Measuring amplifier GSV-5A6 Contents 6-channel measuring amplifier GSV-5A6...................3 Connection diagram M8 Round plug for supply voltage............4 Connection diagram for output socket: 15-pin Sub-D..............5 Connection diagram for input socket: 44-pin Sub-D HD............6 LEDs and buttons..........................7 Function of the push buttons.....................7 Further functions of all buttons:....................8...
All voltage outputs are calibrated and can be used synchronously. The GSV-5A6 is configured with 4 push buttons and the current state of the 6 channels can be seen on 14 LEDs.
The colors refer to the wire colors of the included 3 m cable with the 15-pin plug (SubD15). ME-Meßsysteme GmbH Eduard-Maurer Str. 9 Tel.: +49 3302 89824 60 Mail: info@me-systeme.de 16761 Hennigsdorf Fax: +49 3302 89824 69...
Currently unused pins: 7; 14; 15; 22; 29; 30; 37; 44 When connecting sensors using four-wire technology, the pins +UF must be bridged to +US and -UF must be bridged to -US. ME-Meßsysteme GmbH Eduard-Maurer Str. 9 Tel.: +49 3302 89824 60 Mail: info@me-systeme.de...
(Only possible if a channel has been previously selected) If one or all channels have been selected, a new input sensitivity can be set with “IN”. This needs to be ME-Meßsysteme GmbH Eduard-Maurer Str. 9 Tel.: +49 3302 89824 60 Mail: info@me-systeme.de...
“CH”. Shunt active mode: If the shunt resistor has been activated (visually recognizable by the flashing of LED 12), it can be disabled by pressing any button. The GSV-5A6 can then be used as usual. ME-Meßsysteme GmbH Eduard-Maurer Str. 9 Tel.: +49 3302 89824 60...
Technical data Inputs GSV-5A6 Unit Input sensitivity (factory setting) ±2,0 mV/V Can be switched to ±4,0; ±1,0; ±0,5 mV/V Analog outputs Number Analog voltage output ±10 Output Impedance Cutoff frequency 250 / 2500 (Option) Supply Supply voltage 10 … 28...
Case dimensions GSV-5A6 Change log Version Date Changes 2024 Initial release 05/2024 SW: Minor changes, Zero routine time corrected Subject to change. All information describes our products in general terms. They do not represent a guarantee of properties within the meaning of §459 Para. 2, BGB (German Civil Code) and do not constitute liability.
Need help?
Do you have a question about the GSV-5A6 and is the answer not in the manual?
Questions and answers