Operating Instructions
Proline Promag H 100
HART
Electromagnetic flowmeter
Solutions
Services

Advertisement

loading
Need help?

Need help?

Do you have a question about the Proline Promag H 100 HART and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Endress+Hauser Proline Promag H 100 HART

  • Page 1 Products Solutions Services BA01171D/06/EN/03.24-00 71689937 2024-12-20 Valid as of version 01.00.zz (Device firmware) Operating Instructions Proline Promag H 100 HART Electromagnetic flowmeter...
  • Page 2 • The manufacturer reserves the right to modify technical data without prior notice. Your Endress+Hauser Sales Center will supply you with current information and updates to these Instructions. Endress+Hauser...
  • Page 3: Table Of Contents

    Proline Promag H 100 HART Table of contents Table of contents Mounting the measuring device ... 23 About this document ....6 6.2.1...
  • Page 4 10.5.6 Using parameters for device 13.3 Endress+Hauser services ....102 administration ....
  • Page 5 Proline Promag H 100 HART Table of contents 16.3 Input ......108 16.4 Output .
  • Page 6: About This Document

    About this document Proline Promag H 100 HART About this document Document function These Operating Instructions contain all the information that is required in various phases of the life cycle of the device: from product identification, incoming acceptance and storage, to mounting, connection, operation and commissioning through to troubleshooting, maintenance and disposal.
  • Page 7: Symbols For Certain Types Of Information

    • The W@M Device Viewer : Enter the serial number from the nameplate (www.endress.com/deviceviewer) • The Endress+Hauser Operations App: Enter the serial number from the nameplate or scan the 2-D matrix code (QR code) on the nameplate. For a detailed list of the individual documents along with the documentation code...
  • Page 8: Standard Documentation

    About this document Proline Promag H 100 HART 1.3.1 Standard documentation Document type Purpose and content of the document Technical Information Planning aid for your device The document contains all the technical data on the device and provides an overview of the accessories and other products that can be ordered for the device.
  • Page 9: Basic Safety Instructions

    Proline Promag H 100 HART Basic safety instructions Basic safety instructions Requirements for the personnel The personnel for installation, commissioning, diagnostics and maintenance must fulfill the following requirements: ‣ Trained, qualified specialists must have a relevant qualification for this specific function and task.
  • Page 10: Workplace Safety

    Verification for borderline cases: ‣ For special fluids and fluids for cleaning, Endress+Hauser is glad to provide assistance in verifying the corrosion resistance of fluid-wetted materials, but does not accept any warranty or liability as minute changes in the temperature, concentration or level of contamination in the process can alter the corrosion resistance properties.
  • Page 11: Security

    Proline Promag H 100 HART Basic safety instructions IT security We only provide a warranty if the device is installed and used as described in the Operating Instructions. The device is equipped with security mechanisms to protect it against any inadvertent changes to the device settings.
  • Page 12: Product Description

    Product description Proline Promag H 100 HART Product description The device consists of a transmitter and a sensor. The device is available as a compact version: The transmitter and sensor form a mechanical unit. Product design 3.1.1 Device version with HART communication type A0023153 ...
  • Page 13: 4 Incoming Acceptance And Product

    A0028673 and documents present? • If one of the conditions is not satisfied, contact your Endress+Hauser Sales Center. • Depending on the device version, the CD-ROM might not be part of the delivery! The Technical Documentation is available via the Internet or via the Endress+Hauser Operations App, see the "Product identification"...
  • Page 14: Transmitter Nameplate

    "Supplementary device-dependent documentation" →  8 • The W@M Device Viewer: Enter the serial number from the nameplate (www.endress.com/deviceviewer) • The Endress+Hauser Operations App: Enter the serial number from the nameplate or scan the 2-D matrix code (QR code) on the nameplate. 4.2.1...
  • Page 15: Sensor Nameplate

    Proline Promag H 100 HART Incoming acceptance and product identification 4.2.2 Sensor nameplate Order code: Ser. no.: Ext. ord. cd.: Date: A0029204  3 Example of sensor nameplate Name of the sensor Manufacturing location Order code Serial number (ser. no.) Extended order code (Ext.
  • Page 16: Symbols On Measuring Device

    Incoming acceptance and product identification Proline Promag H 100 HART 4.2.3 Symbols on measuring device Symbol Meaning WARNING! This symbol alerts you to a dangerous situation. Failure to avoid this situation can result in serious or fatal injury. Reference to documentation Refers to the corresponding device documentation.
  • Page 17: Storage And Transport

    Proline Promag H 100 HART Storage and transport Storage and transport Storage conditions Observe the following notes for storage: ‣ Store in the original packaging to ensure protection from shock. ‣ Do not remove protective covers or protective caps installed on process connections.
  • Page 18: Measuring Devices With Lifting Lugs

