USER MANUAL
CF-260 Centrifuge
Version: B
Date: July 2024
Document Number: 03-2020-260
For research only. Not for use in diagnostic procedures

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the CF-260 and is the answer not in the manual?

Questions and answers

Summary of Contents for revvity CF-260

  • Page 1 USER MANUAL CF-260 Centrifuge Version: B Date: July 2024 Document Number: 03-2020-260 For research only. Not for use in diagnostic procedures...
  • Page 2 Copyright This manual is published by Revvity 935 Cobb Place Blvd, Kennesaw, GA 30144 USA. Copyright Revvity, Inc. All rights reserved, including but not limited to those rights to reproduce this publication or parts thereof. Reproduction of this publication or parts thereof, or the products it describes, by any means or in any form is expressly prohibited without the written permission of Revvity.
  • Page 3 This product is warranted to be free from defects in material and workmanship for a period of ONE YEAR from the date of delivery. Revvity will repair or replace and return free of charge any part which is returned to its factory within said period, transportation prepaid by user, and which is found upon inspection to have been defective in materials or workmanship.
  • Page 4: Table Of Contents

    Table of Contents Table of Symbols ................................... 5 Safety Information ................................6 Intended Use ..................................7 Specifications ..................................7 Installation ....................................8 Unpacking ....................................8 Installation Location ................................8 Connecting the Power Cord ............................... 8 Instructions for Use ................................9 Overall View ...................................
  • Page 5: Table Of Symbols

    Table of Symbols Symbol Description Symbol Description Caution. Refer to the On (power). User documentation (IEC 60417-5007) (ISO 7000-0434B) Hazardous voltage; risk Off (power). of electric shock. (IEC 60417-5008) (IEC 60417-6042) Consult Instructions for Fuse. (IEC 60417-5016) Use. (ISO 7000-1641) WEEE symbol CE Compliance Mark (EN50419:2005)
  • Page 6: Safety Information

    Safety Information Before using the machine, make sure to read and understand this manual thoroughly. Keep the manual close to the machine, easily accessible to all the users. Improper operation can cause injury to persons or damage to the equipment. In the interest of your own personal safety, always observe the following safety instruc- tions: Do not use rotors and buckets, which show clear signs of corrosion or mechanical defects.
  • Page 7: Intended Use

    Intended Use The CF-260 Centrifuge is a universal laboratory centrifuge. It is designed for use in medical, scientific and industrial laboratories, for separating the substances with different specific densities by centrifugal force. In particular, it is intended for preparation of human samples (body fluids), in approved test tubes, before fur- ther analysis.
  • Page 8: Installation

    Retain the packaging material for any subsequent transport or storage, which are allowed only in the original packaging. The CF-260 Centrifuge consists of the following: Description Quantity...
  • Page 9: Instructions For Use

    Instructions for Use Overall View Window Knob Control Panel Main Switch Rotor Key Magnets Name Label Auxiliary Lid Release Mains Socket Fuse Compartment Ventilator Foot...
  • Page 10: Control Panel Description

    Control Panel Description Before using the centrifuge for the first time, familiarize yourself with the control panel and the functions of the knob and keys. Indicator for Selection of Rotational Speed (RPM) Display for Rotational Speed (RPM) or Centrifugal Force (RCF) Indicator for Selection of Centrifugal Force (RCF) SPEED Key Display for Time (TIME), Acceleration Level ( ) and Braking Level ( )
  • Page 11: Knob And Button Functions

    Knob and Button Functions By rotating the knob, you change the values of the parameters on display. By rotating the knob clockwise, the value increases, and by rotating it counter-clockwise, the value decreases. By pressing on the knob, you confirm the new value for the parameter. By pressing this key, you stop the run of the centrifuge.
  • Page 12: Turning On The Instrument

    Turning ON the Instrument Turn on the main switch, which is located on the front of the right side of the centrifuge. All segments (eights) are displayed on all displays at first, and all the keys and indicators light up simultaneously. Then centrifuge model and program version (X.XX) are displayed, then lines, and finally the values of operation parameters are displayed.
  • Page 13: Rotor Loading

