PROFINET over Ethernet-APL Potential equalization via grounding electrodes on process connection A0028972 Hexagonal-headed bolts of process connection Integrated grounding electrodes O-ring seal Sensor Special connection instructions 7.6.1 Connection examples PROFINET with Ethernet-APL A0047536  19...
  • Page 63: Current Output 4-20 Ma

    Proline Promag H 500 PROFINET over Ethernet-APL Electrical connection Current output 4-20 mA 4...20 mA A0028758  20 Connection example for 4-20 mA current output (active) Automation system with current input (e.g. PLC) Analog display unit: observe maximum load Transmitter 4...20 mA...
  • Page 64: Switch Output

    Electrical connection Proline Promag H 500 PROFINET over Ethernet-APL Switch output A0028760  23 Connection example for switch output (passive) Automation system with switch input (e.g. PLC with a 10 kΩ pull-up or pull-down resistor) Power supply Transmitter: observe input values →  214...
  • Page 65: Hardware Settings

    Proline Promag H 500 PROFINET over Ethernet-APL Electrical connection Status input A0028764  26 Connection example for status input Automation system with status output (e.g. PLC) Power supply Transmitter Hardware settings 7.7.1 Setting the device name A measuring point can be quickly identified within a plant on the basis of the tag name.
  • Page 66 Electrical connection Proline Promag H 500 PROFINET over Ethernet-APL DIP switch ON/OFF Device name – 3…7 – Serial number of the device: EH-PROMAG500-065 Setting the device name: Proline 500 - digital Risk of electric shock when opening the transmitter housing.
  • Page 67: Activating The Default Ip Address

    Proline Promag H 500 PROFINET over Ethernet-APL Electrical connection A0034498 1. Depending on the housing version, loosen the securing clamp or fixing screw of the housing cover. 2. Depending on the housing version, unscrew or open the housing cover and disconnect the local display from the main electronics module where necessary .
  • Page 68 Electrical connection Proline Promag H 500 PROFINET over Ethernet-APL A0034500 1. Loosen the 4 fixing screws on the housing cover. 2. Open the housing cover. 3. Fold open the terminal cover. 4. Set DIP switch no. 2 on the I/O electronics module from OFF → ON.
  • Page 69: Ensuring The Degree Of Protection

    Proline Promag H 500 PROFINET over Ethernet-APL Electrical connection Ensuring the degree of protection The measuring instrument fulfills all the requirements for the degree of protection IP66/67, Type 4X enclosure. To guarantee the degree of protection IP66/67, Type 4X enclosure, carry out the following steps after the electrical connection: 1.
  • Page 70: Operation Options

    Operation options Proline Promag H 500 PROFINET over Ethernet-APL Operation options Overview of operation options A0046226 Local operation via display module Computer with web browser or with operating tool (e.g FieldCare, DeviceCare, SIMATIC PDM) Field Xpert SMT70 Mobile handheld terminal Automation system (e.g.
  • Page 71: Structure And Function Of The Operating

    Proline Promag H 500 PROFINET over Ethernet-APL Operation options Structure and function of the operating menu 8.2.1 Structure of the operating menu For an overview of the operating menu for experts: see the "Description of Device Parameters" document supplied with the device →  237...
  • Page 72: Operating Philosophy

    Operation options Proline Promag H 500 PROFINET over Ethernet-APL 8.2.2 Operating philosophy The individual parts of the operating menu are assigned to certain user roles (e.g. operator, maintenance etc.). Each user role contains typical tasks within the device life cycle.
  • Page 73: Access To Operating Menu Via Local Display

    Proline Promag H 500 PROFINET over Ethernet-APL Operation options Access to operating menu via local display 8.3.1 Operational display X X X X X X X 1120.50 A0029346 Operational display Device tag Status area Display range for measured values (up to 4 lines) Operating elements →...
  • Page 74 Operation options Proline Promag H 500 PROFINET over Ethernet-APL Input Symbol Meaning Status input Measurement channel numbers Symbol Meaning Measurement channel 1 to 4  The measurement channel number is displayed only if more than one channel is present for the same measured variable type (e.g. totalizer 1 to 3).
  • Page 75: Navigation View

    Proline Promag H 500 PROFINET over Ethernet-APL Operation options 8.3.2 Navigation view In the submenu In the wizard /../Operation 0091-1 /../Curr. output 1 Access stat.disp Assign curr. Operator Volume flow Locking status Display A0013993-EN A0016327-EN Navigation view Navigation path to current position...
  • Page 76: Locking Procedure

    Operation options Proline Promag H 500 PROFINET over Ethernet-APL Setup Is displayed: • In the menu next to the "Setup" selection • At the left in the navigation path in the Setup menu Diagnosis Is displayed: • In the menu next to the "Diagnostics" selection •...
  • Page 77: Editing View

    Proline Promag H 500 PROFINET over Ethernet-APL Operation options 8.3.3 Editing view Numeric editor +0.000 Xx – A0034250  28 For entering values in parameters (e.g. limit values) Entry display area Input screen Confirm, delete or reject entry Operating elements...
  • Page 78 Operation options Proline Promag H 500 PROFINET over Ethernet-APL Operating key Meaning Enter key • Pressing the key briefly confirms your selection. • Pressing the key for 2 s confirms your entry. Escape key combination (press keys simultaneously) Close the editing view without accepting a change.
  • Page 79: Operating Elements

    Proline Promag H 500 PROFINET over Ethernet-APL Operation options 8.3.4 Operating elements Operating key Meaning Minus key In menu, submenu Moves the selection bar upwards in a picklist In wizards Goes to previous parameter In the text and numeric editor Move the entry position to the left.
  • Page 80 Operation options Proline Promag H 500 PROFINET over Ethernet-APL Calling up and closing the context menu The user is in the operational display. 1. Press the  and  keys for longer than 3 seconds.  The context menu opens.
  • Page 81: Navigating And Selecting From List

    Proline Promag H 500 PROFINET over Ethernet-APL Operation options 8.3.6 Navigating and selecting from list Different operating elements are used to navigate through the operating menu. The navigation path is displayed on the left in the header. Icons are displayed in front of the individual menus.
  • Page 82: Calling Up Help Text

    Operation options Proline Promag H 500 PROFINET over Ethernet-APL The direct access code consists of a 5-digit number (at maximum) and the channel number, which identifies the channel of a process variable: e.g. 00914-2. In the navigation view, this appears on the right-hand side in the header of the selected parameter.
  • Page 83: User Roles And Related Access Authorization

    Proline Promag H 500 PROFINET over Ethernet-APL Operation options Ent. access code Invalid or out of range input value Min:0 Max:9999 A0014049-EN For a description of the editing view - consisting of the text editor and numeric editor - with symbols →  77, for a description of the operating elements →  79 8.3.10...
  • Page 84: Enabling And Disabling The Keypad

    Operation options Proline Promag H 500 PROFINET over Ethernet-APL 2. Enter the access code.  The -symbol in front of the parameters disappears; all previously write- protected parameters are now re-enabled. 8.3.12 Enabling and disabling the keypad lock The keypad lock makes it possible to block access to the entire operating menu via local operation.
  • Page 85: Requirements