    Storage and transport Proline Promag H 100 HART 5.2.2 Measuring devices with lifting lugs CAUTION Special transportation instructions for devices with lifting lugs ‣ Only use the lifting lugs fitted on the device or flanges to transport the device. ‣...
  • Page 19: Installation

    Proline Promag H 100 HART Installation Installation Installation conditions 6.1.1 Mounting position Mounting location A0029343 Preferably install the sensor in an ascending pipe, and ensure a sufficient distance to the next pipe elbow: h ≥ 2 × DN Installation in down pipes Install a siphon with a vent valve downstream of the sensor in down pipes whose length h ≥...
  • Page 20: Inlet And Outlet Runs

    Installation Proline Promag H 100 HART Orientation The direction of the arrow on the sensor nameplate helps you to install the sensor according to the flow direction (direction of medium flow through the piping). Orientation Recommendation Vertical orientation A0015591 Horizontal orientation, transmitter at...
  • Page 21: Requirements From Environment And Process

    Proline Promag H 100 HART Installation ≥ 5 × DN ≥ 2 × DN A0028997  5 Order code for "Design", option A "Insertion length short, ISO/DVGW until DN400, DN450-2000 1:1" and order code for "Design", option B "Insertion length long, ISO/DVGW until DN400, DN450-2000 1:1.3"...
  • Page 22 Installation Proline Promag H 100 HART Never install the sensor on the pump suction side in order to avoid the risk of low pressure, and thus damage to the liner. Furthermore, install pulse dampers if reciprocating, diaphragm or peristaltic pumps are used.
  • Page 23: Mounting The Measuring Device

    Proline Promag H 100 HART Installation [mbar] 8 m/s 7 m/s 6 m/s 5 m/s 4 m/s 3 m/s max. 8° 2 m/s 1 m/s d / D A0029002 Mounting the measuring device 6.2.1 Required tools For sensor For flanges and other process connections: Corresponding mounting tools 6.2.2...
  • Page 24: Turning The Display Module

    Installation Proline Promag H 100 HART 5. Install the measuring device or turn the transmitter housing so that the cable entries do not point upwards.  A0029263 Mounting the seals CAUTION An electrically conductive layer could form on the inside of the measuring tube! Risk of measuring signal short circuit.
  • Page 25: Post-Installation Check

    Proline Promag H 100 HART Installation Compact and ultra-compact housing version, hygienic, stainless 8 mm A0023195 Post-installation check Is the device undamaged (visual inspection)?  Does the measuring device conform to the measuring point specifications? For example: • Process temperature ...
  • Page 26: Electrical Connection

    Electrical connection Proline Promag H 100 HART Electrical connection WARNING Live parts! Incorrect work performed on the electrical connections can result in an electric shock. ‣ Set up a disconnecting device (switch or power-circuit breaker) to easily disconnect the device from the supply voltage.
  • Page 27: Terminal Assignment

    Proline Promag H 100 HART Electrical connection Cable diameter • Cable glands supplied: M20 × 1.5 with cable ⌀ 6 to 12 mm (0.24 to 0.47 in) • Spring terminals: Wire cross-sections 0.5 to 2.5 mm (20 to 14 AWG) 7.2.3...
  • Page 28: Pin Assignment, Device Plug

    Electrical connection Proline Promag H 100 HART Terminal number Order code for Power supply Output 1 Output 2 "Output" 2 (L-) 1 (L+) 27 (–) 26 (+) 25 (–) 24 (+) Option B DC 24 V 4-20 mA HART (active)
  • Page 29: Connecting The Device

    Proline Promag H 100 HART Electrical connection Connecting the device NOTICE An incorrect connection compromises electrical safety! ‣ Only properly trained specialist staff may perform electrical connection work. ‣ Observe applicable federal/national installation codes and regulations. ‣ Comply with local workplace safety regulations.
  • Page 30 Electrical connection Proline Promag H 100 HART 8 mm 8 mm 3 mm 10 (0.4) mm (in) A0017844  11 Device versions with connection examples Cable Device plug for signal transmission Device plug for supply voltage For device version with device plug: follow step 6 only.
  • Page 31: Ensuring Potential Equalization

    • Grounding rings can be ordered separately as accessory DK5HR* from Endress+Hauser (does not contain any seals). When ordering make sure that the grounding rings are...
  • Page 32 Electrical connection Proline Promag H 100 HART Potential equalization via additional grounding ring A0028971 Hexagonal-headed bolts of process connection O-ring seals Plastic disk (spacer) or grounding ring Sensor Potential equalization via grounding electrodes on process connection A0028972 Hexagonal-headed bolts of process connection...
  • Page 33: Special Connection Instructions

