★ READ CAREFULLY THIS INSTRUCTION MANUAL AND THE INSTRUCTIONS OF THE
ASPIRATING PUMP PRIOR TO USING THIS PRODUCT.
★ DO NOT DISCARD THIS INSTRUCTION MANUAL UNTIL ALL THE TUBES IN THIS BOX ARE
USED UP.
1. PERFORMANCE:
Measuring Range
and Pump Stroke
(*)Graduations on the detector tube are based on 2 pump strokes.
Sampling Time
Colour Change
Detectable Limit
Operating Temperature
Operating Humidity
Aspirating Pump
1. THE DETECTOR TUBE AND PRETREAT TUBE CONTAIN CHEMICAL REAGENTS.
2. DO NOT TOUCH THESE REAGENTS DIRECTLY ONCE TUBES WERE BROKEN.
3. KEEP THE TUBES OUT OF THE REACH OF CHILDREN.
NOTICE
1. USE ONLY WITH PUMP MODELS AP-20, AP-20S, 400B, AP-1, AP-1S OR 400A.
OTHERWISE, CONSIDERABLE ERROR IN INDICATION MAY OCCUR.
2. BEFORE TESTING, CHECK THE ASPIRATING PUMP FOR LEAKS. ( REFER TO ITEM 8.
INSPECTION OF ASPIRATING PUMP) ANY PUMPS SHOWING SIGNS OF LEAKAGE
SHOULD BE CORRECTED BEFORE USE.
3. DO NOT USE THIS TUBE OUTSIDE THE STATED OPERATING TEMPERATURE RANGE.
4. STORE TUBES IN A REFRIGERATED PLACE ( 0-10 ℃/32-50°F), AND USE BEFORE
EXPIRATION DATE PRINTED ON THE TOP OF THE BOX.
5. PRIOR TO USE, READ CAREFULLY ITEM 9. USER RESPONSIBILITY.
6. READ THE CONCENTRATION IMMEDIATELY AFTER MEASUREMENT.
2. SAMPLING AND MEASUREMENT:
① Break both ends of the detector tube (Ⅱ) and pretreat tube (Ⅰ), and connect each end of the detector tube and
pretreat tube with connecting tube as shown in Fig.1.
CAUTION
FROM SPLINTERING GLASS.
② Insert the detector tube into the aspirating pump securely as shown in Fig.1. (Arrow mark shall point to the
pump.)
③ Align the guide marks on the shaft and stopper of the aspirating pump.
④ Pull the pump handle at a full stroke until it locks and wait for 1.5 minutes or until the completion of sampling
is confirmed with the flow indicator of the pump. ( See descriptions about the flow indicator in the instruction
manual of the pump.)
⑤ Push back the handle without removing the detector tube from the rubber tube connector so that air in the pump
will be discharged perfectly. Then repeat the step ③ ~④ once more.
⑥ On completion of sampling, read the scale at the maximum point of the stained layer.
⑦ When the concentration is below the scale range, 3 or 4 pump strokes can be used to determine concentrations
of 35 to 250 ppm or 23 to 167 ppm. Then multiply the reading value by 1/2 or 1/3 respectively after
temperature correction.
CHLOROFORM DETECTOR TUBE
: 70 - 500 ppm(*)
: 2 pump strokes
: 3 minutes
: White → Yellowish Orange
: 20 ppm (4 pump strokes)
: 10 - 40 ℃ (50-104°F) (Temperature correction is necessary.)
: No correction is necessary up to 80%R.H. at 30 ℃ .
: Model AP-20, AP-20S, 400B, AP-1, AP-1S or 400A
SAFETY GLASSES AND GLOVES SHOULD BE WORN TO PREVENT INJURY
INSTRUCTION MANUAL
35 - 250 ppm
3 pump strokes
4.5 minutes
CAUTION
Fig.1
23 - 167 ppm
4 pump strokes
6 minutes
No.152S