    Proline Promag H 500 PROFINET over Ethernet-APL Operation options 8.4.2 Requirements Computer hardware Hardware Interface CDI-RJ45 WLAN Interface The computer must have a RJ45 The operating unit must have a WLAN interface. interface. Connection Standard Ethernet cable Connection via Wireless LAN.
  • Page 86: Connecting The Device

    Operation options Proline Promag H 500 PROFINET over Ethernet-APL Measuring device: Via CDI-RJ45 service interface Device CDI-RJ45 service interface Measuring device The measuring device has an RJ45 interface. Web server Web server must be enabled; factory setting: ON  For information on enabling the Web server →  90...
  • Page 87 Proline Promag H 500 PROFINET over Ethernet-APL Operation options 3. Connect the computer to the RJ45 plug via the standard Ethernet cable →  92. 4. If a 2nd network card is not used, close all the applications on the notebook.
  • Page 88: Logging On

    Operation options Proline Promag H 500 PROFINET over Ethernet-APL 2. Enter the IP address of the web server in the address line of the web browser: 192.168.1.212  The login page appears. 2 3 4 Device name: Device tag: Signal Status:...
  • Page 89: User Interface

    Proline Promag H 500 PROFINET over Ethernet-APL Operation options 8.4.5 User interface A0029418 Function row Local display language Navigation area Header The following information appears in the header: • Device name • Device tag • Device status with status signal →  173 •...
  • Page 90: Disabling The Web Server

    Operation options Proline Promag H 500 PROFINET over Ethernet-APL Working area Depending on the selected function and the related submenus, various actions can be performed in this area: • Configuring parameters • Reading measured values • Calling up help text •...
  • Page 91: Access To The Operating Menu Via The Operating Tool

    Proline Promag H 500 PROFINET over Ethernet-APL Operation options 3. If no longer needed: Reset the modified properties of the Internet protocol (TCP/IP) →  86. If communication with the web server was established via the default IP address 192.168.1.212, DIP switch no. 10 must be reset (from ON → OFF). Afterwards, the IP address of the device is active again for network communication.
  • Page 92: Service Interface

    Operation options Proline Promag H 500 PROFINET over Ethernet-APL Service interface Via service interface (CDI-RJ45) A point-to-point connection can be established to configure the device onsite. With the housing open, the connection is established directly via the service interface (CDI-RJ45) of the device.
  • Page 93 Proline Promag H 500 PROFINET over Ethernet-APL Operation options A0034569 Transmitter with integrated WLAN antenna Transmitter with external WLAN antenna LED lit constantly: WLAN reception is enabled on measuring device LED flashing: WLAN connection established between operating unit and measuring device Computer with WLAN interface and web browser (e.g.
  • Page 94: Fieldcare

    FieldCare Function range FDT-based (Field Device Technology) plant asset management tool from Endress+Hauser. It can configure all smart field units in a system and helps you manage them. By using the status information, it is also a simple but effective way of checking their status and condition.
  • Page 95: Devicecare

    Proline Promag H 500 PROFINET over Ethernet-APL Operation options 2. In the network: Add a device.  The Add device window opens. 3. Select the CDI Communication TCP/IP option from the list and press OK to confirm. 4. Right-click CDI Communication TCP/IP and select the Add device option in the context menu that opens.
  • Page 96: Simatic Pdm

    Operation options Proline Promag H 500 PROFINET over Ethernet-APL The fastest way to configure Endress+Hauser field devices is with the dedicated "DeviceCare" tool. Together with the device type managers (DTMs) it presents a convenient, comprehensive solution. Innovation brochure IN01047S Source for device description files →  97 8.5.4...
  • Page 97: System Integration

    Proline Promag H 500 PROFINET over Ethernet-APL System integration System integration Overview of device description files 9.1.1 Current version data for the device Firmware version 01.00.zz • On the title page of the manual • On the transmitter nameplate • Firmware version Diagnostics →...
  • Page 98: File Name Of The Manufacturer-Specific Device Master File (Gsd)

    System integration Proline Promag H 500 PROFINET over Ethernet-APL 9.2.1 File name of the manufacturer-specific device master file (GSD) Example of the name of a device master file: GSDML-V2.43-EH-PROMAG_300_500_APL_yyyymmdd.xml GSDML Description language V2.43 Version of the PROFINET specification Endress+Hauser PROMAG...
  • Page 99: Cyclic Data Transmission

    Proline Promag H 500 PROFINET over Ethernet-APL System integration Cyclic data transmission 9.3.1 Overview of the modules The following graphic shows which modules are available to the device for cyclic data transfer. Cyclic data transfer is performed with an automation system.
  • Page 100: Data Structure

    System integration Proline Promag H 500 PROFINET over Ethernet-APL Selection: input variable Slot Sub-slot Input variables Volume flow • Volume flow • Mass flow • Corrected volume flow • Flow velocity • Temperature • Electronics temperature • Buildup index • Current input 1 •...
  • Page 101 Proline Promag H 500 PROFINET over Ethernet-APL System integration Selection: device function, binary input, slot 81 Slot Sub-slot Device function Status (meaning) Partially filled pipe detection • 0 (device function not active) • 1 (device function active) Low flow cut off...
  • Page 102 System integration Proline Promag H 500 PROFINET over Ethernet-APL Selection: input variable Slot Sub-slot Input variables Volume Data structure Volume Totalizer Control input data Byte 1 Byte 2 Byte 3 Byte 4 Byte 5 Measured value: floating point number (IEEE 754) Status Status coding →...
  • Page 103 Proline Promag H 500 PROFINET over Ethernet-APL System integration Totalizer Control module Transmit totalizer value from the measuring device to the automation system. The Totalizer Control module cyclically transmits a selected totalizer value, including the status, from the measuring device to the automation system. The totalizer value is depicted in the first four bytes in the form of a floating point number as per the IEEE 754 standard.
  • Page 104: Binary Output Module

    System integration Proline Promag H 500 PROFINET over Ethernet-APL Slot Sub-slot Compensation value Temperature Density Data structure Output data of Analog Output Byte 1 Byte 2 Byte 3 Byte 4 Byte 5 Measured value: floating point number (IEEE 754) Status Status coding →...
  • Page 105: Status Coding

    Proline Promag H 500 PROFINET over Ethernet-APL System integration Slot Sub-slot Device function Status (meaning) Reserved Reserved Only available with the Heartbeat application package Selection: device function, binary output, slot 211 Slot Sub-slot Device function Status (meaning) Flow override • 0 (disable device function) •...
  • Page 106: Factory Setting

    System integration Proline Promag H 500 PROFINET over Ethernet-APL Status Coding (hex) Meaning GOOD - Maintenance 0xA4 to 0xA7 The measured value is valid. required Maintenance of the device due in the near future. GOOD - Maintenance 0xA8 to 0xAB The measured value is valid.
  • Page 107: System Redundancy S2

    Proline Promag H 500 PROFINET over Ethernet-APL System integration System redundancy S2 A redundant layout with two automation systems is necessary for processes that are in continuous operation. If one system fails the second system guarantees continued, uninterrupted operation. The measuring device supports S2 system redundancy and can communicate with both automation systems simultaneously.
  • Page 108: Commissioning