    Proline Promag H 100 HART Electrical connection Special connection instructions 7.5.1 Connection examples Current output 4 to 20 mA HART 4...20 mA A0029055  12 Connection example for 4 to 20 mA HART current output (active) Automation system with current input (e.g. PLC) Cable shield provided at one end.
  • Page 34: Pulse/Frequency Output

    Electrical connection Proline Promag H 100 HART Pulse/frequency output 12345 A0028761  14 Connection example for pulse/frequency output (passive) Automation system with pulse/frequency input (e.g. PLC with 10 kΩ pull-up or pull-down resistor) Power supply Transmitter: observe input values Switch output A0028760 ...
  • Page 35: Ensuring The Degree Of Protection

    Proline Promag H 100 HART Electrical connection 4...20 mA 4...20 mA A0019830  17 Connection example for HART input (master mode) via current output (active) Automation system with current input (e.g. PLC). Prerequisite: automation system with HART version 6, HART commands 113 and 114 can be processed.
  • Page 36 Electrical connection Proline Promag H 100 HART Are all cable glands installed, securely tightened and leak-tight? Cable run with "water trap"  →  35? Depending on the device version:  Are all connectors securely tightened →  29? Does the supply voltage match the specifications on the transmitter nameplate →  113? ...
  • Page 37: Operation Options

    Proline Promag H 100 HART Operation options Operation options Overview of operating options A0019598 Computer with Web browser (e.g. Internet Explorer) or with operating tool (e.g. FieldCare, AMS Device Manager, SIMATIC PDM) Field Xpert SFX350 or SFX370 Field Communicator 475 Control system (e.g.
  • Page 38: Structure And Function Of The Operating Menu

    Operation options Proline Promag H 100 HART Structure and function of the operating menu 8.2.1 Structure of the operating menu For an overview of the operating menu for experts: "Description of Device Parameters" document supplied with the device Operating menu for operators and maintenances...
  • Page 39: Operating Philosophy

    Proline Promag H 100 HART Operation options 8.2.2 Operating philosophy The individual parts of the operating menu are assigned to certain user roles (operator, maintenance etc.). Each user role contains typical tasks within the device lifecycle. Menu/parameter User role and tasks...
  • Page 40: Access To The Operating Menu Via The Web Browser

    Operation options Proline Promag H 100 HART Access to the operating menu via the web browser 8.3.1 Function range Thanks to the integrated Web server, the device can be operated and configured via a Web browser and via a service interface (CDI-RJ45) . In addition to the measured values, status information on the device is also displayed and allows the user to monitor the status of the device.
  • Page 41: Establishing A Connection

    Proline Promag H 100 HART Operation options 8.3.3 Establishing a connection Via service interface (CDI-RJ45) Preparing the measuring device Configuring the Internet protocol of the computer The following information refers to the default Ethernet settings of the device. IP address of the device: 192.168.1.212 (factory setting) 1.
  • Page 42: Logging On

    Operation options Proline Promag H 100 HART 2. Enter the IP address of the Web server in the address line of the Web browser: 192.168.1.212  The login page appears. 2 3 4 A0029417 Picture of device Device name Device tag (→  54)
  • Page 43: User Interface

    Proline Promag H 100 HART Operation options 8.3.5 User interface A0032879 Picture of device Device name Device tag Status signal Current measured values Navigation area Local display language Header The following information appears in the header: • Device tag • Device status with status signal →  89 •...
  • Page 44: Disabling The Web Server

    Operation options Proline Promag H 100 HART Navigation area If a function is selected in the function bar, the submenus of the function open in the navigation area. The user can now navigate through the menu structure. Working area Depending on the selected function and the related submenus, various actions can be performed in this area: •...
  • Page 45: Access To The Operating Menu Via The Operating Tool

    VIATOR Bluetooth modem with connecting cable Transmitter Via service interface (CDI) A0014019 Service interface (CDI = Endress+Hauser Common Data Interface) of the measuring device Commubox FXA291 Computer with FieldCare operating tool with COM DTM CDI Communication FXA291 Via service interface (CDI-RJ45)
  • Page 46: Field Xpert Sfx350, Sfx370

    FieldCare Function scope FDT-based plant asset management tool from Endress+Hauser. It can configure all smart field devices in a system and helps you manage them. By using the status information, it is also a simple but effective way of checking their status and condition.
  • Page 47 Proline Promag H 100 HART Operation options Establishing a connection 1. Start FieldCare and launch the project. 2. In the network: Add a device.  The Add device window opens. 3. Select the CDI Communication TCP/IP option from the list and press OK to confirm.
  • Page 48: Devicecare

