Instruction Manual
Non-contact Infrared Thermometer
Applicable for Model: JXB-195
NOTE:
You must read this instruction manual thoroughly before use and follow all instructions.
Keep the instructions for later use. You must check the device functions according to this
instruction manual before use each time and only use this device if it is in proper working
condition.
1

Advertisement

loading
Need help?

Need help?

Do you have a question about the JXB-195 and is the answer not in the manual?

Questions and answers

walter dereck forsyth
March 9, 2025

how to change to Fahrenheit scale forjxb-195

1 comments:
Mr. Anderson
March 9, 2025

To change the Berrcom JXB-195 thermometer to the Fahrenheit scale, press the “On/Scan” button to select Fahrenheit (℉), then press the “MODE” button to confirm.

This answer is automatically generated

Subscribe to Our Youtube Channel

Summary of Contents for Berrcom JXB-195

  • Page 1 Instruction Manual Non-contact Infrared Thermometer Applicable for Model: JXB-195 NOTE: You must read this instruction manual thoroughly before use and follow all instructions. Keep the instructions for later use. You must check the device functions according to this instruction manual before use each time and only use this device if it is in proper working...
  • Page 2: Intended Use

    I. INTENDED USE The device is an infrared thermometer intended to measure the forehead temperature of infants and adults without contacting the human body. It can be used by consumers in a household environment and doctors in the clinic as a reference. WARNINGS AND PRECAUTIONS Only use this thermometer once you have read and understood the instruction for use.
  • Page 3: Product Description

     to environment, please follow the local law and regulation on its disposal. You contact Berrcom ( “manufacture” ) or your dealer on the following issues:  If an error occurs or if the thermometer is damaged. DO NOT attempt to repair the thermometer yourself, as this will void the warranty.
  • Page 4 V. HOW TO USE YOUR NON-CONTACT INFRARED THERMOMETER 1. Install two batteries according to the manual provided in Section VIII 2. For the first time use after installing batteries, wait 10~15 minutes for the device to warm up and adjust to the room temperature. 3.
  • Page 5 compress on his or her forehead. . Do NOT take an individual’s temperature when there has been a significant change in ambient conditions. Allow the device to adjust to the significant change in the ambient temperature due to a change in environment for at least 30 minutes before taking the measurement.
  • Page 6: Technical Specifications

    to 38℃, the product will beep continuously. 6. Buzzer ON/OFF In the switch on state, press the “MODE” button for 3 second and the screen will display F1, press “MODE” button for three enter to F5 function, press the “On/Scan” button the sound will open and display “ON”...
  • Page 7: Disposing Of Batteries

    changed. To change the batteries, open the lid on the back of the device (as fig.2) and ensure the batteries are correctly positioned. Improperly installing the batteries could cause damage to the device. Do NOT use rechargeable batteries. Use single-use batteries only. fig.2 2.
  • Page 8 is seen. If the device is determined not to be visually clean at the end of the cleaning step, you should either repeat the relevant previous cleaning steps or safely dispose of the device so that a visibly soiled device is not used again. Disinfection: 1) Disinfect the thermometer casing and the probe for 30 seconds with a cotton swab or non-dust cloth dipped in 75% ethanol.
  • Page 9: Troubleshooting

    XI. TROUBLESHOOTING If you have any problem while using your thermometer, please refer to this guide to help resolve the problem. If the problem cannot be resolved, please contact our customer service. THE SCREEN DISPLAYS THE MESSAGE HI In Body Temperature Mode, the thermometer will display “HI” if the measured object temperature is higher than 43°C (109.4°F).
  • Page 10: Emc Declaration

    1, All necessary instructions for maintaining BASIC SAFETY and ESSENTIAL PERFORMANCE with regard to electromagnetic disturbances for the excepted service life. Portable and mobile RF communication equipment may affect the performance of the JXB-195, avoiding strong electromagnetic interference when used, such as near mobile phones, microwave ovens, etc.
  • Page 11 Guidance and manufacturer’s declaration -electromagnetic emissions and immunity Table 1 Guidance and manufacturer’s declaration - electromagnetic emissions Emissions test Compliance RF emissions Group 1 CISPR 11 RF emissions Class B CISPR 11 Harmonic emissions No aplicación IEC 61000-3-2 Voltage fluctuations/ flicker emissions No aplicación IEC 61000-3-3...
  • Page 12 Power frequency 30 A/m 30 A/m magnetic field 50Hz/60Hz 50Hz/60Hz IEC 61000-4-8 Conducted RF No application No application IEC61000-4-6 10 V/m 10 V/m Radiated RF 80 MHz – 2, 7 GHz 80 MHz – 2, 7 GHz IEC61000-4-3 80 % AM at 1 kHz 80 % AM at 1 kHz NOTE U is the a.c.
  • Page 13 The following warranty terms apply: 1. The warranty period for BERRCOM products is 18 months from the date of purchase. In the case of a warranty claim, the date of purchase has to be proven using the sales receipt or invoice.
  • Page 14 ANNEX IV of Directive 2012/19/EU. Before disposing of your Non-contact Infrared Thermometer (Model: JXB-195), please take away the batteries from Non-contact Infrared Thermometer (Model: JXB-195) and dispose of the batteries and Non-contact Infrared Thermometer (Model: JXB-195) separately.
  • Page 15 Other countries For country-specific information on disposal, contact your local dealers or importer. Guangzhou Berrcom Medical Device Co., Ltd. Address: No.38 Huanzhen Xi Road, Dagang Town, Nansha, 511470 Guangzhou, PEOPLE’S REPUBLIC OF CHINA Tel: +86 (20) 34938449 Fax: +86 (20) 34936960 Shanghai International Holding Corp.