    Commissioning Proline Promag H 500 PROFINET over Ethernet-APL Commissioning 10.1 Post-mounting and post-connection check Before commissioning the device: ‣ Make sure that the post-installation and post-connection checks have been performed successfully. • Checklist for "Post-mounting" check→  35 • Checklist for "Post-connection check" →  69 10.2...
  • Page 109: Configuring The Measuring Instrument

    Proline Promag H 500 PROFINET over Ethernet-APL Commissioning 10.5 Configuring the measuring instrument The Setup menu with its guided wizards contains all the parameters needed for standard operation. X X X X X X X 20.50 Main menu Display language English Display/operat.
  • Page 110: Defining The Tag Name

    Commissioning Proline Promag H 500 PROFINET over Ethernet-APL ‣ Pulse/frequency/switch output →  122 1 to n ‣ Relay output 1 to n →  128 ‣ →  130 Low flow cut off ‣ Empty pipe detection →  131 ‣...
  • Page 111: Parameter Overview With Brief Description

    Proline Promag H 500 PROFINET over Ethernet-APL Commissioning "APL port" submenu Navigation "Setup" menu → Communication → APL port ‣ APL port →  111 IP address (7263) Subnet mask (7265) →  111 Default gateway (7264) →  111 MAC address (7262) →...
  • Page 112: Setting The System Units

    Commissioning Proline Promag H 500 PROFINET over Ethernet-APL Parameter overview with brief description Parameter Description User entry / User interface Factory setting IP address Enter the IP address of the measuring 4 octet: 0 to 255 (in the 192.168.1.212 device. particular octet) Subnet mask Displays the subnet mask.
  • Page 113 Proline Promag H 500 PROFINET over Ethernet-APL Commissioning Conductivity unit →  113 Temperature unit →  113 →  113 Mass flow unit Mass unit →  113 Density unit →  113 Corrected volume flow unit →  114 →...
  • Page 114 Commissioning Proline Promag H 500 PROFINET over Ethernet-APL Parameter Prerequisite Description Selection Factory setting Corrected volume flow unit – Select corrected volume flow Unit choose list Country-specific: unit. • Nl/h • Sft³/h Result The selected unit applies to: Corrected volume flow parameter (→...
  • Page 115: 10.5.4 Configuration Of The Analog Inputs

    Proline Promag H 500 PROFINET over Ethernet-APL Commissioning 10.5.4 Configuration of the Analog Inputs The Analog inputs submenu guides the user systematically to the individual Analog input 1 to n submenu. From here you get to the parameters of the individual analog input.
  • Page 116: Displaying The I/O Configuration

    Commissioning Proline Promag H 500 PROFINET over Ethernet-APL 10.5.5 Displaying the I/O configuration The I/O configuration submenu guides the user systematically through all the parameters in which the configuration of the I/O modules is displayed. Navigation "Setup" menu → I/O configuration ‣...
  • Page 117 Proline Promag H 500 PROFINET over Ethernet-APL Commissioning Navigation "Setup" menu → Current input ‣ Current input 1 to n Terminal number →  117 Signal mode →  117 →  117 0/4 mA value 20 mA value →  117 Current span →...
  • Page 118: Configuring The Status Input

    Commissioning Proline Promag H 500 PROFINET over Ethernet-APL 10.5.7 Configuring the status input The Status input submenu guides the user systematically through all the parameters that have to be set for configuring the status input. Navigation "Setup" menu → Status input 1 to n ‣...
  • Page 119 Proline Promag H 500 PROFINET over Ethernet-APL Commissioning Navigation "Setup" menu → Current output ‣ Current output 1 to n Terminal number →  119 Signal mode →  119 →  120 Process variable current output Current range output →...
  • Page 120 Commissioning Proline Promag H 500 PROFINET over Ethernet-APL Parameter Prerequisite Description User interface /  Factory setting Selection / User entry Process variable current output – Select the process variable for • Off Volume flow the current output. • Volume flow • Mass flow •...
  • Page 121 Proline Promag H 500 PROFINET over Ethernet-APL Commissioning Parameter Prerequisite Description User interface /  Factory setting Selection / User entry Damping current output A process variable is selected Set reaction time for output 0.0 to 999.9 s 1.0 s in the Assign current output signal to fluctuations in the parameter (→...
  • Page 122: Configuring The Pulse/Frequency

    Commissioning Proline Promag H 500 PROFINET over Ethernet-APL 10.5.9 Configuring the pulse/frequency/switch output The Pulse/frequency/switch output wizard guides you systematically through all the parameters that can be set for configuring the selected output type. Navigation "Setup" menu → Advanced setup → Pulse/frequency/switch output ‣...
  • Page 123 Proline Promag H 500 PROFINET over Ethernet-APL Commissioning Parameter overview with brief description Parameter Prerequisite Description Selection / User Factory setting interface / User entry Operating mode – Define the output as a pulse, • Pulse Pulse frequency or switch output. • Frequency • Switch Terminal number –...
  • Page 124 Commissioning Proline Promag H 500 PROFINET over Ethernet-APL Assign frequency output →  124 Minimum frequency value →  125 →  125 Maximum frequency value Measuring value at minimum →  125 frequency Measuring value at maximum →  125...
  • Page 125 Proline Promag H 500 PROFINET over Ethernet-APL Commissioning Parameter Prerequisite Description Selection / User Factory setting interface / User entry Minimum frequency value The Frequency option is Enter minimum frequency. 0.0 to 10 000.0 Hz 0.0 Hz selected in the Operating mode parameter (→  122)
  • Page 126 Commissioning Proline Promag H 500 PROFINET over Ethernet-APL Configuring the switch output Navigation "Setup" menu → Pulse/frequency/switch output ‣ Pulse/frequency/switch output 1 to n Operating mode →  126 Terminal number →  126 Signal mode →  126 Switch output function →...
  • Page 127 Proline Promag H 500 PROFINET over Ethernet-APL Commissioning Parameter Prerequisite Description Selection / User Factory setting interface / User entry Switch output function The Switch option is selected Select function for switch • Off in the Operating mode output. • On parameter. • Diagnostic behavior •...
  • Page 128: 10.5.10 Configuring The Relay Output

    Commissioning Proline Promag H 500 PROFINET over Ethernet-APL Parameter Prerequisite Description Selection / User Factory setting interface / User entry Switch-on delay • The Switch option is Define delay for the switch-on 0.0 to 100.0 s 0.0 s selected in the Operating of status output.
  • Page 129 Proline Promag H 500 PROFINET over Ethernet-APL Commissioning Switch state →  130 Powerless relay status →  130 Parameter overview with brief description Parameter Prerequisite Description User interface /  Factory setting Selection / User entry Terminal number – Shows the terminal numbers •...
  • Page 130: 10.5.11 Configuring The Low Flow Cut Off

    Commissioning Proline Promag H 500 PROFINET over Ethernet-APL Parameter Prerequisite Description User interface /  Factory setting Selection / User entry Switch-on delay In the Relay output function Define delay for the switch-on 0.0 to 100.0 s 0.0 s parameter, the Limit option is of status output.
  • Page 131: 10.5.12 Configuring Empty Pipe Detection