    DeviceCare Function scope Tool to connect and configure Endress+Hauser field devices. The fastest way to configure Endress+Hauser field devices is with the dedicated "DeviceCare" tool. Together with the device type managers (DTMs) it presents a convenient, comprehensive solution. For details, see Innovation Brochure IN01047S Source for device description files See information →...
  • Page 49: System Integration

    Proline Promag H 100 HART System integration System integration Overview of device description files 9.1.1 Current version data for the device Firmware version 01.01.zz • On the title page of the Operating instructions • On the transmitter nameplate • Firmware version Diagnostics →...
  • Page 50: Device Variables

    System integration Proline Promag H 100 HART Dynamic variables Measured variables (HART device variables) Primary dynamic variable (PV) Volume flow Secondary dynamic variable (SV) Totalizer 1 Tertiary dynamic variable (TV) Totalizer 2 Quaternary dynamic variable (QV) Totalizer 3 The assignment of the measured variables to the dynamic variables can be modified and...
  • Page 51: Other Settings

    Proline Promag H 100 HART System integration Other settings 9.3.1 Burst mode functionality in accordance with HART 7 Specification Navigation "Expert" menu → Communication → HART output → Burst configuration → Burst configuration 1 to n ‣ Burst configuration ‣...
  • Page 52: Parameter Overview With Brief Description

    System integration Proline Promag H 100 HART Parameter overview with brief description Parameter Description Selection / User entry Burst mode 1 to n Activate the HART burst mode for burst message X. • Off • On Burst command 1 to n Select the HART command that is sent to the HART master.
  • Page 53: Commissioning

    Proline Promag H 100 HART Commissioning Commissioning 10.1 Function check Before commissioning the measuring device: ‣ Make sure that the post-installation and post-connection checks have been performed. • "Post-installation check" checklist→  25 • "Post-connection check" checklist →  35 10.2...
  • Page 54: Defining The Tag Name

    Commissioning Proline Promag H 100 HART 10.4.1 Defining the tag name To enable fast identification of the measuring point within the system, you can enter a unique designation using the Device tag parameter and thus change the factory setting. Enter the tag name in the "FieldCare" operating tool →  47 Navigation "Setup"...
  • Page 55: Configuring The Current Output

    Proline Promag H 100 HART Commissioning 10.4.2 Configuring the current output The Current output submenu guides you systematically through all the parameters that have to be set for configuring the current output. Navigation "Setup" menu → Current output 1 Structure of the submenu ‣...
  • Page 56: Configuring The Pulse/Frequency

    Commissioning Proline Promag H 100 HART Parameter Prerequisite Description Selection / User Factory setting entry 20 mA value One of the following options is Enter 20 mA value. Signed floating-point Depends on country selected in the Current span number and nominal parameter (→  55): diameter •...
  • Page 57 Proline Promag H 100 HART Commissioning Assign flow direction check →  61 Assign status →  61 →  58 Value per pulse Pulse width →  58 Failure mode →  58 Minimum frequency value →  59 →...
  • Page 58 Commissioning Proline Promag H 100 HART Parameter Prerequisite Description Selection / User Factory setting entry Value per pulse In the Operating mode Enter measured value at which Signed floating-point Depends on country parameter, the Pulse option is a pulse is output. number...
  • Page 59 Proline Promag H 100 HART Commissioning Parameter Prerequisite Description Selection / User Factory setting entry Minimum frequency value In the Operating mode Enter minimum frequency. 0.0 to 10 000.0 Hz – parameter (→  57), the Frequency option is selected, and one of the following...
  • Page 60 Commissioning Proline Promag H 100 HART Parameter Prerequisite Description Selection / User Factory setting entry Damping output In the Operating mode Set reaction time for output 0 to 999.9 s – parameter (→  57) the signal to fluctuations in the Frequency option is selected measured value.
  • Page 61 Proline Promag H 100 HART Commissioning Parameter overview with brief description Parameter Prerequisite Description Selection / User Factory setting entry Operating mode – Define the output as a pulse, • Pulse – frequency or switch output. • Frequency • Switch Switch output function...
  • Page 62: Configuring The Local Display

    Commissioning Proline Promag H 100 HART Parameter Prerequisite Description Selection / User Factory setting entry Switch-off value • In the Operating mode Enter measured value for the Signed floating-point Country-specific: parameter, the Switch switch-off point. number • 0 l/h option is selected.
  • Page 63: Configuring The Output Conditioning

    Proline Promag H 100 HART Commissioning Parameter overview with brief description Parameter Prerequisite Description Selection / User Factory setting entry Format display A local display is provided. Select how measured values • 1 value, max. size – are shown on the display.
  • Page 64: Output Conditioning

