Medical IT System Intelligent Insulation Monitoring Devices (4-Piece Set) Installation and Operation Manual V2.8 Acrel Co., Ltd.
Page 2
Declaration Please read this instruction carefully before using this product. All pictures, logos and symbols involved are owned by Acrel Co., Ltd. All or part of the content shall not be reproduced publicly without written authorization by non-company personnel. Please read the instructions and precautions in this operation manual carefully before using this series of products.
Page 3
Modified Records Time Versions Reasons for revision On the basis of the original insulation monitoring 2016.01.20 V2.0 products, the contents of all five pieces of products are integrated to replace the instructions of each sub-product. 2016.10.25 V2.1 Some errors have been fixed The overview added "products conform …...
Contents 1 Introduction .......................... 1 2 Function Features ......................... 2 2.1 Function Features of AITR series Medical Isolation Transformer ........ 2 2.2 AIM-M10 Medical Intelligent Insulation Monitor ............2 2.3 Function Features of AID10/150 ................... 2 2.4 Function Features of AKH-0.66P26 Current Transformer ..........3 3 Reference Standard ......................
Products comply with the enterprise standard Q31/0114000129C013-2016 IT System Insulation Monitor provisions. Insulation monitoring products of medical IT system (4-piece set) include AITR series medical isolation transformer, AIM-M10 medical intelligent insulation monitor, AKH-0.66P26 current transformer and AID series (AID10, AID150) external alarm and display instrument. as shown in Table 1.
2 Function Features 2.1 Function Features of AITR series Medical Isolation Transformer ➢ Core component of a medical IT system for converting TN into IT system (ungrounded system); ➢ The transformation ratio between the primary and secondary windings is 1:1;...
Special requirements for isolation transformers for power supply in medical locations. 4 Technical Parameters 4.1 Technical Parameters of AITR series Medical Isolation Transformer Refer to Table 2. Table 2 Technical Parameters of AITR Series of Medical Isolation Transformer Type AITR10000S AITR8000S...
5 Installation and Wiring 5.1 Shape and Mounting Hole Size 5.1.1 External dimensions of AITR series medical isolation transformer (unit: mm) Shape and size of AITR series medical isolation transformer are shown as below and in Table 6. Front view Side view...
Acrel AID150 Comm Fault Alarm Test Menu Mute AID150 Front view AID150 Side view AID150 Hole size 5.1.5 External dimensions of AKH-0.66P26 current transformer (unit: mm) Front view Side view Bottom view 5.2 Installation Method In addition to the AID series external alarm and display instrument, the M10 five-piece set insulation monitoring devices of medical IT system should be installed in the distribution cabinet (isolated power cabinet).
Page 12
5.2.2 Installation mode of AID series alarm and display instrument (1) If AID10 is embedded into the wall, the standard 86*86 mounting hole shall be reserved. The installation diagram is as follows: 86 boxes and walls Lower shell assembly 1) Separate the upper and 2) Connect the connecting wire AID10 shape lower shell of AID10...
Page 13
Front view Product shape 1. Size of wall 2. Insert the shell into openings the wall Acrel AID150 Attention : Comm Error Alarm As our products are TEST MENU constantly updated, this drawing is for Screw reference only until it has been 3.
5.3 Wiring Method 5.3.1 Wiring mode of AITR series medical isolation transformer The input terminals at the transformer terminal blocks are labeled with , in which two terminals "PM" 0 and 230 are connected to the input 220V single-phase AC. The output terminals are labeled with , in which the output voltage of two terminals 0 and 230 is AC 220V and is connected to external "SEC"...
Page 15
5.3.2 Wiring mode of AIM-M10 Lower row terminals: FE, KE (1,2) are connected to the field equipotential grounding terminal block as instrument functional grounding. A, B (3,4) are communication terminals connected with external alarm and display instrument.+24V, 0V (5,6) are DC power output to supply power for alarm and display instrument.
When an AID150 is required to access multiple AIM-M10 Medical Intelligent Insulation Monitors, all instrument communication A and B communication terminals should be connected by hand, and if the communication distance is too long or the communication is abnormal, a 120Ω matching resistor should be accessed in parallel at the first and the last ends of the communication, as shown in the following figure: The 24V power supply can be connected by multiple copper wires of 2*1.5mm...
Isolation transformer type Selected line diameter AITR3150S 3*4 mm AITR5000S/AITR6300S 3*6 mm AITR8000S/AITR10000S 3*10 mm (2) AIM-M10 insulation monitor terminals 11, 12 for auxiliary power and IT system insulation monitoring, can be connected to the back end of the medical single-phase isolation transformer, it is appropriate to series 6A rated fuse protection.
use the pressure connection with the needle-type fittings, and then insert into the corresponding terminal of the instrument and tighten the screws to avoid the abnormal work conditions of instrument caused by loose connection. (3) The grounding wire of the instrument and the transformer shall be reliably connected with the equipotential terminals in the field.