    Proline Promag H 500 PROFINET over Ethernet-APL Commissioning 10.5.12 Configuring empty pipe detection • The measuring instruments are calibrated with water (approx. 500 µS/cm) at the factory. For liquids with a lower conductivity, it is advisable to perform a new full pipe adjustment onsite.
  • Page 132: 10.5.13 Configuring Flow Damping

    Commissioning Proline Promag H 500 PROFINET over Ethernet-APL 10.5.13 Configuring flow damping The Configure flow damping wizard guides the user systematically through the parameters, depending on the selected scenario: • Configuration of damping for the application To configure flow damping for the specific requirements of the process application.
  • Page 133: 10.5.14 "Build-Up Index Adjustment" Wizard

    Proline Promag H 500 PROFINET over Ethernet-APL Commissioning Parameter overview with brief description Parameter Description Selection / User interface Factory setting Scenario Select the applicable scenario. • Replace old device Configure damping for • Configure damping for application application • Restore factory settings Old device Select the measuring device to replace.
  • Page 134 Commissioning Proline Promag H 500 PROFINET over Ethernet-APL Navigation "Expert" menu → Sensor → Build-up index adjustment ‣ Build-up index adjustment Prerequisites →  134 Progress →  134 →  134 Build-up index reference value E 1 Signal to noise ratio →...
  • Page 135: Advanced Settings

    Proline Promag H 500 PROFINET over Ethernet-APL Commissioning 10.6 Advanced settings The Advanced setup submenu with its submenus contains parameters for specific settings. Navigation to the "Advanced setup" submenu X X X X X X X 20.50 0104-1 Main menu...
  • Page 136: Using The Parameter To Enter The Access Code

    Commissioning Proline Promag H 500 PROFINET over Ethernet-APL ‣ WLAN settings →  141 ‣ Electrode cleaning cycle →  143 ‣ →  144 Heartbeat setup ‣ Configuration backup →  144 ‣ Administration →  146 10.6.1 Using the parameter to enter the access code Navigation "Setup"...
  • Page 137 Proline Promag H 500 PROFINET over Ethernet-APL Commissioning Navigation "Setup" menu → Advanced setup → Totalizer 1 to n ‣ Totalizer 1 to n Assign process variable →  137 Unit totalizer →  137 →  137 Totalizer operation mode Failure mode →...
  • Page 138: Carrying Out Additional Display Configurations

    Commissioning Proline Promag H 500 PROFINET over Ethernet-APL 10.6.4 Carrying out additional display configurations In the Display submenu you can set all the parameters associated with the configuration of the local display. Navigation "Setup" menu → Advanced setup → Display ‣...
  • Page 139 Proline Promag H 500 PROFINET over Ethernet-APL Commissioning Parameter overview with brief description Parameter Prerequisite Description Selection / User Factory setting entry Format display A local display is provided. Select how measured values • 1 value, max. size 1 value, max. size are shown on the display.
  • Page 140 Commissioning Proline Promag H 500 PROFINET over Ethernet-APL Parameter Prerequisite Description Selection / User Factory setting entry 0% bargraph value 3 A selection was made in the Enter 0% value for bar graph Signed floating-point Country-specific: Value 3 display parameter. display. number •...
  • Page 141: Wlan Configuration

    Proline Promag H 500 PROFINET over Ethernet-APL Commissioning Parameter Prerequisite Description Selection / User Factory setting entry Separator A local display is provided. Select decimal separator for • . (point) . (point) displaying numerical values. • , (comma) Backlight One of the following Switch the local display •...
  • Page 142 Commissioning Proline Promag H 500 PROFINET over Ethernet-APL Parameter overview with brief description Parameter Prerequisite Description Selection / User Factory setting entry / User interface WLAN – Switch WLAN on and off. • Disable Enable • Enable WLAN mode – Select WLAN mode. • WLAN access...
  • Page 143: Performing Electrode Cleaning

    Proline Promag H 500 PROFINET over Ethernet-APL Commissioning 10.6.6 Performing electrode cleaning The Electrode cleaning cycle submenu contains the parameters that must be set for the configuration of electrode cleaning. The submenu is only available if the device was ordered with electrode cleaning.
  • Page 144: Performing Heartbeat Basic Setup

    Commissioning Proline Promag H 500 PROFINET over Ethernet-APL 10.6.7 Performing Heartbeat basic setup Heartbeat setup submenu guides the user systematically through all the parameters that can be used for the Heartbeat basic setup. The wizard only appears if the device has the Heartbeat Verification +Monitoring application package.
  • Page 145: Function Scope Of The "Configuration Management" Parameter

    Proline Promag H 500 PROFINET over Ethernet-APL Commissioning Navigation "Setup" menu → Advanced setup → Configuration backup ‣ Configuration backup Operating time →  145 Last backup →  145 →  145 Configuration management Backup state →  145 Comparison result →...
  • Page 146: Using Parameters For Device Administration

    Commissioning Proline Promag H 500 PROFINET over Ethernet-APL Options Description Compare The device configuration saved in the device memory is compared with the current device configuration of the HistoROM backup. Clear backup data The backup copy of the device configuration is deleted from the memory of the device.
  • Page 147: Simulation

    Proline Promag H 500 PROFINET over Ethernet-APL Commissioning Using the parameter to reset the access code Navigation "Setup" menu → Advanced setup → Administration → Reset access code ‣ Reset access code →  147 Operating time Reset access code →...
  • Page 148 Commissioning Proline Promag H 500 PROFINET over Ethernet-APL Navigation "Diagnostics" menu → Simulation ‣ Simulation Assign simulation process variable →  149 Process variable value →  149 →  150 Current input 1 to n simulation Value current input 1 to n →...
  • Page 149 Proline Promag H 500 PROFINET over Ethernet-APL Commissioning Parameter overview with brief description Parameter Prerequisite Description Selection / User Factory setting entry Assign simulation process variable – Select a process variable for • Off the simulation process that is • Volume flow activated.
  • Page 150: Protecting Settings From Unauthorized Access

    Commissioning Proline Promag H 500 PROFINET over Ethernet-APL Parameter Prerequisite Description Selection / User Factory setting entry Diagnostic event simulation – Select a diagnostic event to • Off simulate this event. • Diagnostic event picklist (depends on the category selected) Current input 1 to n simulation –...
  • Page 151 Via Web browser, FieldCare, DeviceCare (via CDI-RJ45 service interface), fieldbus You can only obtain a reset code from your local Endress+Hauser service organization. The code must be calculated explicitly for every device. 1. Note down the serial number of the device.
  • Page 152: Write Protection Via Write Protection Switch

    Commissioning Proline Promag H 500 PROFINET over Ethernet-APL 10.8.2 Write protection via write protection switch Unlike parameter write protection via a user-specific access code, this allows the user to lock write access to the entire operating menu - apart from the "Contrast display"...
  • Page 153 Proline Promag H 500 PROFINET over Ethernet-APL Commissioning NOTICE Excessive tightening torque applied to the fixing screws! Risk of damaging the plastic transmitter. ‣ Tighten the fixing screws as per the tightening torque: 2.5 Nm (1.8 lbf ft) Tighten the fixing screws.
  • Page 154: Operation

