– they are not a substitute for a medical Best regards, examination! Discuss your results with the doctor. Never Your Beurer Team use them to make medical decisions independently (e.g. regarding medication and dosage thereof). 1. Getting to know your instrument •...
Page 10
• You may only operate the blood pressure monitor with • Batteries should not be charged or reactivated with any batteries. other means, nor should they be taken apart, thrown in • The automatic switch-off function switches off the blood the fire or short-circuited.
3. Unit description / Icons in the display 4. Prepare measurement Inserting battery • Remove the battery compartment lid on the lower left side of the device. • Insert two 1.5 V micro batteries (alka- mmHg line, type LR 03). Make sure that the batteries are insert- ed with the correct polarity, according mmHg...
Page 12
Positioning cuff there may be serious divergences. Relax your arm and the palms of your hands. • Bare your left wrist, making sure that the circulation in the • In order not to distort the result, it is important to keep still arm is not restricted by any clothes etc.
Page 13
Evaluating results Repeat the measurement if the flashing icon is dis- played after the measurement. Please note that you should Cardiac arrhythmia: rest for 5 minutes between measurements and not talk or This instrument can identify possible cardiac arrhythmia dis- move during the measurement.
• Never place any heavy objects on the instrument. normal in the example described). 8. Technical details 6. Error messages/troubleshooting Model no. BC 44 In case of faults, the “Er_” message appears in the display. Measurement Oscillating, non-invasive blood Error messages can occur when...
Page 15
Battery life For approx. 250 measurements, de- • If you are using the instrument for commercial or eco- pending on blood pressure level or nomic purposes, you must check measurement precision inflation pressure regularly in accordance with the “Operators’ Ordinance Accessories Storage box, instructions for use, on Medical Products”.
Page 16
O n e - b u t t o n B l o o d P r e s s u r e Mo n i t o r...
Page 17
• The readings that you take may only be used for informa- Best regards, tion purposes – they are not a substitute for a medical Your Beurer Team examination! Discuss your results with the doctor. Never use them to make medical decisions independently (e.g.
Page 18
• Remove the batteries from the instrument if they are worn Storage and Care out or if you are not going to use the instrument for any • The blood pressure monitor is made up of precision elec- length of time. This prevents any damage as a result of tronic components.
Page 19
3. Unit description/Icons in the display mmHg mmHg /min 1. Mains adapter port 2. Systolic pressure 3. Diastolic pressure 4. Pulse 5. Pulse symbol 1. Cuff tube 6. Battery compartment 2. Cuff 7. Start/stop button 3. Cuff connector 8. Battery replacement symbol 9.
Page 20
4. Prepare measurement mains adapter is available from retailers or from the service address under order no. 071.19. The blood pressure moni- Inserting battery tor should only be used with the mains adapters specified • Remove the battery cover here. The mains adapter should only be connected to the from the back of the monitor.
Page 21
Insert the cuff tubing into the Measuring blood pressure socket for the cuff attachment. • Switch on the blood pressure monitor with the start/stop button • Before the measurement, the last saved test result is briefly displayed. If there is no measurement in the memo- ry, the instrument always displays the value “0”.
Page 22
Evaluating results move during the measurement. If the icon appears often, please contact your doctor. Any self-diagnosis and Cardiac arrhythmia: treatment based on the test results may be dangerous. It is This instrument can identify possible cardiac arrhythmia dis- vital to follow your doctor’s instructions. orders during measurement and if necessary indicates the measurement with the flashing icon WHO classification:...
Page 23
The bar graph in the display and the scale on the unit indicate 7. Cleaning and care the range of the blood pressure which has been recorded. • Clean your blood pressure computer carefully using only a If the values for systolic and diastolic pressure are in two slightly moistened cloth.
Page 24
European standards EN1060-1 (non-invasive blood Output 6 V DC, 600mA, only in connection pressure monitors, Part 1: general require Electromechan- with beurer blood pressure monitor ical Blood Pressure Measurement Systems). Supplier Friwo Gerätebau GmbH • If you are using the instrument for commercial or eco-...