    Commissioning Proline Promag H 100 HART Navigation "Setup" menu → Output conditioning Structure of the"Output conditioning" submenu ‣ Output conditioning Assign current output →  64 Damping output 1 →  64 Measuring mode output 1 →  64 →  64...
  • Page 65: Configuring The Low Flow Cut Off

    Proline Promag H 100 HART Commissioning Parameter Prerequisite Description Selection / User entry Measuring mode output 1 – Select measuring mode for output. • Forward flow • Forward/Reverse flow • Reverse flow • Reverse flow compensation Assign pulse output The Pulse option is selected in the Select process variable for pulse output.
  • Page 66 Commissioning Proline Promag H 100 HART Parameter Prerequisite Description Selection / User Factory setting entry Off value low flow cutoff One of the following options is Enter off value for low flow cut 0 to 100.0 % – selected in the Assign process off.
  • Page 67: Configuring Empty Pipe Detection

    Proline Promag H 100 HART Commissioning 10.4.7 Configuring empty pipe detection The Empty pipe detection submenu contains parameters that must be configured for the configuration of empty pipe detection. Navigation "Setup" menu → Empty pipe detection ‣ Empty pipe detection Empty pipe detection →...
  • Page 68 Commissioning Proline Promag H 100 HART Device type →  68 Manufacturer ID →  68 →  68 Burst command Slot number →  68 Timeout →  68 Failure mode →  69 →  69 Failure value...
  • Page 69 Proline Promag H 100 HART Commissioning Parameter Prerequisite Description Selection / User Factory setting entry Failure mode In the Capture mode Define behavior if external • Alarm – parameter, the Burst network process variable is missed. • Last valid value option or Master network •...
  • Page 70: Advanced Settings

    Commissioning Proline Promag H 100 HART 10.5 Advanced settings The Advanced setup submenu together with its submenus contains parameters for specific settings. Navigation "Setup" menu → Advanced setup ‣ Advanced setup Enter access code ‣ System units →  70 ‣...
  • Page 71 Proline Promag H 100 HART Commissioning Corrected volume flow unit →  71 Corrected volume unit →  71 Parameter overview with brief description Parameter Prerequisite Description Selection Factory setting Volume flow unit – Select volume flow unit. Unit choose list Country-specific: •...
  • Page 72: Carrying Out A Sensor Adjustment

    Commissioning Proline Promag H 100 HART 10.5.2 Carrying out a sensor adjustment The Sensor adjustment submenu contains parameters that pertain to the functionality of the sensor. Navigation "Setup" menu → Advanced setup → Sensor adjustment ‣ Sensor adjustment Installation direction →...
  • Page 73 Proline Promag H 100 HART Commissioning Parameter Prerequisite Description Selection Factory setting Totalizer operation mode One of the following options is Select totalizer calculation • Net flow total – selected in the Assign process mode. • Forward flow total variable parameter (→  72) •...
  • Page 74: Carrying Out Additional Display Configurations

    Commissioning Proline Promag H 100 HART 10.5.4 Carrying out additional display configurations In the Display submenu you can set all the parameters associated with the configuration of the local display. Navigation "Setup" menu → Advanced setup → Display ‣ Display Format display →...
  • Page 75 Proline Promag H 100 HART Commissioning Parameter overview with brief description Parameter Prerequisite Description Selection / User Factory setting entry Format display A local display is provided. Select how measured values • 1 value, max. size – are shown on the display.
  • Page 76: Performing Electrode Cleaning

    Commissioning Proline Promag H 100 HART Parameter Prerequisite Description Selection / User Factory setting entry Decimal places 4 A measured value is specified Select the number of decimal • x – in the Value 4 display places for the display value. • x.x parameter.
  • Page 77: Using Parameters For Device Administration

    Proline Promag H 100 HART Commissioning ECC duration →  77 ECC recovery time →  77 →  77 ECC cleaning cycle ECC Polarity →  77 Parameter overview with brief description Parameter Prerequisite Description Selection / User Factory setting entry / User...
  • Page 78: Simulation

    Commissioning Proline Promag H 100 HART Parameter overview with brief description Parameter Description User entry / Selection Define access code Define release code for write access to parameters. 0 to 9 999 Device reset Reset the device configuration - either entirely or in part - to a •...
  • Page 79 Proline Promag H 100 HART Commissioning Parameter overview with brief description Parameter Prerequisite Description Selection / User entry Assign simulation process variable – Select a process variable for the • Off simulation process that is activated. • Volume flow • Mass flow •...
  • Page 80: Protecting Settings From Unauthorized Access