Type Acrel AIM-M10 Insulation Overload Overtemp Test (2) AID10 External Alarm and Display Panel AID10 (3) AID150 External Alarm and Display Panel Type Acrel AID150 Comm Fault Alarm Test Menu Mute 6.2 LED Indicator Instructions 6.2.1 AIM-M10 Indicator Instructions When the instrument operation is normal, the indicator light flashes, with the flashing frequency of about one time per second.
6.2.2 AID10 Indicator Instructions When the device is in normal operation, the indicator flashes, and the flickering frequency is about once a second. When the insulation resistance exceeds the alarm value, the indicator light flashes to Insulation alarm. When load current exceeds the total load current of transformer, the indicator light Overload flashes to alarm.
"Mute" button When there is alarm, press this button to eliminate the alarm sound. "▲" button In programming mode, used to increase or decrease the value. "▼" button In non-programming mode, press this button to enter the programming mode; " Menu/ " button In programming mode, used as the Enter button.
Page 22
℃ KΩ TIME (3) Exit programming mode In the programming mode, enter the "save parameter" page through the Test key, select "yes" and "no" by Up and Down keys, and the flashing indicates the selection. Press the " " key to save the parameters or not, exit the programming mode and enter the operation mode.
Page 23
When the insulation monitor or residual current monitor detects the fault, AID150 displays the corresponding alarm interface and sends out the corresponding sound and light alarm. Normal system fault indication(AIM-M10) fault indication(AIM-R100) (2) Fault record viewing interface operations and descriptions -------------------- -------NO.01-------- --------------------...
Page 24
-------------------- -------------------- Press Up and Down to input password Input password: Press MENU System normal 0000 Press Enter to confirm -------------------- 2015-07-02 12:30:45 -------------------- Password Time Press Erasure to return Slave Comm Clear Firmware -------------------- -------------------- Save param settings? Press Up and Down to select YES or NO Press Enter to confirm -------------------- --------------------...
7 Communication Protocol 7.1 Modbus-RTU Communication Protocol Meter RS485 interface adopts Modbus-RTU communication protocol, which defines the address, function code, data, check code in detail. It is the necessary content to complete the data exchange between the host and slave machine. 7.2 Introduction of Function Code 7.2.1 Function code 03H or 04H: Read the registers This function allows the user to acquire the data collected and recorded by equipment and the...
Number of High byte Number of High byte registers Low byte registers Low byte Number of registers Low byte High byte check code High byte 0004H data Low byte High byte 0005H data Low byte High byte 0006H data Low byte Low byte check code High byte...
000BH low Fault type bit7~bit6: Reserved bit5: 0 normal; 1 Temperature sensor disconnection bit4: 0 normal; 1 PE or KE disconnection fault bit3: Reserved bit2: 0 normal; 1 Transformer overheat fault Bit1: 0 normal; 1 Overload fault Bit0: 0 normal; 1 Insulation resistance fault e.g.
Page 28
iC60L MA 16A/2P OR isolation power cabinet Wall socket1 (special for surgery) iC60L MA 16A/2P Wall socket2 (special for surgery) iC60L MA 16A/2P AC220V/AC220V 8kVA Wall socket3 (special for surgery) Isolation transformer Current Transformer iC60L MA 16A/2P Wall socket4 (special for surgery) AKH-0.66P26 C65 ICB 40A/2P iC60L MA 16A/2P...
9 Power On and Debugging Instructions 9.1 Wiring Check For each set of IT system, the wiring check should be conducted before power on, mainly checking whether there is wrong, missed or short connection. The examination can be conducted sequentially in the following order according to the wiring diagrams shown in section 5.4 of this manual: (1) Check whether each four-piece set constitutes an independent IT distribution system, and ensure that the current, resistance and temperature signals monitored by each insulation monitor are connected...
9.3 Settings and Debugging (1) Acrel medical IT products need to enter a password when entering the menu settings, and the initial password for medical IT products is 0001. (2) Rated current setting. After the system is powered on, set the AIM-M10 load current alarm value according to the capacity of the isolation transformer.
Page 31
(4) System number setting. When AID150 is connected to N insulation monitors, it is necessary to set the "number of systems" in the AID150 settings to N. When the number of insulation monitors is more than 16, AID150 should be increased for network monitoring. (5) Communication view and debugging.