Kitagawa 134SA Instruction Manual

Trichloroethylene detector tube

Advertisement

Quick Links

★ READ CAREFULLY THIS INSTRUCTION MANUAL AND THE INSTRUCTIONS OF THE ASPIRATING PUMP
PRIOR TO USING THIS PRODUCT.
★ DO NOT DISCARD THIS INSTRUCTION MANUAL UNTIL ALL THE TUBES IN THIS BOX ARE USED UP.
1. PERFORMANCE:
Measuring Range
and Pump Stroke
(*)Graduations on the detector tube are based on 1 pump stroke.
Sampling Time
Colour Change
Detectable Limit
Operating Temperature
Aspirating Pump
1. THE DETECTOR TUBE CONTAINS CHEMICAL REAGENTS.
2. DO NOT TOUCH THESE REAGENTS DIRECTLY ONCE TUBES WERE BROKEN.
3. KEEP THE TUBES OUT OF THE REACH OF CHILDREN.
NOTICE
1. USE ONLY WITH PUMP MODELS AP-20, AP-20S, 400B, AP-1, AP-1S OR 400A. OTHERWISE,
CONSIDERABLE ERROR IN INDICATION MAY OCCUR.
2. BEFORE TESTING, CHECK THE ASPIRATING PUMP FOR LEAKS. (REFER TO ITEM 8.
INSPECTION OF ASPIRATING PUMP.) ANY PUMPS SHOWING SIGNS OF LEAKAGE SHOULD
BE CORRECTED BEFORE USE.
3. DO NOT USE THIS TUBE OUTSIDE THE STATED OPERATING TEMPERATURE RANGE.
4. STORE TUBES IN A REFRIGERATED PLACE (0-10 ℃/32-50°F), AND USE BEFORE EXPIRATION
DATE PRINTED ON THE TOP OF THE BOX.
5. PRIOR TO USE, READ CAREFULLY ITEM 9. USER RESPONSIBILITY.
6. READ THE CONCENTRATION IMMEDIATELY AFTER MEASUREMENT.
2. SAMPLING AND MEASUREMENT:
① Break both ends of the detector tube.
CAUTION
SAFETY GLASSES AND GLOVES SHOULD BE WORN TO PREVENT INJURY
FROM SPLINTERING GLASS.
② Insert the detector tube into the aspirating pump securely as shown in Fig.1. (Arrow mark shall point to the
pump.)
③ Align the guide marks on the shaft and stopper of the aspirating pump.
④ Pull the pump handle at a full stroke until it locks and wait for 2 minutes or until the completion of
sampling is confirmed with the flow indicator of the pump. (See descriptions about the flow indicator in the
instruction manual of the pump.)
⑤ On completion of sampling, read the scale at the maximum point of the stained layer.
⑥ When the concentration is over the scale range, a 1/2 pump stroke can be used to determine concentrations
of 10 to 300 ppm.
1)Remove the detector tube from the pump.
2)Turn the pump handle right or left by 1/4 (90
3)Insert the new tube into the aspirating pump.
4)Pull the pump handle at a 1/2 strokes until it locks and wait for 1 minute or until the completion of sampling is
INSTRUCTION MANUAL
TRICHLOROETHYLENE DETECTOR TUBE
: 5 - 150 ppm(*)
: 1 pump stroke
: 2 minutes
Yellow → Red
: 1ppm (1 pump stroke)
: 0 - 40 ℃ (32-104°F) (Temperature correction is necessary.)
: Model AP-20, AP-20S, 400B, AP-1, AP-1S or 400A
o
), push it toward to the pump.
10 - 300 ppm
1/2 pump stroke
: 1 minute
CAUTION
Fig.1
No.134SA

Advertisement

loading
Need help?

Need help?

Do you have a question about the 134SA and is the answer not in the manual?

Questions and answers

Summary of Contents for Kitagawa 134SA

  • Page 1 INSTRUCTION MANUAL TRICHLOROETHYLENE DETECTOR TUBE No.134SA ★ READ CAREFULLY THIS INSTRUCTION MANUAL AND THE INSTRUCTIONS OF THE ASPIRATING PUMP PRIOR TO USING THIS PRODUCT. ★ DO NOT DISCARD THIS INSTRUCTION MANUAL UNTIL ALL THE TUBES IN THIS BOX ARE USED UP.
  • Page 2 confirmed with the flow indicator of the pump. 5)On completion of sampling, read the scale at the maximum point of the stained layer. fter temperature correction undermentioned. ( 6)Then multiply the reading value by 2 a REFER TO ITEM 3. CORRECTION FOR AMBIENT CONDITIONS.) SPECIAL NOTE: Ⅰ.