    Commissioning Proline Promag H 100 HART 10.7 Protecting settings from unauthorized access The following options exist for protecting the configuration of the measuring device from unintentional modification after commissioning: • Write protection via access code for Web browser →  80 •...
  • Page 81 Proline Promag H 100 HART Commissioning 2 = Not used 1 = Write protection A0022571 Disconnect the T-DAT from the main electronics module. 4. Setting the write protection switch on the main electronics module to the ON position enables the hardware write protection. Setting the write protection switch on the main electronics module to the OFF position (factory setting) disables the hardware write protection.
  • Page 82: Operation

    Operation Proline Promag H 100 HART Operation 11.1 Reading the device locking status Device active write protection: Locking status parameter Navigation "Operation" menu → Locking status Function scope of "Locking status" parameter Options Description Hardware locked The locking switch (DIP switch) for locking the hardware is activated on the main electronic module.
  • Page 83: Totalizer" Submenu

    Proline Promag H 100 HART Operation Corrected volume flow →  83 Temperature →  83 →  83 Corrected conductivity Parameter overview with brief description Parameter Prerequisite Description User interface Volume flow – Displays the volume flow currently Signed floating-point measured.
  • Page 84: Output Values

    Operation Proline Promag H 100 HART Navigation "Diagnostics" menu → Measured values → Totalizer ‣ Totalizer Totalizer value 1 to n →  84 Totalizer overflow 1 to n →  84 Parameter overview with brief description Parameter Prerequisite Description...
  • Page 85: Adapting The Measuring Device To The Process Conditions

    Proline Promag H 100 HART Operation Parameter overview with brief description Parameter Prerequisite Description User interface Output current 1 – Displays the current value currently 3.59 to 22.5 mA calculated for the current output. Measured current 1 – Displays the current value currently 0 to 30 mA measured for the current output.
  • Page 86: Function Scope Of The "Control Totalizer" Parameter

    Operation Proline Promag H 100 HART Parameter overview with brief description Parameter Prerequisite Description Selection / User entry Control Totalizer 1 to n One of the following options is selected Control totalizer value. • Totalize in the Assign process variable • Reset + hold parameter (→...
  • Page 87: Diagnostics And Troubleshooting

    Proline Promag H 100 HART Diagnostics and troubleshooting Diagnostics and troubleshooting 12.1 General troubleshooting For local display Error Possible causes Solution Local display dark and no output Supply voltage does not match the Apply the correct supply voltage signals value indicated on the nameplate.
  • Page 88: Diagnostic Information Via Light Emitting Diodes

    Diagnostics and troubleshooting Proline Promag H 100 HART Error Possible causes Solution No connection via HART protocol Commubox Observe the documentation for the • Connected incorrectly Commubox. • Configured incorrectly  FXA195 HART: Document • Drivers not installed correctly "Technical Information"...
  • Page 89: Diagnostic Information In The Web Browser

    Proline Promag H 100 HART Diagnostics and troubleshooting 12.3 Diagnostic information in the Web browser 12.3.1 Diagnostic options Any faults detected by the measuring device are displayed in the Web browser on the home page once the user has logged on.
  • Page 90: Calling Up Remedy Information

    Diagnostics and troubleshooting Proline Promag H 100 HART Diagnostic information Diagnostic code Status signal Diagnostic number Short text ↓ ↓ ↓ Example Process limit A0013958 3-digit number 12.3.2 Calling up remedy information Remedy information is provided for every diagnostic event to ensure that problems can be rectified quickly.
  • Page 91: Calling Up Remedy Information

    Proline Promag H 100 HART Diagnostics and troubleshooting Status signals The status signals provide information on the state and reliability of the device by categorizing the cause of the diagnostic information (diagnostic event). Symbol Meaning Failure A device error has occurred. The measured value is no longer valid.
  • Page 92: Adapting The Diagnostic Information

    Diagnostics and troubleshooting Proline Promag H 100 HART 12.5 Adapting the diagnostic information 12.5.1 Adapting the diagnostic behavior Each item of diagnostic information is assigned a specific diagnostic behavior at the factory. The user can change this assignment for specific diagnostic information in the Diagnostic behavior submenu.
  • Page 93: Overview Of Diagnostic Information

    Proline Promag H 100 HART Diagnostics and troubleshooting 12.6 Overview of diagnostic information The amount of diagnostic information and the number of measured variables affected increase if the measuring device has one or more application packages. In the case of some items of diagnostic information, the status signal and the diagnostic behavior can be changed.
  • Page 94 Diagnostics and troubleshooting Proline Promag H 100 HART Diagnostic Short text Remedy instructions Status Diagnostic number signal behavior [from the [from the factory] factory] Electronic failure 1. Reset device Alarm 2. Contact service Electronic failure 1. Do not reset device Warning 2.
  • Page 95: Pending Diagnostic Events