    Operation Proline Promag H 500 PROFINET over Ethernet-APL Operation 11.1 Reading off the device locking status Device active write protection: Locking status parameter Operation → Locking status Function scope of the "Locking status" parameter Options Description None The access authorization displayed in the Access status parameter applies →...
  • Page 155: Process Variables" Submenu

    Proline Promag H 500 PROFINET over Ethernet-APL Operation 11.4.1 "Process variables" submenu The Process variables submenu contains all the parameters needed to display the current measured values for each process variable. Navigation "Diagnostics" menu → Measured values → Process variables ‣...
  • Page 156: Totalizer

    Operation Proline Promag H 500 PROFINET over Ethernet-APL Parameter Prerequisite Description User interface Corrected conductivity One of the following conditions is met: Displays the conductivity currently Positive floating-point • Order code for "Sensor option", option corrected. number CI "Medium temperature Dependency measurement"...
  • Page 157: Input Values" Submenu

    Proline Promag H 500 PROFINET over Ethernet-APL Operation 11.4.3 "Input values" submenu The Input values submenu guides you systematically to the individual input values. Navigation "Diagnostics" menu → Measured values → Input values ‣ Input values ‣ Current input 1 to n →...
  • Page 158: Output Values

    Operation Proline Promag H 500 PROFINET over Ethernet-APL Parameter overview with brief description Parameter Description User interface Value status input Shows the current input signal level. • High • Low 11.4.4 Output values The Output values submenu contains all the parameters needed to display the current measured values for every output.
  • Page 159 Proline Promag H 500 PROFINET over Ethernet-APL Operation Navigation "Diagnostics" menu → Measured values → Output values → Pulse/frequency/switch output 1 to n ‣ Pulse/frequency/switch output 1 to n →  159 Output frequency Pulse output 1 to n →  159 Switch state →...
  • Page 160: Adapting The Measuring Device To The Process Conditions

    Operation Proline Promag H 500 PROFINET over Ethernet-APL 11.5 Adapting the measuring device to the process conditions The following are available for this purpose: • Basic settings using theSetup menu (→  109) • Advanced settings using theAdvanced setup submenu (→  135) 11.6...
  • Page 161: Function Range Of "Reset All Totalizers" Parameter

    Proline Promag H 500 PROFINET over Ethernet-APL Operation 11.6.2 Function range of "Reset all totalizers" parameter Options Description Cancel No action is executed and the user exits the parameter. Reset + totalize Resets all totalizers to 0 and restarts the totaling process. This deletes all the previously aggregated flow values.
  • Page 162 Operation Proline Promag H 500 PROFINET over Ethernet-APL Data logging →  163 Logging delay →  163 →  163 Data logging control Data logging status →  163 Entire logging duration →  163 Parameter overview with brief description...
  • Page 163 Proline Promag H 500 PROFINET over Ethernet-APL Operation Parameter Prerequisite Description Selection / User Factory setting entry / User interface Assign channel 4 The Extended HistoROM Assign a process variable to For the picklist, see application package is logging channel. Assign channel 1 available.
  • Page 164: Diagnostics And Troubleshooting

    Diagnostics and troubleshooting Proline Promag H 500 PROFINET over Ethernet-APL Diagnostics and troubleshooting 12.1 General troubleshooting For local display Error Possible causes Remedial action Local display is dark, but signal output is within The cable of the display module is not plugged in...
  • Page 165 Proline Promag H 500 PROFINET over Ethernet-APL Diagnostics and troubleshooting For access Error Possible causes Remedial action Write access to parameter not possible. Hardware write protection is enabled. Set the write protection switch on the main electronics module to the OFF position →...
  • Page 166: Diagnostic Information Via Light Emitting Diodes

    Diagnostics and troubleshooting Proline Promag H 500 PROFINET over Ethernet-APL For system integration Error Possible causes Remedy The PROFINET device name is not A device name containing one or Specify a correct device name displayed correctly and contains more underscores has been (without underscores) via the coding.
  • Page 167 Proline Promag H 500 PROFINET over Ethernet-APL Diagnostics and troubleshooting Color Meaning Flashing red A diagnostic event with "Warning" diagnostic behavior has occurred. A diagnostic event with "Alarm" diagnostic behavior has occurred. Flashing red/green The device restarts/self-test. Flashing/network status Green Cyclic data exchange is active.
  • Page 168: Sensor Connection Housing

    Diagnostics and troubleshooting Proline Promag H 500 PROFINET over Ethernet-APL Color Meaning A diagnostic event with "Alarm" diagnostic behavior has occurred. Flashing red/green The device restarts/self-test. Flashing/network status Green Cyclic data exchange is active. Flashing green Following request from automation system:...
  • Page 169 Proline Promag H 500 PROFINET over Ethernet-APL Diagnostics and troubleshooting Color Meaning Supply voltage Green Supply voltage is ok. Supply voltage is off or too low. Endress+Hauser...
  • Page 170: Diagnostic Information On Local Display

    Diagnostics and troubleshooting Proline Promag H 500 PROFINET over Ethernet-APL 12.3 Diagnostic information on local display 12.3.1 Diagnostic message Faults detected by the self-monitoring system of the measuring device are displayed as a diagnostic message in alternation with the operational display.
  • Page 171: Diagnostic Information

    Proline Promag H 500 PROFINET over Ethernet-APL Diagnostics and troubleshooting Diagnostic behavior Symbol Meaning Alarm • Measurement is interrupted. • Signal outputs and totalizers assume the defined alarm condition. • A diagnostic message is generated. Warning • Measurement is resumed.
  • Page 172: Calling Up Remedial Measures

    Diagnostics and troubleshooting Proline Promag H 500 PROFINET over Ethernet-APL 12.3.2 Calling up remedial measures X X X X X X X X X X X X X X 20.50 S801 Supply voltage Menu Diagnostic list Diagnostics 1 S801 Supply voltage...
  • Page 173: Calling Up Remedy Information

    Proline Promag H 500 PROFINET over Ethernet-APL Diagnostics and troubleshooting A0031056 Status area with status signal Diagnostic information Remedial measures with service ID In addition, diagnostic events which have occurred can be shown in the Diagnostics menu: • Via parameter →  195 •...
  • Page 174: Calling Up Remedy Information

    Diagnostics and troubleshooting Proline Promag H 500 PROFINET over Ethernet-APL Xxxxxx/…/…/ Device name: Xxxxxxx Mass flow: kg/h 12.34 Device tag: Xxxxxxx Volume flow: 12.34 m /h ³ Status signal: Function check (C) Xxxxxx Diagnostics 1: C485 Simu... Remedy information: Deactivate...
  • Page 175: Adapting The Diagnostic Information

    Proline Promag H 500 PROFINET over Ethernet-APL Diagnostics and troubleshooting 12.6 Adapting the diagnostic information 12.6.1 Adapting the diagnostic behavior Each item of diagnostic information is assigned a specific diagnostic behavior at the factory. The user can change this assignment for specific diagnostic information in the Diagnostic behavior submenu.
  • Page 176: Overview Of Diagnostic Information

