8.5 Automatic defrost, disinfection and power saving settings

Software settings:
Press MENU ENTER, to open the settings menu.
Select program settings and press MENU ENTER, to
open the menu for setting automatic defrost and power
saving times.
Enter PIN code 5792. PIN code is usually restricted to
service technicians or device administrators.
Select defrosting program 1 to 3 and set a time for the
automatic defrosting program using the + and – but-
tons (e.g., each day at 12:00 h and at 23:00 h).
Set a starting time for the automatic power saving
mode using the + and – buttons (e.g., start at 21:00 h).
Select temperature increase end and set an end time for
the automatic power saving mode using the + and –
buttons (e.g., end at 04:00 h).
On non-working days, the energy-saving mode can be
executed for the whole day (for 24 hours). Use the +
and – buttons to select the non-working days on which
the energy-saving mode is to be executed (e.g. Satur-
day, Sunday).
1-Monday 2-Tuesday 3-Wednesday 4-Thursday 5-Friday 6-Saturday 7-Sunday
24
>
Chamber temp
Quickfreeze
UV disinfection
Light
TIME
TUE 13:05:05
Language
Display version
>
Program settings
Enter PIN
0000
>
>
Defr. Start 1
Defr. Start 2
Defr. Start 3
Disinf. Start
Disinf. day
>
Incr. start
Incr. end
Incr. days
Disinf. day
Incr. start
>
Incr. end
Incr. days
Disinf. day
Incr. start
Incr. end
>
Incr. days
Manual_MEV_MEV+_2024-06-001.00_EN.docx
-20 °C
Off
Auto
English
5 µm
12:00
23:00
__:__
__:__
_ _ _ _ _ _ _
21:00
04:00
_ _ _ _ _ _ _
21:00
04:00
_ _ _ _ _ 6 7
_ _ _ _ _ _ _
21:00
04:00
_ _ _ _ _ 6 7