    Proline Promag H 100 HART Diagnostics and troubleshooting Diagnostic Short text Remedy instructions Status Diagnostic number signal behavior [from the [from the factory] factory] Empty pipe detection Execute EPD adjustment Warning Configuration 1. Check IP addresses in network Warning 2. Change IP address...
  • Page 96: Diagnostic List

    Diagnostics and troubleshooting Proline Promag H 100 HART Previous diagnostics →  96 Operating time from restart →  96 →  96 Operating time Parameter overview with brief description Parameter Prerequisite Description User interface Actual diagnostics A diagnostic event has occurred.
  • Page 97: Filtering The Event Logbook

    Proline Promag H 100 HART Diagnostics and troubleshooting In addition to the operation time of its occurrence, each event is also assigned a symbol that indicates whether the event has occurred or is ended: • Diagnostic event • : Occurrence of the event •...
  • Page 98: 12.10 Resetting The Measuring Device

    Diagnostics and troubleshooting Proline Promag H 100 HART Info number Info name I1335 Firmware changed I1351 Empty pipe detection adjustment failure I1353 Empty pipe detection adjustment ok I1361 Web server: login failed I1397 Fieldbus: access status changed I1398 CDI: access status changed...
  • Page 99 Proline Promag H 100 HART Diagnostics and troubleshooting Device name →  99 Order code →  99 →  99 Extended order code 1 Extended order code 2 →  100 Extended order code 3 →  100 ENP version →...
  • Page 100 Diagnostics and troubleshooting Proline Promag H 100 HART Parameter Description User interface Factory setting Extended order code 2 Shows the 2nd part of the extended order Character string – code.  The extended order code can also be found on the nameplate of the sensor and transmitter in the "Ext.
  • Page 101: 12.12 Firmware History

    "Manufacturer' s information" document. The manufacturer' s information is available: • In the Download Area of the Endress+Hauser web site: www.endress.com → Downloads • Specify the following details: •...
  • Page 102: Maintenance

    Replacement seals (accessory part) →  122 13.2 Measuring and test equipment Endress+Hauser offers a wide variety of measuring and test equipment, such as W@M or device tests. Your Endress+Hauser Sales Center can provide detailed information on the services. List of some of the measuring and testing equipment: →  105 13.3...
  • Page 103: Repairs

    • The measuring devices have a modular design. • Spare parts are grouped into logical kits with the associated Installation Instructions. • Repairs are carried out by Endress+Hauser Service or by appropriately trained customers. • Certified devices can only be converted to other certified devices by Endress+Hauser Service or at the factory.
  • Page 104: Disposal

    Repairs Proline Promag H 100 HART 14.5 Disposal 14.5.1 Removing the measuring device 1. Switch off the device. WARNING Danger to persons from process conditions. ‣ Beware of hazardous process conditions such as pressure in the measuring device, high temperatures or aggressive fluids.
  • Page 105: Accessories

    Various accessories, which can be ordered with the device or subsequently from Endress +Hauser, are available for the device. Detailed information on the order code in question is available from your local Endress+Hauser sales center or on the product page of the Endress+Hauser website: www.endress.com.
  • Page 106: Service-Specific Accessories

    Accessories Description Applicator Software for selecting and sizing Endress+Hauser measuring devices: • Choice of measuring devices for industrial requirements • Calculation of all the necessary data for identifying the optimum flowmeter: e.g. nominal diameter, pressure loss, flow velocity and accuracy.
  • Page 107: System Components

    Proline Promag H 100 HART Accessories 15.4 System components Accessories Description Memograph M graphic The Memograph M graphic data manager provides information on all the relevant data manager measured variables. Measured values are recorded correctly, limit values are monitored and measuring points analyzed. The data are stored in the 256 MB internal memory and also on a SD card or USB stick.
  • Page 108: Technical Data

    Technical data Proline Promag H 100 HART Technical data 16.1 Application The measuring device is only suitable for flow measurement of liquids with a minimum conductivity of 5 µS/cm. Depending on the version ordered, the measuring device can also measure potentially explosive, flammable, poisonous and oxidizing media.
  • Page 109 Proline Promag H 100 HART Technical data Nominal Recommended Factory settings diameter flow Full scale value current min./max. full scale value Pulse value Low flow cut off output (v ~ 0.3/10 m/s) (~ 2 pulse/s) (v ~ 0.04 m/s) (v ~ 2.5 m/s)
  • Page 110: Output

    • Operating pressure to increase accuracy (Endress+Hauser recommends the use of a pressure measuring device for absolute pressure, e.g. Cerabar M or Cerabar S) • Medium temperature to increase accuracy (e.g. iTEMP) •...
  • Page 111 Proline Promag H 100 HART Technical data Assignable measured • Volume flow variables • Mass flow • Corrected volume flow Frequency output Output frequency Adjustable: 0 to 10 000 Hz Damping Adjustable: 0 to 999 s Pulse/pause ratio Assignable measured •...
  • Page 112: Local Display