    Diagnostics and troubleshooting Proline Promag H 500 PROFINET over Ethernet-APL Supported status information Status Coding (hex) BAD - Maintenance alarm 0x24 to 0x27 BAD - Process related 0x28 to 0x2B BAD - Function check 0x3C to 0x3F UNCERTAIN - Initial value...
  • Page 177 Proline Promag H 500 PROFINET over Ethernet-APL Diagnostics and troubleshooting Diagnostic information Remedy instructions Influenced measured variables Short text 083 Memory content inconsistent 1. Restart device • Conductivity 2. Restore S-DAT data • Corrected conductivity Measured variable status 3. Replace S-DAT •...
  • Page 178: Diagnostic Of Electronic

    Diagnostics and troubleshooting Proline Promag H 500 PROFINET over Ethernet-APL Diagnostic information Remedy instructions Influenced measured variables Short text 170 Coil resistance faulty Check ambient and process temperature • Conductivity • Corrected conductivity Measured variable status • Measured values • Density...
  • Page 179 Proline Promag H 500 PROFINET over Ethernet-APL Diagnostics and troubleshooting Diagnostic information Remedy instructions Influenced measured variables Short text 242 Firmware incompatible 1. Check firmware version • Conductivity 2. Flash or replace electronic module • Corrected conductivity Measured variable status •...
  • Page 180 Diagnostics and troubleshooting Proline Promag H 500 PROFINET over Ethernet-APL Diagnostic information Remedy instructions Influenced measured variables Short text 271 Main electronics faulty 1. Restart device • Conductivity 2. Replace main electronic module • Corrected conductivity Measured variable status • Measured values •...
  • Page 181 Proline Promag H 500 PROFINET over Ethernet-APL Diagnostics and troubleshooting Diagnostic information Remedy instructions Influenced measured variables Short text 276 I/O module faulty 1. Restart device • Conductivity 2. Change I/O module • Corrected conductivity Measured variable status • Measured values •...
  • Page 182 Diagnostics and troubleshooting Proline Promag H 500 PROFINET over Ethernet-APL Diagnostic information Remedy instructions Influenced measured variables Short text 311 Sensor electronics (ISEM) faulty Maintenance required! • Conductivity Do not reset device • Corrected conductivity Measured variable status • Measured values •...
  • Page 183 Proline Promag H 500 PROFINET over Ethernet-APL Diagnostics and troubleshooting Diagnostic information Remedy instructions Influenced measured variables Short text 361 I/O module 1 to n faulty 1. Restart device • Conductivity 2. Check electronic modules • Corrected conductivity Measured variable status 3.
  • Page 184 Diagnostics and troubleshooting Proline Promag H 500 PROFINET over Ethernet-APL Diagnostic information Remedy instructions Influenced measured variables Short text 376 Sensor electronics (ISEM) faulty 1. Replace sensor electronic module • Conductivity (ISEM) • Corrected conductivity Measured variable status [from the factory] 2.
  • Page 185: Diagnostic Of Configuration

    Proline Promag H 500 PROFINET over Ethernet-APL Diagnostics and troubleshooting Diagnostic information Remedy instructions Influenced measured variables Short text 383 Memory content Reset device • Conductivity • Corrected conductivity Measured variable status • Measured values • Density Quality Good • Electronics temperature Quality substatus •...
  • Page 186 Diagnostics and troubleshooting Proline Promag H 500 PROFINET over Ethernet-APL Diagnostic information Remedy instructions Influenced measured variables Short text 431 Trim 1 to n required Carry out trim – Measured variable status Quality Good Quality substatus Coding (hex) 0x80 to 0x83...
  • Page 187 Proline Promag H 500 PROFINET over Ethernet-APL Diagnostics and troubleshooting Diagnostic information Remedy instructions Influenced measured variables Short text 442 Frequency output 1 saturated 1. Check frequency output settings – 2. Check process Measured variable status Quality Good Quality substatus...
  • Page 188 Diagnostics and troubleshooting Proline Promag H 500 PROFINET over Ethernet-APL Diagnostic information Remedy instructions Influenced measured variables Short text 484 Failure mode simulation active Deactivate simulation • Conductivity • Corrected conductivity Measured variable status • Density • Electronics temperature Quality Good •...
  • Page 189 Proline Promag H 500 PROFINET over Ethernet-APL Diagnostics and troubleshooting Diagnostic information Remedy instructions Influenced measured variables Short text 492 Frequency output 1 to n simulation active Deactivate simulation frequency output – Measured variable status Quality Good Quality substatus Coding (hex)
  • Page 190 Diagnostics and troubleshooting Proline Promag H 500 PROFINET over Ethernet-APL Diagnostic information Remedy instructions Influenced measured variables Short text 496 Status input 1 to n simulation active Deactivate simulation status input – Measured variable status Quality Good Quality substatus Coding (hex)
  • Page 191 Proline Promag H 500 PROFINET over Ethernet-APL Diagnostics and troubleshooting Diagnostic information Remedy instructions Influenced measured variables Short text 530 Electrode cleaning active Switch off electrode cleaning • Conductivity • Corrected conductivity Measured variable status • Measured values • Density...
  • Page 192: Diagnostic Of Process

    Diagnostics and troubleshooting Proline Promag H 500 PROFINET over Ethernet-APL 12.7.4 Diagnostic of process Diagnostic information Remedy instructions Influenced measured variables Short text 803 Loop current 1 faulty 1. Check wiring – 2. Change I/O module Measured variable status Quality...
  • Page 193 Proline Promag H 500 PROFINET over Ethernet-APL Diagnostics and troubleshooting Diagnostic information Remedy instructions Influenced measured variables Short text 834 Process temperature too high Reduce process temperature • Conductivity • Corrected conductivity Measured variable status [from the factory] • Density •...
  • Page 194 Diagnostics and troubleshooting Proline Promag H 500 PROFINET over Ethernet-APL Diagnostic information Remedy instructions Influenced measured variables Short text 937 Sensor symmetry 1. Eliminate external magnetic field near • Conductivity sensor • Corrected conductivity Measured variable status [from the factory] 2.
  • Page 195: Pending Diagnostic Events

    Proline Promag H 500 PROFINET over Ethernet-APL Diagnostics and troubleshooting Diagnostic information Remedy instructions Influenced measured variables Short text 962 Pipe empty 1. Perform full pipe adjustment • Conductivity 2. Perform empty pipe adjustment • Corrected conductivity Measured variable status [from the factory] 3.
  • Page 196: Diagnostics List

    Diagnostics and troubleshooting Proline Promag H 500 PROFINET over Ethernet-APL Parameter overview with brief description Parameter Prerequisite Description User interface Actual diagnostics A diagnostic event has occurred. Shows the current occured diagnostic Symbol for diagnostic event along with its diagnostic behavior, diagnostic code information.
  • Page 197: 12.10.2 Filtering The Event Logbook

    Proline Promag H 500 PROFINET over Ethernet-APL Diagnostics and troubleshooting / ../Eventlist I1091 Config. change I1157 Mem.err. ev.list 0d01h19m10s F311 Electr. failure A0014008-EN  40 Using the example of the local display • A maximum of 20 event messages can be displayed in chronological order.
  • Page 198 Diagnostics and troubleshooting Proline Promag H 500 PROFINET over Ethernet-APL Info number Info name I1092 HistoROM backup deleted I1137 Electronics changed I1151 History reset I1155 Reset electronics temperature I1156 Memory error trend I1157 Memory error event list I1256 Display: access status changed...
  • Page 199: 12.11 Resetting The Measuring Device

