BC 44
EN Blood pressure monitor
Instructions for use ..................................................................2

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the BC 44 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Beurer BC 44

  • Page 1 BC 44 EN Blood pressure monitor Instructions for use ..............2...
  • Page 2: Table Of Contents

    ENGLISH Content 1. Getting to know your instrument ........2 7. Evaluating results ............10 2. Important information ............2 8. Error messages/troubleshooting ........11 3. Unit description ...............7 9. Cleaning and storing the device and cuff ......11 4. Icons in the display ............7 10.
  • Page 3 Attention Manufacturer Permissible storage and transport tempera- Storage/Transport Note ture and humidity Note on important information Observe the instructions for use Permissible operating temperature and Operating humidity Application part, type BF Device protected against foreign objects ≥ 12.5 mm and against water dripping at IP22 Direct current an angle...
  • Page 4 • Do not take a measurement within 30 minutes after eat- to use the device. Supervise children around the device to ing, drinking, smoking or exercising. ensure they do not play with it. • Before the initial blood pressure measurement, make sure •...
  • Page 5 • Do not use the cuff on people who have undergone a Notes on handling batteries mastectomy. • Do not place the cuff over wounds as this may cause fur- • If your skin or eyes come into contact with battery fluid, ther injury.
  • Page 6 • Never open the instrument. If these instructions are not • Avoid using this device directly next to other devices or heeded, the warranty will be null and void. stacked on top of other devices, as this could lead to •...
  • Page 7: Unit Description

    3. Unit description 4. Icons in the display mmHg mmHg /min 1. Systolic pressure 2. Diastolic pressure 3. Pulse 4. Pulse symbol 1. Display 5. Battery replacement symbol 2. Risk indicator 6. Pump up, release air (arrow) 3. Wrist cuff 7.
  • Page 8: Prepare Measurement

    5. Prepare measurement 6. Measuring blood pressure Inserting battery Ensure the device is at room temperature before measur- ing. The measurement can be performed on the left or right • Remove the battery compartment lid on wrist. the lower left side of the device. •...
  • Page 9 If the values between the two wrists are significantly differ- Measuring blood pressure ent, please consult your doctor to determine which wrist • Start the blood pressure monitor with the start/stop but- should be used for the measurement. . A beep and a brief display of all the symbols con- firm that the device is switched on.
  • Page 10: Evaluating Results

    7. Evaluating results However, these standard values serve only as a general guideline, as the individual blood pressure varies in different Cardiac arrhythmia: people and different age groups etc. This instrument can identify possible cardiac arrhythmia dis- It is important to consult your doctor regularly for advice. orders during measurement and if necessary indicates the Your doctor will tell you your individual values for normal measurement with the flashing icon...
  • Page 11: Error Messages/Troubleshooting

    In such cases, repeat the measurement. Ensure that you do Blood pressure Systole Diastole Action not move or speak during the measurement. If necessary, value category (in mmHg) (in mmHg) reinsert or replace the batteries. regular monito- High normal 130 – 139 85 – 89 ring by doctor Technical alarm –...
  • Page 12: Technical Details

    2 x 1,5 V AAA batteries 10. Technical details Battery life For approx. 300 measurements, depend- Model no. BC 44 ing on the blood pressure level and/or pump pressure Measurement Oscillometric, non-invasive blood pres- method sure measurement on the wrist...
  • Page 13: Warranty / Service

    electro-mechanical blood pressure measuring systems) and IEC 80601-2-30 (Medical electrical equipment – Part 2 – 30: Particular requirements for the safety and essential performance of automated non-invasive blood pressure monitors). • The accuracy of this blood pressure monitor has been carefully checked and developed with regard to a long useful life.