Instruction Manual MicroCapStar End-tidal CO Analyzer for mice, rats, & larger animals Read instructions carefully before operating this device. This device is not to be used for Human Life Support applications. To avoid possible electrical shock, do not operate this device if is wet or has had liquids spilled onto it.
Page 2
TEL (610)642-7719 LIFE SUPPORT POLICY Instruments manufactured by CWE, Incorporated are not authorized for use as critical components in human life support devices or systems. "Life support devices or systems", as used herein, are devices or systems whose failure to perform, whether through misuse, failure, or proper operation, can reasonably be expected to result in significant injury to the operator or subject persons.
MicroCapStar CO Analyzer Instruction Manual 1.0 INTRODUCTION The MicroCapStar is a micro-sampling respiratory gas analyzer for monitoring the expired carbon dioxide gas concentration and respiratory rate of small animals, including mice. CO is measured using a temperature-controlled, infrared optical bench, which features fast response, excellent accuracy, and long-term stability.
4.0 OPERATION 4.1 CALIBRATION The CO sensor used in the MicroCapStar is highly linear, so a two-point calibration is all that is required. Zero adjust: With the sample tubing set open to room air, press the rocker switch to the position and hold it there.
CO 8.0% (0 – 60 mmHg). 4.3 END-TIDAL OR FAST CO DISPLAY SWITCH The LCD display on the MicroCapStar can report CO concentration in either a continuous reading ( ), or as the peak expired value ( FAST ), depending on the setting of this switch.
The LCD panel will provide a description of the condition. 4.7 SETTING AND USING THE ETCO ALARM This MicroCapStar provides a user-selectable alarm for ETCO . When this alarm is turned on, and an out of range ETCO measurement is encountered, an alarm will sound and the LED will flash.
MicroCapStar CO Analyzer Instruction Manual 5.0 MAINTENANCE The MicroCapStar monitor requires no routine maintenance, other than periodically replacing the in-line hydrophobic filter. This filter is inserted in the port SAMPLE INLET on the front panel of the instrument. These filters are included in the Accessories Kit included with the instrument, and are available separately from CWE, Inc.
MicroCapStar Quick Start Guide Setup: Connect the MicroCapStar sample tubing set using the attached diagram as a guide. Keep the tubing volumes as small as possible. Try to make the connection between the distal “T” connector and the other tracheal tube connections easily detachable for calibration purposes.