Instruction Manual
Non-contact Infrared Thermometer
Applicable for Model: JXB-182
NOTE:
You must read this instruction manual thoroughly before use and follow all instructions.
Keep the instructions for later use. You must check the device functions according to this
instruction manual before use each time and only use this device if it is in proper working
condition.
1

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the JXB-182 and is the answer not in the manual?

Questions and answers

Summary of Contents for Berrcom JXB-182

  • Page 1 Instruction Manual Non-contact Infrared Thermometer Applicable for Model: JXB-182 NOTE: You must read this instruction manual thoroughly before use and follow all instructions. Keep the instructions for later use. You must check the device functions according to this instruction manual before use each time and only use this device if it is in proper working...
  • Page 2: Intended Use

    I. INTENDED USE The device is an infrared thermometer intended to measure the forehead temperature of infants and adults without contacting the human body. It can be used by consumers in a household environment and doctors in the clinic as a reference. WARNINGS AND PRECAUTIONS Only use this thermometer once you have read and understood the instruction for use.
  • Page 3: Product Description

     to environment, please follow the local law and regulation on its disposal. You contact Berrcom ( “manufacture” ) or your dealer on the following issues:  If an error occurs or if the thermometer is damaged. DO NOT attempt to repair the thermometer yourself, as this will void the warranty.
  • Page 4 V. HOW TO USE YOUR NON-CONTACT INFRARED THERMOMETER 1. Install two batteries according to the manual provided in Section VIII 2. For the first time use after installing batteries, wait 10~15 minutes for the device to warm up and adjust to the room temperature. 3.
  • Page 5 fans) . . Do NOT take an individual’s temperature shortly after he or she participated in vigorous physical activity, bathed, took medicine, ate or drank, or had a cold or warm compress on his or her forehead. . Do NOT take an individual’s temperature when there has been a significant change in ambient conditions.
  • Page 6: Technical Specifications

    In the switch on state, press "MEM" button and hold for 5 seconds, and all data in memory will be deleted. Then press "MEM” button again, the display will show "CLr". 5. High-temperature prompt function Under the body temperature mode, if the measured body temperature is higher than or equal to 38℃, the product will beep continuously.
  • Page 7: Disposing Of Batteries

    Product Size Approx. 175 x 50 x 52 mm (L x W x H) Consumption ≤300mW Expected Service Life 5 years Automatic power-off less than 30 secs Measuring site Forehead Reference body site Axillary Mode of operation Adjusted mode VIII. BATTERIES 1.
  • Page 8: Cleaning And Disinfection

    Non-contact Infrared Thermometer 1pc Instruction Manual X. CLEANING AND DISINFECTION The thermometer is a multi-patient reusable device, and the cleaning and disinfection instructions were validated. Please follow the instructions to clean/disinfect your thermometer prior to use. Cleaning: 1) Use a cotton swab moistened with hot water or neutral detergent to throughout clean the casing of thermometer and measuring probe to remove stain or oil.
  • Page 9: Troubleshooting

    Do not expose the thermometer to sunlight or water.  Do not use the thermometer rudely. Dropping, crashing and severe shock will damage  the product Calibration: This thermometer is calibrated at the time of manufactured, If the thermometer is used according to the instruction, periodic recalibration is not required.
  • Page 10: Emc Declaration

    Trade mark Indicates the need for the user to consult the instructions for use for crucial cautionary information such as warnings and precautions that can not, for a variety of reasons, be presented on the medical device itself. Please read the instructions carefully. Application part, Type BF Direct current IP22...
  • Page 11 1, All necessary instructions for maintaining BASIC SAFETY and ESSENTIAL PERFORMANCE with regard to electromagnetic disturbances for the excepted service life. Portable and mobile RF communication equipment may affect the performance of the JXB-182, avoiding strong electromagnetic interference when used, such as near mobile phones, microwave ovens, etc.
  • Page 12 IEC 61000-4-4 Surge No application No application IEC 61000-4-5 Voltage dips, short interruptions voltage variations No application No application on power supply input lines IEC 61000-4-11 Power frequency 30 A/m 30 A/m magnetic field 50Hz/60Hz 50Hz/60Hz IEC 61000-4-8 Conducted RF No application No application IEC61000-4-6...
  • Page 13 The following warranty terms apply: 1. The warranty period for BERRCOM products is 18 months from the date of purchase. In the case of a warranty claim, the date of purchase has to be proven using the sales receipt or invoice.
  • Page 14 ANNEX IV of Directive 2012/19/EU. Before disposing of your Non-contact Infrared Thermometer (Model: JXB-182), please take away the batteries from Non-contact Infrared Thermometer (Model: JXB-182) and dispose of the batteries and Non-contact Infrared Thermometer (Model: JXB-182) separately.
  • Page 15: People's Republic Of China

    authorities and treated according to 2006/66/EC. The batteries and recyclable electronic waste shall be disposed of in a separate waste container. Packaging materials must be disposed of according to local regulations. Consult the authorized collection points for more information. In Germany For the recyclable electronic waste disposal, you can search the authorized collection points at the following website: https: //www.ear-system.de/ear-verzeichnis/sammel-und-ruecknahmestellen...