    Proline Promag H 500 PROFINET over Ethernet-APL Diagnostics and troubleshooting Info number Info name I1712 New flash file received I1725 Sensor electronic module (ISEM) changed I1726 Configuration backup failed 12.11 Resetting the measuring device The entire device configuration or some of the configuration can be reset to a defined state with the Device reset parameter (→...
  • Page 200 Diagnostics and troubleshooting Proline Promag H 500 PROFINET over Ethernet-APL Parameter overview with brief description Parameter Description User interface Factory setting Device tag Shows name of measuring point. Character string comprising Promag numbers, letters and special characters Serial number Shows the serial number of the measuring Max.
  • Page 201: 12.13 Firmware History

    "Manufacturer' s information" document. The manufacturer' s information is available: • In the Download Area of the Endress+Hauser web site: www.endress.com → Downloads • Specify the following details: • Product root: e.g. 5H5B The product root is the first part of the order code: see the nameplate on the device.
  • Page 202: Maintenance

    Replacement seals (accessory part) →  236 13.2 Measuring and test equipment Endress+Hauser offers a variety of measuring and testing equipment, such as Netilion or device tests. Your Endress+Hauser Sales Center can provide detailed information on the services. List of some of the measuring and testing equipment: →  207 13.3...
  • Page 203: Repair

    • The measuring devices have a modular design. • Spare parts are grouped into logical kits with the associated Installation Instructions. • Repairs are carried out by Endress+Hauser Service or by appropriately trained customers. • Certified devices can only be converted to other certified devices by Endress+Hauser Service or at the factory.
  • Page 204: Disposal

    Repair Proline Promag H 500 PROFINET over Ethernet-APL 14.5 Disposal If required by the Directive 2012/19/EU on waste electrical and electronic equipment (WEEE), the product is marked with the depicted symbol in order to minimize the disposal of WEEE as unsorted municipal waste. Do not dispose of products bearing this marking as unsorted municipal waste.
  • Page 205: Accessories

    Various accessories, which can be ordered with the device or subsequently from Endress +Hauser, are available for the device. Detailed information on the order code in question is available from your local Endress+Hauser sales center or on the product page of the Endress+Hauser website: www.endress.com.
  • Page 206: For The Sensor

    Accessories Proline Promag H 500 PROFINET over Ethernet-APL Display guard Is used to protect the display against impact or scoring, for example from sand in Proline 500 – digital desert areas.  Order number: 71228792  Installation Instructions EA01093D Connecting cable The connecting cable can be ordered directly with the measuring device (order code Proline 500 –...
  • Page 207: Communication-Specific Accessories

    • As a downloadable DVD for local PC installation. Netilion lloT ecosystem: Unlock knowledge Endress+Hauser ' s Netilion lloT ecosystem enables you to optimize your plant performance, digitize workflows, share knowledge and improve collaboration. Based on decades of experience in process automation, Endress+Hauser offers the process industry an lloT ecosystem that enables you to gain useful insights from data.
  • Page 208: System Components

    Accessories Proline Promag H 500 PROFINET over Ethernet-APL Accessories Description FieldCare FDT-based plant asset management tool from Endress+Hauser. It can configure all smart field units in your system and helps you manage them. By using the status information, it is also a simple but effective way of checking their status and condition.
  • Page 209: Technical Data

    Proline Promag H 500 PROFINET over Ethernet-APL Technical data Technical data 16.1 Application The measuring device is intended only for the flow measurement of liquids with a minimum conductivity of 5 µS/cm. Depending on the version ordered, the measuring device can also measure potentially explosive, flammable, poisonous and oxidizing media.
  • Page 210 Technical data Proline Promag H 500 PROFINET over Ethernet-APL Recommended Nominal diameter Factory settings flow min./max. full Full scale value Pulse value Low flow cut off scale value current output (~ 2 pulse/s) (v ~ 0.04 m/s) (v ~ 0.3/10 m/s) (v ~ 2.5 m/s)
  • Page 211 Proline Promag H 500 PROFINET over Ethernet-APL Technical data Recommended Nominal diameter Factory settings flow Low flow min./max. full Full scale value cut off Pulse value scale value current output (v ~ (~ 2 pulse/s) (v ~ 0.3/10 m/s) (v ~ 2.5 m/s) 0.04 m/s...
  • Page 212: Status Input

    Technical data Proline Promag H 500 PROFINET over Ethernet-APL Open-circuit voltage ≤ 28.8 V (active) Possible input variables • Temperature • Density Status input Maximum input values • DC –3 to 30 V • If status input is active (ON): R >3 kΩ...
  • Page 213: Output

    Proline Promag H 500 PROFINET over Ethernet-APL Technical data 16.4 Output Output signal PROFINET with Ethernet-APL Device use Device connection to an APL field switch The device may only be operated according to the following APL port classifications: • If used in hazardous areas: SLAA or SLAC •...
  • Page 214 Technical data Proline Promag H 500 PROFINET over Ethernet-APL Damping Configurable: 0 to 999.9 s Assignable measured • Volume flow variables • Mass flow • Corrected volume flow • Flow velocity • Conductivity • Corrected conductivity • Temperature • Electronics temperature...
  • Page 215 Proline Promag H 500 PROFINET over Ethernet-APL Technical data Switching delay Configurable: 0 to 100 s Number of switching Unlimited cycles Assignable functions • Disable • On • Diagnostic behavior • Limit value: • Disable • Volume flow • Mass flow •...
  • Page 216 Technical data Proline Promag H 500 PROFINET over Ethernet-APL The following inputs and outputs are available for assignment: • Choice of current output: 4 to 20 mA (active), 0/4 to 20 mA (passive) • Pulse/frequency/switch output • Choice of current input: 4 to 20 mA (active), 0/4 to 20 mA (passive) •...
  • Page 217 Proline Promag H 500 PROFINET over Ethernet-APL Technical data Local display Plain text display With information on cause and remedial measures Backlight Red lighting indicates a device error. Status signal as per NAMUR recommendation NE 107 Interface/protocol • Via digital communication: PROFINET with Ethernet-APL •...
  • Page 218: Power Supply

    Technical data Proline Promag H 500 PROFINET over Ethernet-APL System redundancy support System redundancy S2 (2 AR with 1 NAP) Device profile PROFINET PA profile 4 (Application interface identifier API: 0x9700) Manufacturer ID Device type ID 0xA43C Device description files (GSD,...
  • Page 219 Proline Promag H 500 PROFINET over Ethernet-APL Technical data Power consumption Transmitter Max. 10 W (active power) switch-on current Max. 36 A (<5 ms) as per NAMUR Recommendation NE 21 Current consumption Transmitter • Max. 400 mA (24 V) • Max. 200 mA (110 V, 50/60 Hz; 230 V, 50/60 Hz) Power supply failure •...
  • Page 220: Performance Characteristics