    Technical data Proline Promag H 100 HART Pulse/frequency/switch output Pulse output Failure mode Choose from: • Actual value • No pulses Frequency output Failure mode Choose from: • Actual value • 0 Hz • Defined value: 0 to 12 500 Hz...
  • Page 113: Power Supply

    Proline Promag H 100 HART Technical data Protocol-specific data Protocol-specific data • For information on the device description files →  49 • For information on the dynamic variables and measured variables (HART device variables) →  49 16.5 Power supply Terminal assignment →...
  • Page 114: Performance Characteristics

    Technical data Proline Promag H 100 HART 16.6 Performance characteristics Reference operating • Error limits following DIN EN 29104, in future ISO 20456 conditions • Water, typically +15 to +45 °C (+59 to +113 °F); 0.5 to 7 bar (73 to 101 psi) •...
  • Page 115: Installation

    Proline Promag H 100 HART Technical data Volume flow Max. ±0.1 % o.r. ± 0.5 mm/s (0.02 in/s) Temperature ±0.5 °C (±0.9 °F) Electrical conductivity Max. ±5 % o.r. Temperature measurement < 15 s response time Influence of ambient Current output temperature o.r.
  • Page 116: Interior Cleaning

    Technical data Proline Promag H 100 HART Vibration resistance • Vibration, sinusoidal according to IEC 60068-2-6 • 2 to 8.4 Hz, 3.5 mm peak • 8.4 to 2 000 Hz, 1 g peak • Vibration broad-band random, according to IEC 60068-2-64 •...
  • Page 117: Pressure-Temperature Ratings

    Proline Promag H 100 HART Technical data Conductivity ≥ 5 μS/cm for liquids in general. Stronger filter damping is required for very low conductivity values. Pressure-temperature An overview of the pressure-temperature ratings for the process connections is ratings provided in the "Technical Information" document...
  • Page 118: 16.10 Mechanical Construction

    Technical data Proline Promag H 100 HART [psi] [mbar] DN250 DN125 DN150 DN200 (10") (5") (6") (8") DN100 (4") DN300 (12") 800 1000 1200 1400 1600 1800 [m /h] ³ 1000 2000 3000 4000 5000 6000 7000 8000 [gal/min] A0032668-EN ...
  • Page 119 Proline Promag H 100 HART Technical data "Compact, aluminum coated" housing version The local display is plugged onto the main electronics module. The electronic connection between the local display and main electronics module is established via a connecting cable. For some work performed on the measuring device (e.g. electrical connection), it is advisable to disconnect the local display from the main electronics module: 1.
  • Page 120: 16.12 Certificates And Approvals

    EU Directives. These are listed in the corresponding EU Declaration of Conformity along with the standards applied. Endress+Hauser confirms successful testing of the device by affixing to it the CE mark. C-Tick symbol The measuring system meets the EMC requirements of the "Australian Communications and Media Authority (ACMA)".
  • Page 121: 16.13 Application Packages

    The application packages can be ordered with the device or subsequently from Endress+Hauser. Detailed information on the order code in question is available from your local Endress+Hauser sales center or on the product page of the Endress+Hauser website: www.endress.com.
  • Page 122: 16.14 Accessories

    • The W@M Device Viewer : Enter the serial number from the nameplate (www.endress.com/deviceviewer) • The Endress+Hauser Operations App: Enter the serial number from the nameplate or scan the 2-D matrix code (QR code) on the nameplate. Standard documentation...
  • Page 123 Proline Promag H 100 HART Technical data Description of device parameters Measuring device Documentation code HART PROFIBUS DP Modbus RS485 EtherNet/IP PROFINET Promag 100 GP01038D GP01039D GP01040D GP01041D GP01042D Supplementary device- Safety Instructions dependent documentation Contents Documentation code ATEX/IECEx Ex nA...
  • Page 124: Index

    Index Proline Promag H 100 HART Index Device locking, status ......82 Device name Adapters ........22 Sensor .
  • Page 125 Proline Promag H 100 HART Index Storage temperature ..... . 115 Incoming acceptance ......13 Vibration resistance .
  • Page 126 Index Proline Promag H 100 HART Menu Totalizer handling (Submenu) ....85 Diagnostics ......95 Web server (Submenu) .
  • Page 127 Proline Promag H 100 HART Index Operating language ..... . . 53 Temperature measurement response time ..115 Output conditioning .
  • Page 128 *71689937* 71689937 www.addresses.endress.com...