Instruction Manual GEMINI End-tidal O & CO Analyzer Read instructions carefully before operating this device. This device is not to be used for Human Life Support applications. To avoid possible electrical shock, do not operate this device if is wet or has had liquids spilled onto it.
Page 2
LIFE SUPPORT POLICY Instruments manufactured by CWE, Incorporated are not authorized for use as critical components in human life support devices or systems. "Life support devices or systems", as used herein, are devices or systems whose failure to perform, whether through misuse, failure, or proper operation, can reasonably be expected to result in significant injury to the operator or subject persons.
A USB port is provided for connection to an external computer, whereby serial data can be independently collected and analyzed. A special feature of the GEMINI is Metabolic Mode, where the response time is slower, but with a 10X increase in gas measurement resolution. This is especially useful for making oxygen consumption measurements with a metabolic chamber where breath-by-breath response is not necessary.
Analyzer Instruction Manual 2.0 SETTING UP THE GEMINI Place the GEMINI close to the subject or animal being monitored. Note that the output signals are redundantly located on the front and rear panels. When monitoring small animals, it is desirable to locate the unit as close to the animal as possible to minimize sample tubing volume.
LED’s, and adjustment knobs at the right. These controls and functions are described in the following sections. 2: GEMINI R IGURE ESPIRATORY NALYZER CONTROLS The normal operating display screen is shown below (Figure 3).
. This gas mix is commercially available in SPAN ZERO small cannisters. NOTE: When calibrating with any compressed gas, allow the GEMINI to draw in the gas from a stream. NEVER apply compressed gas directly to the SAMPLE INLET 3.21 GAS ZERO CALIBRATIONS...
ISPLAY SCREEN DURING A CALIBRATION 3.3 ALARM SETTINGS The GEMINI has both ETCO2 and respiratory rate ( ) user- programmable alarms. If an alarm condition is detected, the source of the alarm will be shown in the STATUS field on the display: LOW FLOW -–...
UNCTIONS DISPLAY SCREEN SHOWING A SELECTED FUNCTION AND VALUE 3.61 METABOLIC MODE The GEMINI is optimized for breath-by-breath CO and O measurements. However, in some cases, speed of response is not as important as extended precision and smoothed response. For...
ODE DISPLAY SHOWING ADDITIONAL DIGITS OF RESOLUTION 3.62 SERIAL OUTPUT FORMAT Besides the standard analog signal outputs, the GEMINI has a serial data output via its built-in USB port (rear panel). The format of this data is set using the AUXILIARY FUNCTIONS menu. Except...
NEVER OPERATE THE INSTRUMENT WITHOUT THE IN- LINE HYDROPHOBIC FILTER. THIS FILTER PREVENTS LIQUIDS AND OTHER CONTAMINENTS FROM ENTERING THE SAMPLE CELLS. 7.0 ORDERING INFORMATION Accessories and replacement parts for the GEMINI are available from CWE, Inc. The following table describes these items. PART No. DESCRIPTION...