    Technical data Proline Promag H 500 PROFINET over Ethernet-APL 16.6 Performance characteristics Reference operating • Error limits following DIN EN 29104, in future ISO 20456 conditions • Water, typically: +15 to +45 °C (+59 to +113 °F); 0.5 to 7 bar (73 to 101 psi) •...
  • Page 221: Electrical Conductivity

    Proline Promag H 500 PROFINET over Ethernet-APL Technical data Conductivity Nominal diameter Measurement error [µS/cm] [%] of reading [mm] [in] > 10 000 to 20 000 2…150 ¹⁄₁₂ to 6 ± 10% > 20 000 to 100 000 2…150 ¹⁄₁₂ to 6 ±...
  • Page 222: Mounting

    The device is suitable for use outdoors and indoors with a relative humidity of 4 to 95 %. Operating height According to EN 61010-1 • ≤ 2 000 m (6 562 ft) • > 2 000 m (6 562 ft) with additional overvoltage protection (e.g. Endress+Hauser HAW Series) Degree of protection Transmitter •...
  • Page 223: Process

    Proline Promag H 500 PROFINET over Ethernet-APL Technical data Sensor • IP66/67, Type 4X enclosure, suitable for pollution degree 4 • When the housing is open: IP20, Type 1 enclosure, suitable for pollution degree 2 External WLAN antenna IP67 Vibration-resistance and...
  • Page 224 Technical data Proline Promag H 500 PROFINET over Ethernet-APL [°F] [°C] 40 60 80 100 120 140 160 [°C] [°F] A0027806  44 Promag 500 – digital Ambient temperature range Fluid temperature [°F] [°C] 40 60 80 100 120 140 160 [°C]...
  • Page 225: 16.10 Mechanical Construction

    Proline Promag H 500 PROFINET over Ethernet-APL Technical data Pressure-temperature For an overview of the pressure-temperature ratings for the process connections, see ratings the Technical Information Pressure tightness Liner: PFA Nominal diameter Limit values for absolute pressure in [mbar] ([psi]) for medium temperatures:...
  • Page 226: Specification

    Technical data Proline Promag H 500 PROFINET over Ethernet-APL Nominal diameter Weight [mm] [in] [kg] [lbs] 5/16 2.00 4.41 ½ 1.90 4.19 2.80 6.17 1 ½ 4.10 9.04 4.60 10.1 – 5.40 11.9 6.00 13.2 7.30 16.1 12.7 28.0 15.1 33.3...
  • Page 227 Proline Promag H 500 PROFINET over Ethernet-APL Technical data Window material Order code for "Transmitter housing": • Option A "Aluminum, coated": glass • Option D "Polycarbonate": plastic Sensor connection housing Order code for "Sensor connection housing": • Option A "Aluminum coated": aluminum, AlSi10Mg, coated •...
  • Page 228 Technical data Proline Promag H 500 PROFINET over Ethernet-APL Electrodes Standard: 1.4435 (316L) Seals • O-ring seal, DN 2 to 25 (1/12 to 1"): EPDM, FKM , Kalrez • Aseptic gasket seal, DN 2 to 150 (1/12 to 6"): EPDM, FKM...
  • Page 229: 16.11 Display And User Interface

    Proline Promag H 500 PROFINET over Ethernet-APL Technical data (All data refer to parts in contact with the medium) Liner with PFA: ≤ 0.4 µm (15.7 µin) (All data refer to parts in contact with the medium) Stainless steel process connections: •...
  • Page 230 Technical data Proline Promag H 500 PROFINET over Ethernet-APL Display elements • 4-line, illuminated, graphic display • White background lighting; switches to red in event of device errors • Format for displaying measured variables and status variables can be individually...
  • Page 231: Historom Data Management

    Proline Promag H 500 PROFINET over Ethernet-APL Technical data device health. Furthermore the device data can be managed and the network parameters can be configured. Access to the network is required for the Ethernet-APL connection. A device that has a WLAN interface (can be ordered as an option) is required for the WLAN connection: order code for "Display;...
  • Page 232 Technical data Proline Promag H 500 PROFINET over Ethernet-APL Data backup Automatic • The most important device data (sensor and transmitter) are automatically saved in the DAT modules • If the transmitter or measuring device is replaced: once the T-DAT containing the...
  • Page 233: 16.12 Certificates And Approvals

    The device meets the legal requirements of the applicable EU Directives. These are listed in the corresponding EU Declaration of Conformity along with the standards applied. Endress+Hauser confirms successful testing of the device by affixing to it the CE mark. UKCA marking The device meets the legal requirements of the applicable UK regulations (Statutory Instruments).
  • Page 234: Pharmaceutical Compatibility

    Directive a) PED/G1/x (x = category) or b) PESR/G1/x (x = category) on the sensor nameplate, Endress+Hauser confirms compliance with the "Essential Safety Requirements" a) specified in Annex I of the Pressure Equipment Directive 2014/68/EU or b) Schedule 2 of Statutory Instruments 2016 No. 1105.
  • Page 235: 16.13 Application Packages

    The application packages can be ordered with the device or subsequently from Endress+Hauser. Detailed information on the order code in question is available from your local Endress+Hauser sales center or on the product page of the Endress+Hauser website: www.endress.com.
  • Page 236: Heartbeat Verification

    For an overview of the scope of the associated Technical Documentation, refer to the following: • Device Viewer (www.endress.com/deviceviewer): Enter the serial number from the nameplate • Endress+Hauser Operations app: Enter serial number from nameplate or scan matrix code on nameplate. Standard documentation Brief Operating Instructions...
  • Page 237: Technical Information

    Proline Promag H 500 PROFINET over Ethernet-APL Technical data Brief Operating Instructions for the transmitter Measuring device Documentation code Proline 500 KA01518D Proline 500 – digital KA01519D Technical Information Measuring device Documentation code Promag H 500 TI01225D Description of Device Parameters...
  • Page 238: Installation Instructions

    Technical data Proline Promag H 500 PROFINET over Ethernet-APL Installation instructions Contents Note Installation instructions for spare part sets and • Access the overview of all the available spare part accessories sets via Device Viewer →  203 • Accessories available for order with Installation Instructions →...
  • Page 239: Index

    Proline Promag H 500 PROFINET over Ethernet-APL Index Index Terminal assignment of Proline 500 - digital ..45 Connecting the measuring instrument Access authorization to parameters Proline 500 ......54 Read access .
  • Page 240 Index Proline Promag H 500 PROFINET over Ethernet-APL Web browser ......172 Field of application Diagnostic message .
  • Page 241 Proline Promag H 500 PROFINET over Ethernet-APL Index see Operational display Mounting location ......21 Text editor .
  • Page 242 Index Proline Promag H 500 PROFINET over Ethernet-APL Configure flow damping (Wizard) ... . 132 Fluid temperature ..... . . 223 Current input .
  • Page 243 Proline Promag H 500 PROFINET over Ethernet-APL Index Resetting the totalizer ....160 Totalizer ......156 Sensor adjustment .
  • Page 244 Index Proline Promag H 500 PROFINET over Ethernet-APL Troubleshooting General ....... 164 TSE/BSE Certificate of Suitability .
  • Page 246 *71661683* 71661683 www.addresses.endress.com...