Kitagawa 121SE Instruction Manual

Kitagawa 121SE Instruction Manual

Phosphine detector tube

Advertisement

Quick Links

★ READ THIS INSTRUCTION MANUAL AND THE INSTRUCTIONS OF THE ASPIRATING PUMP PRIOR
TO USING THIS PRODUCT.
★ DO NOT DISCARD CAREFULLY THIS INSTRUCTION MANUAL UNTIL ALL THE TUBES IN THIS BOX
ARE USED UP.

1. PERFORMANCE:

Measuring Range
and Pump Stroke
(*)Graduations on the detector tube are based on 1 pump stroke.
Sampling Time
Colour Change
Detectable Limit
Operating Temperature
Aspirating Pump
1. THE DETECTOR TUBE CONTAINS CHEMICAL REAGENTS.
2. DO NOT TOUCH THESE REAGENTS DIRECTLY ONCE TUBES WERE BROKEN.
3. KEEP THE TUBES OUT OF THE REACH OF CHILDREN.
NOTICE
1. USE ONLY WITH PUMP MODELS AP-20, AP-20S, 400B. OTHERWISE, CONSIDERABLE ERROR
IN INDICATION MAY OCCUR.
2. BEFORE TESTING, CHECK THE ASPIRATING PUMP FOR LEAKS. (REFER TO ITEM 8.
Inspection Of Aspirating Pump.) ANY PUMPS SHOWING SIGNS OF LEAKAGE SHOULD
BE CORRECTED BEFORE USE.
3. DO NOT USE THIS TUBE OUTSIDE THE STATED OPERATING TEMPERATURE RANGE.
4. STORE TUBES IN A COOL AND DARK PLACE (0-25 ℃/32-77°F), AND USE BEFORE
EXPIRATION DATE PRINTED ON TOP OF THE BOX.
5. PRIOR TO USE, READ ITEM 9 USER RESPONSIBILITY CAREFULLY.
6. READ THE CONCENTRATION IMMEDIATELY AFTER DRAWING THE SAMPLE.

2. SAMPLING AND MEASUREMENT:

① Break both ends of the detector tube.
CAUTION SAFETY GLASSES AND GLOVES SHOULD BE WORN TO PREVENT INJURY
FROM SPLINTERING GLASS.
② Insert the detector tube into the aspirating pump securely as shown in Fig.1. (Arrow mark shall point to
the pump.)
③ Align the guide marks on the shaft and stopper of the aspirating pump.
④ Pull the pump handle at a full stroke until it locks and wait for 1 minute or until the completion of sampling is
confirmed with the flow indicator of the pump. (See descriptions about the flow indicator in the instruction
manual of the pump.)
⑤ On completion of sampling, read the scale at the maximum point of the stained layer.
⑥ When the concentration is over the scale range, a 1/2 pump stroke can be used to determine concentrations of
20 to 1400 ppm.
1)Remove the detector tube from the pump.
INSTRUCTION MANUAL
PHOSPHINE DETECTOR TUBE
: 10 - 700 ppm
20 - 1400 ppm
: 1 pump stroke
: 1 minute
: White → Brown
: 1 ppm(1 pump stroke)
: 0 - 40 ℃ (32-104°F) (No correction is necessary.)
: Model AP-20, AP-20S, 400B
1/2 pump stroke
30 seconds
CAUTION
Fig.1
No.121SE

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the 121SE and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Kitagawa 121SE

  • Page 1 INSTRUCTION MANUAL PHOSPHINE DETECTOR TUBE No.121SE ★ READ THIS INSTRUCTION MANUAL AND THE INSTRUCTIONS OF THE ASPIRATING PUMP PRIOR TO USING THIS PRODUCT. ★ DO NOT DISCARD CAREFULLY THIS INSTRUCTION MANUAL UNTIL ALL THE TUBES IN THIS BOX ARE USED UP.
  • Page 2 2)Turn the pump handle right or left by 1/4 (90 ), push it toward to the pump. 3)Insert the new tube into the aspirating pump. 4)Pull the pump handle at a 1/2 stroke until it locks and wait for 30 seconds or until the completion of sampling is confirmed with the flow indicator of the pump.

Table of Contents