    Rotor Loading You must always load the rotor with test tubes symmetrically. You may only use approved test tubes. Weight difference of the samples in test tubes should be as low as possible in order to avoid potential damages to the motor and to minimize running noise and vibration.
  • Page 14: Setting The Rotational Speed And Centrifugal Force

    Setting the Rotational Speed and Centrifugal Force By pressing the SPEED key, you can set the rotational speed (RPM) and centrifugal force (RCF). The key lights up. By repeated pressings on this key, you select which parameter will be set. When indicator RPM is lit, you can set the rotational speed, and when indicator RCF is lit, you can set the centrifugal force.
  • Page 15: Program Setting

    NOTE: All the operation parameters can be changed during the operation of the centrifuge, and it starts to work with the new settings after that. • Changing the run time during operation You should be aware, that when you extend run time, the difference between newly set and originally set time is added to current time, and when you shorten run time, the difference is subtracted.
  • Page 16: Starting And Stopping The Centrifuge

    The following table shows factory preset values of operation parameters for all 100 programs: Accelera- Rotational Braking Run Time tion Program Speed Level (Min) Level (RPM) (brX) (AcX) 13,000 10,000 1,500 3,000 4,000 15,000 4,00 15,000 5,000 16,000 1,200 3,000 3,000 3,000 Starting and Stopping the Centrifuge...
  • Page 17: Rotor Recognition

    Rotor Recognition Rotor recognition is executed automatically, every time when the centrifuge starts. After rotor change, the centrifuge stops and display shows ROTOR CHG. Press the STOP key to clear this message. New maximum ro- tational speed is automatically set, to correspond with the inserted rotor. To start the centrifuge again, press the START key.
  • Page 18: Auxiliary Lid Release

    Auxiliary Lid Release In case of power failure, the lid can be opened manually. If power failure occurs during the operation of the centrifuge, the rotor may continue rotating for several minutes, before it stops. WARNING: Before auxiliary lid release, turn off the main switch of the centrifuge and wait until rotor fully stops.
  • Page 19: Maintenance And Cleaning Of The Centrifuge

    Maintenance and Cleaning of the Centrifuge Regular Service We recommend having the centrifuge and associated rotors checked by authorized service at least once a year. You must thoroughly clean and disinfect the centrifuge prior to service. Cleaning the Centrifuge After every centrifugation, please remove any condensed water from the rotor chamber using a soft, ab- sorbent cloth.
  • Page 20: Troubleshooting

    Troubleshooting If an error occurs during the operation of the centrifuge, an error message appears on the display, and the centrifuge stops automatically. The display shows message ERROR and number of the error EXX (XX is the error number). Display Responsi- Problem Solution...
  • Page 21 Display Problem Solution Responsi- Speed Time Reduce braking level User Voltage overload on the DC Error on frequency regulator Service link Error on braking resistor Service Voltage too low on the DC Check power supply Service link Repeat run User Reduce acceleration level User DRIVE...
  • Page 22: Users Parameters

    Users Parameters Accessing Parameters With user’s parameters, you can check and set some of the operational properties of the centrifuge. Entry -For entering to user’s parameters, the centrifuge must be in stand-by mode. Simultaneously press the SPEED and PROGRAM keys and hold them for about 5 seconds. When the first parameter is displayed and the TIME key lights up, release the keys.
  • Page 23: Replacing The Centrifuge Fuses

    Replacing the Centrifuge Fuses The following fuses are required for CF-260 Centrifuge: 2 x 6.3AT 250V (230V) 2 x 10AT 250V (120V) • Unplug mains plug from the mains socket. • By pressing the locking device on the left side of the fuse compartment, the fuse holder is released and you can pull it out.
  • Page 24: Appendix

    Appendix Calculation of Centrifugal Force For the calculation of the centrifugal force (RCF), stated as a multiple of the gravitational force “g”, use the following formula: RCF ..Relative centrifugal force ( x g) RCF = 11.18 x r x ( n / 1000 )2 r ....
  • Page 26 Users should exercise caution when handling materials as they may present unknown hazards. Revvity shall not be liable for any damages or losses resulting from handling or contact with the product, as Revvity cannot control actual methods, volumes, or conditions of use. Users are responsible for ensuring the product’s suitability for their specific application.